Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor. recall (Philips North America Llc)

Medical deviceClass IIOngoing

Summary: Philips North America Llc is recalling Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.. Reason: It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.

What is being recalled

Philips Smart-hopping 2.0 AP 1.4 GHz. Patient Monitor.

Why

It was found that the MX40 device could not reconnect to the PIC iX when moving between Radiohead and Trident 1.4 GHz access points if the signal strength changed quickly.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

Model No. 867216; UDI [Serial No.]: , (01)00884838099876(21)DK32800791[DK32800791], (01)00884838099876(21)DK32800825[DK32800825], (01)00884838099876(21)DK32800918[DK32800918], (01)00884838099876(21)DK32800947[DK32800947], (01)00884838099876(21)DK32801055[DK32801055], (01)00884838099876(21)DK32801076[DK32801076], (01)00884838099876(21)DK32801098[DK32801098], (01)00884838099876(21)DK32801120[DK32801120], (01)00884838099876(21)DK32801188[DK32801188], (01)00884838099876(21)DK32801197[DK32801197], (01)00884838099876(21)DK32801225[DK32801225], (01)00884838099876(21)DK32801235[DK32801235], (01)00884838099876(21)DK32801252[DK32801252], (01)00884838099876(21)DK32801306[DK32801306], (01)00884838099876(21)DK32801338[DK32801338], (01)00884838099876(21)DK33402127[DK33402127], (01)00884838099876(21)DK33402141[DK33402141], (01)00884838099876(21)DK33402210[DK33402210], (01)00884838099876(21)DK33402281[DK33402281], (01)00884838099876(21)DK33402309[DK33402309], (01)00884838099876(21)DK33402345[DK33402345], (01)00884838099876(21)DK33402358[DK33402358], (01)00884838099876(21)DK33402363[DK33402363], (01)00884838099876(21)DK33402364[DK33402364], (01)00884838099876(21)DK33402437[DK33402437], (01)00884838099876(21)DK33402526[DK33402526], (01)00884838099876(21)DK33402553[DK33402553], (01)00884838099876(21)DK33402569[DK33402569], (01)00884838099876(21)DK33402630[DK33402630], (01)00884838099876(21)DK33402653[DK33402653], (01)00884838099876(21)DK33402657[DK33402657], (01)00884838099876(21)DK33402737[DK33402737], (01)00884838099876(21)DK33402805[DK33402805], (01)00884838099876(21)DK33402894[DK33402894], (01)00884838099876(21)DK33402955[DK33402955], (01)00884838099876(21)DK33403058[DK33403058], (01)00884838099876(21)DK33403144[DK33403144], (01)00884838099876(21)DK33403145[DK33403145], (01)00884838099876(21)DK33403209[DK33403209], (01)00884838099876(21)DK33403222[DK33403222], (01)00884838099876(21)DK33403235[DK33403235], (01)00884838099876(21)DK33603488[DK33603488], (01)00884838099876(21)DK33603527[DK33603527], (01)00884838099876(21)DK33603582[DK33603582], (01)00884838099876(21)DK33603584[DK33603584], (01)00884838099876(21)DK33603673[DK33603673], (01)00884838099876(21)DK33603678[DK33603678], (01)00884838099876(21)DK33603746[DK33603746], (01)00884838099876(21)DK33603796[DK33603796], (01)00884838099876(21)DK33603971[DK33603971], (01)00884838099876(21)DK33603989[DK33603989], (01)00884838099876(21)DK33603999[DK33603999], (01)00884838099876(21)DK33604020[DK33604020], (01)00884838099876(21)DK33604046[DK33604046], (01)00884838099876(21)DK33604057[DK33604057], (01)00884838099876(21)DK33604109[DK33604109], (01)00884838099876(21)DK33604128[DK33604128], (01)00884838099876(21)DK33604162[DK33604162], (01)00884838099876(21)DK33604229[DK33604229], (01)00884838099876(21)DK33604235[DK33604235], (01)00884838099876(21)DK33604249[DK33604249], (01)00884838099876(21)DK33604252[DK33604252], (01)00884838099876(21)DK33604424[DK33604424], (01)00884838099876(21)DK33604425[DK33604425], (01)00884838099876(21)DK33604524[DK33604524], (01)00884838099876(21)DK33604571[DK33604571], (01)00884838099876(21)DK33604588[DK33604