Philips Azurion 7 M20. Model Numbers: 722079 recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)
Medical deviceClass IIOngoing
Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Philips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.. Reason: Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
What is being recalled
Philips Azurion 7 M20. Model Numbers: 722079, 722224, 722282.
Why
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.
How to tell if yours is affected
1. Model Number: 722079. UDI Numbers: (01)00884838085268(21)1023, (01)00884838085268(21)646, (01)00884838085268(21)703904, (01)00884838085268(21)890, (01)00884838085268(21)1137, (01)00884838085268(21)1048, (01)00884838085268(21)401, (01)00884838085268(21)536, (01)00884838085268(21)1054, (01)00884838085268(21)1010, (01)00884838085268(21)266, (01)00884838085268(21)617, (01)00884838085268(21)246, (01)00884838085268(21)905, (01)00884838085268(21)809, (01)00884838085268(21)848, (01)00884838085268(21)1033, (01)00884838085268(21)836, (01)00884838085268(21)466, (01)00884838085268(21)1155, (01)00884838085268(21)673, (01)00884838085268(21)564, (01)00884838085268(21)1032, (01)00884838085268(21)1117, (01)00884838085268(21)820, (01)00884838085268(21)1211, (01)00884838085268(21)1213, (01)00884838085268(21)47, (01)00884838085268(21)1365, (01)00884838085268(21)881, (01)00884838085268(21)774, (01)00884838085268(21)955, (01)00884838085268(21)633, (01)00884838085268(21)703540, (01)00884838085268(21)296, (01)00884838085268(21)574, (01)00884838085268(21)842, (01)00884838085268(21)884, (01)00884838085268(21)991, (01)00884838085268(21)703336, (01)00884838085268(21)753, (01)00884838085268(21)703493, (01)00884838085268(21)790, (01)00884838085268(21)481, (01)00884838085268(21)407, (01)00884838085268(21)408, (01)00884838085268(21)877, (01)00884838085268(21)703395, (01)00884838085268(21)703323, (01)00884838085268(21)483, (01)00884838085268(21)241, (01)00884838085268(21)1108, (01)00884838085268(21)1363, (01)00884838085268(21)1414, (01)00884838085268(21)703404, (01)00884838085268(21)703387, (01)00884838085268(21)1355, (01)00884838085268(21)863, (01)00884838085268(21)350, (01)00884838085268(21)908, (01)00884838085268(21)1378, (01)00884838085268(21)262, (01)00884838085268(21)598, (01)00884838085268(21)275, (01)00884838085268(21)520, (01)00884838085268(21)593, (01)00884838085268(21)1018, (01)00884838085268(21)1432, (01)00884838085268(21)594, (01)00884838085268(21)999, (01)00884838085268(21)361, (01)00884838085268(21)526, (01)00884838085268(21)1021, (01)00884838085268(21)608, (01)00884838085268(21)611, (01)00884838085268(21)413, (01)00884838085268(21)895, (01)00884838085268(21)428, (01)00884838085268(21)883, (01)00884838085268(21)758, (01)00884838085268(21)274, (01)00884838085268(21)525, (01)00884838085268(21)885, (01)00884838085268(21)843, (01)00884838085268(21)754, (01)00884838085268(21)703528, (01)00884838085268(21)599, (01)00884838085268(21)1171, (01)00884838085268(21)817, (01)00884838085268(21)766, (01)00884838085268(21)878, (01)00884838085268(21)1170, (01)00884838085268(21)703310, (01)00884838085268(21)944, (01)00884838085268(21)703305, (01)00884838085268(21)1285, (01)00884838085268(21)854, (01)00884838085268(21)730, (01)00884838085268(21)703497, (01)00884838085268(21)957, (01)00884838085268(21)658, (01)00884838085268(21)1022, (01)00884838085268(21)621, (01)00884838085268(21)628, (01)00884838085268(21)795, (01)00884838085268(21)1007, (01)00884838085268(21)1289, (01)00884838085268(21)605, (01)00884838085268(21)65, (01)00884838085268(21)34, (01)00884838085268(21)1222, (01)00884838085268(21)769, (01)00884838085268(21)938, (01)00884838085268(21)655, (01)00884838085268(21)1311, (01)00884838085268(21)538, (01)00884838085268(21)989, (01)00884838085268(21)946, (01)00884838085268(21)888, (01)00884838085268(21)402, (01)00884838085268(21)1042, (01)00884838085268(21)374, (01)00884838085268(21)459, (01)00884838085268(21)1282, (01)00884838085268(21)841, (01)00884838085268(21)703532, (01)00884838085268(21)1360, (01)00884838085268(21)1359, (01)00884838085268(21)1358, (01)00884838085268(21)50, (01)00884838085268(21)326, (01)00884838085268(21)216, (01)00884838085268(21)927, (01)00884838085268(21)981, (01)00884838085268(21)923, (01)00884838085268(21)1163, (01)00884838085268(21)416, (01)00884838085268(21)424, (01)00884838085268(21)802, (01)00884838085268(21)751, (01)00884838085268(21)436, (01)00884838085268(21)910, (01)00884838085268(21)1242, (01)00884838085268(21)251, (01)00884838085268(21)1156, (01)00884838085268(21)1049, (01)00884838085268(21)1103, (01)00884838085268(21)1361, (01)00884838085268(21)1041, (01)00884838085268(21)29, (01)00884838085268(21)930, (01)00884838085268(21)228, (01)00884838085268(21)30, (01)00884838085268(21)366, (01)00884838085268(21)543, (01)00884838085268(21)921, (01)00884838085268(21)1200, (01)00884838085268(21)359, (01)00884838085268(21)355, (01)00884838085268(21)357, (01)00884838085268(21)356, (01)00884838085268(21)1090, (01)00884838085268(21)801, (01)00884838085268(21)474, (01)00884838085268(21)1187, (01)00884838085268(21)247, (01)00884838085268(21)70, (01)00884838085268(21)742, (01)00884838085268(21)62, (01)00884838085268(21)222, (01)00884838085268(21)537, (01)00884838085268(21)351, (01)00884838085268(21)777, (01)00884838085268(21)898, (01)00884838085268(21)703317, (01)00884838085268(21)653, (01)00884838085268(21)684, (01)00884838085268(21)746, (01)00884838085268(21)998, (01)00884838085268(21)668, (01)00884838085268(21)706, (01)00884838085268(21)847, (01)00884838085268(21)879, (01)00884838085268(21)703308, (01)00884838085268(21)837, (01)00884838085268(21)387, (01)00884838085268(21)417, (01)00884838085268(21)1393, (01)00884838085268(21)212, (01)00884838085268(21)290, (01)00884838085268(21)669, (01)00884838085268(21)239, (01)00884838085268(21)800, (01)00884838085268(21)56, (01)00884838085268(21)703584, (01)00884838085268(21)595, (01)00884838085268(21)64, (01)00884838085268(21)1059, (01)00884838085268(21)1382, (01)00884838085268(21)1381, (01)00884838085268(21)747, (01)00884838085268(21)663, (01)00884838085268(21)703333, (01)00884838085268(21)862, (01)00884838085268(21)703345, (01)00884838085268(21)427, (01)00884838085268(21)60, (01)00884838085268(21)1100, (01)00884838085268(21)327, (01)00884838085268(21)602, (01)00884838085268(21)937, (01)00884838085268(21)1201, (01)00884838085268(21)702, (01)00884838085268(21)333, (01)00884838085268(21)352, (01)00884838085268(21)284, (01)00884838085268(21)679, (01)00884838085268(21)1374, (01)00884838085268(21)750, (01)00884838085268(21)783, (01)00884838085268(21)703813, (01)00884838085268(21)704301, (01)00884838085268(21)780, (01)00884838085268(21)704114, (01)00884838085268(21)703719, (01)00884838085268(21)217, (01)00884838085268(21)816, (01)00884838085268(21)214, (01)00884838085268(21)1280, (01)00884838085268(21)703565, (01)00884838085268(21)704304, (01)00884838085268(21)719, (01)00884838085268(21)704059, (01)00884838085268(21)729, (01)00884838085268(21)943, (01)00884838085268(21)337, (01)00884838085268(21)704184, (01)00884838085268(21)704240, (01)00884838085268(21)704242, (01)00884838085268(21)703662, (01)00884838085268(21)1340, (01)00884838085268(21)703823, (01)00884838085268(21)1026, (01)00884838085268(21)377, (01)00884838085268(21)380, (01)00884838085268(21)379, (01)00884838085268(21)662, (01)00884838085268(21)703449, (01)00884838085268(21)258, (01)00884838085268(21)260, (01)00884838085268(21)451, (01)00884838085268(21)703699, (01)00884838085268(21)259, (01)00884838085268(21)703355, (01)00884838085268(21)703654, (01)00884838085268(21)703870, (01)00884838085268(21)395, (01)00884838085268(21)433, (01)00884838085268(21)1405, (01)00884838085268(21)703999, (01)00884838085268(21)1047, (01)00884838085268(21)1241, (01)00884838085268(21)752, (01)00884838085268(21)703375, (01)00884838085268(21)1319, (01)00884838085268(21)703712, (01)00884838085268(21)31, (01)00884838085268(21)703326, (01)00884838085268(21)703655, (01)00884838085268(21)703702, (01)00884838085268(21)703939, (01)00884838085268(21)531, (01)00884838085268(21)703802, (01)00884838085268(21)703897, (01)00884838085268(21)704203, (01)00884838085268(21)1389, (01)00884838085268(21)703953, (01)00884838085268(21)1053, (01)00884838085268(21)703600, (01)00884838085268(21)570, (01)00884838