Medline Convenience Kits: 1) SPINAL CDS recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CD. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W; 10) TOTAL HIP PROCEDURE, Model Number: CDS980865X; 11) TOTAL HIP PROCEDURE, Model Number: CDS980865Y; 12) TOTAL KNEE CDS, Model Number: CDS981454K; 13) TOTAL KNEE CDS, Model Number: CDS981454L; 14) KNEE ARTHROSCOPY CDS, Model Number: CDS981502K; 15) TOTAL KNEE, Model Number: CDS981505O; 16) TOTAL KNEE, Model Number: CDS981505P; 17) TOTAL KNEE, Model Number: CDS981505Q; 18) TOTAL KNEE, Model Number: CDS981505R; 19) OR LAMINECTOMY CDS, Model Number: CDS981858F; 20) LAMI FUSION, Model Number: CDS982035W; 21) ARMC TOTAL KNEE CDS, Model Number: CDS982281O; 22) ARMC TOTAL KNEE CDS, Model Number: CDS982281P; 23) ARMC TOTAL KNEE CDS, Model Number: CDS982281S; 24) TOTAL HIP CDS, Model Number: CDS983036K; 25) TOTAL KNEE CDS, Model Number: CDS983042K; 26) LOWER EXTREMITY CDS, Model Number: CDS983049K; 27) ZALE SPINE CDS, Model Number: CDS983182G; 28) TOTAL KNEE CDS, Model Number: CDS983297M; 29) SITTING CERVICAL CDS, Model Number: CDS983480G; 30) SHOULDER CDS, Model Number: CDS983581K; 31) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983838G; 32) TOTAL HIP CDS, Model Number: CDS983840I; 33) TOTAL JOINT CDS-LF, Model Number: CDS983863D; 34) SHOULDER ARTHROSCOPY CDS, Model Number: CDS983879C; 35) TOTAL JOINT, Model Number: CDS983903D; 36) GB TOTAL KNEE, Model Number: CDS984147K; 37) GB LOWER EXTREMITY, Model Number: CDS984156L; 38) LOWER EXTREMTIY, Model Number: CDS984254K; 39) LOWER EXTREMTIY, Model Number: CDS984254L; 40) LOWER EXTREMTIY, Model Number: CDS984254M; 41) LAMI MICRODISC-LF, Model Number: CDS984527I; 42) EXTREMITY CDS-LF, Model Number: CDS984607G; 43) EXTREMITY CDS-LF, Model Number: CDS984607I; 44) HANA/FRACTURE TABLE CDS-LF, Model Number: CDS984610C; 45) TOTAL JOINT CDS, Model Number: CDS985202B; 46) TOTAL JOINT CDS, Model Number: CDS985202C; 47) TRAUMA-LF, Model Number: CDS985279K; 48) ARTHROSCOPY CDS, Model Number: CDS985316A; 49) EXTREMITY CDS, Model Number: CDS985317C; 50) EXTREMITY CDS, Model Number: CDS985317D; 51) EXTREMITY CDS, Model Number: CDS985317F; 52) EXTREMITY CDS, Model Number: CDS985317G; 53) KNEE ARTHROSCOPY CDS, Model Number: CDS985425J; 54) TOTAL HIP PACK (CDSOPM834)-LF, Model Number: CDSOPM834W; 55) JOINT ASPIRATION KIT, Model Number: DYKM1161; 56) TOTAL KNEE SUPPLEMENT PACK NS, Model Number: DYKM2106; 57) KIT ORTHRO TOTAL HIP DR. NANDI, Model Number: DYKMBNDL137; 58) KIT NEUR FUS LUM POST CRANDALL, Model Number: DYKMBNDL138A; 59) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141; 60) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141A; 61) KIT NEURO CRANIOTOMYANEURYSM R, Model Number: DYKMBNDL141B; 62) KIT NEURO FUS LUM POST SANSUR, Model Number: DYKMBNDL142C; 63) KIT NEURO CRANIOTOMY ACOUSTIC, Model Number: DYKMBNDL143; 64) KIT NEURO CRANIOTOMY FOR TUMOR, Model Number: DYKMBNDL144; 65
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide.
How to tell if yours is affected
1) CDS940087AG, UDI-DI: 10193489847260(each), 40193489847261(case), Lot Number: 21EBC900; 2) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22GBE442; 3) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBG548; 4) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBF695; 5) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBA274; 6) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22BBY060; 7) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21KBH163; 8) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21JBF936; 9) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21IBJ260; 10) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21GBI183; 11) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21FBS254; 12) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21BBL223; 13) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21LBO219; 14) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBQ650; 15) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBB625; 16) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBV695; 17) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBU197; 18) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21GBR895; 19) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23JBI054; 20) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23GBP400; 21) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23EBW653; 22) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23DBR887; 23) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23CBG614; 24) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBK758; 25) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBJ844; 26) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBK666; 27) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBI093; 28) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22JBH355; 29) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22HBS531; 30) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBM598; 31) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBL549; 32) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK859; 33) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK858; 34) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBL968; 35) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBC616; 36) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21KBF393; 37) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21JBO077; 38) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBV804; 39) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBB808; 40) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23EBK295; 41) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23CBT556; 42) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23BBP326; 43) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABH483; 44) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABG351; 45) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22IBE748; 46) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22GBS173; 47) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22GBS172; 48) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22FBD583; 49) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22DBN516; 50) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 21LBW480; 51) CDS980865V, UDI-DI: 10193489304206(each), 40193489304207(case), Lot Number: 21FBH003; 52) CDS980865V, UDI-DI: 10193489304206(each), 40193489304207(case), Lot Number: 21ABX303; 53) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22OBL581; 54) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22IMH323; 55) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22GBM603; 56) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22FBO257; 57) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22DBD515; 58) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21VBC540; 59) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21JBA989; 60) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21JBA988; 61) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21IBP619; 62) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23FMI799; 63) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23EME858; 64) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23DMF336; 65) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23DMB940; 66) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23CMA547; 67) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24IMH073; 68) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24HMH806; 69) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24FMG472; 70) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24EMC301; 71) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24CMC328; 72) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24AMG267; 73) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23KMH550; 74) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23IMA365; 75) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23HMF744; 76) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24DMH818; 77) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24CMC092; 78) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24BMF152; 79) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24AMB397; 80) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23FME101; 81) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23EMG643; 82) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23DMH779; 83) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23CMF017; 84) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23CMB403; 85) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 25BMJ035; 86) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 25AMJ535; 87) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24LMC844; 88) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24KMB678; 89) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24JMB660; 90) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24GMI703; 91) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 24CMF795; 92) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 23JMI916; 93) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22KBM385; 94) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22JBK515; 95) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22IBJ349; 96) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22HBQ624; 97) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22HBN671; 98) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22FBS293; 99) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22ABS105; 100) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 24BBH054; 101) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 24ABU182; 102) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23KBQ766; 103) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23KBQ737; 104) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23GBG193; 105) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23DBL133; 106) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23CBU775; 107) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23CBI326; 108) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23ABG957; 109) CDS981505Q, UDI-DI: 10198459028991(each), 40198459028992(case), Lot Number: 24IBD132; 110) CDS981505Q, UDI-DI: 10198459028991(each), 40198459028992(case), Lot Number: 24GBJ138; 111) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 25DBO814; 112) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 25BBD996; 113) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 24LBR473; 114) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 24KBP956; 115) CDS981858F, UDI-DI: 10889942973682(each), 40889942973683(case), Lot Number: 21ABH636; 116) CDS982035W, UDI-DI: 10193489999259(each), 40193489999250(case), Lot Number: 22HBP524; 117) CDS982035W, UDI-DI: 10193489999259(each), 40193489999250(case), Lot Number: 22GBI202; 118) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 23AMB059; 119) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 22LMB455; 120) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 22CME430; 121) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 25BMC093; 122) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24FMA426; 123) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24EMD974; 124) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24CMB637; 125) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24AMD955; 126) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 23IMF292; 127) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 23HME111; 128) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25LME952; 129) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25LMD470; 130) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25KMJ877; 131) CDS983036K, UDI-DI: 10193489842012(each), 40193489842013(case), Lot Number: 23KBN289; 132) CDS983042K, UDI-DI: 10193489842029(each), 40193489842020(case), Lot Number: 23KBN273; 133) CDS983049K, UDI-DI: 10193489841657(each), 40193489841658(case), Lot Number: 23JBY028; 134) CDS983049K, UDI-DI: 10193489841657(each), 40193489841658(case), Lot Number: 23JBX571; 135) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21JBY974; 136) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21JBD011; 137) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21HBK744; 138) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21DBH504; 139) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21CBT358; 140) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 20LBW136; 141) CDS983297M, UDI-DI: 10193489425857(each), 40193489425858(case), Lot Number: 21FDA407; 142) CDS983480G, UDI-DI: 10195327497842(each), 40195327497843(case), Lot Number: 24CDB716; 143) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 25BMI188; 144) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 24CMA539; 145) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 24AME511; 146) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 23IMA904; 147) CDS983838G, UDI-DI: 10193489600360(each), 40193489600361(case), Lot Number: 21ADB812; 148) CDS983840I, UDI-DI: 10193489405668(each), 40193489405669(case), Lot Number: 21FDA409; 149) CDS983863D, UDI-DI: 10889942570980(each), 40889942570981(case), Lot Number: 22KDB188; 150) CDS983863D, UDI-DI: 10889942570980(each), 40889942570981(case), Lot Number: 22JDB465; 151) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 23EBK642; 152) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 23ABM823; 153) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22LBO557; 154) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22KBS560; 155) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22HBT635; 156) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22EBV412; 157) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22ABS153; 158) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22ABP297; 159) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21JBK128; 160) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21IBA056; 161) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21DBU793; 162) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21CBF370; 163) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21CBC111; 164) CDS983903D, UDI-DI: 10198459053474(each), 40198459053475(case), Lot Number: 24HBQ596; 165) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21EBR750; 166) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21DBI790; 167) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21CBJ921; 168) CDS984156L, UDI-DI: 10193489245370(each), 40193489245371(case), Lot Number: 21DBQ627; 169) CDS984156L, UDI-DI: 10193489245370(each), 40193489245371(case), Lot Number: 21CBJ888; 170) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22KBP500; 171) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22HBP143; 172) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22GBO260; 173) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22EBJ581; 174) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22CBW600; 175) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22BBX356; 176) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21KBY807; 177) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21HBM083; 178) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21GBU901; 179) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21FBL924; 180) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21CBD679; 181) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23EBJ573; 182) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23DBD812; 183) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23CBR791; 184) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23BBS246; 185) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23ABK592; 186) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25KBK734; 187) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25IBW208; 188) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25GBG607; 189) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24KBJ956; 190) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24IBP093; 191) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24IBB732; 192) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24GBU116; 193) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24GBB449; 194) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24EBF862; 195) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24DBS099; 196) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24BBS781; 197) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24ABT691; 198) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 23JBT902; 199) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 23HBV822; 200) CDS984527I, UDI-DI: 10193489376739(each), 40193489376730(case), Lot Number: 21BDA355; 201) CDS984607G, UDI-DI: 10889942719389(each), 40889942719380(case), Lot Number: 21EBC029; 202) CDS984607G, UDI-DI: 10889942719389(each), 40889942719380(case), Lot Number: 21ABV551; 203) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23DBK552; 204) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23BBU944; 205) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23ABE701; 206) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23ABA744; 207) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22JBQ863; 208) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22JBH295; 209) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22IBP797; 210) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22IBO809; 211) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22HBJ382; 212) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22GBF580; 213) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBS549; 214) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBR470; 215) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBQ393; 216) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22CBP419; 217) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21VBC530; 218) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21LBE361; 219) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21JBM719; 220) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21IBK997; 221) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 23BBG636; 222) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 23ABF816; 223) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22OBI284; 224) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22KBB895; 225) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22GBU468; 226) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22FBK619; 227) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22BBL459; 228) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 21LBQ977; 229) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 21IBV618; 230) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 24BBB251; 231) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23LBV047; 232) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23LBI902; 233) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23KBB103; 234) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23HBH071; 235) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23FBV622; 236) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23FBE479; 237) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23BBS726; 238) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 25BBD069; 239) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24LBI139; 240) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24JBU521; 241) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24JBM801; 242) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24EBR586; 243) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24DBD819; 244) CDS985279K, UDI-DI: 10195327497460(each), 40195327497461(case), Lot Number: 24CDA763; 245) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23GBT765; 246) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23DBD828; 247) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23CBP302; 248) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23ABC226; 249) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22NBE575; 250) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22LBT331; 251) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22JBR384; 252) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22GBU431; 253) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBQ646; 254) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBM692; 255) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBB121; 256) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21LBI665; 257) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21LBH556; 258) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21GBL257; 259) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21CBM829; 260) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21ABP208; 261) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24IBU020; 262) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24GBW059; 263) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24FBN572; 264) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24DBF304; 265) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24BBO443; 266) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24ABT609; 267) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 23JBS895; 268) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 23IBA494; 269) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 25CBV211; 270) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 25BBD080; 271) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 24LBD251; 272) CDS985317F, UDI-DI: 10198459309311(each), 40198459309312(case), Lot Number: 25HBD051; 273) CDS985317F, UDI-DI: 10198459309311(each), 40198459309312(case), Lot Number: 25FBK413; 274) CDS985317G, UDI-DI: 10198459509063(each), 40198459509064(case), Lot Number: 25LBC622; 275) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 23ABE885; 276) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 22KBC308; 277) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 21CBU636; 278) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 21CBH306; 279) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25KDA137; 280) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25GDA910; 281) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25DDB938; 282) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25ADB448; 283) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24IDA344; 284) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24HDA224; 285) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24ADA064; 286) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23JDB638; 287) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23CDB828; 288) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23BDB741; 289) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23ADB631; 290) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22JDB343; 291) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22FDB552; 292) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22EDA009; 293) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21HDC208; 294) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21ADB694; 295) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21ADA639; 296) DYKM1161, UDI-DI: 10888277838529(each), 40888277838520(case), Lot Number: 23ABS895; 297) DYKM2106, UDI-DI: 10195327020736(each), 40195327020737(case), Lot Number: 22FBT157; 298) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 25CBT503; 299) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24HBH168; 300) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24GBC828; 301) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24FBB726; 302) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24BBJ839; 303) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24ABH888; 304) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22KBO875; 305) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22JBP430; 306) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22IBS840; 307) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22FBI537; 308) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22FBF302; 309) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22DBM212; 310) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22CBB535; 311) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21HBG593; 312) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21HBF503; 313) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21EBE678; 314) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21ABM611; 315) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25GBP632; 316) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25FBU008; 317) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25FBK659; 318) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25EBK537; 319) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25BBI466; 320) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25ABS680; 321) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25ABQ126; 322) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24KBQ931; 323) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24IBE136; 324) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24GBV669; 325) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24GBH578; 326) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23LBO287; 327) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23KBV904; 328) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23JBB704; 329) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23HBY495; 330) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23HBH773; 331) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23GBJ242; 332) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23FBE416; 333) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23EBL742; 334) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23DBL148; 335) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23ABK726; 336) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 22JBG015; 337) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 22JBF639; 338) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22HBQ237; 339) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22FBS797; 340) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBS811; 341) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBQ939; 342) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBQ935; 343) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBF963; 344) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22CBV519; 345) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22BBR667; 346) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22ABL303; 347) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21VBB503; 348) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21IBK151; 349) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21IBJ315; 350) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21GBO489; 351) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21GBJ360; 352) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21FBF908; 353) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21EBM705; 354) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBJ601; 355) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBF326; 356) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBB270; 357) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21BBF813; 358) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21ABL153; 359) DYKMBNDL141A, UDI-DI: 1019532724776 8(each), 40195327247769(case), Lot Number: 23LBA277; 360) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23JBK019; 361) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23IBA610; 362) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23GBP488; 363) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23FBI710; 364) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22LBL868; 365) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22KBD324; 366) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22JBD221; 367) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22IBK345; 368) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25HBA792; 369) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25GBF719; 370) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25FBQ856; 371) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25EBR525; 372) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25DBI356; 373) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25CBI784; 374) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25BBS248; 375) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25ABL332; 376) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24JBF929; 377) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBV609; 378) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBG682; 379) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBA088; 380) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24GBK949; 381) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25LBL432; 382) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25FBC129; 383) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25EBN171; 384) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25BBH467; 385) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21CBT399; 386) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21CBS017; 387) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21BBN599; 388) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21ABU298; 389) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22HLA742; 390) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22FLA327; 391) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22ELB110; 392) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22DLA896; 393) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22DLA727; 394) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22CLA771; 395) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22BLA947; 396) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21LLA651; 397) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21LLA431; 398) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21KLB158; 399) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21JLA461; 400) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ILA240; 401) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21HLA660; 402) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21HLA178; 403) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21GLA871; 404) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21GLA640; 405) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21FLA052; 406) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21DLA534; 407) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21DLA128; 408) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA448; 409) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA414; 410) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA295; 411) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CBH443; 412) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21BLA414; 413) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ABU821; 414) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ABI784; 415) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23GLA491; 416) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23FLA815; 417) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23DLA832; 418) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23BLA654; 419) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22LLA124; 420) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22JLB035; 421) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22ILA386; 422) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25HLA637; 423) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25GLA271; 424) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25FLA991; 425) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25DLA246; 426) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25CLA509; 427) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25ALA559; 428) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24LLA520; 429) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24KLA175; 430) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ILA591; 431) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24HLA916; 432) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24GLA978; 433) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24FLA724; 434) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24DLA205; 435) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALB064; 436) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALA759; 437) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALA066; 438) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 23JLB143; 439) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 23HLA606; 440) DYKMBNDL145A, UDI-DI: 10193489846454(each), 40193489846455(case), Lot Number: 21DBK033; 441) DYKMBNDL145A, UDI-DI: 10193489846454(each), 40193489846455(case), Lot Number: 21CBF520; 442) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLB013; 443) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLA750; 444) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLA145; 445) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22KLA317; 446) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22ILB139; 447) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22ILA954; 448) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22FLA364; 449) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22DLA064; 450) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22CLA549; 451) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22BLB218; 452) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21LLA721; 453) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21KLA020; 454) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21JLA429; 455) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21JLA153; 456) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ILA264; 457) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21HLA172; 458) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21GLB130; 459) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21GLB011; 460) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ELA429; 461) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21DLA040; 462) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21CLA198; 463) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ALA530; 464) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22EBH424; 465) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22CBZ658; 466) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBR619; 467) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBA086; 468) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBA085; 469) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22ABD080; 470) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21JBM753; 471) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21FBR384; 472) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21EBK914; 473) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21ABI979; 474) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23KBJ631; 475) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBY496; 476) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBN659; 477) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBH774; 478) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBH227; 479) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23ABV362; 480) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 22BBN302; 481) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 22BBA226; 482) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21JBQ867; 483) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21GBR322; 484) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21CBM832; 485) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21CBE515; 486) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21ABL344; 487) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 22GBJ697; 488) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 21ABM598; 489) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 21ABM596; 490) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 24JBH678; 491) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 24CBA390; 492) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 22KBA921; 493) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25LBB412; 494) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25FBF593; 495) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBR608; 496) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBA208; 497) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBA179; 498) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBU036; 499) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBT803; 500) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBE721; 501) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24GBV124; 502) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24EBE950; 503) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBM619; 504) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBK954; 505) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBJ797; 506) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24ABJ776; 507) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23KBX885; 508) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23JBY280; 509) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23IBK826; 510) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23GBJ244; 511) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23GBE575; 512) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23EBC184; 513) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23EBB390; 514) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23ABK213; 515) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22KBR441; 516) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22JBO386; 517) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22HBM469; 518) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22GBO583; 519) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22FBX130; 520) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22EBQ040; 521) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22EBH625; 522) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22DBP875; 523) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22CBT349; 524) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21JBM262; 525) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21HBP647; 526) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21HBD948; 527) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21GBG496; 528) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBR810; 529) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBO357; 530) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBC342; 531) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21BBK654; 532) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21ABJ952; 533) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24IBB922; 534) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24DBB757; 535) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24CBB397; 536) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24BBO524; 537) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23KBA330; 538) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23IBV210; 539) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23FBA473; 540) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23CBS133; 541) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23ABK704; 542) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23ABK212; 543) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 22JBW065; 544) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 22JBV648; 545) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22HBA912; 546) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22GBW967; 547) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22GBG264; 548) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22FBH375; 549) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22ABL294; 550) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22ABB447; 551) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21HBK339; 552) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21DBI983; 553) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21CBM818; 554) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21ABP202; 555) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21ABI053; 556) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 20LBY961; 557) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25GBD166; 558) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25GBA035; 559) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 24EBB702; 560) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 23HBW431; 561) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 22JBM009; 562) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 22JBD779; 563) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22HBQ415; 564) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22GBV355; 565) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22ABO110; 566) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22ABL305; 567) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21KBA483; 568) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21GBO491; 569) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21DBH744; 570) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21CBZ652; 571) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21ABR425; 572) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 23HBD671; 573) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22JBW781; 574) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22JBV591; 575) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22IBD939; 576) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 22IBC493; 577) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 22HBS997; 578) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21JBJ745; 579) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21GBR325; 580) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21GBR317; 581) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21DBJ864; 582) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21BBH640; 583) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21BBF814; 584) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21ABJ588; 585) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 24HBS364; 586) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 24HBA089; 587) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23LBN530; 588) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23KBK074; 589) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23EBT501; 590) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 22KBB454; 591) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 22JBV772; 592) DYKMBNDL41B, UDI-DI: 10198459194221(each), 40198459194222(case), Lot Number: 25FBA949; 593) DYKMBNDL41B, UDI-DI: 10198459194221(each), 40198459194222(case), Lot Number: 24KBO741; 594) DYKMBNDL44, UDI-DI: 10193489478907(each), 40193489478908(case), Lot Number: 21BBB228; 595) DYKMBNDL44, UDI-DI: 10193489478907(each), 40193489478908(case), Lot Number: 21ABI958; 596) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22HBN316; 597) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22HBB977; 598) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22GBW926; 599) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22GBS120; 600) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22FBS026; 601) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22EBH425; 602) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22DBS009; 603) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22CBZ659; 604) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22CBP632; 605) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21LBL111; 606) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21KBC792; 607) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21JBZ344; 608) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21IBL058; 609) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21HBC599; 610) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21FBP768; 611) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21EBH613; 612) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21CBM521; 613) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24KBH583; 614) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBR177; 615) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBH752; 616) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBE585; 617) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24IBT870; 618) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24HBA816; 619) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24FBL254; 620) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24FBG799; 621) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24CBU335; 622) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23LBD748; 623) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23KBA266; 624) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23JBN677; 625) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23IBP913; 626) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23DBM061; 627) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23DBD291; 628) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23ABP556; 629) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 22JBM605; 630) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 22JBC209; 631) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21IBO778; 632) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21HBE131; 633) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21GBC744; 634) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21GBA955; 635) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21FBH155; 636) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21DBG619; 637) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBS659; 638) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBM480; 639) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBF517; 640) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21BBN603; 641) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21ABJ546; 642) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21ABC436; 643) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 20LBK848; 644) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22HBK580; 645) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22HBB979; 646) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22GBW927; 647) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22GBG265; 648) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22EBV520; 649) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22DBN318; 650) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22DBB583; 651) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22CBB053; 652) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22ABL284; 653) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22ABK146; 654) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21LBL251; 655) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21JBX804; 656) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21IBO769; 657) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21HBK334; 658) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21GBU931; 659) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21GBN132; 660) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21FBO536; 661) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21FBF707; 662) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21DBL007; 663) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21BBP819; 664) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21ABT642; 665) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21ABI788; 666) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25HBE319; 667) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25GBY473; 668) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25CBU771; 669) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25ABH726; 670) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24LBR198; 671) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24LBH373; 672) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24HBT009; 673) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24EBU783; 674) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23JBK351; 675) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23CBP087; 676) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23CBM230; 677) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23BBS867; 678) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 22LBL894; 679) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 22JBW403; 680) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22OBI934; 681) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22GBQ603; 682) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22FBC970; 683) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22EBA077; 684) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22DBM671; 685) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22DBE579; 686) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22CBX205; 687) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22ABG023; 688) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21LBD743; 689) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21KBC488; 690) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21EBG521; 691) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21CBD336; 692) DYNJ0130759J, UDI-DI: 10195327001438(each), 40195327001439(case), Lot Number: 23BMG804; 693) DYNJ0151234D, UDI-DI: 10193489207439(each), 40193489207430(case), Lot Number: 21ADA115; 694) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21GBF067; 695) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21DKA745; 696) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21DKA380; 697) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21CKA772; 698) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 20LKB286; 699) DYNJ0192492, UDI-DI: 10889942053728(each), 40889942053729(case), Lot Number: 21ADB354; 700) DYNJ0395166K, UDI-DI: 10195327292942(each), 40195327292943(case), Lot Number: 23DBE442; 701) DYNJ0395879J, UDI-DI: 10193489220261(each), 40193489220262(case), Lot Number: 21ADB355; 702) DYNJ0395959T, UDI-DI: 10195327206314(each), 40195327206315(case), Lot Number: 23JDA178; 703) DYNJ0415915K, UDI-DI: 10193489803105(each), 40193489803106(case), Lot Number: 23CMA970; 704) DYNJ0537209Y, UDI-DI: 10195327418229(each), 40195327418220(case), Lot Number: 23IBO050; 705) DYNJ0665273F, UDI-DI: 10193489394931(each), 40193489394932(case), Lot Number: 23EBS696; 706) DYNJ0773815M, UDI-DI: 10193489270976(each), 40193489270977(case), Lot Number: 23JBU875; 707) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24CBN277; 708) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 23LBC731; 709) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 23HMH357; 710) DYNJ0843201G, UDI-DI: 10193489345421(each), 40193489345422(case), Lot Number: 21ADA098; 711) DYNJ15364J, UDI-DI: 10193489707557(each), 40193489707558(case), Lot Number: 21BBD339; 712) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 21EMD491; 713) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 21CMC537; 714) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 20LMF531; 715) DYNJ17452S, UDI-DI: 10193489448726(each), 40193489448727(case), Lot Number: 23CBV576; 716) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23FMF588; 717) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23DMF905; 718) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23AMD537; 719) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22LMA654; 720) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22IMG491; 721) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22GMG472; 722) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22EME207; 723) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22DMA855; 724) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22BMB087; 725) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22AMA539; 726) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21JMH116; 727) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21IMG131; 728) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21HME696; 729) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21GMF267; 730) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21FMB644; 731) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21CMB012; 732) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21BMA742; 733) DYNJ19733C, UDI-DI: 10889942757794(each), 40889942757795(case), Lot Number: 22HMD917; 734) DYNJ19993O, UDI-DI: 10889942854110(each), 40889942854111(case), Lot Number: 23HBI690; 735) DYNJ21811I, UDI-DI: 10889942525959(each), 40889942525950(case), Lot Number: 23BBR769; 736) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21EBD225; 737) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21CBY834; 738) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21BBJ353; 739) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24JBS253; 740) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24IBA042; 741) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24GBQ410; 742) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24GBF658; 743) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24CBH730; 744) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24BBF877; 745) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23LBA701; 746) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23JBF848; 747) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23HBO965; 748) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23GBN955; 749) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23GBA464; 750) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23FBD594; 751) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23DBH622; 752) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23BBJ010; 753) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23BBE709; 754) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22JBG214; 755) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22JBB065; 756) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22BBO487; 757) DYNJ22359D, UDI-DI: 10193489552997(each), 40193489552998(case), Lot Number: 21HDA942; 758) DYNJ22359D, UDI-DI: 10193489552997(each), 40193489552998(case), Lot Number: 21ADC025; 759) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 23CBE334; 760) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 22BBS420; 761) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 22ABM026; 762) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 21LBU859; 763) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 21EBN507; 764) DYNJ22662L, UDI-DI: 10193489650808(each), 40193489650809(case), Lot Number: 21CMB744; 765) DYNJ24231C, UDI-DI: 10198459357626(each), 40198459357627(case), Lot Number: 25LDB494; 766) DYNJ24636I, UDI-DI: 10193489982145(each), 40193489982146(case), Lot Number: 23CBT957; 767) DYNJ24715L, UDI-DI: 10195327165772(each), 40195327165773(case), Lot Number: 23LMD460; 768) DYNJ24716G, UDI-DI: 10193489506389(each), 40193489506380(case), Lot Number: 23DMB573; 769) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 22BBB879; 770) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21KBQ972; 771) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21JBX549; 772) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21JBM247; 773) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21IBK016; 774) DYNJ25094Q, UDI-DI: 10889942372393(each), 40889942372394(case), Lot Number: 21OBA672; 775) DYNJ25796I, UDI-DI: 10889942415670(each), 40889942415671(case), Lot Number: 22LBR363; 776) DYNJ26277A, UDI-DI: 10080196979752(each), 40080196979753(case), Lot Number: 22HMC684; 777) DYNJ26425K, UDI-DI: 10193489341034(each), 40193489341035(case), Lot Number: 21DDB987; 778) DYNJ26425K, UDI-DI: 10193489341034(each), 40193489341035(case), Lot Number: 21DDB922; 779) DYNJ27524G, UDI-DI: 10193489549010(each), 40193489549011(case), Lot Number: 23GBJ533; 780) DYNJ28570I, UDI-DI: 10888277742277(each), 40888277742278(case), Lot Number: 22IMB678; 781) DYNJ30429K, UDI-DI: 10193489281682(each), 40193489281683(case), Lot Number: 21AKA573; 782) DYNJ30429L, UDI-DI: 10193489844221(each), 40193489844222(case), Lot Number: 22HMG195; 783) DYNJ30430O, UDI-DI: 10195327091132(each), 40195327091133(case), Lot Number: 22IMH835; 784) DYNJ30493C, UDI-DI: 10193489442250(each), 40193489442251(case), Lot Number: 22HME641; 785) DYNJ30880J, UDI-DI: 10195327223748(each), 40195327223749(case), Lot Number: 22KMB747; 786) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21LMD700; 787) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21JMG523; 788) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21IME223; 789) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21HMC610; 790) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21DME178; 791) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21CMA929; 792) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21BMC333; 793) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21AMB065; 794) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMJ727; 795) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMF470; 796) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMA381; 797) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23IMG579; 798) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23IMB296; 799) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23GMI083; 800) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23EMJ424; 801) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23CMH896; 802) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23CMG182; 803) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23BMA655; 804) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22LMF819; 805) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22KMC453; 806) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22JMD006; 807) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22IME048; 808) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22HMG533; 809) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22HMB139; 810) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22EMI458; 811) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22CME381; 812) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22BME125; 813) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22AMB915; 814) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25JMK962; 815) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25FMF966; 816) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25EMF404; 817) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25DMH101; 818) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25BMI478; 819) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25AMI942; 820) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24KMG666; 821) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24JME593; 822) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24IME048; 823) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24HMI467; 824) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24HMB126; 825) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24GMA034; 826) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24EMD713; 827) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24DMG373; 828) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24CMD605; 829) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24BMK012; 830) DYNJ31287G, UDI-DI: 10193489468724(each), 40193489468725(case), Lot Number: 21BMB509; 831) DYNJ31287G, UDI-DI: 10193489468724(each), 40193489468725(case), Lot Number: 21BMA106; 832) DYNJ32200F, UDI-DI: 10889942415649(each), 40889942415640(case), Lot Number: 22LBR364; 833) DYNJ32554, UDI-DI: 10884389217726(each), 40884389217727(case), Lot Number: 22JMB735; 834) DYNJ32724M, UDI-DI: 10193489734584(each), 40193489734585(case), Lot Number: 21ADB802; 835) DYNJ33670F, UDI-DI: 10193489220698(each), 