588], (01)00884838099876(21)DK33604612[DK33604612], (01)00884838099876(21)DK33904737[DK33904737], (01)00884838099876(21)DK33904802[DK33904802], (01)00884838099876(21)DK33904818[DK33904818], (01)00884838099876(21)DK33904830[DK33904830], (01)00884838099876(21)DK33904834[DK33904834], (01)00884838099876(21)DK33904949[DK33904949], (01)00884838099876(21)DK33904961[DK33904961], (01)00884838099876(21)DK33905053[DK33905053], (01)00884838099876(21)DK33905076[DK33905076], (01)00884838099876(21)DK33905149[DK33905149], (01)00884838099876(21)DK34305222[DK34305222], (01)00884838099876(21)DK34305227[DK34305227], (01)00884838099876(21)DK34305256[DK34305256], (01)00884838099876(21)DK34405461[DK34405461], (01)00884838099876(21)DK34405464[DK34405464], (01)00884838099876(21)DK34405673[DK34405673], (01)00884838099876(21)DK34405777[DK34405777], (01)00884838099876(21)DK34405778[DK34405778], (01)00884838099876(21)DK34405795[DK34405795], (01)00884838099876(21)DK34706009[DK34706009], (01)00884838099876(21)DK34706048[DK34706048], (01)00884838099876(21)DK34706100[DK34706100], (01)00884838099876(21)DK34706116[DK34706116], (01)00884838099876(21)DK34706181[DK34706181], (01)00884838099876(21)DK34706271[DK34706271], (01)00884838099876(21)DK34706308[DK34706308], (01)00884838099876(21)DK34706385[DK34706385], (01)00884838099876(21)DK34706394[DK34706394], (01)00884838099876(21)DK34706485[DK34706485], (01)00884838099876(21)DK40306620[DK40306620], (01)00884838099876(21)DK40306621[DK40306621], (01)00884838099876(21)DK40306622[DK40306622], (01)00884838099876(21)DK40306681[DK40306681], (01)00884838099876(21)DK40306690[DK40306690], (01)00884838099876(21)DK40306709[DK40306709], (01)00884838099876(21)DK40306749[DK40306749], (01)00884838099876(21)DK40306765[DK40306765], (01)00884838099876(21)DK40306766[DK40306766], (01)00884838099876(21)DK40306781[DK40306781], (01)00884838099876(21)DK40306796[DK40306796], (01)00884838099876(21)DK40306831[DK40306831], (01)00884838099876(21)DK40306838[DK40306838], (01)00884838099876(21)DK40306870[DK40306870], (01)00884838099876(21)DK40306898[DK40306898], (01)00884838099876(21)DK40306958[DK40306958], (01)00884838099876(21)DK40307021[DK40307021], (01)00884838099876(21)DK40307048[DK40307048], (01)00884838099876(21)DK40307129[DK40307129], (01)00884838099876(21)DK40707316[DK40707316], (01)00884838099876(21)DK40707339[DK40707339], (01)00884838099876(21)DK40707345[DK40707345], (01)00884838099876(21)DK40707357[DK40707357], (01)00884838099876(21)DK40707358[DK40707358], (01)00884838099876(21)DK40707371[DK40707371], (01)00884838099876(21)DK40707379[DK40707379], (01)00884838099876(21)DK40707386[DK40707386], (01)00884838099876(21)DK40707419[DK40707419], (01)00884838099876(21)DK40707604[DK40707604], (01)00884838099876(21)DK40707668[DK40707668], (01)00884838099876(21)DK40707670[DK40707670], (01)00884838099876(21)DK40707685[DK40707685], (01)00884838099876(21)DK40707703[DK40707703], (01)00884838099876(21)DK40707802[DK40707802], (01)00884838099876(21)DK40707892[DK40707892], (01)00884838099876(21)DK40707904[DK40707904], (01)00884838099876(21)DK40707917[DK40707917], (01)00884838099876(21)DK40907983[DK40907983], (01)00884838099876(21)DK40908048[DK40908048], (01)00884838099876(21)DK40908094[DK40908094], (01)00884838099876(21)DK40908104[DK40908104], (01)00884838099876(21)DK40908186[DK40908186], (01)00884838099876(21)DK40908203[DK40908203], (01)00884838099876(21)DK40908211[DK40908211], (01)00884838099876(21)DK40908212[DK40908212], (01)00884838099876(21)DK40908228[DK40908228], (01)00884838099876(21)DK40908234[DK40908234], (01)00884838099876(21)DK40908246[DK40908246], (01)00884838099876(21)DK40908253[DK40908253], (01)00884838099876(21)DK40908259[DK40908259], (01)00884838099876(21)DK40908260