085268(21)703769, (01)00884838085268(21)704026, (01)00884838085268(21)703753, (01)00884838085268(21)703967, (01)00884838085268(21)704112, (01)00884838085268(21)984, (01)00884838085268(21)703807, (01)00884838085268(21)703529, (01)00884838085268(21)1073, (01)00884838085268(21)704154, (01)00884838085268(21)778, (01)00884838085268(21)548, (01)00884838085268(21)798, (01)00884838085268(21)704037, (01)00884838085268(21)316, (01)00884838085268(21)703669, (01)00884838085268(21)1429, (01)00884838085268(21)49, (01)00884838085268(21)704209, (01)00884838085268(21)722, (01)00884838085268(21)703619, (01)00884838085268(21)704142, (01)00884838085268(21)600, (01)00884838085268(21)825, (01)00884838085268(21)452, (01)00884838085268(21)969, (01)00884838085268(21)1020, (01)00884838085268(21)703358, (01)00884838085268(21)1095, (01)00884838085268(21)703487, (01)00884838085268(21)704092, (01)00884838085268(21)461, (01)00884838085268(21)765, (01)00884838085268(21)578, (01)00884838085268(21)703572, (01)00884838085268(21)982, (01)00884838085268(21)703462, (01)00884838085268(21)703583, (01)00884838085268(21)1106, (01)00884838085268(21)1044, (01)00884838085268(21)703855, (01)00884838085268(21)703348, (01)00884838085268(21)703872, (01)00884838085268(21)855, (01)00884838085268(21)1030, (01)00884838085268(21)1120, (01)00884838085268(21)703959, (01)00884838085268(21)703489, (01)00884838085268(21)913, (01)00884838085268(21)304, (01)00884838085268(21)303, (01)00884838085268(21)704101, (01)00884838085268(21)305, (01)00884838085268(21)1194, (01)00884838085268(21)703961, (01)00884838085268(21)1337, (01)00884838085268(21)993, (01)00884838085268(21)347, (01)00884838085268(21)74, (01)00884838085268(21)1074, (01)00884838085268(21)230, (01)00884838085268(21)293, (01)00884838085268(21)421, (01)00884838085268(21)434, (01)00884838085268(21)961, (01)00884838085268(21)44, (01)00884838085268(21)1185, (01)00884838085268(21)703477, (01)00884838085268(21)257, (01)00884838085268(21)661, (01)00884838085268(21)704146, (01)00884838085268(21)703982, (01)00884838085268(21)1442, (01)00884838085268(21)1441, (01)00884838085268(21)1440, (01)00884838085268(21)704211, (01)00884838085268(21)703812, (01)00884838085268(21)703423, (01)00884838085268(21)1317, (01)00884838085268(21)1264, (01)00884838085268(21)1278, (01)00884838085268(21)703313, (01)00884838085268(21)703544, (01)00884838085268(21)703749, (01)00884838085268(21)1347, (01)00884838085268(21)958, (01)00884838085268(21)951, (01)00884838085268(21)484, (01)00884838085268(21)703470, (01)00884838085268(21)542, (01)00884838085268(21)235, (01)00884838085268(21)486, (01)00884838085268(21)700, (01)00884838085268(21)348, (01)00884838085268(21)703867, (01)00884838085268(21)1154, (01)00884838085268(21)703965, (01)00884838085268(21)704239, (01)00884838085268(21)703993, (01)00884838085268(21)703542, (01)00884838085268(21)551, (01)00884838085268(21)550, (01)00884838085268(21)703526, (01)00884838085268(21)704085, (01)00884838085268(21)704079, (01)00884838085268(21)703485, (01)00884838085268(21)547, (01)00884838085268(21)371, (01)00884838085268(21)485, (01)00884838085268(21)596, (01)00884838085268(21)1052, (01)00884838085268(21)703804, (01)00884838085268(21)703906, (01)00884838085268(21)703596, (01)00884838085268(21)703937, (01)00884838085268(21)1288, (01)00884838085268(21)703954, (01)00884838085268(21)1183, (01)00884838085268(21)368, (01)00884838085268(21)704262, (01)00884838085268(21)703556, (01)00884838085268(21)902, (01)00884838085268(21)1422, (01)00884838085268(21)571, (01)00884838085268(21)278, (01)00884838085268(21)703754, (01)00884838085268(21)457, (01)00884838085268(21)722079_703569, (01)00884838085268(21)704081, (01)00884838085268(21)704103, (01)00884838085268(21)704019, (01)00884838085268(21)1119, (01)00884838085268(21)960, (01)00884838085268(21)703424, (01)00884838085268(21)704023, (01)00884838085268(21)704119, (01)00884838085268(21)1015, (01)00884838085268(21)1351, (01)00884838085268(21)1076, (01)00884838085268(21)1129, (01)00884838085268(21)704030, (01)00884838085268(21)704126, (01)00884838085268(21)956, (01)00884838085268(21)703819, (01)00884838085268(21)704036, (01)00884838085268(21)703690, (01)00884838085268(21)703648, (01)00884838085268(21)703709, (01)00884838085268(21)704207, (01)00884838085268(21)703728, (01)00884838085268(21)703826, (01)00884838085268(21)381, (01)00884838085268(21)703794, (01)00884838085268(21)703607, (01)00884838085268(21)703574, (01)00884838085268(21)762, (01)00884838085268(21)703743, (01)00884838085268(21)703632, (01)00884838085268(21)703913, (01)00884838085268(21)703889, (01)00884838085268(21)703700, (01)00884838085268(21)704246, (01)00884838085268(21)704054, (01)00884838085268(21)703606, (01)00884838085268(21)703473, (01)00884838085268(21)703346, (01)00884838085268(21)521, (01)00884838085268(21)703750, (01)00884838085268(21)703990, (01)00884838085268(21)703834, (01)00884838085268(21)703727, (01)00884838085268(21)586, (01)00884838085268(21)965, (01)00884838085268(21)498, (01)00884838085268(21)789, (01)00884838085268(21)728, (01)00884838085268(21)291, (01)00884838085268(21)703482, (01)00884838085268(21)703693, (01)00884838085268(21)703533, (01)00884838085268(21)73, (01)00884838085268(21)967, (01)00884838085268(21)475, (01)00884838085268(21)704067, (01)00884838085268(21)349, (01)00884838085268(21)784, (01)00884838085268(21)704274, (01)00884838085268(21)703506, (01)00884838085268(21)66, (01)00884838085268(21)233, (01)00884838085268(21)704144, (01)00884838085268(21)704167, (01)00884838085268(21)704168, (01)00884838085268(21)698, (01)00884838085268(21)643, (01)00884838085268(21)642, (01)00884838085268(21)1210, (01)00884838085268(21)435, (01)00884838085268(21)704070, (01)00884838085268(21)1077, (01)00884838085268(21)450, (01)00884838085268(21)42, (01)00884838085268(21)703614, (01)00884838085268(21)703973, (01)00884838085268(21)704110, (01)00884838085268(21)703701, (01)00884838085268(21)703628, (01)00884838085268(21)703948, (01)00884838085268(21)703955, (01)00884838085268(21)703665, (01)00884838085268(21)1367, (01)00884838085268(21)703899, (01)00884838085268(21)704202, (01)00884838085268(21)703641, (01)00884838085268(21)703711, (01)00884838085268(21)704173, (01)00884838085268(21)703856, (01)00884838085268(21)703842, (01)00884838085268(21)703651, (01)00884838085268(21)704186, (01)00884838085268(21)704269, (01)00884838085268(21)704074, (01)00884838085268(21)704108, (01)00884838085268(21)703821, (01)00884838085268(21)704134, (01)00884838085268(21)704029, (01)00884838085268(21)703601, (01)00884838085268(21)703979, (01)00884838085268(21)703777, (01)00884838085268(21)703741, (01)00884838085268(21)704041, (01)00884838085268(21)703778, (01)00884838085268(21)703670, (01)00884838085268(21)703650, (01)00884838085268(21)703646, (01)00884838085268(21)703846, (01)00884838085268(21)703945, (01)00884838085268(21)704021, (01)00884838085268(21)703909, (01)00884838085268(21)703908, (01)00884838085268(21)704031, (01)00884838085268(21)704271, (01)00884838085268(21)703756, (01)00884838085268(21)703618, (01)00884838085268(21)703724, (01)00884838085268(21)704060, (01)00884838085268(21)704120, (01)00884838085268(21)704017, (01)00884838085268(21)703921, (01)00884838085268(21)703816, (01)00884838085268(21)703988, (01)00884838085268(21)703573, (01)00884838085268(21)703997, (01)00884838085268(21)703674, (01)00884838085268(21)703673, (01)00884838085268(21)703591, (01)00884838085268(21)704002, (01)00884838085268(21)703733, (01)00884838085268(21)703735, (01)00884838085268(21)703986, (01)00884838085268(21)703940, (01)00884838085268(21)703757, (01)00884838085268(21)703751, (01)00884838085268(21)703883, (01)00884838085268(21)704139, (01)00884838085268(21)704157, (01)00884838085268(21)704265, (01)00884838085268(21)704279, (01)00884838085268(21)704289, (01)00884838085268(21)704280, (01)00884838085268(21)704024, (01)00884838085268(21)703911, (01)00884838085268(21)703912, (01)00884838085268(21)703605, (01)00884838085268(21)703797, (01)00884838085268(21)704022, (01)00884838085268(21)704039, (01)00884838085268(21)703471, (01)00884838085268(21)704095, (01)00884838085268(21)703663, (01)00884838085268(21)704138, (01)00884838085268(21)703604, (01)00884838085268(21)704247, (01)00884838085268(21)703707, (01)00884838085268(21)704008, (01)00884838085268(21)704221, (01)00884838085268(21)703799, (01)00884838085268(21)703576, (01)00884838085268(21)704009, (01)00884838085268(21)704069, (01)00884838085268(21)703840, (01)00884838085268(21)704080, (01)00884838085268(21)703998, (01)00884838085268(21)704012, (01)00884838085268(21)703603, (01)00884838085268(21)704130, (01)00884838085268(21)703786, (01)00884838085268(21)704131, (01)00884838085268(21)704078, (01)00884838085268(21)703630, (01)00884838085268(21)704066, (01)00884838085268(21)703983, (01)00884838085268(21)703507, (01)00884838085268(21)704001, (01)00884838085268(21)703645, (01)00884838085268(21)703919, (01)00884838085268(21)704196, (01)00884838085268(21)704197, (01)00884838085268(21)703616, (01)00884838085268(21)704201, (01)00884838085268(21)704133, (01)00884838085268(21)703796, (01)00884838085268(21)703857, (01)00884838085268(21)703859, (01)00884838085268(21)703863, (01)00884838085268(21)703858, (01)00884838085268(21)703860, (01)00884838085268(21)703780, (01)00884838085268(21)704166, (01)00884838085268(21)703642, (01)00884838085268(21)703640, (01)00884838085268(21)704093, (01)00884838085268(21)704259, (01)00884838085268(21)703680, (01)00884838085268(21)703878, (01)00884838085268(21)703901, (01)00884838085268(21)704020, (01)00884838085268(21)703884, (01)00884838085268(21)703958, (01)00884838085268(21)704091, (01)00884838085268(21)1376, (01)00884838085268(21)977, (01)00884838085268(21)1093, (01)00884838085268(21)703975, (01)00884838085268(21)704113, (01)00884838085268(21)704090, (01)00884838085268(21)704147, (01)00884838085268(21)703571, (01)00884838085268(21)703703, (01)00884838085268(21)703887, (01)00884838085268(21)703923, (01)00884838085268(21)704145, (01)00884838085268(21)704215, (01)00884838085268(21)704111, (01)00884838085268(21)704217, (01)00884838085268(21)704124, (01)00884838085268(21)703585, (01)00884838085268(21)704094, (01)00884838085268(21)703759, (01)00884838085268(21)703639, (01)00884838085268(21)703647, (01)00884838085268(21)703627, (01)00884838085268(21)704064, (01)00884838085268(21)704050, (01)00884838085268(21)704255, (01)00884838085268(21)704175, (01)00884838085268(21)703697, (01)00884838085268(21)1263, (01)00884838085268(21)703894, (01)00884838085268(21)703895, (01)00884838085268(21)703891, (01)00884838085268(21)704181, (01)00884838085268(21)704180, (01)00884838085268(21)704007, (01)00884838085268(21)703789, (01)00884838085268(21)704075, (01)00884838085268(21)703827, (01)00884838085268(21)703587, (01)00884838085268(21)704056, (01)00884838085268(21)704122, (01)00884838085268(21)703876, (01)00884838085268(21)703989, (01)00884838085268(21)704046, (01)00884838085268(21)703679, (01)00884838085268(21)703924, (01)00884838085268(21)704223, (01)00884838085268(21)704051, (01)00884838085268(21)703934, (01)00884838085268(21)1287, (01)00884838085268(21)704216, (01)00884838085268(21)704267, (01)00884838085268(21)704177, (01)00884838085268(21)703825, (01)00884838085268(21)704115, (01)00884838085268(21)703836, (01)00884838085268(21)703932, (01)00884838085268(21)703661, (01)00884838085268(21)704152, (01)00884838085268(21)703831, (01)00884838085268(21)704132, (01)00884838085268(21)703548, (01)00884838085268(21)704200, (01)00884838085268(21)703666, (01)00884838085268(21)703730, (01)00884838085268(21)703837, (01)00884838085268(21)1167, (01)00884838085268(21)703798, (01)00884838085268(21)703818, (01)00884838085268(21)704096, (01)00884838085268(21)704151, (01)00884838085268(21)703725, (01)00884838085268(21)704178, (01)00884838085268(21)704161, (01)00884838085268(21)704014, (01)00884838085268(21)703977, (01)00884838085268(21)704082, (01)00884838085268(21)704089, (01)00884838085268(21)703888, (01)00884838085268(21)703930, (01)00884838085268(21)703928, (01)00884838085268(21)703929, (01)00884838085268(21)703931, (01)00884838085268(21)703974, (01)00884838085268(21)703881, (01)00884838085268(21)704252, (01)00884838085268(21)703941, (01)00884838085268(21)703917, (01)00884838085268(21)703667, (01)00884838085268(21)703668, (01)00884838085268(21)704153, (01)00884838085268(21)703617, (01)00884838085268(21)704169, (01)00884838085268(21)704219, (01)00884838085268(21)704275, (01)00884838085268(21)703625, (01)00884838085268(21)703660, (01)00884838085268(21)704104, (01)00884838085268(21)704128, (01)00884838085268(21)704065, (01)00884838085268(21)1228, (01)00884838085268(21)704116, (01)00884838085268(21)704102, (01)00884838085268(21)703898, (01)00884838085268(21)703636, (01)00884838085268(21)703843, (01)00884838085268(21)704264, (01)00884838085268(21)703692, (01)00884838085268(21)703664, (01)00884838085268(21)703555, (01)00884838085268(21)704218, (01)00884838085268(21)704109, (01)00884838085268(21)1224, (01)00884838085268(21)1225, (01)00884838085268(21)704125, (01)00884838085268(21)703684, (01)00884838085268(21)703742, (01)00884838085268(21)703723, (01)00884838085268(21)703635, (01)00884838085268(21)704206, (01)00884838085268(21)703910, (01)00884838085268(21)703553, (01)00884838085268(21)704189, (01)00884838085268(21)703951, (01)00884838085268(21)704098, (01)00884838085268(21)703539, (01)00884838085268(21)703541, (01)00884838085268(21)703678, (01)00884838085268(21)1034, (01)00884838085268(21)703900, (01)00884838085268(21)703893, (01)00884838085268(21)703761, (01)00884838085268(21)756, (01)00884838085268(21)786, (01)00884838085268(21)390, (01)00884838085268(21)398, (01)00884838085268(21)703653, (01)00884838085268(21)1046, (01)00884838085268(21)1255, (01)00884838085268(21)665, (01)00884838085268(21)667, (01)00884838085268(21)317, (01)00884838085268(21)703715, (01)00884838085268(21)1253, (01)00884838085268(21)703414, (01)00884838085268(21)456, (01)00884838085268(21)941, (01)00884838085268(21)703505, (01)00884838085268(21)392, (01)00884838085268(21)544, (01)00884838085268(21)323, (01)00884838085268(21)703658, (01)00884838085268(21)704135, (01)00884838085268(21)704149, (01)00884838085268(21)1332, (01)00884838085268(21)703873, (01)00884838085268(21)704010, (01)00884838085268(21)77, (01)00884838085268(21)704057, (01)00884838085268(21)1357, (01)00884838085268(21)703459, (01)00884838085268(21)703944, (01)00884838085268(21)952, (01)00884838085268(21)676, (01)00884838085268(21)703720, (01)00884838085268(21)703892, (01)00884838085268(21)410, (01)00884838085268(21)75, (01)00884838085268(21)703985, (01)00884838085268(21)338, (01)00884838085268(21)906, (01)00884838085268(21)1349, (01)00884838085268(21)703579, (01)00884838085268(21)703810, (01)00884838085268(21)704303, (01)00884838085268(21)703694, (01)00884838085268(21)703874, (01)00884838085268(21)1188, (01)00884838085268(21)764, (01)00884838085268(21)759, (01)00884838085268(21)704015, (01)00884838085268(21)703675, (01)00884838085268(21)703970, (01)00884838085268(21)703877, (01)00884838085268(21)704117, (01)00884838085268(21)703956, (01)00884838085268(21)704084, (01)00884838085268(21)704213, (01)00884838085268(21)467, (01)00884838085268(21)276, (01)00884838085268(21)670, (01)00884838085268(21)1384, (01)00884838085268(21)704188, (01)00884838085268(21)703717, (01)00884838085268(21)703922, (01)00884838085268(21)796, (01)00884838085268(21)704016, (01)00884838085268(21)703729, (01)00884838085268(21)704243, (01)00884838085268(21)695, (01)00884838085268(21)616, (01)00884838085268(21)1139, (01)00884838085268(21)775, (01)00884838085268(21)703372, (01)00884838085268(21)703440, (01)00884838085268(21)703502, (01)00884838085268(21)582, (01)00884838085268(21)494, (01)00884838085268(21)703377, (01)00884838085268(21)405, (01)00884838085268(21)562, (01)00884838085268(21)24, (01)00884838085268(21)703952, (01)00884838085268(21)704018, (01)00884838085268(21)703957, (01)00884838085268(21)704033, (01)00884838085268(21)704032, (01)00884838085268(21)703695, (01)00884838085268(21)703475, (01)00884838085268(21)703947, (01)00884838085268(21)704063, (01)00884838085268(21)703890, (01)00884838085268(21)703589, (01)00884838085268(21)704204, (01)00884838085268(21)703536, (01)00884838085268(21)703849, (01)00884838085268(21)703659, (01)00884838085268(21)704025, (01)00884838085268(21)703926, (01)00884838085268(21)703978, (01)00884838085268(21)703991, (01)00884838085268(21)704260, (01)00884838085268(21)703987, (01)00884838085268(21)704250, (01)00884838085268(21)704195, (01)00884838085268(21)703672, (01)00884838085268(21)703468, (01)00884838085268(21)703