VANTAGE GALAN 3T Model MRT-3020/MEXL-3020 recall (Canon Medical System, USA, INC.)

Medical deviceClass IIOngoing

Summary: Canon Medical System, USA, INC. is recalling VANTAGE GALAN 3T Model MRT-3020/MEXL-3020. Reason: There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

What is being recalled

VANTAGE GALAN 3T Model MRT-3020/MEXL-3020

Why

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI. (states named: California, Colorado, Florida, Iowa, Illinois, Indiana, Kansas, Maryland, Missouri, Montana, Nevada, New York, Ohio, Oregon, Pennsylvania, Puerto Rico, Texas, Virginia, Wisconsin)

How to tell if yours is affected

Model Number: MRT-3020/MEXL-3020 UDI-DI code: 04987670102615 CMSC Model Number/Model Number/Serial Numbers/SID/UDI code: MRT-3020/4A MEXL-3020/5D 5DB2152002 30058153 (01)04987670102615(21)4AA2152001 MRT-3020/4A MEXL-3020/5D 5DC2262005 30075456 (01)04987670102615(21)4AB2262002 MRT-3020/4A MEXL-3020/5D 5DE2452007 30097122 (01)04987670102615(21)4AC2452003 MRT-3020/5A MEXL-3020/5D 5DB2172003 30047848 (01)04987670102615(21)5AA2172001 MRT-3020/6A MEXL-3020/7D 7DE2482014 30102813 (01)04987670102615(21)6AA2482002 MRT-3020/6A MEXL-3020/7D 7DE2472012 30108012 (01)04987670102615(21)6AA2472001 MRT-3020/7A MEXL-3020/7D 7DD2392008 30100563 (01)04987670102615(21)7AA2392001 MRT-3020/A4 MEXL-3020/D5 D5B2152013 30060527 (01)04987670102615(21)A4D2152004 MRT-3020/A4 MEXL-3020/D5 D5B2212019 30064817 (01)04987670102615(21)A4D2212006 MRT-3020/A4 MEXL-3020/D5 D5A2082005 30069306 (01)04987670102615(21)A4C2082002 MRT-3020/A4 MEXL-3020/D5 D5A2092007 30071480 (01)04987670102615(21)A4C2092003 MRT-3020/A4 MEXL-3020/D5 D5E24X2043 30072643 (01)04987670102615(21)A4F24X2010 MRT-3020/A4 MEXL-3020/D5 D5B2192016 30085151 (01)04987670102615(21)A4D2192005 MRT-3020/A4 MEXL-3020/D5 D5C2262022 30098384 (01)04987670102615(21)A4E2262008 MRT-3020/A4 MEXL-3020/D5 D5C2262021 30100803 (01)04987670102615(21)A4E2262007 MRT-3020/A4 MEXL-3020/D5 D5E2492042 30109383 (01)04987670102615(21)A4F2492009 MRT-3020/A5 MEXL-3020/S5 S5B18X2046 30007763 (01)04987670102615(21)A5B18X2011 MRT-3020/A5 MEXL-3020/S5 S5A17X2033 30031740 (01)04987670102615(21)A5A17X2008 MRT-3020/A5 MEXL-3020/S5 S5B17X2034 30034832 (01)04987670102615(21)A5B1882010 MRT-3020/A5 MEXL-3020/S5 S5A1732019 30036503 (01)04987670102615(21)A5A1732004 MRT-3020/A5 MEXL-3020/S5 S5A1742021 30040917 (01)04987670102615(21)A5A1742005 MRT-3020/A5 MEXL-3020/S5 S5C1972054 30041962 (01)04987670102615(21)A5C1972014 MRT-3020/A5 MEXL-3020/S5 S5A1732018 30042870 (01)04987670102615(21)A5A1732003 MRT-3020/A5 MEXL-3020/S5 S5B1842043 30043510 (01)04987670102615(21)A5B18X2012 MRT-3020/A5 MEXL-3020/D5 D5A2062003 30053897 (01)04987670102615(21)A5D2062015 MRT-3020/A5 MEXL-3020/S5 S5C18Y2048 30056987 (01)04987670102615(21)A5C18Y2013 MRT-3020/A5 MEXL-3020/D5 D5D23X2032 30087647 (01)04987670102615(21)A5G23X2017 MRT-3020/A5 MEXL-3020/D5 D5C2292024 30087649 (01)04987670102615(21)A5F2292016 MRT-3020/A5 MEXL-3020/S5.004 S5A1662001 30092401 (01)04987670102615(21)A5A1662001 MRT-3020/A6 MEXL-3020/D7 D7B2152015 30056208 (01)04987670102615(21)A6D2152001 MRT-3020/A7 MEXL-3020/S7 S7B1892004 30044596 (01)04987670102615(21)A7B1892001 MRT-3020/A7 MEXL-3020/D7 D7B2182017 30081608 (01)04987670102615(21)A7D2182004 MRT-3020/A7 MEXL-3020/S7.007 S7B1962013 30094084 (01)04987670102615(21)A7B1962002 MRT-3020/A7 MEXL-3020/S7.007 S7B1992017 30119571 (01)04987670102615(21)A7B1992003 MRT-3020/A7 MEXL-3020/S7 S7B1892004 30044596 MRT-3020/A7 MEXL-3020/D7 D7B2182017 30081608 MRT-3020/A7 MEXL-3020/S7.007 S7B1962013 30094084 MRT-3020/A7 MEXL-3020/S7.007 S7B1992017 30119571

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1017-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 19, 2025
ReportedJanuary 21, 2026
CompanyCanon Medical System, USA, INC.
Company locationTustin, CA, United States
Initiated byVoluntary: Firm initiated
Quantity34 systems