Techno-Path Multichem AE Unassayed Level 1 REF:AE621X recall (TECHNO-PATH MANUFACTURING LTD.)

Medical deviceClass IIOngoing

Summary: TECHNO-PATH MANUFACTURING LTD. is recalling Techno-Path Multichem AE Unassayed Level 1 REF:AE621X. Reason: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

What is being recalled

Techno-Path Multichem AE Unassayed Level 1 REF:AE621X

Why

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the state of Massachusetts. (states named: Massachusetts)

How to tell if yours is affected

UDI: (01)05391523442800/Lot # 65705261

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0205-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedAugust 14, 2026
ReportedOctober 28, 2026
CompanyTECHNO-PATH MANUFACTURING LTD.
Company locationBallina Tipperary, Ireland
Initiated byVoluntary: Firm initiated
Quantity832 units