[DK40908260], (01)00884838099876(21)DK40908346[DK40908346], (01)00884838099876(21)DK40908356[DK40908356], (01)00884838099876(21)DK40908363[DK40908363], (01)00884838099876(21)DK40908364[DK40908364], (01)00884838099876(21)DK40908406[DK40908406], (01)00884838099876(21)DK40908432[DK40908432], (01)00884838099876(21)DK40908475[DK40908475], (01)00884838099876(21)DK40908486[DK40908486], (01)00884838099876(21)DK43608549[DK43608549], (01)00884838099876(21)DK43608580[DK43608580], (01)00884838099876(21)DK43608655[DK43608655], (01)00884838099876(21)DK43608703[DK43608703], (01)00884838099876(21)DK43608714[DK43608714], (01)00884838099876(21)DK43608770[DK43608770], (01)00884838099876(21)DK43608787[DK43608787], (01)00884838099876(21)DK43608789[DK43608789], (01)00884838099876(21)DK43608793[DK43608793], (01)00884838099876(21)DK43608797[DK43608797], (01)00884838099876(21)DK43608819[DK43608819], (01)00884838099876(21)DK43608903[DK43608903], (01)00884838099876(21)DK43608917[DK43608917], (01)00884838099876(21)DK43708977[DK43708977], (01)00884838099876(21)DK43708990[DK43708990], (01)00884838099876(21)DK43708991[DK43708991], (01)00884838099876(21)DK43708995[DK43708995], (01)00884838099876(21)DK43708997[DK43708997], (01)00884838099876(21)DK43709095[DK43709095], (01)00884838099876(21)DK43709121[DK43709121], (01)00884838099876(21)DK43709225[DK43709225], (01)00884838099876(21)DK43809353[DK43809353], (01)00884838099876(21)DK43809399[DK43809399], (01)00884838099876(21)DK43809411[DK43809411], (01)00884838099876(21)DK43809424[DK43809424], (01)00884838099876(21)DK43809436[DK43809436], (01)00884838099876(21)DK43809471[DK43809471], (01)00884838099876(21)DK43809563[DK43809563], (01)00884838099876(21)DK43809564[DK43809564], (01)00884838099876(21)DK43809618[DK43809618], (01)00884838099876(21)DK43809707[DK43809707], (01)00884838099876(21)DK43809713[DK43809713], (01)00884838099876(21)DK43809714[DK43809714], (01)00884838099876(21)DK43809723[DK43809723], (01)00884838099876(21)DK43809730[DK43809730], (01)00884838099876(21)DK43909810[DK43909810], (01)00884838099876(21)DK43909928[DK43909928], (01)00884838099876(21)DK43910123[DK43910123], (01)00884838099876(21)DK50310146[DK50310146], (01)00884838099876(21)DK50310151[DK50310151], (01)00884838099876(21)DK50310168[DK50310168], (01)00884838099876(21)DK50310194[DK50310194], (01)00884838099876(21)DK50310216[DK50310216], (01)00884838099876(21)DK50310233[DK50310233], (01)00884838099876(21)DK50310234[DK50310234], (01)00884838099876(21)DK50310246[DK50310246], (01)00884838099876(21)DK50310279[DK50310279], (01)00884838099876(21)DK50310280[DK50310280], (01)00884838099876(21)DK50310290[DK50310290], (01)00884838099876(21)DK50310297[DK50310297], (01)00884838099876(21)DK50310305[DK50310305], (01)00884838099876(21)DK50310358[DK50310358], (01)00884838099876(21)DK50310359[DK50310359], (01)00884838099876(21)DK50310363[DK50310363], (01)00884838099876(21)DK50310366[DK50310366], (01)00884838099876(21)DK50310390[DK50310390], (01)00884838099876(21)DK50310392[DK50310392], (01)00884838099876(21)DK50310490[DK50310490], (01)00884838099876(21)DK50310501[DK50310501], (01)00884838099876(21)DK50310512[DK50310512], (01)00884838099876(21)DK50310544[DK50310544], (01)00884838099876(21)DK50310552[DK50310552], (01)00884838099876(21)DK50310557[DK50310557], (01)00884838099876(21)DK50310629[DK50310629], (01)00884838099876(21)DK50310640[DK50310640], (01)00884838099876(21)DK50310688[DK50310688], (01)00884838099876(21)DK50310689[DK50310689], (01)00884838099876(21)DK50310696[DK50310696], (01)00884838099876(21)DK50310703[DK50310703], (01)00884838099876(21)DK50310725[DK50310725], (01)00884838099876(21)DK50310842[DK50310842], (01)00884838099876(21)DK50310857[DK50310857], (01)00884838099876(21)DK50310861[DK50310861], (01)00884838099876(21)DK50310914[DK50310914], (01)00884838099876(21)DK50310929[DK50310929], (01)00884838099876(21)DK50410998[DK50410998], (01)00884838099876(21)DK50411020[DK50411020], (01)00884838099876(21)DK50411031[DK50411031], (01)00884838099876(21)DK50411044[DK50411044], (01)00884838099876(21)DK50411061[DK50411061], (01)00884838099876(21)DK50411062[DK50411062], (01)00884838099876(21)DK50411075[DK50411075], (01)00884838099876(21)DK50411122[DK50411122], (01)00884838099876(21)DK50411153[DK50411153], (01)00884838099876(21)DK50411170[DK50411170], (01)00884838099876(21)DK50411198[DK50411198], (01)00884838099876(21)DK50411207[DK50411207], (01)00884838099876(21)DK50411209[DK50411209], (01)00884838099876(21)DK50411214[DK50411214], (01)00884838099876(21)DK50411216[DK50411216], (01)00884838099876(21)DK50411217[DK50411217], (01)00884838099876(21)DK50411237[DK50411237], (01)00884838099876(21)DK50411264[DK50411264], (01)00884838099876(21)DK50411273[DK50411273], (01)00884838099876(21)DK50411303[DK50411303], (01)00884838099876(21)DK50411320[DK50411320], (01)00884838099876(21)DK50411356[DK50411356], (01)00884838099876(21)DK50411368[DK50411368], (01)00884838099876(21)DK50411375[DK50411375], (01)00884838099876(21)DK50511417[DK50511417], (01)00884838099876(21)DK50511422[DK50511422], (01)00884838099876(21)DK50511520[DK50511520], (01)00884838099876(21)DK50511536[DK50511536], (01)00884838099876(21)DK50511547[DK50511547], (01)00884838099876(21)DK50511572[DK50511572], (01)00884838099876(21)DK50511590[DK50511590], (01)00884838099876(21)DK50511639[DK50511639], (01)00884838099876(21)DK50511645[DK50511645], (01)00884838099876(21)DK50511648[DK50511648], (01)00884838099876(21)DK50511657[DK50511657], (01)00884838099876(21)DK50511675[DK50511675], (01)00884838099876(21)DK50511678[DK50511678], (01)00884838099876(21)DK50511681[DK50511681], (01)00884838099876(21)DK50511828[DK50511828], (01)00884838099876(21)DK50811982[DK50811982], (01)00884838099876(21)DK50812076[DK50812076], (01)00884838099876(21)DK50812086[DK50812086], (01)00884838099876(21)DK50812087[DK50812087], (01)00884838099876(21)DK50812097[DK50812097], (01)00884838099876(21)DK50812122[DK50812122], (01)00884838099876(21)DK50812220[DK50812220], (01)00884838099876(21)DK50812303[DK50812303], (01)00884838099876(21)DK50812360[DK50812360], (01)00884838099876(21)DK50812406[DK50812406], (01)00884838099876(21)DK50812438[DK50812438], (01)00884838099876(21)DK50812480[DK50812480], (01)00884838099876(21)DK51112549[DK51112549], (01)00884838099876(21)DK51112584[DK51112584], (01)00884838099876(21)DK51112818[DK51112818], (01)00884838099876(21)DK51112885[DK51112885], (01)00884838099876(21)DK51212997[DK51212997], (01)00884838099876(21)DK51213045[DK51213045], (01)00884838099876(21)DK51313240[DK51313240], (01)00884838099876(21)DK51313250[DK51313250], (01)00884838099876(21)DK51313275[DK51313275], (01)00884838099876(21)DK51313276[DK51313276], (01)00884838099876(21)DK51313330[DK51313330], (01)00884838099876(21)DK51313342[DK51313342], (01)00884838099876(21)DK51313388[DK51313388], (01)00884838099876(21)DK51313389[DK51313389], (01)00884838099876(21)DK51313498[DK51313498], (01)00884838099876(21)DK51313526[DK51313526], (01)00884838099876(21)DK51313533[DK51313533], (01)00884838099876(21)DK51313613[DK51313613], (01)00884838099876(21)DK51313618[DK51313618], (01)00884838099876(21)DK51313621[DK51313621], (01)00884838099876(21)DK51313681[DK51313681], (01)00884838099876(21)DK51313712[DK51313712], (01)00884838099876(21)DK51313782[DK51313782], (01)00884838099876(21)DK51313853[DK51313853], (01)00884838099876(21)DK51813974[DK51813974], (01)00884838099876(21)DK51814002[DK51814002], (01)00884838099876(21)DK51814017[DK51814017], (01)00884838099