40193489220699(case), Lot Number: 21ADA278; 836) DYNJ34683D, UDI-DI: 10193489508284(each), 40193489508285(case), Lot Number: 21ADC027; 837) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25GMF870; 838) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25EMG346; 839) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25CMF473; 840) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25BMI129; 841) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25AMB882; 842) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 24LMB128; 843) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 24JMF348; 844) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25JML524; 845) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25FMA192; 846) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25EMD264; 847) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25DMC946; 848) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25CMI416; 849) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25BME369; 850) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25AMI234; 851) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25AMG411; 852) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 24JME432; 853) DYNJ35512F, UDI-DI: 10889942930517(each), 40889942930518(case), Lot Number: 21ABR627; 854) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24FBK949; 855) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24EBO185; 856) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24BBU887; 857) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23LBG099; 858) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23JBD889; 859) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23GBH314; 860) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22JBK351; 861) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22FBC561; 862) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22EBQ927; 863) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22DBB588; 864) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22ABU211; 865) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21LBI599; 866) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21IBB712; 867) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21GBS876; 868) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21FBS907; 869) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21DBO076; 870) DYNJ35835G, UDI-DI: 10195327412005(each), 40195327412006(case), Lot Number: 24FMB572; 871) DYNJ36637N, UDI-DI: 10193489796933(each), 40193489796934(case), Lot Number: 23CMH794; 872) DYNJ36638O, UDI-DI: 10193489796834(each), 40193489796835(case), Lot Number: 23CMH795; 873) DYNJ36639L, UDI-DI: 10193489796940(each), 40193489796941(case), Lot Number: 23CMB824; 874) DYNJ36812F, UDI-DI: 10195327371180(each), 40195327371181(case), Lot Number: 23HBP365; 875) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 24EBK442; 876) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 24DBC101; 877) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23IBS456; 878) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23IBB162; 879) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23HBP073; 880) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23GBD566; 881) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23EBU238; 882) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 24HBT248; 883) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 23KBA024; 884) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25KBL533; 885) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25BBB514; 886) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25ABI114; 887) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24KBM633; 888) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24JBE066; 889) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24HBD378; 890) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24EBK443; 891) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23LBI544; 892) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23JBQ413; 893) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23IBV399; 894) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23IBK102; 895) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23EBD178; 896) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23DBM582; 897) DYNJ37123D, UDI-DI: 10193489459302(each), 40193489459303(case), Lot Number: 22BBM514; 898) DYNJ37428F, UDI-DI: 10889942164516(each), 40889942164517(case), Lot Number: 21CBQ635; 899) DYNJ37635L, UDI-DI: 10195327176303(each), 40195327176304(case), Lot Number: 23FBL430; 900) DYNJ37689J, UDI-DI: 10198459439056(each), 40198459439057(case), Lot Number: 25GMJ466; 901) DYNJ37696G, UDI-DI: 10193489294866(each), 40193489294867(case), Lot Number: 21DMB435; 902) DYNJ37696G, UDI-DI: 10193489294866(each), 40193489294867(case), Lot Number: 21AMA190; 903) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 23AMB157; 904) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22GMF679; 905) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22DMH440; 906) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22BMB144; 907) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21LMF755; 908) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21JMA643; 909) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21JMA051; 910) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21IMD490; 911) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21GMC277; 912) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21FMB538; 913) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21EMC373; 914) DYNJ37883F, UDI-DI: 10195327527068(each), 40195327527069(case), Lot Number: 23KMJ277; 915) DYNJ38234G, UDI-DI: 10195327327651(each), 40195327327652(case), Lot Number: 23EBR720; 916) DYNJ38247C, UDI-DI: 10888277746299(each), 40888277746290(case), Lot Number: 22HMC530; 917) DYNJ38397J, UDI-DI: 10193489646719(each), 40193489646710(case), Lot Number: 23CBR314; 918) DYNJ38397J, UDI-DI: 10193489646719(each), 40193489646710(case), Lot Number: 21CBJ903; 919) DYNJ38404J, UDI-DI: 10195327373955(each), 40195327373956(case), Lot Number: 23IME110; 920) DYNJ38406I, UDI-DI: 10195327373900(each), 40195327373901(case), Lot Number: 23IMA431; 921) DYNJ38529D, UDI-DI: 10195327525545(each), 40195327525546(case), Lot Number: 24DBC120; 922) DYNJ38529D, UDI-DI: 10195327525545(each), 40195327525546(case), Lot Number: 24BBJ214; 923) DYNJ38529G, UDI-DI: 10198459222429(each), 40198459222420(case), Lot Number: 25ABK835; 924) DYNJ39881F, UDI-DI: 10193489704068(each), 40193489704069(case), Lot Number: 23GMI050; 925) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22EBK821; 926) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22BBU071; 927) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22BBO207; 928) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21IBR355; 929) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21HBH566; 930) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21HBA944; 931) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBS723; 932) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBS675; 933) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBL153; 934) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24DBM519; 935) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24DBA107; 936) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24BBP203; 937) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24ABR822; 938) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23JBC963; 939) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23HBO846; 940) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23GBJ167; 941) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23FBL144; 942) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23BBQ109; 943) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23ABM892; 944) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 22HBJ054; 945) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 22FBS460; 946) DYNJ40309L, UDI-DI: 10193489913811(each), 40193489913812(case), Lot Number: 22BBK400; 947) DYNJ40557, UDI-DI: 10884389868850(each), 40884389868851(case), Lot Number: 21CBG563; 948) DYNJ40557, UDI-DI: 10884389868850(each), 40884389868851(case), Lot Number: 21ABN203; 949) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 22DBK226; 950) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 22BBT297; 951) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 21IBX396; 952) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 21IBX363; 953) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 23ABB877; 954) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 22GBG734; 955) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 22EBR223; 956) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 24BBB096; 957) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23LBA343; 958) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23JBX673; 959) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23IBH883; 960) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23HBP214; 961) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25IBF313; 962) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25HBL160; 963) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25GBP428; 964) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25BBL392; 965) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24LBL572; 966) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24KBQ488; 967) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24HBL231; 968) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24HBE197; 969) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24GBJ691; 970) DYNJ40557F, UDI-DI: 10198459589157(each), 40198459589158(case), Lot Number: 25KBR002; 971) DYNJ40597C, UDI-DI: 10193489974225(each), 40193489974226(case), Lot Number: 22DBK958; 972) DYNJ41030K, UDI-DI: 10195327156763(each), 40195327156764(case), Lot Number: 23GMB280; 973) DYNJ41030K, UDI-DI: 10195327156763(each), 40195327156764(case), Lot Number: 22HME079; 974) DYNJ41394C, UDI-DI: 10888277937208(each), 40888277937209(case), Lot Number: 21ABW205; 975) DYNJ41692F, UDI-DI: 10193489662993(each), 40193489662994(case), Lot Number: 22LBR516; 976) DYNJ41808C, UDI-DI: 10195327565664(each), 40195327565665(case), Lot Number: 24ABF458; 977) DYNJ41808C, UDI-DI: 10195327565664(each), 40195327565665(case), Lot Number: 23KBT722; 978) DYNJ41857C, UDI-DI: 10193489662658(each), 40193489662659(case), Lot Number: 22IBW059; 979) DYNJ42051F, UDI-DI: 10193489705317(each), 40193489705318(case), Lot Number: 21DDA607; 980) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 23DBB335; 981) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22LBP906; 982) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22KBR014; 983) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22IBU352; 984) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22HBZ293; 985) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22GBD911; 986) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22CBZ149; 987) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22BBJ794; 988) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21LBN641; 989) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21JBZ483; 990) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21GBB636; 991) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21ABU488; 992) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25LBM037; 993) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25JBB592; 994) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25GBF601; 995) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25FBS104; 996) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25FBG165; 997) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25DBB075; 998) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25CBP208; 999) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25BBC870; 1000) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24KBM235; 1001) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24JBL213; 1002) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24IBA105; 1003) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24HBF332; 1004) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24GBM991; 1005) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24FBK367; 1006) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24BBI050; 1007) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24ABT170; 1008) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24ABK305; 1009) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23LBB918; 1010) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23JBT608; 1011) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23IBG121; 1012) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23GBL372; 1013) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23EBD832; 1014) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23EMJ755; 1015) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23CMH177; 1016) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23BMG639; 1017) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22KMD575; 1018) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22HMH141; 1019) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22GMA359; 1020) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22DMI958; 1021) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22CME903; 1022) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22AMD130; 1023) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21LMD228; 1024) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21KMG901; 1025) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21KME766; 1026) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21JMG016; 1027) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21HMC587; 1028) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21CMD909; 1029) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21BMD156; 1030) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21BMA680; 1031) DYNJ42888B, UDI-DI: 10193489335002(each), 40193489335003(case), Lot Number: 21BBB540; 1032) DYNJ42944C, UDI-DI: 10193489660593(each), 40193489660594(case), Lot Number: 21CBJ906; 1033) DYNJ43006I, UDI-DI: 10193489927160(each), 40193489927161(case), Lot Number: 22JMG800; 1034) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23IMD412; 1035) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23GMG538; 1036) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23GMB626; 1037) DYNJ43186C, UDI-DI: 10193489475210(each), 40193489475211(case), Lot Number: 21CBO118; 1038) DYNJ43186C, UDI-DI: 10193489475210(each), 40193489475211(case), Lot Number: 21BBC547; 1039) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 22FBB767; 1040) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 22BBL717; 1041) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 21KBL792; 1042) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 21IBR381; 1043) DYNJ43186F, UDI-DI: 10195327186982(each), 40195327186983(case), Lot Number: 23BBC787; 1044) DYNJ43186F, UDI-DI: 10195327186982(each), 40195327186983(case), Lot