Allura Xper FD10 recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)

Medical deviceClass IIOngoing

Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Allura Xper FD10; Model Number: 722010;. Reason: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

What is being recalled

Allura Xper FD10; Model Number: 722010;

Why

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

How to tell if yours is affected

Model Number: 722010; System Product Name: Allura Xper FD10; UDI-DI: 00884838059030; Serial Numbers: 764, 1396, 728, 980, 782, 125, 1096, 661, 225, 1055, 1185, 368, 743, 1006, 570, 901, 552, 973, 795, 888, 1140, 781, 1060, 354, 186, 516, 517, 64, 386, 387, 530, 996, 964, 1265, 525, 307, 1189, 98, 774, 935, 721, 254, 1008, 1213, 1273, 66, 156, 304, 1422, 388, 235, 236, 1066, 1090, 549, 689;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1068-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 15, 2025
ReportedJanuary 21, 2026
CompanyPHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Company locationBest, Netherlands
Initiated byVoluntary: Firm initiated
Quantity56 units