Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI... recall (Wilson-Cook Medical Inc.)
Medical deviceClass IIOngoing
Summary: Wilson-Cook Medical Inc. is recalling Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also a. Reason: Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
What is being recalled
Teslatome Bipolar Sphincterotomes REF: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 These devices are used for cannulation of the ductal system and for sphincterotomy. If preloaded, the device also aids in bridging difficult strictures during ERCP. It is indicated for sphincterotome-aided, wire-guided selective cannulation of the biliary ducts.
Why
Due to product that may contain a void in the material that separates the wire guide and cutting wire lumens, which may cause an inappropriate path of current during cutting.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
U.S. Distribution to states of: CA, CO, IN, and NC. (states named: California, Colorado, Indiana, North Carolina)
How to tell if yours is affected
RPN: TESLA-B2535-260 Model Number: G60784 UDI-DI code: 00827002607842 Work Order Number / RPN W4954835 / TESLA-B2535-260 W4954836 / TESLA-B2535-260 W4957208 / TESLA-B2535-260 W4958992 / TESLA-B2535-260 W4960895 / TESLA-B2535-260 W4960896 / TESLA-B2535-260 W4961670 / TESLA-B2535-260 W4965708 / TESLA-B2535-260 W4966040 / TESLA-B2535-260 W4966803 / TESLA-B2535-260 W4967866 / TESLA-B2535-260 W4968515 / TESLA-B2535-260 W4968867 / TESLA-B2535-260 W4969292 / TESLA-B2535-260 W4969674 / TESLA-B2535-260 W4969687 / TESLA-B2535-260 W4970315 / TESLA-B2535-260 W4970662 / TESLA-B2535-260 W4971561 / TESLA-B2535-260 W4972693 / TESLA-B2535-260 W4973085 / TESLA-B2535-260 W4974435 / TESLA-B2535-260 W4974588 / TESLA-B2535-260 W4975407 / TESLA-B2535-260 W4975688 / TESLA-B2535-260 W4977514 / TESLA-B2535-260 W4977824 / TESLA-B2535-260 W4979182 / TESLA-B2535-260 W4979183 / TESLA-B2535-260 W4979665 / TESLA-B2535-260 W4982425 / TESLA-B2535-260 W4982436 / TESLA-B2535-260 REF: TESLA-B2535-450 Model Number: G60785 UDI-DI code: 00827002607859 Work Order Number / RPN: W4962311 / TESLA-B2535-450 W4966039 / TESLA-B2535-450 W4968222 / TESLA-B2535-450 W4969306 / TESLA-B2535-450 W4970322 / TESLA-B2535-450 W4971499 / TESLA-B2535-450 W4975687 / TESLA-B2535-450 W4977254 / TESLA-B2535-450 W4982013 / TESLA-B2535-450
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1362-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | December 18, 2025 |
| Reported | February 25, 2026 |
| Company | Wilson-Cook Medical Inc. |
| Company location | Winston Salem, NC, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 153 |
More recalls from Wilson-Cook Medical Inc. · all medical device recalls · recalls from December 2025