467, (01)00884838085268(21)703652, (01)00884838085268(21)703714, (01)00884838085268(21)703481, (01)00884838085268(21)419, (01)00884838085268(21)724, (01)00884838085268(21)1057, (01)00884838085268(21)771, (01)00884838085268(21)1339, (01)00884838085268(21)1176, (01)00884838085268(21)1160, (01)00884838085268(21)703490, (01)00884838085268(21)703631, (01)00884838085268(21)703549, (01)00884838085268(21)703443, (01)00884838085268(21)1025, (01)00884838085268(21)1336, (01)00884838085268(21)703339, (01)00884838085268(21)1161, (01)00884838085268(21)703352, (01)00884838085268(21)1131, (01)00884838085268(21)1091, (01)00884838085268(21)1390, (01)00884838085268(21)1178, (01)00884838085268(21)703649, (01)00884838085268(21)703531, (01)00884838085268(21)703762, (01)00884838085268(21)703763, (01)00884838085268(21)703534, (01)00884838085268(21)703705, (01)00884838085268(21)703706, (01)00884838085268(21)703704, (01)00884838085268(21)703554, (01)00884838085268(21)703444, (01)00884838085268(21)703353, (01)00884838085268(21)1058, (01)00884838085268(21)703787, (01)00884838085268(21)1075, (01)00884838085268(21)703431, (01)00884838085268(21)827, (01)00884838085268(21)1024, (01)00884838085268(21)703410, (01)00884838085268(21)844, (01)00884838085268(21)1094, (01)00884838085268(21)1069, (01)00884838085268(21)703522, (01)00884838085268(21)703354, (01)00884838085268(21)703590, (01)00884838085268(21)1180, (01)00884838085268(21)1067, (01)00884838085268(21)723, (01)00884838085268(21)1159, (01)00884838085268(21)703795, (01)00884838085268(21)703570, (01)00884838085268(21)703454, (01)00884838085268(21)703721, (01)00884838085268(21)703722, (01)00884838085268(21)1173, (01)00884838085268(21)703422, (01)00884838085268(21)703518, (01)00884838085268(21)1404, (01)00884838085268(21)1437, (01)00884838085268(21)1092, (01)00884838085268(21)703484, (01)00884838085268(21)703325, (01)00884838085268(21)568, (01)00884838085268(21)1207, (01)00884838085268(21)703865, (01)00884838085268(21)703403, (01)00884838085268(21)1068, (01)00884838085268(21)1099, (01)00884838085268(21)703519, (01)00884838085268(21)1126, (01)00884838085268(21)703916, (01)00884838085268(21)703907, (01)00884838085268(21)703451, (01)00884838085268(21)1071, (01)00884838085268(21)1307, (01)00884838085268(21)612, (01)00884838085268(21)703373, (01)00884838085268(21)277, (01)00884838085268(21)917, (01)00884838085268(21)225, (01)00884838085268(21)256, (01)00884838085268(21)918, (01)00884838085268(21)640, (01)00884838085268(21)864, (01)00884838085268(21)288, (01)00884838085268(21)794, (01)00884838085268(21)692, (01)00884838085268(21)703363, (01)00884838085268(21)980, (01)00884838085268(21)703452, (01)00884838085268(21)463, (01)00884838085268(21)541, (01)00884838085268(21)893, (01)00884838085268(21)1145, (01)00884838085268(21)373, (01)00884838085268(21)929, (01)00884838085268(21)609. Serial Numbers: (1023, 646, 703904, 890, 1137, 1048, 401, 536, 1054, 1010, 266, 617, 246, 905, 809, 848, 1033, 836, 466, 1155, 673, 564, 1032, 1117, 820, 1211, 1213, 47, 1365, 881, 774, 955, 633, 703540, 296, 574, 842, 884, 991, 703336, 753, 703493, 790, 481, 407, 408, 877, 703395, 703323, 483, 241, 1108, 1363, 1414, 703404, 703387, 1355, 863, 350, 908, 1378, 262, 598, 275, 520, 593, 1018, 1432, 594, 999, 361, 526, 1021, 608, 611, 413, 895, 428, 883, 758, 274, 525, 885, 843, 754, 703528, 599, 1171, 817, 766, 878, 1170, 703310, 944, 703305, 1285, 854, 730, 703497, 957, 658, 1022, 621, 628, 795, 1007, 1289, 605, 65, 34, 1222, 769, 938, 655, 1311, 538, 989, 946, 888, 402, 1042, 374, 459, 1282, 841, 703532, 1360, 1359, 1358, 50, 326, 216, 927, 981, 923, 1163, 416, 424, 802, 751, 436, 910, 1242, 251, 1156, 1049, 1103, 1361, 1041, 29, 930, 228, 30, 366, 543, 921, 1200, 359, 355, 357, 356, 1090, 801, 474, 1187, 247, 70, 742, 62, 222, 537, 351, 777, 898, 703317, 653, 684, 746, 998, 668, 706, 847, 879, 703308, 837, 387, 417, 1393, 212, 290, 669, 239, 800, 56, 703584, 595, 64, 1059, 1382, 1381, 747, 663, 703333, 862, 703345, 427, 60, 1100, 327, 602, 937, 1201, 702, 333, 352, 284, 679, 1374, 750, 783, 703813, 704301, 780, 704114, 703719, 217, 816, 214, 1280, 703565, 704304, 719, 704059, 729, 943, 337, 704184, 704240, 704242, 703662, 1340, 703823, 1026, 377, 380, 379, 662, 703449, 258, 260, 451, 703699, 259, 703355, 703654, 703870, 395, 433, 1405, 703999, 1047, 1241, 752, 703375, 1319, 703712, 31, 703326, 703655, 703702, 703939, 531, 703802, 703897, 704203, 1389, 703953, 1053, 703600, 570, 703769, 704026, 703753, 703967, 704112, 984, 703807, 703529, 1073, 704154, 778, 548, 798, 704037, 316, 703669, 1429, 49, 704209, 722, 703619, 704142, 600, 825, 452, 969, 1020, 703358, 1095, 703487, 704092, 461, 765, 578, 703572, 982, 703462, 703583, 1106, 1044, 703855, 703348, 703872, 855, 1030, 1120, 703959, 703489, 913, 304, 303, 704101, 305, 1194, 703961, 1337, 993, 347, 74, 1074, 230, 293, 421, 434, 961, 44, 1185, 703477, 257, 661, 704146, 703982, 1442, 1441, 1440, 704211, 703812, 703423, 1317, 1264, 1278, 703313, 703544, 703749, 1347, 958, 951, 484, 703470, 542, 235, 486, 700, 348, 703867, 1154, 703965, 704239, 703993, 703542, 551, 550, 703526, 704085, 704079, 703485, 547, 371, 485, 596, 1052, 703804, 703906, 703596, 703937, 1288, 703954, 1183, 368, 704262, 703556, 902, 1422, 571, 278, 703754, 457, 722079_703569, 704081, 704103, 704019, 1119, 960, 703424, 704023, 704119, 1015, 1351, 1076, 1129, 704030, 704126, 956, 703819, 704036, 703690, 703648, 703709, 704207, 703728, 703826, 381, 703794, 703607, 703574, 762, 703743, 703632, 703913, 703889, 703700, 704246, 704054, 703606, 703473, 703346, 521, 703750, 703990, 703834, 703727, 586, 965, 498, 789, 728, 291, 703482, 703693, 703533, 73, 967, 475, 704067, 349, 784, 704274, 703506, 66, 233, 704144, 704167, 704168, 698, 643, 642, 1210, 435, 704070, 1077, 450, 42, 703614, 703973, 704110, 703701, 703628, 703948, 703955, 703665, 1367, 703899, 704202, 703641, 703711, 704173, 703856, 703842, 703651, 704186, 704269, 704074, 704108, 703821, 704134, 704029, 703601, 703979, 703777, 703741, 704041, 703778, 703670, 703650, 703646, 703846, 703945, 704021, 703909, 703908, 704031, 704271, 703756, 703618, 703724, 704060, 704120, 704017, 703921, 703816, 703988, 703573, 703997, 703674, 703673, 703591, 704002, 703733, 703735, 703986, 703940, 703757, 703751, 703883, 704139, 704157, 704265, 704279, 704289, 704280, 704024, 703911, 703912, 703605, 703797, 704022, 704039, 703471, 704095, 703663, 704138, 703604, 704247, 703707, 704008, 704221, 703799, 703576, 704009, 704069, 703840, 704080, 703998, 704012, 703603, 704130, 703786, 704131, 704078, 703630, 704066, 703983, 703507, 704001, 703645, 703919, 704196, 704197, 703616, 704201, 704133, 703796, 703857, 703859, 703863, 703858, 703860, 703780, 704166, 703642, 703640, 704093, 704259, 703680, 703878, 703901, 704020, 703884, 703958, 704091, 1376, 977, 1093, 703975, 704113, 704090, 704147, 703571, 703703, 703887, 703923, 704145, 704215, 704111, 704217, 704124, 703585, 704094, 703759, 703639, 703647, 703627, 704064, 704050, 704255, 704175, 703697, 1263, 703894, 703895, 703891, 704181, 704180, 704007, 703789, 704075, 703827, 703587, 704056, 704122, 703876, 703989, 704046, 703679, 703924, 704223, 704051, 703934, 1287, 704216, 704267, 704177, 703825, 704115, 703836, 703932, 703661, 704152, 703831, 704132, 703548, 704200, 703666, 703730, 703837, 1167, 703798, 703818, 704096, 704151, 703725, 704178, 704161, 704014, 703977, 704082, 704089, 703888, 703930, 703928, 703929, 703931, 703974, 703881, 704252, 703941, 703917, 703667, 703668, 704153, 703617, 704169, 704219, 704275, 703625, 703660, 704104, 704128, 704065, 1228, 704116, 704102, 703898, 703636, 703843, 704264, 703692, 703664, 703555, 704218, 704109, 1224, 1225, 704125, 703684, 703742, 703723, 703635, 704206, 703910, 703553, 704189, 703951, 704098, 703539, 703541, 703678, 1034, 703900, 703893, 703761, 756, 786, 390, 398, 703653, 1046, 1255, 665, 667, 317, 703715, 1253, 703414, 456, 941, 703505, 392, 544, 323, 703658, 704135, 704149, 1332, 703873, 704010, 77, 704057, 1357, 703459, 703944, 952, 676, 703720, 703892, 410, 75, 703985, 338, 906, 1349, 703579, 703810, 704303, 703694, 703874, 1188, 764, 759, 704015, 703675, 703970, 703877, 704117, 703956, 704084, 704213, 467, 276, 670, 1384, 704188, 703717, 70 3922, 796, 704016, 703729, 704243, 695, 616, 1139, 775, 703372, 703440, 703502, 582, 494, 703377, 405, 562, 24, 703952, 704018, 703957, 704033, 704032, 703695, 703475, 703947, 704063, 703890, 703589, 704204, 703536, 703849, 703659, 704025, 703926, 703978, 703991, 704260, 703987, 704250, 704195, 703672, 703468, 703467, 703652, 703714, 703481, 419, 724, 1057, 771, 1339, 1176, 1160, 703490, 703631, 703549, 703443, 1025, 1336, 703339, 1161, 703352, 1131, 1091, 1390, 1178, 703649, 703531, 703762, 703763, 703534, 703705, 703706, 703704, 703554, 703444, 703353, 1058, 703787, 1075, 703431, 827, 1024, 703410, 844, 1094, 1069, 703522, 703354, 703590, 1180, 1067, 723, 1159, 703795, 703570, 703454, 703721, 703722, 1173, 703422, 703518, 1404, 1437, 1092, 703484, 703325, 568, 1207, 703865, 703403, 1068, 1099, 703519, 1126, 703916, 703907, 703451, 1071, 1307, 612, 703373, 277, 917, 225, 256, 918, 640, 864, 288, 794, 692, 703363, 980, 703452, 463, 541, 893, 1145, 373, 929, 609. 