Number: 22LBR903; 1045) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23GBO024; 1046) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23DBQ604; 1047) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23CBX218; 1048) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23CBB473; 1049) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23ABO776; 1050) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 24CBF071; 1051) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 24ABR199; 1052) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23LBU954; 1053) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23KBN758; 1054) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23IBM911; 1055) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23GBV872; 1056) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23FBT488; 1057) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23EBW551; 1058) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23DBB880; 1059) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23DBB718; 1060) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23BBS734; 1061) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23BBC162; 1062) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22OBI176; 1063) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22OBB442; 1064) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22LBN719; 1065) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22HBW601; 1066) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBQ330; 1067) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBE821; 1068) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBB221; 1069) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22EBH854; 1070) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22BBU712; 1071) DYNJ43421B, UDI-DI: 10193489 901412(each), 40193489901413(case), Lot Number: 22ABH986; 1072) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21KBL461; 1073) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21HBT086; 1074) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21GBR214; 1075) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21IME539; 1076) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21HMB776; 1077) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21FMD820; 1078) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22BMC971; 1079) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMB868; 1080) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMA613; 1081) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMA059; 1082) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 21LMD524; 1083) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23EMI628; 1084) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23DMG739; 1085) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23BMC726; 1086) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23BMC530; 1087) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22KMD710; 1088) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22IMH856; 1089) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22FMH503; 1090) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22FMH502; 1091) DYNJ44020A, UDI-DI: 10193489806649(each), 40193489806640(case), Lot Number: 23DBI587; 1092) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22OBM108; 1093) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22OBD081; 1094) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBQ924; 1095) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBK861; 1096) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBC461; 1097) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22GBC896; 1098) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22BBG493; 1099) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21LBQ228; 1100) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21KBA071; 1101) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21GBT043; 1102) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21GBC952; 1103) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21EBN450; 1104) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21DBC978; 1105) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21BBR226; 1106) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21ABB322; 1107) DYNJ45062C, UDI-DI: 10193489733389(each), 40193489733380(case), Lot Number: 21KBP326; 1108) DYNJ45145F, UDI-DI: 10193489824889(each), 40193489824880(case), Lot Number: 21BBA717; 1109) DYNJ45892C, UDI-DI: 10193489272017(each), 40193489272018(case), Lot Number: 22HMD408; 1110) DYNJ46534J, UDI-DI: 10193489337778(each), 40193489337779(case), Lot Number: 21FDC712; 1111) DYNJ46534J, UDI-DI: 10193489337778(each), 40193489337779(case), Lot Number: 21FDB447; 1112) DYNJ46980D, UDI-DI: 10889942852895(each), 40889942852896(case), Lot Number: 23EBB740; 1113) DYNJ47104C, UDI-DI: 10889942489190(each), 40889942489191(case), Lot Number: 21ABW902; 1114) DYNJ47608B, UDI-DI: 10193489731019(each), 40193489731010(case), Lot Number: 23ABH658; 1115) DYNJ47715I, UDI-DI: 10193489887525(each), 40193489887526(case), Lot Number: 22HME316; 1116) DYNJ47784G, UDI-DI: 10193489887532(each), 40193489887533(case), Lot Number: 22HMF248; 1117) DYNJ48063G, UDI-DI: 10193489497830(each), 40193489497831(case), Lot Number: 22JMB326; 1118) DYNJ48091I, UDI-DI: 10195327500122(each), 40195327500123(case), Lot Number: 23JMI555; 1119) DYNJ48151I, UDI-DI: 10195327261870(each), 40195327261871(case), Lot Number: 23HMG449; 1120) DYNJ48178, UDI-DI: 10888277559226(each), 40888277559227(case), Lot Number: 23HME394; 1121) DYNJ48237J, UDI-DI: 10193489988185(each), 40193489988186(case), Lot Number: 22HMD410; 1122) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24GBP662; 1123) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24FBK165; 1124) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24EBU758; 1125) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24CBC160; 1126) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23JBY697; 1127) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23IBJ632; 1128) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23FBQ576; 1129) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23EBN371; 1130) DYNJ48473I, UDI-DI: 10198459349829(each), 40198459349820(case), Lot Number: 25FBB300; 1131) DYNJ49105G, UDI-DI: 10193489927382(each), 40193489927383(case), Lot Number: 23FBA447; 1132) DYNJ49247B, UDI-DI: 10193489504446(each), 40193489504447(case), Lot Number: 22KMB758; 1133) DYNJ49247B, UDI-DI: 10193489504446(each), 40193489504447(case), Lot Number: 22HMD932; 1134) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21DBI826; 1135) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21BBQ667; 1136) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21ABL108; 1137) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21LBG756; 1138) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21KBJ103; 1139) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21FBK536; 1140) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21EBN532; 1141) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 23DMB606; 1142) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 23CME921; 1143) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22LMC345; 1144) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22KMC642; 1145) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22IMA085; 1146) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22GME891; 1147) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22DMC969; 1148) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22DBI721; 1149) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 21LBW673; 1150) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23KMH889; 1151) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23IMA323; 1152) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23GMF313; 1153) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 22EDA621; 1154) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 21IDA623; 1155) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 21HDB682; 1156) DYNJ49782I, UDI-DI: 10195327259105(each), 40195327259106(case), Lot Number: 23CBK571; 1157) DYNJ49782J, UDI-DI: 10195327406240(each), 40195327406241(case), Lot Number: 23EBU897; 1158) DYNJ50357C, UDI-DI: 10193489826791(each), 40193489826792(case), Lot Number: 21BKA485; 1159) DYNJ50360C, UDI-DI: 10193489826821(each), 40193489826822(case), Lot Number: 23HMC675; 1160) DYNJ50360C, UDI-DI: 10193489826821(each), 40193489826822(case), Lot Number: 23HMC674; 1161) DYNJ50429D, UDI-DI: 10195327043506(each), 40195327043507(case), Lot Number: 23BBB098; 1162) DYNJ50450C, UDI-DI: 10193489595307(each), 40193489595308(case), Lot Number: 21DBK486; 1163) DYNJ50450C, UDI-DI: 10193489595307(each), 40193489595308(case), Lot Number: 21BBE826; 1164) DYNJ50622I, UDI-DI: 10193489467000(each), 40193489467001(case), Lot Number: 21ABW206; 1165) DYNJ50710D, UDI-DI: 10193489716689(each), 40193489716680(case), Lot Number: 20KBZ974; 1166) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 22OBJ527; 1167) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 22HBC296; 1168) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 21GBU067; 1169) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 21DBM047; 1170) DYNJ51036B, UDI-DI: 10889942229475(each), 40889942229476(case), Lot Number: 21AMD560; 1171) DYNJ51036B, UDI-DI: 10889942229475(each), 40889942229476(case), Lot Number: 21AMA118; 1172) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21FME965; 1173) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21DMC480; 1174) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21CMB429; 1175) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22JMC109; 1176) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22JMA473; 1177) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22IMD862; 1178) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22DMH641; 1179) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22DMF167; 1180) DYNJ51036G, UDI-DI: 10195327332365(each), 40195327332366(case), Lot Number: 23FBQ872; 1181) DYNJ51088A, UDI-DI: 10193489985313(each), 40193489985314(case), Lot Number: 21IDA625; 1182) DYNJ51091A, UDI-DI: 10193489985306(each), 40193489985307(case), Lot Number: 22DDC213; 1183) DYNJ51095A, UDI-DI: 10193489985436(each), 40193489985437(case), Lot Number: 22DDC333; 1184) DYNJ51096B, UDI-DI: 10193489985290(each), 40193489985291(case), Lot Number: 22EDA486; 1185) DYNJ51176C, UDI-DI: 10889942724642(each), 40889942724643(case), Lot Number: 21ADA263; 1186) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 25CBJ143; 1187) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 25ABK343; 1188) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 24LBN605; 1189) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 24FBF369; 1190) DYNJ51774C, UDI-DI: 10889942816200(each), 40889942816201(case), Lot Number: 21CBI319; 1191) DYNJ51802I, UDI-DI: 10193489476217(each), 40193489476218(case), Lot Number: 21ADA367; 1192) DYNJ51803G, UDI-DI: 10193489476224(each), 40193489476225(case), Lot Number: 21ADA120; 1193) DYNJ52207B, UDI-DI: 10193489635614(each), 40193489635615(case), Lot Number: 23BMH298; 1194) DYNJ52418F, UDI-DI: 10193489249705(each), 40193489249706(case), Lot Number: 21ADA541; 1195) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 22EBP123; 1196) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 22DBA960; 1197) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21JBS706; 1198) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21IBG396; 1199) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21GBK596; 1200) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21DBM019; 1201) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21CBN962; 1202) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21ABE527; 1203) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24DBS978; 1204) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24BBP202; 1205) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24BBE314; 1206) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24ABG587; 1207) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23KBQ185; 1208) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23IBW198; 1209) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23HBC482; 1210) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23DBE733; 1211) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23BBH694; 1212) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 22IBA379; 1213) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 22GBE349; 1214) DYNJ52491F, UDI-DI: 10195327435899(each), 40195327435890(case), Lot Number: 25LME675; 1215) DYNJ52523A, UDI-DI: 10889942502462(each), 40889942502463(case), Lot Number: 21BBK699; 1216) DYNJ52607C, UDI-DI: 10195327134631(each), 40195327134632(case), Lot Number: 23DBG719; 1217) DYNJ52720C, UDI-DI: 10195327480356(each), 40195327480357(case), Lot Number: 23KBX843; 1218) DYNJ52950C, UDI-DI: 10193489745337(each), 40193489745338(case), Lot Number: 21BBK701; 1219) DYNJ52957F, UDI-DI: 10193489315721(each), 40193489315722(case), Lot Number: 23FBR704; 1220) DYNJ52958D, UDI-DI: 10193489740349(each), 40193489740340(case), Lot Number: 21BBA721; 1221) DYNJ53212D, UDI-DI: 10193489946871(each), 40193489946872(case), Lot Number: 22BBE369; 1222) DYNJ53212D, UDI-DI: 10193489946871(each), 40193489946872(case), Lot Number: 21GBD684; 1223) DYNJ53287D, UDI-DI: 10193489552980(each), 40193489552981(case), Lot Number: 21BDA322; 1224) DYNJ53420D, UDI-DI: 10193489975949(each), 40193489975940(case), Lot Number: 23CBE656; 1225) DYNJ53539D, UDI-DI: 10195327158903(each), 40195327158904(case), Lot Number: 23KBK120; 1226) DYNJ53686A, UDI-DI: 10193489386882(each), 40193489386883(case), Lot Number: 21ADA123; 1227) DYNJ53788A, UDI-DI: 10889942909421(each), 40889942909422(case), Lot Number: 21ADB618; 1228) DYNJ53825, UDI-DI: 10889942443963(each), 40889942443964(case), Lot Number: 24DBJ816; 1229) DYNJ53970K, UDI-DI: 10195327532741(each), 40195327532742(case), Lot Number: 23KMF315; 1230) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 22FMH898; 1231) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21KME883; 1232) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21JMH447; 1233) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21JMC220; 1234) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21IME798; 1235) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21HMF506; 1236) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21HMA349; 1237) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23JMD007; 1238) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23HMG008; 1239) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23HME945; 1240) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23FMH056; 1241) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23CMH919; 1242) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23AMB679; 1243) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25FMJ582; 1244) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25EMF753; 1245) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25BME348; 