Recall Ledger

Direxion Torqueable Microcatheter recall (Boston Scientific Cork, Ltd.)

Medical deviceClass IIOngoing

Summary: Boston Scientific Cork, Ltd. is recalling Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280. Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

What is being recalled

Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280

Why

The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution US Nationwide and territory of Puerto Rico. The countries of Albania, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Lithuania, Mexico, Moldova, Netherlands, New Zealand, Norway, Panama, PERU, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tunisia, T¿rkiye, UAE, United Kingdom. (states named: Puerto Rico)

How to tell if yours is affected

Material Number: M001195280; UDI-DI: 08714729839392; Batch Numbers: 38149679, 38323401, 38444072, 38506482, 38506491, 38560458, 38595645, 38810252, 38936816, 39140997, 39141264, 39189785, 39189791, 40482687, 40493382, 40512909;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0231-2027
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedSeptember 16, 2026
ReportedOctober 28, 2026
CompanyBoston Scientific Cork, Ltd.
Company locationCork, Ireland
Initiated byVoluntary: Firm initiated
Quantity381

More recalls from Boston Scientific Cork, Ltd. · all medical device recalls · recalls from September 2026

Official FDA record