Recall Ledger

Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined... recall (Medtronic Sofamor Danek USA Inc)

Medical deviceClass IIOngoing

Summary: Medtronic Sofamor Danek USA Inc is recalling Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder. Reason: Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection.

What is being recalled

Grafton" DBM and Grafton Plus" DBM contain demineralized human bone tissue combined with an inert additive to yield a demineralized bone matrix (DBM) allograft product having a particular physical form and/or handling property. Grafton DBM Gel and Grafton Plus DBM Paste are produced from a powder form of DBM, while Grafton DBM Flex, Putty, Matrix, Crunch" and Orthoblend are produced from a fiber form of DBM Crunch" also contains demineralized bone chips/cubes, while Grafton DBM Orthoblend contains non-demineralized cancellous bone chips in addition to DBM. Grafton DBM and Grafton Plus DBM are malleable and can be molded or cut into various sizes and shapes according to the intended implant site.

Why

Medtronic identified through donor record review that the donor for Medtronic batch A62881 did not meet all Medtronic donor eligibility requirements related to the presence of microbial organisms. While the batch passed finished product sterility testing Medtronic is retrieving any unused product manufactured from the donor tissue. Donor bone tissue tested positive for Clostridium butyricum and while tissue did undergo gamma radiation, if implanted, there is a possibility of clostridium infection

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of OK, VA, CO, AR, FL, TX, MD, DE, IL, MI, KS, AL, GA. (states named: Alabama, Arkansas, Colorado, Delaware, Florida, Georgia, Illinois, Kansas, Maryland, Michigan, Oklahoma, Texas, Virginia)

How to tell if yours is affected

UDI-DI: 00643169123021. Serial Numbers: A62881-034 through A62881-051, specifically: A62881-034, A62881-035, A62881-036, A62881-037, A62881-038, A62881-039, A62881-040, A62881-041, A62881-042, A62881-043, A62881-044, A62881-045, A62881-046, A62881-047, A62881-048, A62881-049, A62881-050, A62881-051

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0137-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 12, 2026
ReportedOctober 21, 2026
CompanyMedtronic Sofamor Danek USA Inc
Company locationMemphis, TN, United States
Initiated byVoluntary: Firm initiated
Quantity18

More recalls from Medtronic Sofamor Danek USA Inc · all medical device recalls · recalls from August 2026