Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934; . Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number: DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number: DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number: DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number: DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number: DYNDH1934; 6) BREAST BIOPSY PACK, Model Number: DYNJ17489I; 7) VOR PACK-LF, Model Number: DYNJ48539; 8) VOR PACK-LF, Model Number: DYNJ48539A; 9) IR BIOPSY PACK, Model Number: DYNJ59897B; 10) BREAST PACK, Model Number: DYNJ82998
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide.
How to tell if yours is affected
1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 23ABK216; 2) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBU924; 3) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 4) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBL439; 5) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBC364; 6) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23ABU170; 7) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBM302; 8) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBK913; 9) DYNDH1143C, UDI-DI: 10193489193336(each), 40193489193337(case), Lot Number: 22GBZ268; 10) DYNDH1873, UDI-DI: 10195327486464(each), 40195327486465(case), Lot Number: 23KBK476; 11) DYNDH1934, UDI-DI: 10195327519728(each), 40195327519729(case), Lot Number: 24FMF173; 12) DYNJ17489I, UDI-DI: 10195327224783(each), 40195327224784(case), Lot Number: 23CBL337; 13) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24GBD309; 14) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24EBQ366; 15) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24DBK656; 16) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24ABL751; 17) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23KBD033; 18) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23HBR688; 19) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23GBI270; 20) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23DMD933; 21) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23AMC919; 22) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22EMD675; 23) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22BMH767; 24) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 21JMC430; 25) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LMF497; 26) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LME482; 27) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25JBA908; 28) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25IBA545; 29) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25HBK043; 30) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25EBA588; 31) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 24IBT900; 32) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23EBQ760; 33) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 22LBH614; 34) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22LMF004; 35) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22JMG410
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1392-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | February 25, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 2740 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026