Recall Ledger

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) OPEN HEART ACCESSORY CDSKit SKU CDS984289O; 2) OPEN HEART CDSKit SKU CDS984355K; 3) BAG A OPEN HEART KITKit SKU DYKM1025D; 4) RR-CABG SUPPLEMENTAL PACKKit SKU DYNJ0394752O; 5) OPEN HEART C. Reason: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

What is being recalled

Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) OPEN HEART ACCESSORY CDSKit SKU CDS984289O; 2) OPEN HEART CDSKit SKU CDS984355K; 3) BAG A OPEN HEART KITKit SKU DYKM1025D; 4) RR-CABG SUPPLEMENTAL PACKKit SKU DYNJ0394752O; 5) OPEN HEART CDSKit SKU DYNJ902879T; 6) BASIC CARDIAC ANESTHESIAKit SKU DYNJ903523F; 7) OPEN HEART CDSKit SKU DYNJ905557B; 8) AN03 HEART/ANESTHESIA KIT-LFKit SKU PHS972012D.

Why

Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

Medline SKU PHS972096014B, UDI/DI 10889942695911 (EA) 40889942695912 (CS), Lot Number 25GBY100; Medline SKU CDS984289O, UDI/DI 10198459418143 (EA) 40198459418144 (CS), Lot Number 25HBI736; Medline SKU DYNJ903523F, UDI/DI 10198459316593 (EA) 40198459316594 (CS), Lot Number 25HBL296; Medline SKU DYNJ903523F, UDI/DI 10198459316593 (EA) 40198459316594 (CS), Lot Number 25HBN030; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25HLA445; Medline SKU DYKM1025D, UDI/DI 10195327482893 (EA) 40195327482894 (CS), Lot Number 25HMA140; Medline SKU DYNJ0394752O, UDI/DI 10193489432794 (EA) 40193489432795 (CS), Lot Number 25IBI197; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25IBL774; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25IMG620; Medline SKU DYNJ0394752O, UDI/DI 10193489432794 (EA) 40193489432795 (CS), Lot Number 25IMH657; Medline SKU PHS972012D, UDI/DI 10193489571875 (EA) 40193489571876 (CS), Lot Number 25IMH924; Medline SKU DYNJ905557B, UDI/DI 10193489468243 (EA) 40193489468244 (CS), Lot Number 25JBH529; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25JBO555; Medline SKU DYNJ903523F, UDI/DI 10198459316593 (EA) 40198459316594 (CS), Lot Number 25JLA937; Medline SKU CDS984355K, UDI/DI 10198459258848 (EA) 40198459258849 (CS), Lot Number 25KBB119; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25KBH512; Medline SKU DYNJ902879T, UDI/DI 10195327461294 (EA) 40195327461295 (CS), Lot Number 25KBH512.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1050-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedNovember 26, 2025
ReportedJanuary 21, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity633 units

More recalls from Medline Industries, LP · all medical device recalls · recalls from November 2025

Official FDA record