Recall Ledger

TS Femoral Prep Tray recall (Howmedica Osteonics Corp.)

Medical deviceClass IIOngoing

Summary: Howmedica Osteonics Corp. is recalling TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.. Reason: Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

What is being recalled

TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.

Why

Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.

How to tell if yours is affected

Catalog Number: 6543-6-507. GTIN: 07613327557138. All lot numbers used prior to the correction date of May 8, 2026 are impacted.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0165-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedSeptember 16, 2026
ReportedOctober 21, 2026
CompanyHowmedica Osteonics Corp.
Company locationMahwah, NJ, United States
Initiated byVoluntary: Firm initiated
Quantity582 trays

More recalls from Howmedica Osteonics Corp. · all medical device recalls · recalls from September 2026

Official FDA record