876(21)DK51814145[DK51814145], (01)00884838099876(21)DK51814159[DK51814159], (01)00884838099876(21)DK51814193[DK51814193], (01)00884838099876(21)DK51814242[DK51814242], (01)00884838099876(21)DK51814253[DK51814253], (01)00884838099876(21)DK51814262[DK51814262], (01)00884838099876(21)DK51814274[DK51814274], (01)00884838099876(21)DK51814295[DK51814295], (01)00884838099876(21)DK51814325[DK51814325], (01)00884838099876(21)DK51814331[DK51814331], (01)00884838099876(21)DK51814345[DK51814345], (01)00884838099876(21)DK51814466[DK51814466], (01)00884838099876(21)DK52014528[DK52014528], (01)00884838099876(21)DK52014543[DK52014543], (01)00884838099876(21)DK52014557[DK52014557], (01)00884838099876(21)DK52014562[DK52014562], (01)00884838099876(21)DK52014623[DK52014623], (01)00884838099876(21)DK52014635[DK52014635], (01)00884838099876(21)DK52014682[DK52014682], (01)00884838099876(21)DK52014762[DK52014762], (01)00884838099876(21)DK52014763[DK52014763], (01)00884838099876(21)DK52014851[DK52014851], (01)00884838099876(21)DK52014854[DK52014854], (01)00884838099876(21)DK52014858[DK52014858], (01)00884838099876(21)DK52014873[DK52014873], (01)00884838099876(21)DK52014876[DK52014876], (01)00884838099876(21)DK52014877[DK52014877], (01)00884838099876(21)DK52014940[DK52014940], (01)00884838099876(21)DK52015039[DK52015039], (01)00884838099876(21)DK52015075[DK52015075], (01)00884838099876(21)DK52015080[DK52015080], (01)00884838099876(21)DK52115152[DK52115152], (01)00884838099876(21)DK52115179[DK52115179], (01)00884838099876(21)DK52115266[DK52115266], (01)00884838099876(21)DK52115272[DK52115272], (01)00884838099876(21)DK52115273[DK52115273], (01)00884838099876(21)DK52115278[DK52115278], (01)00884838099876(21)DK52115302[DK52115302], (01)00884838099876(21)DK52115371[DK52115371], (01)00884838099876(21)DK52115383[DK52115383], (01)00884838099876(21)DK52115584[DK52115584], (01)00884838099876(21)DK52115587[DK52115587], (01)00884838099876(21)DK52115679[DK52115679], (01)00884838099876(21)DK52315787[DK52315787], (01)00884838099876(21)DK52315839[DK52315839], (01)00884838099876(21)DK52315847[DK52315847], (01)00884838099876(21)DK52315891[DK52315891], (01)00884838099876(21)DK52315919[DK52315919], (01)00884838099876(21)DK52315940[DK52315940], (01)00884838099876(21)DK52315942[DK52315942], (01)00884838099876(21)DK52315948[DK52315948], (01)00884838099876(21)DK52315956[DK52315956], (01)00884838099876(21)DK52315965[DK52315965], (01)00884838099876(21)DK52315969[DK52315969], (01)00884838099876(21)DK52315979[DK52315979], (01)00884838099876(21)DK52315988[DK52315988], (01)00884838099876(21)DK52316026[DK52316026], (01)00884838099876(21)DK52316045[DK52316045], (01)00884838099876(21)DK52316053[DK52316053], (01)00884838099876(21)DK52316112[DK52316112], (01)00884838099876(21)DK52316134[DK52316134], (01)00884838099876(21)DK52316156[DK52316156], (01)00884838099876(21)DK52316235[DK52316235], (01)00884838099876(21)DK52316258[DK52316258], (01)00884838099876(21)DK52316272[DK52316272], (01)00884838099876(21)DK52316279[DK52316279], (01)00884838099876(21)DK52316283[DK52316283], (01)00884838099876(21)DK52316292[DK52316292], (01)00884838099876(21)DK52316312[DK52316312], (01)00884838099876(21)DK52316319[DK52316319], (01)00884838099876(21)DK52316322[DK52316322], (01)00884838099876(21)DK52316335[DK52316335], (01)00884838099876(21)DK52316395[DK52316395], (01)00884838099876(21)DK52516487[DK52516487], (01)00884838099876(21)DK52517004[DK52517004].

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1033-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 12, 2025
ReportedJanuary 14, 2026
CompanyPhilips North America Llc
Company locationCambridge, MA, United States
Initiated byVoluntary: Firm initiated
Quantity13,470 Units