Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-PLASTIC CRANIOFACIAL PK-LF, Medline kit SKU DYNJ0395085V; 2) NH PERI DISC PACK, Medline kit SKU DYNJ66311C; 3) DISCECTOMY PACK, Medline kit SKU DYNJ86770; 4) TOTAL KNEE, Medline kit SKU DY. Reason: Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
What is being recalled
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: 1) RR-PLASTIC CRANIOFACIAL PK-LF, Medline kit SKU DYNJ0395085V; 2) NH PERI DISC PACK, Medline kit SKU DYNJ66311C; 3) DISCECTOMY PACK, Medline kit SKU DYNJ86770; 4) TOTAL KNEE, Medline kit SKU DYNJ909643F.
Why
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution.
How to tell if yours is affected
Medline kit SKU DYNJ0395085V,UDI/DI 10198459313745 (EA) 40198459313746 (CS), Lot Number 25HMC337; Medline kit SKU DYNJ66311C,UDI/DI 10198459364556 (EA) 40198459364557 (CS), Lot Number 25KBG419; Medline kit SKU DYNJ86770,UDI/DI 10195327674229 (EA) 40195327674220 (CS), Lot Number 25IBU284; Medline kit SKU DYNJ86770,UDI/DI 10195327674229 (EA) 40195327674220 (CS), Lot Number 25KBJ623; Medline kit SKU DYNJ909643F,UDI/DI 10198459392603 (EA) 40198459392604 (CS), Lot Number 25HBU964; Medline kit SKU DYNJ909643F,UDI/DI 10198459392603 (EA) 40198459392604 (CS), Lot Number 25IBV819; Medline kit SKU DYNJ909643F,UDI/DI 10198459392603 (EA) 40198459392604 (CS), Lot Number 25KBF028.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1060-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | November 26, 2025 |
| Reported | January 21, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 338 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from November 2025