AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product... recall (Steris Corporation)

Medical deviceClass IIOngoing

Summary: Steris Corporation is recalling AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensil. Reason: Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

What is being recalled

AMSCO 7053HP Washer/Disinfector Model/Catalog Number: 7053HP Product Description: The AMSCO 7053HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.

Why

Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia. (states named: Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, North Carolina, New Hampshire, New York, Pennsylvania, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin)

How to tell if yours is affected

Lot Code: UDI: 00724995176662, 00724995176679, 00724995176655 Serial numbers of distributed units: 3619725008, 3619725011, 3619725013, 3621625010, 3621625011, 3621625013, 3621625014, 3621625015, 3621925001, 3621925002, 3621925005, 3621925010, 3622425005, 3622725005, 3622725011, 3622725012, 3623125006, 3623425016, 3623825008, 3624025002, 3624025008, 3624525007

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0910-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedNovember 19, 2025
ReportedDecember 17, 2025
CompanySteris Corporation
Company locationMentor, OH, United States
Initiated byVoluntary: Firm initiated
Quantity22