Recall Ledger

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade... recall (Zimmer, Inc.)

Medical deviceClass IIOngoing

Summary: Zimmer, Inc. is recalling Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Right, Trochanteric Nail Component: N/A. Reason: Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

What is being recalled

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509300 Software Version: N/A Product Description: Affixus 9 mm, Length 300 mm, Right, Trochanteric Nail Component: N/A

Why

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT. (states named: Alabama, Arizona, California, Florida, Indiana, Kentucky, Massachusetts, Missouri, Montana, North Carolina, New Mexico, Rhode Island, Texas, Utah)

How to tell if yours is affected

Lot Code: Model No 815509300 ; UDI-DI (01)00887868143519(17)340823(10)66717550 ; Lot Number 66717550

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-0983-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 2, 2025
ReportedDecember 31, 2025
CompanyZimmer, Inc.
Company locationWarsaw, IN, United States
Initiated byVoluntary: Firm initiated
Quantity11

More recalls from Zimmer, Inc. · all medical device recalls · recalls from December 2025

Official FDA record