Medline Convenience Kits: 1) OPEN HEART CDS recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP CABG CDS, Model Number: CDS840087AK; 5) OFF PUMP CABG CDS, Model Number: CDS840087AL; 6) OPEN HEART, Model Numb. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What is being recalled

Medline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840015AQ; 2) OPEN HEART CDS, Model Number: CDS840023T; 3) OFF PUMP CABG CDS, Model Number: CDS840087AI; 4) OFF PUMP CABG CDS, Model Number: CDS840087AK; 5) OFF PUMP CABG CDS, Model Number: CDS840087AL; 6) OPEN HEART, Model Number: CDS840150S; 7) OPEN HEART, Model Number: CDS840150T; 8) OPEN HEART, Model Number: CDS840150U; 9) OPEN HEART, Model Number: CDS840150V; 10) OPEN HEART, Model Number: CDS840150W; 11) UNIVERSAL HEART CDS, Model Number: CDS840246O; 12) GENERAL AAA #11-RF, Model Number: CDS840261AB; 13) OPEN HEART ADULT, Model Number: CDS840396AA; 14) OPEN HEART ADULT, Model Number: CDS840396AB; 15) OPEN HEART ADULT, Model Number: CDS840396X; 16) OPEN HEART ADULT, Model Number: CDS840396Y; 17) OPEN HEART CDS, Model Number: CDS840428K; 18) OPEN HEART CDS, Model Number: CDS840428L; 19) PACEMAKER CDS, Model Number: CDS980839C; 20) KIT, CVC ADD A PACK HUM-LF, Model Number: CDS982564C; 21) OPEN HEART ADULT HUM, Model Number: CDS982566K; 22) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575K; 23) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575M; 24) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575O; 25) OPEN HEART BOX 1 CDS MRH, Model Number: CDS982575P; 26) OPEN HEART, Model Number: CDS983348S; 27) HEART CABG CDS, Model Number: CDS983376Q; 28) HEART CABG CDS, Model Number: CDS983376T; 29) HEART CABG CDS, Model Number: CDS983376U; 30) HEART CABG CDS, Model Number: CDS983376V; 31) OPEN HEART CDS, Model Number: CDS983513F; 32) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786D; 33) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786F; 34) OPEN HEART ADULT PART 2 CDS, Model Number: CDS983786G; 35) PEDS CARDIAC CDS, Model Number: CDS984117F; 36) OPEN HEART CDS, Model Number: CDS984276P; 37) OPEN HEART ACCESSORY CDS, Model Number: CDS984289J; 38) OPEN HEART ACCESSORY CDS, Model Number: CDS984289K; 39) OPEN HEART ACCESSORY CDS, Model Number: CDS984289L; 40) OPEN HEART ACCESSORY CDS, Model Number: CDS984289M; 41) OPEN HEART ACCESSORY CDS, Model Number: CDS984289N; 42) OPEN HEART ACCESSORY CDS, Model Number: CDS984289O; 43) OPEN HEART CDS, Model Number: CDS984355F; 44) OPEN HEART CDS, Model Number: CDS984355G; 45) OPEN HEART CDS, Model Number: CDS984355I; 46) OPEN HEART CDS, Model Number: CDS984355J; 47) OPEN HEART A, Model Number: CDS985105M; 48) OPEN HEART A, Model Number: CDS985105N; 49) OPEN HEART A, Model Number: CDS985105O; 50) VASCULAR CDS, Model Number: CDS985215C; 51) BASE HEART PACK-LF, Model Number: CVT159AA; 52) ENDO KIT, Model Number: DYKE1580B; 53) KIT VASCULAR MINOR, Model Number: DYKMBNDL11; 54) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118; 55) KIT VASCULAR CAROTID, Model Number: DYKMBNDL118A; 56) KIT VASCULAR BYPASS GRAFT GENE, Model Number: DYKMBNDL120; 57) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14; 58) KIT VASCULAR PERCUTANEOUS TEVA, Model Number: DYKMBNDL14A; 59) KIT VASCULAR VNS, Model Number: DYKMBNDL152; 60) KIT VASCULAR VNS, Model Number: DYKMBNDL152A; 61) KIT THORACIC SYMPHATECTOMY, Model Number: DYKMBNDL187; 62) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189B; 63) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL191; 64) KIT VASCULAR ANGIOGRAM GENERIC, Model Number: DYKMBNDL192; 65) KIT VASCULAR ENDOVENOUS ABLATI, Model Number: DYKMBNDL203; 66) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209A; 67) KIT VASCULAR MEDIAN ARCUATE, Model Number: DYKMBNDL209B; 68) CARDIAC ADULT CABG, Model Number: DYKMBNDL20A; 69) CARDIAC ADULT CABG, Model Number: DYKMBNDL20B; 70) KIT CARDIAC CSICU ADULT ECMO, Model Number: DYKMBNDL21; 71) KIT VASCULAR AAA REPAIR, Model Number: DYKMBNDL23A; 72) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27A; 73) KIT CARDIAC ANEURYSM, Model Number: DYKMBNDL27B; 74) KIT PEDS CARDIAC ADULT, Model Number: DYKMBNDL4

Why

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide.

How to tell if yours is affected

1) CDS840015AQ, UDI-DI: 10195327210144(each), 40195327210145(case), Lot Number: 23FLA191; 2) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23LBM199; 3) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23KBM369; 4) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23JBU616; 5) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23IBJ759; 6) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23IBJ571; 7) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23GBA105; 8) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23CBO830; 9) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23ABM709; 10) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 23ABE738; 11) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22OBM690; 12) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBW565; 13) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBL256; 14) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22KBJ798; 15) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22JBU353; 16) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22JBT770; 17) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22HBU935; 18) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22GBQ917; 19) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22EBK678; 20) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 22CBC309; 21) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21LBV113; 22) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21KBW667; 23) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21IBG586; 24) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21GBH886; 25) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21EBN817; 26) CDS840023T, UDI-DI: 10193489682069(each), 40193489682060(case), Lot Number: 21CKA394; 27) CDS840087AI, UDI-DI: 10193489847413(each), 40193489847414(case), Lot Number: 21ELA419; 28) CDS840087AI, UDI-DI: 10193489847413(each), 40193489847414(case), Lot Number: 21DLA296; 29) CDS840087AK, UDI-DI: 10195327212353(each), 40195327212354(case), Lot Number: 23ELA870; 30) CDS840087AL, UDI-DI: 10195327590406(each), 40195327590407(case), Lot Number: 24KLA701; 31) CDS840150S, UDI-DI: 10193489862409(each), 40193489862400(case), Lot Number: 21EBK732; 32) CDS840150S, UDI-DI: 10193489862409(each), 40193489862400(case), Lot Number: 21DBA306; 33) CDS840150T, UDI-DI: 10193489996197(each), 40193489996198(case), Lot Number: 21JBA896; 34) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22LBS336; 35) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22JBD465; 36) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22GBD762; 37) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22EBL398; 38) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22CBF996; 39) CDS840150U, UDI-DI: 10195327055318(each), 40195327055319(case), Lot Number: 22ABA269; 40) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 24ABJ652; 41) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 23KBC879; 42) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 23KBC675; 43) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 23HBJ716; 44) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 23EBB628; 45) CDS840150V, UDI-DI: 10195327298401(each), 40195327298402(case), Lot Number: 23DBN632; 46) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24FBP521; 47) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24FBP520; 48) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24EBK094; 49) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24DBB189; 50) CDS840150W, UDI-DI: 10195327648046(each), 40195327648047(case), Lot Number: 24CBQ331; 51) CDS840246O, UDI-DI: 10198459095955(each), 40198459095956(case), Lot Number: 25BLA225; 52) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 25EBF565; 53) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 25CBO300; 54) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 25ABU765; 55) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 25ABA041; 56) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24LBK995; 57) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24JBJ413; 58) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24IBM204; 59) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24GBQ632; 60) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24FBS248; 61) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24CBU862; 62) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24CBK620; 63) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24CBF421; 64) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 24ABB897; 65) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23LBK827; 66) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23JBM632; 67) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23GBJ781; 68) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23FBQ953; 69) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23FBF165; 70) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23DBE827; 71) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 23ABG192; 72) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 22OBA486; 73) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 22KBO918; 74) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 22IBE745; 75) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 22FBX514; 76) CDS840261AB, UDI-DI: 10195327104535(each), 40195327104536(case), Lot Number: 22FBL409; 77) CDS840396AA, UDI-DI: 10198459348686(each), 40198459348687(case), Lot Number: 25HBM750; 78) CDS840396AA, UDI-DI: 10198459348686(each), 40198459348687(case), Lot Number: 25FBD477; 79) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 25JBQ605; 80) CDS840396AB, UDI-DI: 10198459468742(each), 40198459468743(case), Lot Number: 25IBD204; 81) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABL893; 82) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABK089; 83) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 24ABC933; 84) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 23GBQ865; 85) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 23GBN891; 86) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 23BBG436; 87) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 23ABG179; 88) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 23ABF773; 89) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22JBG079; 90) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22JBG077; 91) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22JBG072; 92) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22FBF451; 93) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22FBF439; 94) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22DBL431; 95) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22ABE659; 96) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 22ABE658; 97) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 21KBQ156; 98) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 21IBO083; 99) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 21HBO916; 100) CDS840396X, UDI-DI: 10193489461916(each), 40193489461917(case), Lot Number: 21CBG242; 101) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25CBK781; 102) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25CBD493; 103) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25BBP788; 104) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 25BBG163; 105) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24KBK977; 106) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24IBK539; 107) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24FBS930; 108) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24FBJ654; 109) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24CBR417; 110) CDS840396Y, UDI-DI: 10195327574932(each), 40195327574933(case), Lot Number: 24BBO585; 111) CDS840428K, UDI-DI: 10884389911235(each), 40884389911236(case), Lot Number: 21GBS209; 112) CDS840428K, UDI-DI: 10884389911235(each), 40884389911236(case), Lot Number: 21BBH224; 113) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 25DBQ683; 114) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 25CBI244; 115) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 25BBM940; 116) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 25BBF438; 117) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24JBS428; 118) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24HBL828; 119) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24GBU141; 120) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24EBC679; 121) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24CBB817; 122) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 24ABX492; 123) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 23EBU791; 124) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 23EBA214; 125) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 23CBR891; 126) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 22KBW189; 127) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 22IBV957; 128) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 22HBA238; 129) CDS840428L, UDI-DI: 10193489922660(each), 40193489922661(case), Lot Number: 22GLA269; 130) CDS980839C, UDI-DI: 10193489723076(each), 40193489723077(case), Lot Number: 21FBB093; 131) CDS982564C, UDI-DI: 10193489873177(each), 40193489873178(case), Lot Number: 25CMJ833; 132) CDS982564C, UDI-DI: 10193489873177(each), 40193489873178(case), Lot Number: 25BMD423; 133) CDS982564C, UDI-DI: 10193489873177(each), 40193489873178(case), Lot Number: 24JMF683; 134) CDS982564C, UDI-DI: 10193489873177(each), 40193489873178(case), Lot Number: 24HMA337; 135) CDS982564C, UDI-DI: 10193489873177(each), 40193489873178(case), Lot Number: 24EMG641; 136) CDS982566K, UDI-DI: 10193489435597(each), 40193489435598(case), Lot Number: 21CBQ731; 137) CDS982575K, UDI-DI: 10193489734799(each), 40193489734790(case), Lot Number: 21BKA040; 138) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21JBS146; 139) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21JBQ075; 140) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21HBQ086; 141) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21HBF470; 142) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21GBK054; 143) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21EKA064; 144) CDS982575M, UDI-DI: 10193489845754(each), 40193489845755(case), Lot Number: 21DKA457; 145) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22LBS833; 146) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22JBP075; 147) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22IBU776; 148) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22GBM066; 149) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22FBF335; 150) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22EBL281; 151) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22DBI104; 152) CDS982575O, UDI-DI: 10195327101237(each), 40195327101238(case), Lot Number: 22BBQ149; 153) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 24BBB783; 154) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23LBS018; 155) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23KBK940; 156) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23IBP737; 157) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23HBT093; 158) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23HBF574; 159) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23FBU058; 160) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23FBL415; 161) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23CBT572; 162) CDS982575P, UDI-DI: 10195327348908(each), 40195327348909(case), Lot Number: 23CBL429; 163) CDS983348S, UDI-DI: 10193489458992(each), 40193489458993(case), Lot Number: 21FBF717; 164) CDS983348S, UDI-DI: 10193489458992(each), 40193489458993(case), Lot Number: 21EBC025; 165) CDS983348S, UDI-DI: 10193489458992(each), 40193489458993(case), Lot Number: 21DBP216; 166) CDS983348S, UDI-DI: 10193489458992(each), 40193489458993(case), Lot Number: 21CBK379; 167) CDS983376Q, UDI-DI: 10195327379797(each), 40195327379798(case), Lot Number: 23IBR206; 168) CDS983376Q, UDI-DI: 10195327379797(each), 40195327379798(case), Lot Number: 23FBU833; 169) CDS983376T, UDI-DI: 10198459197338(each), 40198459197339(case), Lot Number: 25DMH143; 170) CDS983376T, UDI-DI: 10198459197338(each), 40198459197339(case), Lot Number: 25DME288; 171) CDS983376T, UDI-DI: 10198459197338(each), 40198459197339(case), Lot Number: 25CME797; 172) CDS983376U, UDI-DI: 10198459397585(each), 40198459397586(case), Lot Number: 25HMA016; 173) CDS983376U, UDI-DI: 10198459397585(each), 40198459397586(case), Lot Number: 25GMG706; 174) CDS983376U, UDI-DI: 10198459397585(each), 40198459397586(case), Lot Number: 25GME704; 175) CDS983376V, UDI-DI: 10198459475337(each), 40198459475338(case), Lot Number: 25LMH340; 176) CDS983376V, UDI-DI: 10198459475337(each), 40198459475338(case), Lot Number: 25LMF022; 177) CDS983376V, UDI-DI: 10198459475337(each), 40198459475338(case), Lot Number: 25IMH581; 178) CDS983513F, UDI-DI: 10193489843729(each), 40193489843720(case), Lot Number: 23BBE439; 179) CDS983513F, UDI-DI: 10193489843729(each), 40193489843720(case), Lot Number: 22KBI289; 180) CDS983513F, UDI-DI: 10193489843729(each), 40193489843720(case), Lot Number: 22IBD426; 181) CDS983513F, UDI-DI: 10193489843729(each), 40193489843720(case), Lot Number: 22HBI664; 182) CDS983513F, UDI-DI: 10193489843729(each), 40193489843720(case), Lot Number: 21FBF442; 183) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 24CME689; 184) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 24AMD733; 185) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23KMC560; 186) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23HMC461; 187) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23GMH171; 188) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23GMG766; 189) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23FMI141; 190) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23EMF372; 191) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 23DMH027; 192) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22KMH072; 193) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22JMB563; 194) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22HMF199; 195) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22HMB738; 196) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22FMC683; 197) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 22CMB854; 198) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21LMA822; 199) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21KMA019; 200) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21JMB643; 201) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21IMH924; 202) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21GMC871; 203) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21FMD794; 204) CDS983786D, UDI-DI: 10193489590197(each), 40193489590198(case), Lot Number: 21AKA832; 205) CDS983786F, UDI-DI: 10195327684594(each), 40195327684595(case), Lot Number: 24JMF693; 206) CDS983786F, UDI-DI: 10195327684594(each), 40195327684595(case), Lot Number: 24JMC120; 207) CDS983786F, UDI-DI: 10195327684594(each), 40195327684595(case), Lot Number: 24IMG956; 208) CDS983786F, UDI-DI: 10195327684594(each), 40195327684595(case), Lot Number: 24IMB935; 209) CDS983786G, UDI-DI: 10198459166822(each), 40198459166823(case), Lot Number: 25KMI896; 210) CDS983786G, UDI-DI: 10198459166822(each), 40198459166823(case), Lot Number: 25JME849; 211) CDS983786G, UDI-DI: 10198459166822(each), 40198459166823(case), Lot Number: 25BMD134; 212) CDS983786G, UDI-DI: 10198459166822(each), 40198459166823(case), Lot Number: 24LMC522; 213) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 23BBJ754; 214) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 23ABA881; 215) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 22OBL943; 216) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 22JBU821; 217) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 22EBM615; 218) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21KMH819; 219) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21KMC359; 220) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21IMC057; 221) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21GME785; 222) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21GBA694; 223) CDS984117F, UDI-DI: 10193489520408(each), 40193489520409(case), Lot Number: 21BKA482; 224) CDS984276P, UDI-DI: 10195327446802(each), 40195327446803(case), Lot Number: 24LMC474; 225) CDS984276P, UDI-DI: 10195327446802(each), 40195327446803(case), Lot Number: 24KMC946; 226) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21KLB205; 227) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21ILA628; 228) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21HLA680; 229) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21GLB087; 230) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21GLA312; 231) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21GLA057; 232) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21FLA954; 233) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21EKA147; 234) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21DKA676; 235) CDS984289J, UDI-DI: 10193489263046(each), 40193489263047(case), Lot Number: 21BKA389; 236) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 23FBV540; 237) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 23FBP065; 238) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 23ABP719; 239) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22OBI364; 240) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22LBQ630; 241) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22JBT625; 242) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22IBT454; 243) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22HBL442; 244) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22FBX071; 245) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22EBL445; 246) CDS984289K, UDI-DI: 10195327067243(each), 40195327067244(case), Lot Number: 22ABF746; 247) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24EBT292; 248) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24DBU415; 249) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24CBO822; 250) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 24CBL931; 251) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 23LBB141; 252) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 23KBA264; 253) CDS984289L, UDI-DI: 10195327439064(each), 40195327439065(case), Lot Number: 23JBB129; 254) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24LBA755; 255) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24KBN423; 256) CDS984289M, UDI-DI: 