Recall Ledger

MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed recall (Medline Industries, LP)

Medical deviceClass IOngoing

Summary: Medline Industries, LP is recalling MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L. Reason: Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

What is being recalled

MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003L

Why

Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

US, CA, PA, GUAM, VI (states named: California, Pennsylvania)

How to tell if yours is affected

UDI/DI 40080196294344, All lots

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1015-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedNovember 26, 2025
ReportedJanuary 28, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity35694 units

More recalls from Medline Industries, LP · all medical device recalls · recalls from November 2025

Official FDA record