Recall Ledger

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade... recall (Zimmer, Inc.)

Medical deviceClass IIOngoing

Summary: Zimmer, Inc. is recalling Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609320 Software Version: N/A Product Description: Affixus 9 mm, Length 320 mm, Left, Trochanteric Nail Component: N/A. Reason: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

What is being recalled

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815609320 Software Version: N/A Product Description: Affixus 9 mm, Length 320 mm, Left, Trochanteric Nail Component: N/A

Why

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT. (states named: Alabama, Arizona, California, Florida, Indiana, Kentucky, Massachusetts, Missouri, Montana, North Carolina, New Mexico, Rhode Island, Texas, Utah)

How to tell if yours is affected

Lot Code: Model No 815609320 ; UDI-DI (01)00887868144479(17)340826(10)66717601 ; Lot Number 66717601 Model No 815609320 ; UDI-DI (01)00887868144479(17)340820(10)66892900 ; Lot Number 66892900

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0975-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 2, 2025
ReportedDecember 31, 2025
CompanyZimmer, Inc.
Company locationWarsaw, IN, United States
Initiated byVoluntary: Firm initiated
Quantity23

More recalls from Zimmer, Inc. · all medical device recalls · recalls from December 2025

Official FDA record