2. Model Number: 722224. UDI Numbers: (01)00884838099258(21)750, (01)00884838099258(21)409, (01)00884838099258(21)2744, (01)00884838099258(21)3526, (01)00884838099258(21)1484, (01)00884838099258(21)1083, (01)00884838099258(21)3514, (01)00884838099258(21)2241, (01)00884838099258(21)962, (01)00884838099258(21)288, (01)00884838099258(21)2520, (01)00884838099258(21)91, (01)00884838099258(21)1692, (01)00884838099258(21)2421, (01)00884838099258(21)2432, (01)00884838099258(21)287, (01)00884838099258(21)904, (01)00884838099258(21)3211, (01)00884838099258(21)1874, (01)00884838099258(21)901, (01)00884838099258(21)3067, (01)00884838099258(21)514, (01)00884838099258(21)553, (01)00884838099258(21)2390, (01)00884838099258(21)618, (01)00884838099258(21)2879, (01)00884838099258(21)3642, (01)00884838099258(21)859, (01)00884838099258(21)611, (01)00884838099258(21)2759, (01)00884838099258(21)3490, (01)00884838099258(21)3582, (01)00884838099258(21)1706, (01)00884838099258(21)566, (01)00884838099258(21)565, (01)00884838099258(21)2559, (01)00884838099258(21)2993, (01)00884838099258(21)3020, (01)00884838099258(21)224, (01)00884838099258(21)153, (01)00884838099258(21)656, (01)00884838099258(21)667, (01)00884838099258(21)650, (01)00884838099258(21)609, (01)00884838099258(21)395, (01)00884838099258(21)2723, (01)00884838099258(21)380, (01)00884838099258(21)2911, (01)00884838099258(21)1318, (01)00884838099258(21)1039, (01)00884838099258(21)2649, (01)00884838099258(21)1679, (01)00884838099258(21)3046, (01)00884838099258(21)3048, (01)00884838099258(21)850, (01)00884838099258(21)1117, (01)00884838099258(21)1082, (01)00884838099258(21)273, (01)00884838099258(21)306, (01)00884838099258(21)333, (01)00884838099258(21)3453, (01)00884838099258(21)3283, (01)00884838099258(21)1488, (01)00884838099258(21)966, (01)00884838099258(21)2594, (01)00884838099258(21)1940, (01)00884838099258(21)2656, (01)00884838099258(21)867, (01)00884838099258(21)244, (01)00884838099258(21)2793, (01)00884838099258(21)1975, (01)00884838099258(21)2901, (01)00884838099258(21)105, (01)00884838099258(21)564, (01)00884838099258(21)856, (01)00884838099258(21)1047, (01)00884838099258(21)3373, (01)00884838099258(21)963, (01)00884838099258(21)965, (01)00884838099258(21)494, (01)00884838099258(21)151, (01)00884838099258(21)657, (01)00884838099258(21)3145, (01)00884838099258(21)215, (01)00884838099258(21)516, (01)00884838099258(21)478, (01)00884838099258(21)770, (01)00884838099258(21)3525, (01)00884838099258(21)1175, (01)00884838099258(21)3486, (01)00884838099258(21)3603, (01)00884838099258(21)339, (01)00884838099258(21)103, (01)00884838099258(21)3794, (01)00884838099258(21)3341, (01)00884838099258(21)1606, (01)00884838099258(21)2549, (01)00884838099258(21)2435, (01)00884838099258(21)587, (01)00884838099258(21)112, (01)00884838099258(21)2548, (01)00884838099258(21)507, (01)00884838099258(21)2377, (01)00884838099258(21)730, (01)00884838099258(21)2572, (01)00884838099258(21)492, (01)00884838099258(21)2226, (01)00884838099258(21)1935, (01)00884838099258(21)662, (01)00884838099258(21)78, (01)00884838099258(21)3224, (01)00884838099258(21)1782, (01)00884838099258(21)2452, (01)00884838099258(21)252, (01)00884838099258(21)543, (01)00884838099258(21)351, (01)00884838099258(21)2453, (01)00884838099258(21)300, (01)00884838099258(21)2279, (01)00884838099258(21)2220, (01)00884838099258(21)1526, (01)00884838099258(21)2869, (01)00884838099258(21)1041, (01)00884838099258(21)839, (01)00884838099258(21)3258, (01)00884838099258(21)999, (01)00884838099258(21)982, (01)00884838099258(21)462, (01)00884838099258(21)99, (01)00884838099258(21)3516, (01)00884838099258(21)2655, (01)00884838099258(21)188, (01)00884838099258(21)2369, (01)00884838099258(21)2301, (01)00884838099258(21)2130, (01)00884838099258(21)165, (01)00884838099258(21)1551, (01)00884838099258(21)627, (01)00884838099258(21)1383, (01)00884838099258(21)3026, (01)00884838099258(21)2965, (01)00884838099258(21)2967, (01)00884838099258(21)1999, (01)00884838099258(21)2000, (01)00884838099258(21)28, (01)00884838099258(21)2199, (01)00884838099258(21)746, (01)00884838099258(21)2340, (01)00884838099258(21)3270, (01)00884838099258(21)169, (01)00884838099258(21)3053, (01)00884838099258(21)53, (01)00884838099258(21)332, (01)00884838099258(21)517, (01)00884838099258(21)2427, (01)00884838099258(21)845, (01)00884838099258(21)3579, (01)00884838099258(21)1499, (01)00884838099258(21)3205, (01)00884838099258(21)404, (01)00884838099258(21)2957, (01)00884838099258(21)955, (01)00884838099258(21)253, (01)00884838099258(21)1243, (01)00884838099258(21)1161, (01)00884838099258(21)2367, (01)00884838099258(21)1493, (01)00884838099258(21)64, (01)00884838099258(21)3527, (01)00884838099258(21)3454, (01)00884838099258(21)844, (01)00884838099258(21)3189, (01)00884838099258(21)979, (01)00884838099258(21)787, (01)00884838099258(21)441, (01)00884838099258(21)84, (01)00884838099258(21)949, (01)00884838099258(21)350, (01)00884838099258(21)3673, (01)00884838099258(21)2557, (01)00884838099258(21)581, (01)00884838099258(21)185, (01)00884838099258(21)3401, (01)00884838099258(21)643, (01)00884838099258(21)2733, (01)00884838099258(21)1527, (01)00884838099258(21)503, (01)00884838099258(21)1335, (01)00884838099258(21)1156, (01)00884838099258(21)874, (01)00884838099258(21)875, (01)00884838099258(21)3596, (01)00884838099258(21)495, (01)00884838099258(21)2667, (01)00884838099258(21)3459, (01)00884838099258(21)1622, (01)00884838099258(21)1089, (01)00884838099258(21)352, (01)00884838099258(21)353, (01)00884838099258(21)257, (01)00884838099258(21)1568, (01)00884838099258(21)2852, (01)00884838099258(21)2931, (01)00884838099258(21)2742, (01)00884838099258(21)3474, (01)00884838099258(21)1988, (01)00884838099258(21)1974, (01)00884838099258(21)1851, (01)00884838099258(21)2804, (01)00884838099258(21)2933, (01)00884838099258(21)2777, (01)00884838099258(21)1709, (01)00884838099258(21)1705, (01)00884838099258(21)1942, (01)00884838099258(21)1802, (01)00884838099258(21)3696, (01)00884838099258(21)2772, (01)00884838099258(21)2089, (01)00884838099258(21)1686, (01)00884838099258(21)63, (01)00884838099258(21)1670, (01)00884838099258(21)2186, (01)00884838099258(21)2546, (01)00884838099258(21)3792, (01)00884838099258(21)2513, (01)00884838099258(21)3105, (01)00884838099258(21)942, (01)00884838099258(21)941, (01)00884838099258(21)1609, (01)00884838099258(21)937, (01)00884838099258(21)2244, (01)00884838099258(21)2373, (01)00884838099258(21)3099, (01)00884838099258(21)3264, (01)00884838099258(21)801, (01)00884838099258(21)1330, (01)00884838099258(21)1349, (01)00884838099258(21)1350, (01)00884838099258(21)1369, (01)00884838099258(21)3473, (01)00884838099258(21)1326, (01)00884838099258(21)2245, (01)00884838099258(21)1970, (01)00884838099258(21)776, (01)00884838099258(21)3117, (01)00884838099258(21)799, (01)00884838099258(21)800, (01)00884838099258(21)1747, (01)00884838099258(21)1385, (01)00884838099258(21)2405, (01)00884838099258(21)1812, (01)00884838099258(21)1204, (01)00884838099258(21)1529, (01)00884838099258(21)2060, (01)00884838099258(21)3334, (01)00884838099258(21)2583, (01)00884838099258(21)2479, (01)00884838099258(21)3502, (01)00884838099258(21)2555, (01)00884838099258(21)1005, (01)00884838099258(21)2044, (01)00884838099258(21)3522, (01)00884838099258(21)2025, (01)00884838099258(21)1151, (01)00884838099258(21)1164, (01