10198459118548(each), 40198459118549(case), Lot Number: 24KBA684; 257) CDS984289N, UDI-DI: 10198459256769(each), 40198459256760(case), Lot Number: 25FBD048; 258) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBL186; 259) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBC732; 260) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25IBB599; 261) CDS984289O, UDI-DI: 10198459418143(each), 40198459418144(case), Lot Number: 25GBY100; 262) CDS984355F, UDI-DI: 10888277770348(each), 40888277770349(case), Lot Number: 20LKB341; 263) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23LLA067; 264) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23JLA617; 265) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23ILB099; 266) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23HLA237; 267) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23CLA762; 268) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 23CLA569; 269) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22LLA693; 270) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22LLA544; 271) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22JLB146; 272) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22ILB092; 273) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22ILA721; 274) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22FLA187; 275) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 22BLA378; 276) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21HLA543; 277) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21HLA499; 278) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21GLA223; 279) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21FLA649; 280) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21EKA150; 281) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21DKA538; 282) CDS984355G, UDI-DI: 10193489484632(each), 40193489484633(case), Lot Number: 21DKA352; 283) CDS984355I, UDI-DI: 10195327569518(each), 40195327569519(case), Lot Number: 24ELA326; 284) CDS984355I, UDI-DI: 10195327569518(each), 40195327569519(case), Lot Number: 24BLA231; 285) CDS984355J, UDI-DI: 10198459096228(each), 40198459096229(case), Lot Number: 24LLA698; 286) CDS985105M, UDI-DI: 10193489229950(each), 40193489229951(case), Lot Number: 21FBE255; 287) CDS985105M, UDI-DI: 10193489229950(each), 40193489229951(case), Lot Number: 21BBS510; 288) CDS985105M, UDI-DI: 10193489229950(each), 40193489229951(case), Lot Number: 21ABE734; 289) CDS985105N, UDI-DI: 10193489902297(each), 40193489902298(case), Lot Number: 21LBP381; 290) CDS985105N, UDI-DI: 10193489902297(each), 40193489902298(case), Lot Number: 21JBZ656; 291) CDS985105N, UDI-DI: 10193489902297(each), 40193489902298(case), Lot Number: 21IBO615; 292) CDS985105N, UDI-DI: 10193489902297(each), 40193489902298(case), Lot Number: 21HBD158; 293) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23JBT159; 294) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23HBP383; 295) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23FBV374; 296) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23EBS635; 297) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23DBR889; 298) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 23ABT277; 299) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22LBD272; 300) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22JBP078; 301) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22JBD236; 302) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22IBT102; 303) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22IBE901; 304) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22GBU933; 305) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22DBS062; 306) CDS985105O, UDI-DI: 10195327036249(each), 40195327036240(case), Lot Number: 22BBS888; 307) CDS985215C, UDI-DI: 10193489780109(each), 40193489780100(case), Lot Number: 21DBH711; 308) CVT159AA, UDI-DI: 10193489457513(each), 40193489457514(case), Lot Number: 22DBL279; 309) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 24FBJ229; 310) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 24DBF439; 311) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 24BBT324; 312) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 23JBP216; 313) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 23IBI960; 314) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 23GBI903; 315) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 23EBO066; 316) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 23DBT226; 317) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22LBD173; 318) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22KBA324; 319) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22IBA742; 320) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22GBL773; 321) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22EBS701; 322) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22BBI986; 323) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 22BBF497; 324) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 21KBI013; 325) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 21IBQ531; 326) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 21HBC943; 327) DYKE1580B, UDI-DI: 10193489906011(each), 40193489906012(case), Lot Number: 21FBO379; 328) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21IBW789; 329) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21IBW788; 330) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21IBV599; 331) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21IBK912; 332) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21HBL002; 333) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21GBO480; 334) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21GBJ355; 335) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21FBO542; 336) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21FBF909; 337) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21DBN048; 338) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21DBC538; 339) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21DBA605; 340) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21CBS016; 341) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21CBO740; 342) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21CBM813; 343) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21CBF850; 344) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21BBN597; 345) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21ABO823; 346) DYKMBNDL11, UDI-DI: 10193489478600(each), 40193489478601(case), Lot Number: 21ABI967; 347) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22HLA262; 348) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22FLB050; 349) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22FLA352; 350) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22ELB070; 351) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22ELA030; 352) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22DLA405; 353) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22CLA548; 354) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22BLB034; 355) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22ALA723; 356) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 22ALA370; 357) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21KLA410; 358) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21HLB002; 359) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21GLA991; 360) DYKMBNDL 118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21GLA161; 361) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21FLA215; 362) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21DLA456; 363) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21BLA306; 364) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21BBF731; 365) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 21ABL157; 366) DYKMBNDL118, UDI-DI: 10193489479300(each), 40193489479301(case), Lot Number: 20XBA425; 367) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25LLA510; 368) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25ILA365; 369) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25HLA331; 370) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25GLA109; 371) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25CLA313; 372) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 25BLA585; 373) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 24KLB057; 374) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 24HLA380; 375) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 24CLA833; 376) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23LLA174; 377) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23KLA391; 378) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23ILA666; 379) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23FLA324; 380) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23ELA941; 381) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23ELA838; 382) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23DLA355; 383) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23CLA705; 384) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23BLA828; 385) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 23ALA566; 386) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 22KLA582; 387) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 22ILB112; 388) DYKMBNDL118A, UDI-DI: 10195327240677(each), 40195327240678(case), Lot Number: 22ILA065; 389) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25LLA349; 390) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25JLB053; 391) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25GLA494; 392) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25FLA240; 393) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25DLA352; 394) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25BLA990; 395) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25BLA424; 396) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 25ALB013; 397) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24LLA579; 398) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24KLA949; 399) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24GLA056; 400) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24ELA331; 401) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24ELA077; 402) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24CLA597; 403) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24BLA044; 404) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 24ALB056; 405) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23LLA448; 406) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23KLA994; 407) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23HLB003; 408) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23GLA840; 409) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23FLA967; 410) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23ELA675; 411) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23ELA040; 412) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23DLA682; 413) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23CLA186; 414) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 23BLA245; 415) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22LLA597; 416) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22LLA146; 417) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22JLB091; 418) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22HLB104; 419) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22GLB229; 420) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22GLA229; 421) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22ELB071; 422) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22DLA164; 423) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 22CBD042; 424) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21LLA903; 425) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21LLA708; 426) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21LLA186; 427) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21JLA983; 428) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21JLA460; 429) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21ILA312; 430) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21HLB003; 431) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21GLB132; 432) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21FLA945; 433) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21FLA091; 434) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21ELA354; 435) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21CLA211; 436) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21BLA317; 437) DYKMBNDL120, UDI-DI: 10193489478082(each), 40193489478083(case), Lot Number: 21ABJ093; 438) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23LBP163; 439) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23LBC712; 440) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23JBY285; 441) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23IBG420; 442) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23HBO526; 443) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23HBC121; 444) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23FBQ222; 445) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23EBP454; 446) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23CBP836; 447) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 23BBP965; 448) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22LBK907; 449) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22JBK931; 450) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22JBD609; 451) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22IBC491; 452) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22HBH510; 453) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22HBB974; 454) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22GBS119; 455) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22DBR013; 456) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22DBP495; 457) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22DBL096; 458) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22CBS524; 459) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22BBL863; 460) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 22ABB464; 461) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 21JBD210; 462) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 21HBC607; 463) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 21FBB100; 464) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 21BBG053; 465) DYKMBNDL14, UDI-DI: 10193489478631(each), 40193489478632(case), Lot Number: 21ABN769; 466) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25LBN111; 467) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25GBP609; 468) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25EBN099; 469) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25DBI894; 470) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25DBI893; 471) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25CBR584; 472) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 25ABE334; 473) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24LBC933; 474) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24JBY394; 475) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24HBC859; 476) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24GBQ246; 477) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24EBR754; 478) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24DBR729; 479) DYKMBNDL14A, UDI-DI: 10195327628840(each), 40195327628841(case), Lot Number: 24DBB197; 480) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 23JBT153; 481) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 23FBM174; 482) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 23FBJ607; 483) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 23ABF139; 484) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 22FBI463; 485) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 22FBC067; 486) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 22EBP246; 487) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 21LBY227; 488) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 21HBL119; 489) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 21FBQ540; 490) DYKMBNDL152, UDI-DI: 10193489478433(each), 40193489478434(case), Lot Number: 21BBP835; 491) DYKMBNDL152A, UDI-DI: 10198459000423(each), 40198459000424(case), Lot Number: 24DBS389; 492) DYKMBNDL187, UDI-DI: 10193489478532(each), 40193489478533(case), Lot Number: 22HBN514; 493) DYKMBNDL187, UDI-DI: 10193489478532(each), 40193489478533(case), Lot Number: 21BBG091; 494) DYKMBNDL187, UDI-DI: 10193489478532(each), 40193489478533(case), Lot Number: 21ABI995; 495) DYKMBNDL189B, UDI-DI: 10195327240905(each), 40195327240906(case), Lot Number: 23BLA829; 496) DYKMBNDL189B, UDI-DI: 10195327240905(each), 40195327240906(case), Lot Number: 23ALA700; 497) DYKMBNDL189B, UDI-DI: 10195327240905(each), 40195327240906(case), Lot Number: 22LLA640; 498) DYKMBNDL189B, UDI-DI: 10195327240905(each), 40195327240906(case), Lot Number: 22ILA136; 499) DYKMBNDL189B, UDI-DI: 10195327240905(each), 40195327240906(case), Lot Number: 22ILA034; 500) DYKMBNDL191, UDI-DI: 10193489479119(each), 40193489479110(case), Lot Number: 21ABO840; 501) DYKMBNDL191, UDI-DI: 10193489479119(each), 40193489479110(case), Lot Number: 21ABI997; 502) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 25GBP928; 503) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 25EBV280; 504) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 25EBE324; 505) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 25BBG078; 506) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 25ABA825; 507) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24JBY395; 508) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24IBJ517; 509) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24IBB016; 510) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24GBW445; 511) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24GBB018; 512) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24EBU596; 513) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24EBE951; 514) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24DBR154; 515) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24CBR493; 516) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24BBI231; 517) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24ABX985; 518) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 24ABK350; 519) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23LBO306; 520) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23LBD811; 521) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23KBP813; 522) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23IBP911; 523) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23FBP259; 524) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23DBG689; 525) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23DBD501; 526) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23ABV363; 527) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 23ABH676; 528) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22KBK586; 529) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22IBG488; 530) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22HBB975; 531) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22GBO608; 532) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22FBE110; 533) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22CBU218; 534) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 22ABW479; 535) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21LBO227; 536) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21KBH204; 537) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21JBJ226; 538) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21IBK913; 539) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21GBK377; 540) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21EBM701; 541) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21DBY119; 542) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21DBG176; 543) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21CBR878; 544) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21CBG285; 545) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21BBA034; 546) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21ABO827; 547) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 21ABI787; 548) DYKMBNDL192, UDI-DI: 10193489479355(each), 40193489479356(case), Lot Number: 20LBY955; 549) DYKMBNDL203, UDI-DI: 10193489479461(each), 40193489479462(case), Lot Number: 21ABT656; 550) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 25BBH561; 551) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 25BBC619; 552) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 24JBA492; 553) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 24HBC860; 554) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 24GBP997; 555) DYKMBNDL209A, UDI-DI: 10198459000515(each), 40198459000516(case), Lot Number: 24FBS686; 556) DYKMBNDL209B, UDI-DI: 10198459218361(each), 40198459218362(case), Lot Number: 25GBT327; 557) DYKMBNDL209B, UDI-DI: 10198459218361(each), 40198459218362(case), Lot Number: 25FBQ774; 558) DYKMBNDL209B, UDI-DI: 10198459218361(each), 40198459218362(case), Lot Number: 25DBI895; 559) DYKMBNDL209B, UDI-DI: 10198459218361(each), 40198459218362(case), Lot Number: 25CBI945; 560) DYKMBNDL209B, UDI-DI: 10198459218361(each), 40198459218362(case), Lot Number: 25BBK693; 561) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22OBL116; 562) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22OBK254; 563) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22IBT163; 564) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22IBC492; 565) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22HBQ336; 566) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22DBL978; 567) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22CBC663; 568) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22BBX263; 569) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22BBJ109; 570) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 22ABM690; 571) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21LBW419; 572) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21LBT988; 573) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21LBF033; 574) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21KBU495; 575) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21JBQ840; 576) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21HBO953; 577) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21GBJ356; 578) DYKMBNDL20A, UDI-DI: 10193489893861(each), 40193489893862(case), Lot Number: 21EBK885; 579) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 25GBW134; 580) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 