Intellijoint Navigation System: Workstation G2 Configured recall (Intellijoint Surgical, Inc.)
Medical deviceClass IIOngoing
Summary: Intellijoint Surgical, Inc. is recalling Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249. Reason: A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
What is being recalled
Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249
Why
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain. (states named: California, Mississippi, New Jersey, Texas)
How to tell if yours is affected
REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-0119-2027 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | August 28, 2026 |
| Reported | October 14, 2026 |
| Company | Intellijoint Surgical, Inc. |
| Company location | Kitchener, N/A, Canada |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 8 |
More recalls from Intellijoint Surgical, Inc. · all medical device recalls · recalls from August 2026