)00884838099258(21)2561, (01)00884838099258(21)1027, (01)00884838099258(21)2509, (01)00884838099258(21)2588, (01)00884838099258(21)3347, (01)00884838099258(21)1449, (01)00884838099258(21)765, (01)00884838099258(21)501, (01)00884838099258(21)3428, (01)00884838099258(21)546, (01)00884838099258(21)1475, (01)00884838099258(21)2586, (01)00884838099258(21)1470, (01)00884838099258(21)2496, (01)00884838099258(21)778, (01)00884838099258(21)2464, (01)00884838099258(21)2585, (01)00884838099258(21)2959, (01)00884838099258(21)3597, (01)00884838099258(21)3255, (01)00884838099258(21)828, (01)00884838099258(21)1502, (01)00884838099258(21)2515, (01)00884838099258(21)3631, (01)00884838099258(21)1044, (01)00884838099258(21)1344, (01)00884838099258(21)1619, (01)00884838099258(21)2238, (01)00884838099258(21)720, (01)00884838099258(21)2630, (01)00884838099258(21)1581, (01)00884838099258(21)1138, (01)00884838099258(21)2732, (01)00884838099258(21)1424, (01)00884838099258(21)3653, (01)00884838099258(21)1949, (01)00884838099258(21)675, (01)00884838099258(21)2696, (01)00884838099258(21)3005, (01)00884838099258(21)2307, (01)00884838099258(21)548, (01)00884838099258(21)3546, (01)00884838099258(21)1194, (01)00884838099258(21)1129, (01)00884838099258(21)1067, (01)00884838099258(21)1412, (01)00884838099258(21)3629, (01)00884838099258(21)2467, (01)00884838099258(21)1104, (01)00884838099258(21)499, (01)00884838099258(21)530, (01)00884838099258(21)2560, (01)00884838099258(21)1002, (01)00884838099258(21)1704, (01)00884838099258(21)3479, (01)00884838099258(21)2592, (01)00884838099258(21)3183, (01)00884838099258(21)3641, (01)00884838099258(21)1368, (01)00884838099258(21)3227, (01)00884838099258(21)1522, (01)00884838099258(21)3660, (01)00884838099258(21)2320, (01)00884838099258(21)887, (01)00884838099258(21)687, (01)00884838099258(21)2709, (01)00884838099258(21)1474, (01)00884838099258(21)532, (01)00884838099258(21)596, (01)00884838099258(21)1306, (01)00884838099258(21)1647, (01)00884838099258(21)1153, (01)00884838099258(21)1991, (01)00884838099258(21)2866, (01)00884838099258(21)2923, (01)00884838099258(21)1280, (01)00884838099258(21)2864, (01)00884838099258(21)2865, (01)00884838099258(21)3311, (01)00884838099258(21)1281, (01)00884838099258(21)2566, (01)00884838099258(21)3752, (01)00884838099258(21)62, (01)00884838099258(21)37, (01)00884838099258(21)3598, (01)00884838099258(21)2947, (01)00884838099258(21)3665, (01)00884838099258(21)3411, (01)00884838099258(21)3666, (01)00884838099258(21)3753, (01)00884838099258(21)3383, (01)00884838099258(21)2745, (01)00884838099258(21)2738, (01)00884838099258(21)2527, (01)00884838099258(21)686, (01)00884838099258(21)3225, (01)00884838099258(21)2715, (01)00884838099258(21)3724, (01)00884838099258(21)3754, (01)00884838099258(21)3729, (01)00884838099258(21)1711, (01)00884838099258(21)3570, (01)00884838099258(21)707, (01)00884838099258(21)706, (01)00884838099258(21)708, (01)00884838099258(21)575, (01)00884838099258(21)1785, (01)00884838099258(21)1862, (01)00884838099258(21)1549, (01)00884838099258(21)3216, (01)00884838099258(21)2943, (01)00884838099258(21)1427, (01)00884838099258(21)1406, (01)00884838099258(21)2310, (01)00884838099258(21)171, (01)00884838099258(21)413, (01)00884838099258(21)2347, (01)00884838099258(21)827, (01)00884838099258(21)48, (01)00884838099258(21)1051, (01)00884838099258(21)1052, (01)00884838099258(21)1053, (01)00884838099258(21)658, (01)00884838099258(21)1050, (01)00884838099258(21)1348, (01)00884838099258(21)2277, (01)00884838099258(21)1054, (01)00884838099258(21)1080, (01)00884838099258(21)2751, (01)00884838099258(21)2605, (01)00884838099258(21)1092, (01)00884838099258(21)2987, (01)00884838099258(21)145, (01)00884838099258(21)2534, (01)00884838099258(21)1799, (01)00884838099258(21)35, (01)00884838099258(21)2692, (01)00884838099258(21)2953, (01)00884838099258(21)3236, (01)00884838099258(21)2998, (01)00884838099258(21)3634, (01)00884838099258(21)1840, (01)00884838099258(21)829, (01)00884838099258(21)3111, (01)00884838099258(21)1289, (01)00884838099258(21)264, (01)00884838099258(21)1919, (01)00884838099258(21)3701, (01)00884838099258(21)1667, (01)00884838099258(21)2431, (01)00884838099258(21)3318, (01)00884838099258(21)2713, (01)00884838099258(21)3191, (01)00884838099258(21)3648, (01)00884838099258(21)3197, (01)00884838099258(21)768, (01)00884838099258(21)449, (01)00884838099258(21)448, (01)00884838099258(21)842, (01)00884838099258(21)520, (01)00884838099258(21)1875, (01)00884838099258(21)3533, (01)00884838099258(21)1463, (01)00884838099258(21)3535, (01)00884838099258(21)2097, (01)00884838099258(21)1033, (01)00884838099258(21)1800, (01)00884838099258(21)3259, (01)00884838099258(21)358, (01)00884838099258(21)2224, (01)00884838099258(21)3118, (01)00884838099258(21)489, (01)00884838099258(21)2760, (01)00884838099258(21)1742, (01)00884838099258(21)2476, (01)00884838099258(21)752, (01)00884838099258(21)2429, (01)00884838099258(21)1114, (01)00884838099258(21)1119, (01)00884838099258(21)3232, (01)00884838099258(21)3252, (01)00884838099258(21)3498, (01)00884838099258(21)635, (01)00884838099258(21)2538, (01)00884838099258(21)2157, (01)00884838099258(21)551, (01)00884838099258(21)1365, (01)00884838099258(21)1366, (01)00884838099258(21)3319, (01)00884838099258(21)3799, (01)00884838099258(21)1272, (01)00884838099258(21)460, (01)00884838099258(21)2781, (01)00884838099258(21)3412, (01)00884838099258(21)990, (01)00884838099258(21)355, (01)00884838099258(21)1446, (01)00884838099258(21)716, (01)00884838099258(21)1498, (01)00884838099258(21)2643, (01)00884838099258(21)410, (01)00884838099258(21)459, (01)00884838099258(21)2444, (01)00884838099258(21)1076, (01)00884838099258(21)2286, (01)00884838099258(21)2360, (01)00884838099258(21)2674, (01)00884838099258(21)2485, (01)00884838099258(21)3655, (01)00884838099258(21)2780, (01)00884838099258(21)412, (01)00884838099258(21)2191, (01)00884838099258(21)508, (01)00884838099258(21)804, (01)00884838099258(21)3481, (01)00884838099258(21)2936, (01)00884838099258(21)2490, (01)00884838099258(21)2489, (01)00884838099258(21)3427, (01)00884838099258(21)1040, (01)00884838099258(21)2272, (01)00884838099258(21)416, (01)00884838099258(21)2460, (01)00884838099258(21)1676, (01)00884838099258(21)2845, (01)00884838099258(21)1205, (01)00884838099258(21)1355, (01)00884838099258(21)1761, (01)00884838099258(21)1360, (01)00884838099258(21)1180, (01)00884838099258(21)32, (01)00884838099258(21)3234, (01)00884838099258(21)85, (01)00884838099258(21)985, (01)00884838099258(21)2446, (01)00884838099258(21)156, (01)00884838099258(21)1809, (01)00884838099258(21)347, (01)00884838099258(21)2678, (01)00884838099258(21)1420, (01)00884838099258(21)3790, (01)00884838099258(21)1434, (01)00884838099258(21)2312, (01)00884838099258(21)2402, (01)00884838099258(21)1673, (01)00884838099258(21)2357, (01)00884838099258(21)1665, (01)00884838099258(21)1824, (01)00884838099258(21)1655, (01)00884838099258(21)3263, (01)00884838099258(21)3317, (01)00884838099258(21)2887, (01)00884838099258(21)3586, (01)00884838099258(21)3718, (01)00884838099258(21)1927, (01)00884838099258(21)1553, (01)00884838099258(21)3109, (01)00884838099258(21)2380, (01)00884838099258(21)1413, (01)00884838099258(21)1127, (01)00884838099258(21)2123, (01)00884838099258(21)1732, (01)00884838099258(21)1448, (01)00884838099258(21)2791, (01)00884838099258(21)1788, (01)00884838099258(21)2632, (01)00884838099258(21)1500, (01)00884838099258(21)3583, (01)00884838099258(21)2956, (01)00884838099258(21)3226, (01)00884838099258(21)1513, (01)00884838099258(21)1198, (01)00884838099258(21)2660, (01)00884838099258(21)3517, (01)00884838099258(21)1819, (01)00884838099258(21)1440, (01)00884838099258(21)1285, (01)00884838099258(21)1178, (01)00884838099258(21)1607, (01)00884838099258(21)1418, (01)00884838099258(21)2471, (01)00884838099258(21)1631, (01)00884838099258(21)2115, (01)00884838099258(21)2666, (01)00884838099258(21)1599, (01)00884838099258(21)2940, (01)00884838099258(21)2941, (01)00884838099258(21)1489, (01)00884838099258(21)1497, (01)00884838099258(21)1971, (01)00884838099258(21)1990, (01)00884838099258(21)3002, (01)00884838099258(21)2796, (01)00884838099258(21)2024, (01)00884838099258(21)3056, (01)00884838099258(21)2225, (01)00884838099258(21)2702, (01)00884838099258(21)1674, (01