24GBQ018; 581) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 24CBI595; 582) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 24ABO454; 583) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 24ABG854; 584) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23KBU641; 585) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23JBK348; 586) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23FBK698; 587) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23EBN795; 588) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23CBN839; 589) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 23ABK214; 590) DYKMBNDL20B, UDI-DI: 10195327240950(each), 40195327240951(case), Lot Number: 22JBJ135; 591) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23HBA622; 592) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23GBP601; 593) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23GBM665; 594) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23FBA474; 595) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23EBB668; 596) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23CBN840; 597) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23CBH186; 598) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23ABQ111; 599) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 23ABM977; 600) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22OBG255; 601) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22LBN127; 602) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22IBJ800; 603) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22HBU088; 604) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22HBU087; 605) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22GBZ220; 606) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22EBD893; 607) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22EBD883; 608) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22BBY940; 609) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22BBR580; 610) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 22BBM493; 611) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21LBX392; 612) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21JBQ919; 613) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21IBV208; 614) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21HBJ429; 615) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21GBN254; 616) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21GBB473; 617) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21DBV049; 618) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21CBS044; 619) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21CBG306; 620) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21ABO062; 621) DYKMBNDL21, UDI-DI: 10193489478693(each), 40193489478694(case), Lot Number: 21ABI052; 622) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 25ILA004; 623) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 25DLA429; 624) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 25BLA145; 625) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 25ALA464; 626) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24LLB030; 627) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24LLA606; 628) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24LLA319; 629) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24KLA966; 630) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24ILB092; 631) DYKMBNDL23A, UDI-DI: 10198459000522(each), 40198459000523(case), Lot Number: 24ELA529; 632) DYKMBNDL27A, UDI-DI: 10193489894592(each), 40193489894593(case), Lot Number: 21LBP526; 633) DYKMBNDL27A, UDI-DI: 10193489894592(each), 40193489894593(case), Lot Number: 21FBP769; 634) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22HBO327; 635) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22GBH499; 636) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22FBR367; 637) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22FBP614; 638) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22DBB582; 639) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22CBT350; 640) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22ABN985; 641) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 22ABI604; 642) DYKMBNDL27B, UDI-DI: 10195327059828(each), 40195327059829(case), Lot Number: 21LBW378; 643) DYKMBNDL47C, UDI-DI: 10195327264369(each), 40195327264360(case), Lot Number: 23HLA318; 644) DYKMBNDL47C, UDI-DI: 10195327264369(each), 40195327264360(case), Lot Number: 23HLA157; 645) DYKMBNDL47C, UDI-DI: 10195327264369(each), 40195327264360(case), Lot Number: 23BLA112; 646) DYKMBNDL47C, UDI-DI: 10195327264369(each), 40195327264360(case), Lot Number: 23BLA067; 647) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21IBF812; 648) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21HBF509; 649) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21FBP358; 650) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21DBT811; 651) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21CBH959; 652) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21CBF514; 653) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21BBM368; 654) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21BBE095; 655) DYKMBNDL59, UDI-DI: 10193489479263(each), 40193489479264(case), Lot Number: 21ABP200; 656) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 24CBE718; 657) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 23KBF076; 658) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 23KBA267; 659) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 23JBB706; 660) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 23IBO640; 661) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 23IBA613; 662) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22KBA942; 663) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22HBH513; 664) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22HBB980; 665) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22GBZ221; 666) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22FBR924; 667) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 22EBS814; 668) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21IBP937; 669) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21IBJ339; 670) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21HBL107; 671) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21GBK385; 672) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21FBP748; 673) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21DBJ865; 674) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21CBZ729; 675) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21CBH554; 676) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21CBE516; 677) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21BBG054; 678) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 21ABJ590; 679) DYKMBNDL62, UDI-DI: 10193489477344(each), 40193489477345(case), Lot Number: 20LBY970; 680) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 25HBA877; 681) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 25EBH355; 682) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 25DBQ010; 683) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 25BBO687; 684) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 25BBJ987; 685) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 24GBG844; 686) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 24GBF626; 687) DYKMBNDL62A, UDI-DI: 10198459000591(each), 40198459000592(case), Lot Number: 24FBL346; 688) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 22CLA947; 689) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 22CLA118; 690) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21LLA608; 691) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21KLA810; 692) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21JLA984; 693) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21JLA700; 694) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21DLA476; 695) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21CLA506; 696) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21BLA578; 697) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21ABJ868; 698) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 21ABH909; 699) DYKMBNDL63, UDI-DI: 10193489477665(each), 40193489477666(case), Lot Number: 20LBY938; 700) DYKMBNDL63A, UDI-DI: 10195327241131(each), 40195327241132(case), Lot Number: 23HLA425; 701) DYKMBNDL63A, UDI-DI: 10195327241131(each), 40195327241132(case), Lot Number: 23HLA229; 702) DYKMBNDL63A, UDI-DI: 10195327241131(each), 40195327241132(case), Lot Number: 23FLA828; 703) DYKMBNDL63A, UDI-DI: 10195327241131(each), 40195327241132(case), Lot Number: 23FLA214; 704) DYKMBNDL63A, UDI-DI: 10195327241131(each), 40195327241132(case), Lot Number: 23BLA021; 705) DYKMBNDL72A, UDI-DI: 10195327241162(each), 40195327241163(case), Lot Number: 23BBP960; 706) DYKMBNDL72A, UDI-DI: 10195327241162(each), 40195327241163(case), Lot Number: 23BBO772; 707) DYKMBNDL72A, UDI-DI: 10195327241162(each), 40195327241163(case), Lot Number: 22KBP432; 708) DYKMBNDL72A, UDI-DI: 10195327241162(each), 40195327241163(case), Lot Number: 22IBB259; 709) DYKMBNDL72A, UDI-DI: 10195327241162(each), 40195327241163(case), Lot Number: 22IBA961; 710) DYKMBNDL75A, UDI-DI: 10193489891379(each), 40193489891370(case), Lot Number: 21HBO553; 711) DYKMBNDL75A, UDI-DI: 10193489891379(each), 40193489891370(case), Lot Number: 21HBG642; 712) DYKMBNDL75A, UDI-DI: 10193489891379(each), 40193489891370(case), Lot Number: 21DBL930; 713) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21KBR303; 714) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21JBX708; 715) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21JBJ020; 716) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21JBA903; 717) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21GBP396; 718) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21GBE968; 719) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21FBR394; 720) DYKMBNDL83A, UDI-DI: 10193489893892(each), 40193489893893(case), Lot Number: 21FBB096; 721) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22IBC545; 722) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22HBU467; 723) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22FBT012; 724) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22EBD480; 725) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22CBD044; 726) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22BBR566; 727) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22ABZ390; 728) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 22ABI714; 729) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 21WBA903; 730) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 21LBU984; 731) DYKMBNDL83B, UDI-DI: 10195327050283(each), 40195327050284(case), Lot Number: 21KBU369; 732) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 25GBP929; 733) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 24FBC495; 734) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 24EBU705; 735) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23LBE775; 736) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23KBO528; 737) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23JBT912; 738) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23IBF578; 739) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23GBE627; 740) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23FBN705; 741) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23DBM062; 742) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23CBK438; 743) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 23ABH679; 744) DYKMBNDL83C, UDI-DI: 10195327241186(each), 40195327241187(case), Lot Number: 22IBU720; 745) DYNDA2526, UDI-DI: 10193489191790(each), 40193489191791(case), Lot Number: 24ABW725; 746) DYNJ0130723P, UDI-DI: 10195327000240(each), 40195327000241(case), Lot Number: 22DBD996; 747) DYNJ0321545R, UDI-DI: 10195327311773(each), 40195327311774(case), Lot Number: 24DMB031; 748) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23HMG775; 749) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23FMJ236; 750) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23FME226; 751) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23BMH207; 752) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23AMH832; 753) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 23AMG743; 754) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 22JMC538; 755) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 22IMD310; 756) DYNJ0349727D, UDI-DI: 10195327000158(each), 40195327000159(case), Lot Number: 22FMC741; 757) DYNJ0376608O, UDI-DI: 10195327192051(each), 40195327192052(case), Lot Number: 22JMA708; 758) DYNJ04516N, UDI-DI: 10888277744516(each), 40888277744517(case), Lot Number: 23DMG505; 759) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 25HBR258; 760) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 25EBM022; 761) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 25EBK297; 762) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 25CBG503; 763) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 25BBM272; 764) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 24FBS121; 765) DYNJ0537013AA, UDI-DI: 10195327167677(each), 40195327167678(case), Lot Number: 23CMC628; 766) DYNJ0545079V, UDI-DI: 10193489866278(each), 40193489866279(case), Lot Number: 23HMG638; 767) DYNJ0545079V, UDI-DI: 10193489866278(each), 40193489866279(case), Lot Number: 23HMD010; 768) DYNJ0826299T, UDI-DI: 10193489377989(each), 40193489377980(case), Lot Number: 23JBS475; 769) DYNJ0842825F, UDI-DI: 10195327398392(each), 40195327398393(case), Lot Number: 24ADA007; 770) DYNJ20888I, UDI-DI: 10193489335088(each), 40193489335089(case), Lot Number: 21DKA173; 771) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 23CMB466; 772) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 23BMD920; 773) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 23BMA230; 774) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 22JMF904; 775) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 22HMC821; 776) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 22DMH293; 777) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 22CMD166; 778) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 21LBI035; 779) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 21KBG504; 780) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 21IBP062; 781) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 21GBN644; 782) DYNJ20888J, UDI-DI: 10193489922387(each), 40193489922388(case), Lot Number: 21FBO028; 783) DYNJ22255R, UDI-DI: 10195327194246(each), 40195327194247(case), Lot Number: 24HBS926; 784) DYNJ24524N, UDI-DI: 10195327164829(each), 40195327164820(case), Lot Number: 24JLA043; 785) DYNJ24524N, UDI-DI: 10195327164829(each), 40195327164820(case), Lot Number: 23JLA017; 786) DYNJ27409U, UDI-DI: 10195327104450(each), 40195327104451(case), Lot Number: 23HBP691; 787) DYNJ28805G, UDI-DI: 10193489574326(each), 40193489574327(case), Lot Number: 23DBE702; 788) DYNJ32435I, UDI-DI: 10889942575176(each), 40889942575177(case), Lot Number: 21BBB532; 789) DYNJ33583B, UDI-DI: 10193489655025(each), 40193489655026(case), Lot Number: 21ADA099; 790) DYNJ36600F, UDI-DI: 10889942552566(each), 40889942552567(case), Lot Number: 21JBS052; 791) DYNJ38318F, UDI-DI: 10193489728514(each), 40193489728515(case), Lot Number: 22IME581; 792) DYNJ38396M, UDI-DI: 10195327373788(each), 40195327373789(case), Lot Number: 24FMA213; 793) DYNJ41556F, UDI-DI: 10193489968170(each), 40193489968171(case), Lot Number: 23IMH983; 794) DYNJ41835G, UDI-DI: 10195327049089(each), 40195327049080(case), Lot Number: 23ABB558; 795) DYNJ41839C, UDI-DI: 10195327049119(each), 40195327049110(case), Lot Number: 23ABI719; 796) DYNJ42921F, UDI-DI: 10193489351293(each), 40193489351294(case), Lot Number: 21BBI917; 797) DYNJ42921G, UDI-DI: 10193489981384(each), 40193489981385(case), Lot Number: 23CBD572; 798) DYNJ43399I, UDI-DI: 10195327209384(each), 40195327209385(case), Lot Number: 23CBU835; 799) DYNJ43414C, UDI-DI: 10193489953305(each), 40193489953306(case), Lot Number: 22BBX074; 800) DYNJ43414C, UDI-DI: 10193489953305(each), 40193489953306(case), Lot Number: 22ABU181; 801) DYNJ43414C, UDI-DI: 10193489953305(each), 40193489953306(case), Lot Number: 21LBU243; 802) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 23HMI407; 803) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 23HMF199; 804) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 23GME748; 805) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 23DMH121; 806) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 22KMC257; 807) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 22IMA070; 808) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 22GMA594; 809) DYNJ43414D, UDI-DI: 10195327136680(each), 40195327136681(case), Lot Number: 22EMB590; 810) DYNJ44727B, UDI-DI: 10193489396201(each), 40193489396202(case), Lot Number: 21ADA100; 811) DYNJ44727I, UDI-DI: 10195327625122(each), 40195327625123(case), Lot Number: 25GDA295; 812) DYNJ45146D, UDI-DI: 10195327329280(each), 40195327329281(case), Lot Number: 24FMB575; 813) DYNJ47684C, UDI-DI: 10195327491291(each), 40195327491292(case), Lot Number: 24BBG775; 814) DYNJ50478F, UDI-DI: 10193489493665(each), 40193489493666(case), Lot Number: 21HDA608; 815) DYNJ52363D, UDI-DI: 10195327183806(each), 40195327183807(case), Lot Number: 23EBR722; 816) DYNJ52474I, UDI-DI: 10195327274160(each), 40195327274161(case), Lot Number: 23IMF448; 817) DYNJ52954F, UDI-DI: 10193489976359(each), 40193489976350(case), Lot Number: 21LBC027; 818) DYNJ52954F, UDI-DI: 10193489976359(each), 40193489976350(case), Lot Number: 21KBM611; 819) DYNJ54969B, UDI-DI: 10198459435621(each), 40198459435622(case), Lot Number: 25IBJ616; 820) DYNJ55482F, UDI-DI: 10195327332273(each), 40195327332274(case), Lot Number: 23IBQ266; 821) DYNJ56089, UDI-DI: 10889942603343(each), 40889942603344(case), Lot Number: 23IBI638; 822) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21LDB911; 823) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21KDC120; 824) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21JDC499; 825) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21GDB160; 826) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21EDB243; 827) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21CDC611; 828) DYNJ56282B, UDI-DI: 10193489755329(each), 40193489755320(case), Lot Number: 21ADB595; 829) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 23DDA114; 830) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 23BDC168; 831) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 23ADB688; 832) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22LDA744; 833) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22KDA370; 834) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22HDB845; 835) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22GDB133; 836) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22EDA542; 837) DYNJ56282C, UDI-DI: 10195327093860(each), 40195327093861(case), Lot Number: 22CDB903; 838) DYNJ57597A, UDI-DI: 10195327131982(each), 40195327131983(case), Lot Number: 23IBF558; 839) DYNJ58425A, UDI-DI: 10193489607857(each), 40193489607858(case), Lot Number: 21ADB526; 840) DYNJ58499B, UDI-DI: 10193489817652(each), 40193489817653(case), Lot Number: 21LME012; 841) DYNJ58499B, UDI-DI: 10193489817652(each), 40193489817653(case), Lot Number: 21IMD693; 842) DYNJ58499B, UDI-DI: 10193489817652(each), 40193489817653(case), Lot Number: 21GME120; 843) DYNJ58499B, UDI-DI: 10193489817652(each), 40193489817653(case), Lot Number: 21FMC877; 844) DYNJ58499B, UDI-DI: 10193489817652(each), 40193489817653(case), Lot Number: 21CMC924; 845) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 24KMA127; 846) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 24IMC097; 847) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 24AMH616; 848) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 23FME537; 849) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 23DMI717; 850) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 23CMA172; 851) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 22KMD048; 852) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 22FMA864; 853) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 22EMA454; 854) DYNJ58499C, UDI-DI: 10195327090333(each), 40195327090334(case), Lot Number: 22BMG669; 855) DYNJ58861A, UDI-DI: 10193489749939(each), 40193489749930(case), Lot Number: 23GBG421; 856) DYNJ61124D, UDI-DI: 10195327561109(each), 40195327561100(case), Lot Number: 24DMC274; 857) DYNJ61169C, UDI-DI: 10195327606084(each), 40195327606085(case), Lot Number: 24BBV434; 858) DYNJ61212G, UDI-DI: 10195327090364(each), 40195327090365(case), Lot Number: 22HMD944; 859) DYNJ61212G, UDI-DI: 10195327090364(each), 40195327090365(case), Lot Number: 22EMA897; 860) DYNJ61984A, UDI-DI: 10193489747126(each), 40193489747127(case), Lot Number: 21ADA126; 861) DYNJ63075C, UDI-DI: 10195327102173(each), 40195327102174(case), Lot Number: 23AME711; 862) DYNJ63099C, UDI-DI: 10193489429091(each), 40193489429092(case), Lot Number: 21BBD822; 863) DYNJ63282C, UDI-DI: 10195327297091(each), 40195327297092(case), Lot Number: 24IBD031; 864) DYNJ64143B, UDI-DI: 10195327279585(each), 40195327279586(case), Lot Number: 23FBL851; 865) DYNJ64188A, UDI-DI: 10193489291063(each), 40193489291064(case), Lot Number: 22FBC048; 866) DYNJ64188A, UDI-DI: 10193489291063(each), 40193489291064(case), Lot Number: 22EBL053; 867) DYNJ64188A, UDI-DI: 10193489291063(each), 40193489291064(case), Lot Number: 22ABY741; 868) DYNJ64188A, UDI-DI: 10193489291063(each), 40193489291064(case), Lot Number: 21LBR909; 869) DYNJ64188A, UDI-DI: 10193489291063(each), 40193489291064(case), Lot Number: 21GBN579; 870) DYNJ64832B, UDI-DI: 10195327601546(each), 40195327601547(case), Lot Number: 24ILA258; 871) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24FMH619; 872) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24EMG861; 873) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24EMD877; 874) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24DMI886; 875) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24DMF522; 876) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 24AMI243; 877) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 23IMA649; 878) DYNJ64843A, UDI-DI: 10195327413118(each), 40195327413119(case), Lot Number: 23FMJ611; 879) DYNJ67442B, UDI-DI: 10195327133542(each), 40195327133543(case), Lot Number: 23HBQ817; 880) DYNJ80139A, UDI-DI: 10195327410773(each), 40195327410774(case), Lot Number: 23HME442; 881) DYNJ81119, UDI-DI: 10195327086503(each), 40195327086504(case), Lot Number: 23CMA914; 882) DYNJ81119, UDI-DI: 10195327086503(each), 40195327086504(case), Lot Number: 22CME785; 883) DYNJ81119A, UDI-DI: 10195327375133(each), 