1246) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25AMH782; 1247) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24KMH993; 1248) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24HMI416; 1249) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24FMI100; 1250) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24EME642; 1251) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24DMJ360; 1252) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24AMA137; 1253) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 23KMI515; 1254) DYNJ54912B, UDI-DI: 10193489317077(each), 40193489317078(case), Lot Number: 23IBP189; 1255) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25EBP981; 1256) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25BBM551; 1257) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25ABD151; 1258) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 24JBJ473; 1259) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 24HBT914; 1260) DYNJ54988C, UDI-DI: 10195327007003(each), 40195327007004(case), Lot Number: 23CMC077; 1261) DYNJ55071F, UDI-DI: 10195327410957(each), 40195327410958(case), Lot Number: 23LBC872; 1262) DYNJ55317F, UDI-DI: 10193489692389(each), 40193489692380(case), Lot Number: 23EBU898; 1263) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23JBH543; 1264) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23IBA353; 1265) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23GBC804; 1266) DYNJ55317I, UDI-DI: 10195327534202(each), 40195327534203(case), Lot Number: 23LBO119; 1267) DYNJ55324F, UDI-DI: 10195327120726(each), 40195327120727(case), Lot Number: 23FBA486; 1268) DYNJ55353C, UDI-DI: 10195327173487(each), 40195327173488(case), Lot Number: 23FBR554; 1269) DYNJ55441C, UDI-DI: 10889942828449(each), 40889942828440(case), Lot Number: 23JBB387; 1270) DYNJ55744A, UDI-DI: 10889942688937(each), 40889942688938(case), Lot Number: 21EMA571; 1271) DYNJ56257C, UDI-DI: 10193489825114(each), 40193489825115(case), Lot Number: 23IBE172; 1272) DYNJ56766C, UDI-DI: 10193489648485(each), 40193489648486(case), Lot Number: 21ABG397; 1273) DYNJ56800F, UDI-DI: 10193489353709(each), 40193489353700(case), Lot Number: 21BKA247; 1274) DYNJ56864F, UDI-DI: 10195327464417(each), 40195327464418(case), Lot Number: 23KBV146; 1275) DYNJ56889, UDI-DI: 10889942691739(each), 40889942691730(case), Lot Number: 23JBT092; 1276) DYNJ56889, UDI-DI: 10889942691739(each), 40889942691730(case), Lot Number: 23CBO165; 1277) DYNJ56986F, UDI-DI: 10193489201901(each), 40193489201902(case), Lot Number: 21BBL376; 1278) DYNJ56986F, UDI-DI: 10193489201901(each), 40193489201902(case), Lot Number: 21ABM378; 1279) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 24CMC651; 1280) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 23KMB418; 1281) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22LBR917; 1282) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22KBR892; 1283) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22EBQ164; 1284) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22DBD469; 1285) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22BBT830; 1286) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21VBB773; 1287) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21LBS202; 1288) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21LBS167; 1289) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21HBN583; 1290) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21GBP729; 1291) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21FBQ044; 1292) DYNJ57001A, UDI-DI: 10193489479836(each), 40193489479837(case), Lot Number: 21BBA507; 1293) DYNJ57080A, UDI-DI: 10193489234664(each), 40193489234665(case), Lot Number: 21CMB611; 1294) DYNJ57266F, UDI-DI: 10193489498189(each), 40193489498180(case), Lot Number: 21BBA724; 1295) DYNJ57806B, UDI-DI: 10193489643749(each), 40193489643740(case), Lot Number: 24ABI521; 1296) DYNJ57824G, UDI-DI: 10193489237528(each), 40193489237529(case), Lot Number: 21KDB870; 1297) DYNJ57824G, UDI-DI: 10193489237528(each), 40193489237529(case), Lot Number: 21IDB163; 1298) DYNJ58044B, UDI-DI: 10193489915846(each), 40193489915847(case), Lot Number: 23JBY451; 1299) DYNJ58082A, UDI-DI: 10195327096748(each), 40195327096749(case), Lot Number: 22KDC032; 1300) DYNJ58082A, UDI-DI: 10195327096748(each), 40195327096749(case), Lot Number: 22FDB387; 1301) DYNJ58085C, UDI-DI: 10195327096779(each), 40195327096770(case), Lot Number: 22EDA279; 1302) DYNJ58372A, UDI-DI: 10193489345247(each), 40193489345248(case), Lot Number: 21ADA509; 1303) DYNJ58374F, UDI-DI: 10195327095062(each), 40195327095063(case), Lot Number: 23IBU240; 1304) DYNJ58443C, UDI-DI: 10193489778588(each), 40193489778589(case), Lot Number: 23IMA198; 1305) DYNJ58563C, UDI-DI: 10193489460025(each), 40193489460026(case), Lot Number: 21BBA725; 1306) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21KMG511; 1307) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21IME158; 1308) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21IMD510; 1309) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22JME349; 1310) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22IMG805; 1311) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22GME707; 1312) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22CMA966; 1313) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22AMF288; 1314) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22AMF287; 1315) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 21LMF341; 1316) DYNJ58713, UDI-DI: 10889942880980(each), 40889942880981(case), Lot Number: 21BBB510; 1317) DYNJ58817A, UDI-DI: 10195327312114(each), 40195327312115(case), Lot Number: 23HBQ114; 1318) DYNJ59782, UDI-DI: 10888277783799(each), 40888277783790(case), Lot Number: 21CBE064; 1319) DYNJ59915D, UDI-DI: 10195327042288(each), 40195327042289(case), Lot Number: 23ELA911; 1320) DYNJ59929C, UDI-DI: 10195327036942(each), 40195327036943(case), Lot Number: 23LBF701; 1321) DYNJ60062K, UDI-DI: 10195327322540(each), 40195327322541(case), Lot Number: 23DBG359; 1322) DYNJ60103A, UDI-DI: 10193489321364(each), 40193489321365(case), Lot Number: 22HMB861; 1323) DYNJ60485C, UDI-DI: 10195327493486(each), 40195327493487(case), Lot Number: 23LBQ563; 1324) DYNJ60730A, UDI-DI: 10195327157302(each), 40195327157303(case), Lot Number: 22HBU496; 1325) DYNJ61116B, UDI-DI: 10193489226768(each), 40193489226769(case), Lot Number: 21ABV416; 1326) DYNJ61244A, UDI-DI: 10193489811032(each), 40193489811033(case), Lot Number: 21ABW945; 1327) DYNJ61480A, UDI-DI: 10193489963533(each), 40193489963534(case), Lot Number: 23JBA340; 1328) DYNJ61581C, UDI-DI: 10195327152147(each), 40195327152148(case), Lot Number: 23EME363; 1329) DYNJ61687, UDI-DI: 10193489589702(each), 40193489589703(case), Lot Number: 23IMB431; 1330) DYNJ61692D, UDI-DI: 10195327472580(each), 40195327472581(case), Lot Number: 23KMH106; 1331) DYNJ61735A, UDI-DI: 10193489469387(each), 40193489469388(case), Lot Number: 22LMF337; 1332) DYNJ61739A, UDI-DI: 10193489469417(each), 40193489469418(case), Lot Number: 23CMF304; 1333) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20XBC904; 1334) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20LBX795; 1335) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20LBT867; 1336) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21CBH759; 1337) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21BBS149; 1338) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21BBD938; 1339) DYNJ61836D, UDI-DI: 10193489863345(each), 40193489863346(case), Lot Number: 21DBR351; 1340) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22KBG479; 1341) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22JBI959; 1342) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22HBT804; 1343) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22GBC809; 1344) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22BBO063; 1345) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21LBW709; 1346) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21KBY178; 1347) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21JBT954; 1348) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21JBF773; 1349) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21IBU114; 1350) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23FBC115; 1351) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23DBJ016; 1352) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23BBJ107; 1353) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23ABJ028; 1354) DYNJ61845, UDI-DI: 10193489605358(each), 40193489605359(case), Lot Number: 21DDA680; 1355) DYNJ61846, UDI-DI: 10193489605518(each), 40193489605519(case), Lot Number: 21DDC139; 1356) DYNJ61846, UDI-DI: 10193489605518(each), 40193489605519(case), Lot Number: 21ADB041; 1357) DYNJ61849, UDI-DI: 10193489605181(each), 40193489605182(case), Lot Number: 21DDB382; 1358) DYNJ62201, UDI-DI: 10193489633528(each), 40193489633529(case), Lot Number: 21CDC428; 1359) DYNJ62368C, UDI-DI: 10193489407525(each), 40193489407526(case), Lot Number: 21BBA729; 1360) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23IBN349; 1361) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23HBI392; 1362) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23GBN992; 1363) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23EBE814; 1364) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23CBO687; 1365) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23BBD864; 1366) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23BBD249; 1367) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22IBK987; 1368) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22HBW557; 1369) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22HBM930; 1370) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22GBQ160; 1371) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22GBF339; 1372) DYNJ62668B, UDI-DI: 10193489945690(each), 40193489945691(case), Lot Number: 22DBD976; 1373) DYNJ62743B, UDI-DI: 10193489800784(each), 40193489800785(case), Lot Number: 22IBK765; 1374) DYNJ62999B, UDI-DI: 10193489833461(each), 40193489833462(case), Lot Number: 21GDA095; 1375) DYNJ63001A, UDI-DI: 10193489245639(each), 40193489245630(case), Lot Number: 21ADB042; 1376) DYNJ63016C, UDI-DI: 10193489833515(each), 40193489833516(case), Lot Number: 21CDC230; 1377) DYNJ63030A, UDI-DI: 10193489245820(each), 40193489245821(case), Lot Number: 21BDA062; 1378) DYNJ63031B, UDI-DI: 10193489833492(each), 40193489833493(case), Lot Number: 21FDA942; 1379) DYNJ63031B, UDI-DI: 10193489833492(each), 40193489833493(case), Lot Number: 21EDB802; 1380) DYNJ63035C, UDI-DI: 10193489833539(each), 40193489833530(case), Lot Number: 21EDB803; 1381) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23GMG812; 1382) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23EMI678; 1383) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23CMI070; 1384) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23AMH685; 1385) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22LMJ344; 1386) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22IMA805; 1387) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22IMA803; 1388) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22HMA990; 1389) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22GMC785; 1390) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22CMD192; 1391) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 24BMA651; 1392) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23LMG702; 1393) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23KMH957; 1394) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23KMC330; 1395) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21KBJ343; 1396) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21JBO125; 1397) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21HBM419; 1398) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21FBQ059; 1399) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21CBC907; 1400) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21ABP388; 1401) DYNJ63201B, UDI-DI: 10195327291006(each), 40195327291007(case), Lot Number: 23ABB913; 1402) DYNJ63202B, UDI-DI: 10195327291082(each), 40195327291083(case), Lot Number: 23ABB886; 1403) DYNJ63510A, UDI-DI: 10193489295818(each), 40193489295819(case), Lot Number: 21ADA838; 1404) DYNJ63603B, UDI-DI: 10193489901566(each), 40193489901567(case), Lot Number: 23DBP360; 1405) DYNJ63639B, UDI-DI: 10193489900323(each), 40193489900324(case), Lot Number: 23BBM590; 1406) DYNJ63802A, UDI-DI: 10193489799637(each), 40193489799638(case), Lot Number: 21EBA575; 1407) DYNJ63802A, UDI-DI: 10193489799637(each), 40193489799638(case), Lot Number: 21DBU650; 1408) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24JBC327; 1409) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24FBK285; 1410) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24EBP263; 1411) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24ABK520; 1412) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 23KBU440; 1413) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 23DBC378; 1414) DYNJ63813C, UDI-DI: 10193489914214(each), 40193489914215(case), Lot Number: 23DBG674; 1415) DYNJ63955, UDI-DI: 10193489774276(each), 40193489774277(case), Lot Number: 21CDC326; 1416) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25FDB361; 1417) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25CDB456; 1418) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25ADB757; 1419) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24LDB383; 1420) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24KDB121; 1421) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24HDA013; 1422) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24EDA419; 1423) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24CDB433; 1424) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 23LDB441; 1425) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Numbe r: 23JDB401; 1426) DYNJ64164, UDI-DI: 10193489792034(each), 40193489792035(case), Lot Number: 21ABH841; 1427) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22FMG702; 1428) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22EMA712; 1429) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22DMD228; 1430) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22BBM524; 1431) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22BBJ435; 1432) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21OBA733; 1433) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21IBG214; 1434) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21GBU418; 1435) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21FBK920; 1436) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21DBV412; 1437) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21BBM393; 1438) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23GMH157; 1439) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23FME303; 1440) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23EMB852; 1441) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23BMC271; 1442) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 