)00884838099258(21)2408, (01)00884838099258(21)2822, (01)00884838099258(21)1941, (01)00884838099258(21)2607, (01)00884838099258(21)1825, (01)00884838099258(21)2387, (01)00884838099258(21)2388, (01)00884838099258(21)3552, (01)00884838099258(21)1520, (01)00884838099258(21)1696, (01)00884838099258(21)2216, (01)00884838099258(21)2894, (01)00884838099258(21)2020, (01)00884838099258(21)2019, (01)00884838099258(21)3103, (01)00884838099258(21)1729, (01)00884838099258(21)3765, (01)00884838099258(21)3599, (01)00884838099258(21)3431, (01)00884838099258(21)3080, (01)00884838099258(21)2821, (01)00884838099258(21)2116, (01)00884838099258(21)1979, (01)00884838099258(21)2491, (01)00884838099258(21)1202, (01)00884838099258(21)2372, (01)00884838099258(21)3041, (01)00884838099258(21)1754, (01)00884838099258(21)2333, (01)00884838099258(21)2040, (01)00884838099258(21)1658, (01)00884838099258(21)1152, (01)00884838099258(21)3611, (01)00884838099258(21)2726, (01)00884838099258(21)3421, (01)00884838099258(21)3254, (01)00884838099258(21)3434, (01)00884838099258(21)2808, (01)00884838099258(21)2762, (01)00884838099258(21)2807, (01)00884838099258(21)2761, (01)00884838099258(21)2763, (01)00884838099258(21)1292, (01)00884838099258(21)1384, (01)00884838099258(21)2173, (01)00884838099258(21)1822, (01)00884838099258(21)1641, (01)00884838099258(21)1291, (01)00884838099258(21)1442, (01)00884838099258(21)2697, (01)00884838099258(21)1532, (01)00884838099258(21)2695, (01)00884838099258(21)2118, (01)00884838099258(21)2290, (01)00884838099258(21)1749, (01)00884838099258(21)2109, (01)00884838099258(21)2639, (01)00884838099258(21)2766, (01)00884838099258(21)3773, (01)00884838099258(21)2249, (01)00884838099258(21)2890, (01)00884838099258(21)2469, (01)00884838099258(21)2067, (01)00884838099258(21)2065, (01)00884838099258(21)1752, (01)00884838099258(21)1755, (01)00884838099258(21)2826, (01)00884838099258(21)2382, (01)00884838099258(21)3798, (01)00884838099258(21)3689, (01)00884838099258(21)1996, (01)00884838099258(21)1848, (01)00884838099258(21)1844, (01)00884838099258(21)1989, (01)00884838099258(21)1841, (01)00884838099258(21)2059, (01)00884838099258(21)3731, (01)00884838099258(21)772, (01)00884838099258(21)773, (01)00884838099258(21)1950, (01)00884838099258(21)2273, (01)00884838099258(21)2295, (01)00884838099258(21)3703, (01)00884838099258(21)2023, (01)00884838099258(21)2053, (01)00884838099258(21)1166, (01)00884838099258(21)1070, (01)00884838099258(21)3743, (01)00884838099258(21)2200, (01)00884838099258(21)3783, (01)00884838099258(21)1849, (01)00884838099258(21)2428, (01)00884838099258(21)1503, (01)00884838099258(21)1847, (01)00884838099258(21)1846, (01)00884838099258(21)3024, (01)00884838099258(21)3585, (01)00884838099258(21)2313, (01)00884838099258(21)2992, (01)00884838099258(21)1746, (01)00884838099258(21)3566, (01)00884838099258(21)2470, (01)00884838099258(21)1983, (01)00884838099258(21)2379, (01)00884838099258(21)2934, (01)00884838099258(21)2066, (01)00884838099258(21)2543, (01)00884838099258(21)2292, (01)00884838099258(21)3402, (01)00884838099258(21)1675, (01)00884838099258(21)2256, (01)00884838099258(21)2158, (01)00884838099258(21)2798, (01)00884838099258(21)1135, (01)00884838099258(21)1293, (01)00884838099258(21)2117, (01)00884838099258(21)2824, (01)00884838099258(21)2701, (01)00884838099258(21)3702, (01)00884838099258(21)2815, (01)00884838099258(21)2794, (01)00884838099258(21)1744, (01)00884838099258(21)2461, (01)00884838099258(21)2110, (01)00884838099258(21)1818, (01)00884838099258(21)2041, (01)00884838099258(21)2393, (01)00884838099258(21)2090, (01)00884838099258(21)2463, (01)00884838099258(21)2468, (01)00884838099258(21)2898, (01)00884838099258(21)2966, (01)00884838099258(21)1437, (01)00884838099258(21)2147, (01)00884838099258(21)2136, (01)00884838099258(21)3116, (01)00884838099258(21)1858, (01)00884838099258(21)2551, (01)00884838099258(21)1423, (01)00884838099258(21)1770, (01)00884838099258(21)1944, (01)00884838099258(21)1361, (01)00884838099258(21)3630, (01)00884838099258(21)3260, (01)00884838099258(21)3704, (01)00884838099258(21)3706, (01)00884838099258(21)3705, (01)00884838099258(21)1417, (01)00884838099258(21)2179, (01)00884838099258(21)1441, (01)00884838099258(21)1748, (01)00884838099258(21)1987, (01)00884838099258(21)2922, (01)00884838099258(21)1820, (01)00884838099258(21)2392, (01)00884838099258(21)1531, (01)00884838099258(21)3059, (01)00884838099258(21)1464, (01)00884838099258(21)1867, (01)00884838099258(21)1191, (01)00884838099258(21)1920, (01)00884838099258(21)1695, (01)00884838099258(21)3220, (01)00884838099258(21)2033, (01)00884838099258(21)2108, (01)00884838099258(21)2206, (01)00884838099258(21)3182, (01)00884838099258(21)1616, (01)00884838099258(21)1414, (01)00884838099258(21)1451, (01)00884838099258(21)2895, (01)00884838099258(21)2827, (01)00884838099258(21)1769, (01)00884838099258(21)1778, (01)00884838099258(21)2851, (01)00884838099258(21)2964, (01)00884838099258(21)1137, (01)00884838099258(21)3511, (01)00884838099258(21)3445, (01)00884838099258(21)3567, (01)00884838099258(21)3782, (01)00884838099258(21)3796, (01)00884838099258(21)3797, (01)00884838099258(21)3826, (01)00884838099258(21)3837, (01)00884838099258(21)3822, (01)00884838099258(21)3785, (01)00884838099258(21)3827, (01)00884838099258(21)690, (01)00884838099258(21)660, (01)00884838099258(21)3440, (01)00884838099258(21)1521, (01)00884838099258(21)2684, (01)00884838099258(21)68, (01)00884838099258(21)1016, (01)00884838099258(21)938, (01)00884838099258(21)529, (01)00884838099258(21)3076, (01)00884838099258(21)2848, (01)00884838099258(21)528, (01)00884838099258(21)293, (01)00884838099258(21)2361, (01)00884838099258(21)3381, (01)00884838099258(21)1261, (01)00884838099258(21)2638, (01)00884838099258(21)221, (01)00884838099258(21)1034, (01)00884838099258(21)1671, (01)00884838099258(21)1834, (01)00884838099258(21)3010, (01)00884838099258(21)1868, (01)00884838099258(21)3013, (01)00884838099258(21)3014, (01)00884838099258(21)3266, (01)00884838099258(21)3261, (01)00884838099258(21)1733, (01)00884838099258(21)2814, (01)00884838099258(21)375, (01)00884838099258(21)1617, (01)00884838099258(21)3595, (01)00884838099258(21)2163, (01)00884838099258(21)1530, (01)00884838099258(21)1626, (01)00884838099258(21)2048, (01)00884838099258(21)3339, (01)00884838099258(21)978, (01)00884838099258(21)1401, (01)00884838099258(21)97, (01)00884838099258(21)3574, (01)00884838099258(21)203, (01)00884838099258(21)2344, (01)00884838099258(21)3237, (01)00884838099258(21)2756, (01)00884838099258(21)1284, (01)00884838099258(21)2309, (01)00884838099258(21)3012, (01)00884838099258(21)3321, (01)00884838099258(21)2670, (01)00884838099258(21)393, (01)00884838099258(21)2252, (01)00884838099258(21)2625, (01)00884838099258(21)2716, (01)00884838099258(21)2137, (01)00884838099258(21)1148, (01)00884838099258(21)2568, (01)00884838099258(21)2482, (01)00884838099258(21)2944, (01)00884838099258(21)1248, (01)00884838099258(21)2161, (01)00884838099258(21)1528, (01)00884838099258(21)2455, (01)00884838099258(21)1501, (01)00884838099258(21)405, (01)00884838099258(21)3394, (01)00884838099258(21)1290, (01)00884838099258(21)2422, (01)00884838099258(21)781, (01)00884838099258(21)3368, (01)00884838099258(21)1683, (01)00884838099258(21)1741, (01)00884838099258(21)2757, (01)00884838099258(21)1823, (01)00884838099258(21)2008, (01)00884838099258(21)3739, (01)00884838099258(21)1373, (01)00884838099258(21)1480, (01)00884838099258(21)3749, (01)00884838099258(21)688, (01)00884838099258(21)1227, (01)00884838099258(21)1022, (01)00884838099258(21)697, (01)00884838099258(21)2006, (01)00884838099258(21)699, (01)00884838099258(21)1428, (01)00884838099258(21)3778, (01)00884838099258(21)1242, (01)00884838099258(21)2517, (01)00884838099258(21)743, (01)00884838099258(21)3774, (01)00884838099258(21)2909, (01)00884838099258(21)1207, (01)00884838099258(21)3396, (01)00884838099258(21)705, (01)00884838099258(21)1021, (01)00884838099258(21)1176, (01)00884838099258(21)1465, (01)00884838099258(21)759, (01)00884838099258(21)1552, (01)00884838099258(21)1496, (01)00884838099258(21)1792, (01)00884838099258(21)168, (01)00884838099258(21)3679, (01)00884838099258(21)197, (01)00884838099258(21)83, (01)00884838099258(21)2973, (01)00884838099258(21)889, (01)00884838099258(21)212. Serial