40195327375134(case), Lot Number: 23EMC719; 884) DYNJ81137, UDI-DI: 10195327087401(each), 40195327087402(case), Lot Number: 22CMG961; 885) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 24EBV261; 886) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 24EBL792; 887) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 24CBG996; 888) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 23LBP736; 889) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 23KBI274; 890) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 23IBV864; 891) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 23HBR577; 892) DYNJ81595B, UDI-DI: 10195327421793(each), 40195327421794(case), Lot Number: 23GBO643; 893) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 25HMB789; 894) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 25DMI426; 895) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 25BME563; 896) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 24LMA851; 897) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 24KMJ080; 898) DYNJ83088B, UDI-DI: 10198459190643(each), 40198459190644(case), Lot Number: 24KMI822; 899) DYNJ83719A, UDI-DI: 10195327420550(each), 40195327420551(case), Lot Number: 24DMF633; 900) DYNJ84091A, UDI-DI: 10195327443191(each), 40195327443192(case), Lot Number: 23KBG001; 901) DYNJ84091A, UDI-DI: 10195327443191(each), 40195327443192(case), Lot Number: 23KBF009; 902) DYNJ87479, UDI-DI: 10198459041587(each), 40198459041588(case), Lot Number: 24GBS099; 903) DYNJ87511A, UDI-DI: 10198459382550(each), 40198459382551(case), Lot Number: 25HBF150; 904) DYNJ87511A, UDI-DI: 10198459382550(each), 40198459382551(case), Lot Number: 25HBA633; 905) DYNJ88769, UDI-DI: 10198459202926(each), 40198459202927(case), Lot Number: 25IME191; 906) DYNJ88769, UDI-DI: 10198459202926(each), 40198459202927(case), Lot Number: 25EMA015; 907) DYNJ88769, UDI-DI: 10198459202926(each), 40198459202927(case), Lot Number: 24LMF078; 908) DYNJ900170C, UDI-DI: 10193489629569(each), 40193489629560(case), Lot Number: 21DDB333; 909) DYNJ900170C, UDI-DI: 10193489629569(each), 40193489629560(case), Lot Number: 21CDC595; 910) DYNJ900377O, UDI-DI: 10193489353235(each), 40193489353236(case), Lot Number: 20XBB301; 911) DYNJ901386B, UDI-DI: 10193489436617(each), 40193489436618(case), Lot Number: 23CLB141; 912) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 25CDA194; 913) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 25BDA907; 914) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 23JDA345; 915) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 23HDC577; 916) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 22HDA555; 917) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 21KDB349; 918) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 21JDB453; 919) DYNJ902586A, UDI-DI: 10193489985320(each), 40193489985321(case), Lot Number: 21IDC339; 920) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 25IBD297; 921) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 25DBU815; 922) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 25CBT340; 923) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 25BBE023; 924) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24LBK021; 925) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24KBR909; 926) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24HBM038; 927) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24GBI128; 928) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24EBD860; 929) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24CBS645; 930) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 24CBI435; 931) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 23JBR706; 932) DYNJ902778G, UDI-DI: 10195327462079(each), 40195327462070(case), Lot Number: 23JBE626; 933) DYNJ903524G, UDI-DI: 10195327625542(each), 40195327625543(case), Lot Number: 24HBH453; 934) DYNJ903524G, UDI-DI: 10195327625542(each), 40195327625543(case), Lot Number: 24DBQ829; 935) DYNJ903524G, UDI-DI: 10195327625542(each), 40195327625543(case), Lot Number: 24DBJ940; 936) DYNJ903524I, UDI-DI: 10198459113406(each), 40198459113407(case), Lot Number: 24JBG835; 937) DYNJ903524I, UDI-DI: 10198459113406(each), 40198459113407(case), Lot Number: 24JBG276; 938) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25GBN623; 939) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25FBR252; 940) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25EBO480; 941) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25DBS666; 942) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25DBL980; 943) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25CBD990; 944) DYNJ903524J, UDI-DI: 10198459181009(each), 40198459181000(case), Lot Number: 25BBH114; 945) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21JDA960; 946) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21IDC340; 947) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21HDA088; 948) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21GDA911; 949) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21FDC116; 950) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21FDA591; 951) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21DDC145; 952) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 21DDB310; 953) DYNJ903739D, UDI-DI: 10193489583687(each), 40193489583688(case), Lot Number: 20LDB197; 954) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23HDB853; 955) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23HDB009; 956) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23EDC472; 957) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23CDC565; 958) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23CDA726; 959) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 23ADB082; 960) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 22KDC140; 961) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 22EDB233; 962) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 22DDB442; 963) DYNJ903739G, UDI-DI: 10195327102012(each), 40195327102013(case), Lot Number: 22CDC192; 964) DYNJ904029B, UDI-DI: 10889942702350(each), 40889942702351(case), Lot Number: 23IBE598; 965) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 24EBB968; 966) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 24CBU879; 967) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 24BBQ536; 968) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23LBI661; 969) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23KBI004; 970) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23JBP553; 971) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23JBI204; 972) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23HBT882; 973) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23FBP149; 974) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23FBK961; 975) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23EBR735; 976) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 23CBL806; 977) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22LBB705; 978) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22KBE372; 979) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22IBR068; 980) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22IBE752; 981) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22HBI691; 982) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22HBH292; 983) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22GBS153; 984) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22DBL303; 985) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 22CBN885; 986) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21LBJ305; 987) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21GBL277; 988) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21GBD916; 989) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21FBG341; 990) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21EBO360; 991) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21CBY305; 992) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21CBT332; 993) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 21BBG581; 994) DYNJ904216C, UDI-DI: 10193489249927(each), 40193489249928(case), Lot Number: 20LBT629; 995) DYNJ905226D, UDI-DI: 10193489435375(each), 40193489435376(case), Lot Number: 21DBF379; 996) DYNJ905226G, UDI-DI: 10198459520631(each), 40198459520632(case), Lot Number: 25KBO679; 997) DYNJ905240D, UDI-DI: 10193489309348(each), 40193489309349(case), Lot Number: 21ABV430; 998) DYNJ905240D, UDI-DI: 10193489309348(each), 40193489309349(case), Lot Number: 21ABJ943; 999) DYNJ905240F, UDI-DI: 10193489825800(each), 40193489825801(case), Lot Number: 21FBL635; 1000) DYNJ905240F, UDI-DI: 10193489825800(each), 40193489825801(case), Lot Number: 21DBP339; 1001) DYNJ905285B, UDI-DI: 10193489635706(each), 40193489635707(case), Lot Number: 21CBU587; 1002) DYNJ905290B, UDI-DI: 10193489635775(each), 40193489635776(case), Lot Number: 21DBO193; 1003) DYNJ905748F, UDI-DI: 10195327544287(each), 40195327544288(case), Lot Number: 24HDB035; 1004) DYNJ905748F, UDI-DI: 10195327544287(each), 40195327544288(case), Lot Number: 24GDA066; 1005) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 25DLA006; 1006) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 25CLA904; 1007) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 24LLA602; 1008) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 24HLA810; 1009) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 24FLA072; 1010) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 24ALB054; 1011) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 23JLA832; 1012) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 23HLA336; 1013) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 23DLA451; 1014) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 23BLA005; 1015) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 22LLA418; 1016) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 22FLA761; 1017) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 22FLA126; 1018) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 22CLA919; 1019) DYNJ905980A, UDI-DI: 10195327054892(each), 40195327054893(case), Lot Number: 22BLA199; 1020) DYNJ906225F, UDI-DI: 10195327102197(each), 40195327102198(case), Lot Number: 22FDA421; 1021) DYNJ906225G, UDI-DI: 10195327209773(each), 40195327209774(case), Lot Number: 24ADA475; 1022) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 25IDA895; 1023) DYNJ906225I, UDI-DI: 10195327658328(each), 40195327658329(case), Lot Number: 24EDA698; 1024) DYNJ906346F, UDI-DI: 10195327624309(each), 40195327624300(case), Lot Number: 24HLB134; 1025) DYNJ906715D, UDI-DI: 10193489871975(each), 40193489871976(case), Lot Number: 22HDC485; 1026) DYNJ906715D, UDI-DI: 10193489871975(each), 40193489871976(case), Lot Number: 22FDA438; 1027) DYNJ906878A, UDI-DI: 10193489391305(each), 40193489391306(case), Lot Number: 23KLA328; 1028) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 24BMD704; 1029) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23HME502; 1030) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23GMC253; 1031) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23EMK082; 1032) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23EME402; 1033) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23EMC988; 1034) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23DMF809; 1035) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 23BMC096; 