Medline Convenience Kits: 1) KIT UROLOGY SLING recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) KIT UROLOGY SLING, Model Number: DYKMBNDL106; 2) KIT UROLOGY SLING, Model Number: DYKMBNDL106A; 3) KIT UROLOGY PENILE REPAIR, Model Number: DYKMBNDL188; 4) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189; 5) KIT UROLOGY PERIPHERAL STIMULA, Model Number: DYKMBNDL189A
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide.
How to tell if yours is affected
1) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21BBG085; 2) DYKMBNDL106, UDI-DI: 10193489479294(each), 40193489479295(case), Lot Number: 21ABJ542; 3) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22CLA049; 4) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22BLB059; 5) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 22ALA713; 6) DYKMBNDL106A, UDI-DI: 10193489880816(each), 40193489880817(case), Lot Number: 21KLA631; 7) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21CBS043; 8) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS540; 9) DYKMBNDL188, UDI-DI: 10193489478549(each), 40193489478540(case), Lot Number: 21ABS539; 10) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21BBG092; 11) DYKMBNDL189, UDI-DI: 10193489478457(each), 40193489478458(case), Lot Number: 21ABI996; 12) DYKMBNDL189A, UDI-DI: 10193489880823(each), 40193489880824(case), Lot Number: 21JLA521
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1416-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | February 25, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 59 units |