Numbers: 750, 409, 2744, 3526, 1484, 1083, 3514, 2241, 962, 288, 2520, 91, 1692, 2421, 2432, 287, 904, 3211, 1874, 901, 3067, 514, 553, 2390, 618, 2879, 3642, 859, 611, 2759, 3490, 3582, 1706, 566, 565, 2559, 2993, 3020, 224, 153, 656, 667, 650, 609, 395, 2723, 380, 2911, 1318, 1039, 2649, 1679, 3046, 3048, 850, 1117, 1082, 273, 306, 333, 3453, 3283, 1488, 966, 2594, 1940, 2656, 867, 244, 2793, 1975, 2901, 105, 564, 856, 1047, 3373, 963, 965, 494, 151, 657, 3145, 215, 516, 478, 770, 3525, 1175, 3486, 3603, 339, 103, 3794, 3341, 1606, 2549, 2435, 587, 112, 2548, 507, 2377, 730, 2572, 492, 2226, 1935, 662, 78, 3224, 1782, 2452, 252, 543, 351, 2453, 300, 2279, 2220, 1526, 2869, 1041, 839, 3258, 999, 982, 462, 99, 3516, 2655, 188, 2369, 2301, 2130, 165, 1551, 627, 1383, 3026, 2965, 2967, 1999, 2000, 28, 2199, 746, 2340, 3270, 169, 3053, 53, 332, 517, 2427, 845, 3579, 1499, 3205, 404, 2957, 955, 253, 1243, 1161, 2367, 1493, 64, 3527, 3454, 844, 3189, 979, 787, 441, 84, 949, 350, 3673, 2557, 581, 185, 3401, 643, 2733, 1527, 503, 1335, 1156, 874, 875, 3596, 495, 2667, 3459, 1622, 1089, 352, 353, 257, 1568, 2852, 2931, 2742, 3474, 1988, 1974, 1851, 2804, 2933, 2777, 1709, 1705, 1942, 1802, 3696, 2772, 2089, 1686, 63, 1670, 2186, 2546, 3792, 2513, 3105, 942, 941, 1609, 937, 2244, 2373, 3099, 3264, 801, 1330, 1349, 1350, 1369, 3473, 1326, 2245, 1970, 776, 3117, 799, 800, 1747, 1385, 2405, 1812, 1204, 1529, 2060, 3334, 2583, 2479, 3502, 2555, 1005, 2044, 3522, 2025, 1151, 1164, 2561, 1027, 2509, 2588, 3347, 1449, 765, 501, 3428, 546, 1475, 2586, 1470, 2496, 778, 2464, 2585, 2959, 3597, 3255, 828, 1502, 2515, 3631, 1044, 1344, 1619, 2238, 720, 2630, 1581, 1138, 2732, 1424, 3653, 1949, 675, 2696, 3005, 2307, 548, 3546, 1194, 1129, 1067, 1412, 3629, 2467, 1104, 499, 530, 2560, 1002, 1704, 3479, 2592, 3183, 3641, 1368, 3227, 1522, 3660, 2320, 887, 687, 2709, 1474, 532, 596, 1306, 1647, 1153, 1991, 2866, 2923, 1280, 2864, 2865, 3311, 1281, 2566, 3752, 62, 37, 3598, 2947, 3665, 3411, 3666, 3753, 3383, 2745, 2738, 2527, 686, 3225, 2715, 3724, 3754, 3729, 1711, 3570, 707, 706, 708, 575, 1785, 1862, 1549, 3216, 2943, 1427, 1406, 2310, 171, 413, 2347, 827, 48, 1051, 1052, 1053, 658, 1050, 1348, 2277, 1054, 1080, 2751, 2605, 1092, 2987, 145, 2534, 1799, 35, 2692, 2953, 3236, 2998, 3634, 1840, 829, 3111, 1289, 264, 1919, 3701, 1667, 2431, 3318, 2713, 3191, 3648, 3197, 768, 449, 448, 842, 520, 1875, 3533, 1463, 3535, 2097, 1033, 1800, 3259, 358, 2224, 3118, 489, 2760, 1742, 2476, 752, 2429, 1114, 1119, 3232, 3252, 3498, 635, 2538, 2157, 551, 1365, 1366, 3319, 3799, 1272, 460, 2781, 3412, 990, 355, 1446, 716, 1498, 2643, 410, 459, 2444, 1076, 2286, 2360, 2674, 2485, 3655, 2780, 412, 2191, 508, 804, 3481, 2936, 2490, 2489, 3427, 1040, 2272, 416, 2460, 1676, 2845, 1205, 1355, 1761, 1360, 1180, 32, 3234, 85, 985, 2446, 156, 1809, 347, 2678, 1420, 3790, 1434, 2312, 2402, 1673, 2357, 1665, 1824, 1655, 3263, 3317, 2887, 3586, 3718, 1927, 1553, 3109, 2380, 1413, 1127, 2123, 1732, 1448, 2791, 1788, 2632, 1500, 3583, 2956, 3226, 1513, 1198, 2660, 3517, 1819, 1440, 1285, 1178, 1607, 1418, 2471, 1631, 2115, 2666, 1599, 2940, 2941, 1489, 1497, 1971, 1990, 3002, 2796, 2024, 3056, 2225, 2702, 1674, 2408, 2822, 1941, 2607, 1825, 2387, 2388, 3552, 1520, 1696, 2216, 2894, 2020, 2019, 3103, 1729, 3765, 3599, 3431, 3080, 2821, 2116, 1979, 2491, 1202, 2372, 3041, 1754, 2333, 2040, 1658, 1152, 3611, 2726, 3421, 3254, 3434, 2808, 2762, 2807, 2761, 2763, 1292, 1384, 2173, 1822, 1641, 1291, 1442, 2697, 1532, 2695, 2118, 2290, 1749, 2109, 2639, 2766, 3773, 2249, 2890, 2469, 2067, 2065, 1752, 1755, 2826, 2382, 3798, 3689, 1996, 1848, 1844, 1989, 1841, 2059, 3731, 772, 773, 1950, 2273, 2295, 3703, 2023, 2053, 1166, 1070, 3743, 2200, 3783, 1849, 2428, 1503, 1847, 1846, 3024, 3585, 2313, 2992, 1746, 3566, 2470, 1983, 2379, 2934, 2066, 2543, 2292, 3402, 1675, 2256, 2158, 2798, 1135, 1293, 2117, 2824, 2701, 3702, 2815, 2794, 1744, 2461, 2110, 1818, 2041, 2393, 2090, 2463, 2468, 2898, 2966, 1437, 2147, 2136, 3116, 1858, 2551, 1423, 1770, 1944, 1361, 3630, 3260, 3704, 3706, 3705, 1417, 2179, 1441, 1748, 1987, 2922, 1820, 2392, 1531, 3059, 1464, 1867, 1191, 1920, 1695, 3220, 2033, 2108, 2206, 3182, 1616, 1414, 1451, 2895, 2827, 1769, 1778, 2851, 2964, 1137, 3511, 3445, 3567, 3782, 3796, 3797, 3826, 3837, 3822, 3785, 3827, 690, 660, 3440, 1521, 2684, 68, 1016, 938, 529, 3076, 2848, 528, 293, 2361, 3381, 1261, 2638, 221, 1034, 1671, 1834, 3010, 1868, 3013, 3014, 3266, 3261, 1733, 2814, 375, 1617, 3595, 2163, 1530, 1626, 2048, 3339, 978, 1401, 97, 3574, 203, 2344, 3237, 2756, 1284, 2309, 3012, 3321, 2670, 393, 2252, 2625, 2716, 2137, 1148, 2568, 2482, 2944, 1248, 2161, 1528, 2455, 1501, 405, 3394, 1290, 2422, 781, 3368, 1683, 1741, 2757, 1823, 2008, 3739, 1373, 1480, 3749, 688, 1227, 1022, 697, 2006, 699, 1428, 3778, 1242, 2517, 743, 3774, 2909, 1207, 3396, 705, 1021, 1176, 1465, 759, 1552, 1496, 1792, 168, 3679, 197, 83, 2973, 889, 212. 3. Model Number: 722282. UDI Numbers: (01)00884838110571(21)57, (01)00884838110571(21)54, (01)00884838110571(21)66, (01)00884838110571(21)49, (01)00884838110571(21)154, (01)00884838110571(21)55, (01)00884838110571(21)36, (01)00884838110571(21)133, (01)00884838110571(21)112, (01)00884838110571(21)45, (01)00884838110571(21)116, (01)00884838110571(21)168, (01)00884838110571(21)120, (01)00884838110571(21)118, (01)00884838110571(21)42, (01)00884838110571(21)117, (01)00884838110571(21)31, (01)00884838110571(21)178, (01)00884838110571(21)187, (01)00884838110571(21)26, (01)00884838110571(21)63, (01)00884838110571(21)195, (01)00884838110571(21)194, (01)00884838110571(21)73, (01)00884838110571(21)149, (01)00884838110571(21)99, (01)00884838110571(21)162, (01)00884838110571(21)189, (01)00884838110571(21)173, (01)00884838110571(21)143, (01)00884838110571(21)142, (01)00884838110571(21)130, (01)00884838110571(21)166, (01)00884838110571(21)156, (01)00884838110571(21)84, (01)00884838110571(21)114, (01)00884838110571(21)171, (01)00884838110571(21)44, (01)00884838110571(21)193, (01)00884838110571(21)144, (01)00884838110571(21)28, (01)00884838110571(21)61, (01)00884838110571(21)9, (01)00884838110571(21)165, (01)00884838110571(21)198, (01)00884838110571(21)172, (01)00884838110571(21)16, (01)00884838110571(21)128, (01)00884838110571(21)135, (01)00884838110571(21)136, (01)00884838110571(21)79, (01)00884838110571(21)163, (01)00884838110571(21)191, (01)00884838110571(21)169, (01)00884838110571(21)41, (01)00884838110571(21)109, (01)00884838110571(21)33, (01)00884838110571(21)132, (01)00884838110571(21)160, (01)00884838110571(21)175, (01)00884838110571(21)158, (01)00884838110571(21)11, (01)00884838110571(21)88, (01)00884838110571(21)155, (01)00884838110571(21)71, (01)00884838110571(21)94, (01)00884838110571(21)148, (01)00884838110571(21)92, (01)00884838110571(21)125, (01)00884838110571(21)24, (01)00884838110571(21)139, (01)00884838110571(21)38, (01)00884838110571(21)10, (01)00884838110571(21)75, (01)00884838110571(21)18, (01)00884838110571(21)102, (01)00884838110571(21)151, (01)00884838110571(21)86, (01)00884838110571(21)157, (01)00884838110571(21)161, (01)00884838110571(21)174, (01)00884838110571(21)62, (01)00884838110571(21)70, (01)00884838110571(21)150, (01)00884838110571(21)82, (01)00884838110571(21)52, (01)00884838110571(21)53, (01)00884838110571(21)199, (01)00884838110571(21)15, (01)00884838110571(21)100. Serial Numbers: 57, 54, 66, 49, 154, 55, 36, 133, 112, 45, 116, 168, 120, 118, 42, 117, 31, 178, 187, 26, 63, 195, 194, 73, 149, 99, 162, 189, 173, 143, 142, 130, 166, 156, 84, 114, 171, 44, 193, 144, 28, 61, 9, 165, 198, 172, 16, 128, 135, 136, 79, 163, 191, 169, 41, 109, 33, 132, 160, 175, 15 8, 11, 88, 155, 71, 94, 148, 92, 125, 24, 139, 38, 10, 75, 18, 102, 151, 86, 157, 161, 174, 62, 70, 150, 82, 52, 53, 199, 15, 100.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-0299-2027 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | October 6, 2026 |
| Reported | November 4, 2026 |
| Company | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Company location | Best, Netherlands |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1,896 units |
More recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · all medical device recalls · recalls from October 2026