1036) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22LMD657; 1037) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22JMC898; 1038) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22HBI386; 1039) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22HBA481; 1040) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22GBZ037; 1041) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22GBM635; 1042) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22FBU679; 1043) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22DBN011; 1044) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 22ABS942; 1045) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21KBX807; 1046) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21JBQ089; 1047) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21JBH246; 1048) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21IBP639; 1049) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21HBS192; 1050) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21HBC973; 1051) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21EBC892; 1052) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21CBU163; 1053) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21BBC656; 1054) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21BBB739; 1055) DYNJ907004, UDI-DI: 10193489320091(each), 40193489320092(case), Lot Number: 21ABD399; 1056) DYNJ907040A, UDI-DI: 10193489354027(each), 40193489354028(case), Lot Number: 21CBR866; 1057) DYNJ907040A, UDI-DI: 10193489354027(each), 40193489354028(case), Lot Number: 21CBA219; 1058) DYNJ907857, UDI-DI: 10193489909364(each), 40193489909365(case), Lot Number: 22ABH627; 1059) DYNJ907857, UDI-DI: 10193489909364(each), 40193489909365(case), Lot Number: 21LBJ320; 1060) DYNJ907857, UDI-DI: 10193489909364(each), 40193489909365(case), Lot Number: 21IBX945; 1061) DYNJ907857, UDI-DI: 10193489909364(each), 40193489909365(case), Lot Number: 21GBJ159; 1062) DYNJ907857A, UDI-DI: 10195327133160(each), 40195327133161(case), Lot Number: 22KBM436; 1063) DYNJ907857A, UDI-DI: 10195327133160(each), 40195327133161(case), Lot Number: 22HBK592; 1064) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 23LBG088; 1065) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 23JBT943; 1066) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 23IBS352; 1067) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 23IBE277; 1068) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 23ABM669; 1069) DYNJ907857B, UDI-DI: 10195327256869(each), 40195327256860(case), Lot Number: 22KBN208; 1070) DYNJ907975K, UDI-DI: 101984591082 11(each), 40198459108212(case), Lot Number: 24JLA679; 1071) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22JLA218; 1072) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22HLB053; 1073) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22HLA747; 1074) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22GLA501; 1075) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22FLA101; 1076) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22DLB010; 1077) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22DLA244; 1078) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22CLA903; 1079) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22CLA382; 1080) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 22BLA456; 1081) DYNJ908010, UDI-DI: 10193489961225(each), 40193489961226(case), Lot Number: 21ILA224; 1082) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25GLA823; 1083) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25FLA561; 1084) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25FLA232; 1085) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25ELA641; 1086) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25ELA034; 1087) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 25ALA515; 1088) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24LLA983; 1089) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24LLA353; 1090) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24KLA621; 1091) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24KLA496; 1092) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24GLA795; 1093) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24FLA260; 1094) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24DLB004; 1095) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24BLA934; 1096) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24BLA565; 1097) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24ALA871; 1098) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 24ALA343; 1099) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23LLA191; 1100) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23KLA366; 1101) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23ILA968; 1102) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23ILA658; 1103) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23HLB306; 1104) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23HLA914; 1105) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23GLA437; 1106) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23FLB104; 1107) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23ELA535; 1108) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23ELA088; 1109) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 23CLA155; 1110) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 22LLA687; 1111) DYNJ908010A, UDI-DI: 10195327241360(each), 40195327241361(case), Lot Number: 22KLA529; 1112) DYNJ908121A, UDI-DI: 10195327000578(each), 40195327000579(case), Lot Number: 22OBH193; 1113) DYNJ908121A, UDI-DI: 10195327000578(each), 40195327000579(case), Lot Number: 21JBN851; 1114) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 24CBQ381; 1115) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 24ABB886; 1116) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23LBD780; 1117) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23JBH022; 1118) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23HBS186; 1119) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23EBN792; 1120) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23DBP233; 1121) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23DBE794; 1122) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 23CBU994; 1123) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 22JBF980; 1124) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 22IBK635; 1125) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 22EBK639; 1126) DYNJ908121B, UDI-DI: 10195327095710(each), 40195327095711(case), Lot Number: 22BBH995; 1127) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25JBM106; 1128) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25GBH817; 1129) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25DBN708; 1130) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25CBQ702; 1131) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 25ABC620; 1132) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24LBF355; 1133) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24KBO351; 1134) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24KBE893; 1135) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24JBT362; 1136) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24HBQ664; 1137) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24GBU277; 1138) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24FBS609; 1139) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24FBS600; 1140) DYNJ908121C, UDI-DI: 10198459024863(each), 40198459024864(case), Lot Number: 24FBF789; 1141) DYNJ908179A, UDI-DI: 10195327222574(each), 40195327222575(case), Lot Number: 23ABC799; 1142) DYNJ908179A, UDI-DI: 10195327222574(each), 40195327222575(case), Lot Number: 22IBI258; 1143) DYNJ908609A, UDI-DI: 10195327331184(each), 40195327331185(case), Lot Number: 23JBK170; 1144) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25JBN101; 1145) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25IBM035; 1146) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25HBK349; 1147) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25HBJ871; 1148) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25HBJ863; 1149) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25BBS264; 1150) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25BBD012; 1151) DYNJ908850A, UDI-DI: 10198459181146(each), 40198459181147(case), Lot Number: 25ABC093; 1152) DYNJ909428, UDI-DI: 10195327354190(each), 40195327354191(case), Lot Number: 23EBO817; 1153) DYNJ911360, UDI-DI: 10198459326356(each), 40198459326357(case), Lot Number: 25FBE175; 1154) DYNJ911563, UDI-DI: 10198459402203(each), 40198459402204(case), Lot Number: 25LBO018; 1155) DYNJ911563, UDI-DI: 10198459402203(each), 40198459402204(case), Lot Number: 25KBR482; 1156) DYNJ911563, UDI-DI: 10198459402203(each), 40198459402204(case), Lot Number: 25GBX731; 1157) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22OBA465; 1158) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22IBN892; 1159) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22EBG048; 1160) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22DBH954; 1161) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22ABZ357; 1162) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 22ABC909; 1163) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21JBA003; 1164) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21JBA002; 1165) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21GBU562; 1166) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21FBF477; 1167) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21EBA049; 1168) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 21CKA081; 1169) DYNJ9149605K, UDI-DI: 10193489290714(each), 40193489290715(case), Lot Number: 20LKB195; 1170) DYNJ9149605L, UDI-DI: 10195327235598(each), 40195327235599(case), Lot Number: 22LBJ936; 1171) DYNJ9149605N, UDI-DI: 10195327387136(each), 40195327387137(case), Lot Number: 23FBO505; 1172) DYNJ9149605N, UDI-DI: 10195327387136(each), 40195327387137(case), Lot Number: 23EBH576; 1173) DYNJ9149605N, UDI-DI: 10195327387136(each), 40195327387137(case), Lot Number: 23EBD511; 1174) DYNJ9149605O, UDI-DI: 10195327487102(each), 40195327487103(case), Lot Number: 23KBG171; 1175) DYNJ9149605O, UDI-DI: 10195327487102(each), 40195327487103(case), Lot Number: 23IBD866; 1176) DYNJC2398J, UDI-DI: 10195327131463(each), 40195327131464(case), Lot Number: 23CBL333; 1177) DYNJT1521P, UDI-DI: 10193489263435(each), 40193489263436(case), Lot Number: 23JLA824; 1178) DYNJT3339, UDI-DI: 10198459323577(each), 40198459323578(case), Lot Number: 25EBJ793; 1179) DYNJT3499A, UDI-DI: 10198459377495(each), 40198459377496(case), Lot Number: 25IBV606; 1180) DYNJT6288, UDI-DI: 10198459544255(each), 40198459544256(case), Lot Number: 25JMH882; 1181) DYNJVB91064A, UDI-DI: 10193489996296(each), 40193489996297(case), Lot Number: 22HDA342; 1182) PHS396113012C, UDI-DI: 10193489242683(each), 40193489242684(case), Lot Number: 20LDC155; 1183) PHS807323D, UDI-DI: 10193489927047(each), 40193489927048(case), Lot Number: 23EBJ429; 1184) SA-1090C, UDI-DI: 10080196115563(each), 40080196115564(case), Lot Number: 21LBM246

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1390-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 7, 2026
ReportedFebruary 25, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity39315 units

More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026

Official FDA record