Recall Ledger

Brilliance iCT recall (Philips North America Llc)

Medical deviceClass IIOngoing

Summary: Philips North America Llc is recalling Brilliance iCT; Product Code (REF): 728306;. Reason: As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

What is being recalled

Brilliance iCT; Product Code (REF): 728306;

Why

As part of the CT systems sub assembly, thirty-two (32) fasteners are used to attach the rotor to the bearing within the gantry. One or more of these fasteners may not be torqued to specification. If multiple fasteners are not torqued to specification, the rotor or rotor parts may become unsecured or displaced. While there has been no reported or observed event of expelled parts during gantry rotation, unsecured rotor parts may potentially be expelled.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MD, MN, NC, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, France, Germany, Hong Kong, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Macao, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, R¿union, Russian Fed., Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkmenistan, United Kingdom, Uzbekistan, Vietnam. (states named: Alabama, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Minnesota, North Carolina, New York, Ohio, Oregon, Pennsylvania, Texas, Vermont, Washington, West Virginia)

How to tell if yours is affected

Product Code (REF): 728306; UDI-DI: 00884838059474; Serial Numbers: 87064, 87120, 87121, 87122, 87125, 87126, 87127, 87128, 87129, 87130, 87135, 87143, 87148, 87152, 87153, 87157, 87160, 87162, 87167, 87173, 87174, 87176, 87181, 87197, 87198, 87199, 87200 , 87202, 87203, 87204, 87205, 87206, 87207, 87208, 87209, 87210, 87211, 87212, 87213 , 87214, 87215, 87216, 87217, 87218, 87219, 87220, 87221;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1315-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 7, 2026
ReportedFebruary 18, 2026
CompanyPhilips North America Llc
Company locationCambridge, MA, United States
Initiated byVoluntary: Firm initiated
Quantity47 units

More recalls from Philips North America Llc · all medical device recalls · recalls from January 2026

Official FDA record