Recall Ledger

Medline surgical convenience kits labeled as: CABG AUXILIARY PACK recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline surgical convenience kits labeled as: CABG AUXILIARY PACK, Medline Product Number/SKU DYNJ44925D. Reason: Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.

What is being recalled

Medline surgical convenience kits labeled as: CABG AUXILIARY PACK, Medline Product Number/SKU DYNJ44925D

Why

Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

UDI/DI 10193489461725 (each) 40193489461726 (case), Lot Numbers: 25LMF813, 25IMF341, 25HBR484, 25FBP179, 25EBJ732, 25DBK745, 25CBK191, 25BBE801, 24LBB945, 24IBT658, 24IBD265, 24GBG100, 24EBC752, 24CBQ893, 24BBJ088, 24ABE008, 23JBA374, 23HBV035, 23GBL681, 23EBK074, 23DBL164, 23DBB447, 23ABV418, 22LBG352, 22IBK243, 22HBX829, 22EBA032, 22DBO915, 22BBR573, 21LBB905, 21IBX719, 21HBG260, 21GBH209, 21DBR261, 21BBJ878.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0168-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedSeptember 2, 2026
ReportedOctober 21, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity3125 kits

More recalls from Medline Industries, LP · all medical device recalls · recalls from September 2026