22JMI296; 1443) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 22HMF357; 1444) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 22BMF445; 1445) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 22AMG273; 1446) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21IMC839; 1447) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21HMG002; 1448) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21HME922; 1449) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21GMC645; 1450) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21FME760; 1451) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 22BME410; 1452) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 22AMA774; 1453) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21LMD287; 1454) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21JMG559; 1455) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21IMD744; 1456) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21HMF811; 1457) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21FMF696; 1458) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21FBG721; 1459) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21DBB375; 1460) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21CBS359; 1461) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21BBA813; 1462) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21ABN353; 1463) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 24DMI142; 1464) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 24CMB056; 1465) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23LMD889; 1466) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23GMB321; 1467) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23DMF252; 1468) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23BMI540; 1469) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22JMH309; 1470) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22HMD635; 1471) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22GMC042; 1472) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22EMA913; 1473) DYNJ64510A, UDI-DI: 10193489971057(each), 40193489971058(case), Lot Number: 23CBC285; 1474) DYNJ64527D, UDI-DI: 10198459234729(each), 40198459234720(case), Lot Number: 25FBK195; 1475) DYNJ64527D, UDI-DI: 10198459234729(each), 40198459234720(case), Lot Number: 25DBN471; 1476) DYNJ64603B, UDI-DI: 10195327284749(each), 40195327284740(case), Lot Number: 23CBA277; 1477) DYNJ64733, UDI-DI: 10193489216455(each), 40193489216456(case), Lot Number: 21DDB517; 1478) DYNJ64741A, UDI-DI: 10193489456004(each), 40193489456005(case), Lot Number: 21DDA609; 1479) DYNJ64741A, UDI-DI: 10193489456004(each), 40193489456005(case), Lot Number: 21DDA364; 1480) DYNJ65065A, UDI-DI: 10193489317244(each), 40193489317245(case), Lot Number: 21BBC473; 1481) DYNJ65066B, UDI-DI: 10193489317183(each), 40193489317184(case), Lot Number: 21ABW917; 1482) DYNJ65108A, UDI-DI: 10193489822144(each), 40193489822145(case), Lot Number: 21BBD365; 1483) DYNJ65164C, UDI-DI: 10195327246693(each), 40195327246694(case), Lot Number: 22IBE647; 1484) DYNJ65319B, UDI-DI: 10193489869040(each), 40193489869041(case), Lot Number: 23EBR728; 1485) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23IBG967; 1486) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23HBA343; 1487) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23FBV950; 1488) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23EBJ166; 1489) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23ABT634; 1490) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 25BBC387; 1491) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24LBK259; 1492) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24KBD923; 1493) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24JBJ097; 1494) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24GBS754; 1495) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24EBP995; 1496) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24CBO129; 1497) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24BBP255; 1498) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 23LBT407; 1499) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 22DBB600; 1500) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 22ABV290; 1501) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21KBG570; 1502) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21IBT090; 1503) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21FBI078; 1504) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24FBQ247; 1505) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24CBB587; 1506) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24ABM648; 1507) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23JBA736; 1508) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23IBG477; 1509) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23EBT831; 1510) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23ABG837; 1511) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22IBV229; 1512) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22IBB740; 1513) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22GBW912; 1514) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25JBX795; 1515) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25GLA117; 1516) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLB119; 1517) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLA456; 1518) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLA213; 1519) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25BLA191; 1520) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24ILA811; 1521) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24HLA763; 1522) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24HLA694; 1523) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21HBN607; 1524) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21GBD027; 1525) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21EBE294; 1526) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21ABJ639; 1527) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24FMG725; 1528) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24CMH438; 1529) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23KMB736; 1530) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23EMC290; 1531) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23EMB961; 1532) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23BMH039; 1533) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22OBD050; 1534) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22LMB610; 1535) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22JMD985; 1536) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22ABM177; 1537) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21VBC713; 1538) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21LBU977; 1539) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21ABS730; 1540) DYNJ65730A, UDI-DI: 10193489958270(each), 40193489958271(case), Lot Number: 22OBC403; 1541) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25KMA265; 1542) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25GMJ308; 1543) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25FMC631; 1544) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25EMB299; 1545) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25DMF044; 1546) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25CMG189; 1547) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25BMD281A; 1548) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25BMD281; 1549) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25AMI183; 1550) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24KMF425; 1551) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24JMI243; 1552) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24IMI316; 1553) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24HMH127; 1554) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24GMJ544; 1555) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24GMC894; 1556) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24DMK513; 1557) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24DMD371; 1558) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24CME387; 1559) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24BMJ136; 1560) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24BMC898; 1561) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24AMB121; 1562) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 23KMI663; 1563) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 23JMI202; 1564) DYNJ65934A, UDI-DI: 10193489455359(each), 40193489455350(case), Lot Number: 21BBF325; 1565) DYNJ65958A, UDI-DI: 10193489481624(each), 40193489481625(case), Lot Number: 21BBN445; 1566) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23JBA683; 1567) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23IBT769; 1568) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23HBY749; 1569) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23DBM741; 1570) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23CBB374; 1571) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23BBF990; 1572) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 22LBO308; 1573) DYNJ66138B, UDI-DI: 10198459482359(each), 40198459482350(case), Lot Number: 25HMG871; 1574) DYNJ66572, UDI-DI: 10193489360936(each), 40193489360937(case), Lot Number: 21BBS830; 1575) DYNJ66572, UDI-DI: 10193489360936(each), 40193489360937(case), Lot Number: 21BBS046; 1576) DYNJ66572A, UDI-DI: 10193489869460(each), 40193489869461(case), Lot Number: 22CBF350; 1577) DYNJ66572A, UDI-DI: 10193489869460(each), 40193489869461(case), Lot Number: 21LBJ616; 1578) DYNJ66572B, UDI-DI: 10195327244187(each), 40195327244188(case), Lot Number: 23CBL488; 1579) DYNJ66572B, UDI-DI: 10195327244187(each), 40195327244188(case), Lot Number: 23ABS704; 1580) DYNJ66682, UDI-DI: 10193489370294(each), 40193489370295(case), Lot Number: 21CBJ914; 1581) DYNJ66940B, UDI-DI: 10195327305253(each), 40195327305254(case), Lot Number: 23IBT566; 1582) DYNJ66940B, UDI-DI: 10195327305253(each), 40195327305254(case), Lot Number: 23IBB170; 1583) DYNJ66945B, UDI-DI: 10195327305246(each), 40195327305247(case), Lot Number: 23EBR636; 1584) DYNJ66945B, UDI-DI: 10195327305246(each), 40195327305247(case), Lot Number: 23CBD537; 1585) DYNJ67141A, UDI-DI: 10195327024215(each), 40195327024216(case), Lot Number: 23HBS080; 1586) DYNJ67241, UDI-DI: 10193489392227(each), 40193489392228(case), Lot Number: 23LBD765; 1587) DYNJ67241, UDI-DI: 10193489392227(each), 40193489392228(case), Lot Number: 23HBH208; 1588) DYNJ67278, UDI-DI: 10193489394757(each), 40193489394758(case), Lot Number: 23HBM356; 1589) DYNJ67303A, UDI-DI: 10195327015664(each), 40195327015665(case), Lot Number: 23JBV806; 1590) DYNJ67371, UDI-DI: 10193489889758(each), 40193489889759(case), Lot Number: 22DBE238; 1591) DYNJ67444B, UDI-DI: 10195327134648(each), 40195327134649(case), Lot Number: 23HBS081; 1592) DYNJ67502, UDI-DI: 10193489404173(each), 40193489404174(case), Lot Number: 21BBL566; 1593) DYNJ67509, UDI-DI: 10193489404388(each), 40193489404389(case), Lot Number: 21BBP851; 1594) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23HMG482; 1595) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23HME139; 1596) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23EMI872; 1597) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24FMJ102; 1598) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24CMF453; 1599) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24BMD330; 1600) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 23KMF780; 1601) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 25DMA359; 1602) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 25AMF026; 1603) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 24LMA719; 1604) DYNJ67814B, UDI-DI: 10195327135676(each), 40195327135677(case), Lot Number: 23KBT429; 1605) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25LBN320; 1606) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25JBX784; 1607) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25IBJ392; 1608) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25HBK166; 1609) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25GBL311; 1610) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25BBT615; 1611) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25BBQ467; 1612) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24KBB774; 1613) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24JBK383; 1614) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24FBS410; 1615) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24EBD903; 1616) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24CBD349; 1617) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24ABV046; 1618) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 23LBU562; 1619) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 23KBH486; 1620) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21CBO800; 1621) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21CBK754; 1622) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21BBB519; 1623) DYNJ68052B, UDI-DI: 10195327588441(each), 40195327588442(case), Lot Number: 24CLA962; 1624) DYNJ68308, UDI-DI: 10193489821413(each), 40193489821414(case), Lot Number: 23IBD739; 1625) DYNJ68320, UDI-DI: 10193489822991(each), 40193489822992(case), Lot Number: 21BBB618; 1626) DYNJ68430, UDI-DI: 10193489465211(each), 40193489465212(case), Lot Number: 23IBB159; 1627) DYNJ68720, UDI-DI: 10193489843071(each), 40193489843072(case), Lot Number: 23GBG686; 1628) DYNJ68980, UDI-DI: 10193489870817(each), 40193489870818(case), Lot Number: 23CBU626; 1629) DYNJ69131B, UDI-DI: 10195327173678(each), 40195327173679(case), Lot Number: 23EBP597; 1630) DYNJ69317A, UDI-DI: 10195327224332(each), 40195327224333(case), Lot Number: 23BMH343; 1631) DYNJ69371, UDI-DI: 10193489935028(each), 40193489935029(case), Lot Number: 23KBN032; 1632) DYNJ69915B, UDI-DI: 10195327597184(each), 40195327597185(case), Lot Number: 24JMI703; 1633) DYNJ69934, UDI-DI: 10193489997811(each), 40193489997812(case), Lot Number: 22HME118; 1634) DYNJ80071, UDI-DI: 10195327010126(each), 40195327010127(case), Lot Number: 23CBV608; 1635) DYNJ80085, UDI-DI: 10195327010065(each), 40195327010066(case), Lot Number: 23CBV727; 1636) DYNJ80130A, UDI-DI: 10195327410896(each), 40195327410897(case), Lot Number: 23GMF021; 1637) DYNJ80192, UDI-DI: 10195327014247(each), 40195327014248(case), Lot Number: 21KBI313; 1638) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 25CBT101; 1639) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 25ABB722; 1640) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24KBU010; 1641) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24IBQ276; 1642) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24HBF212; 1643) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24DBD641; 1644) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24ABB561; 1645) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23LBP849; 1646) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23KBR833; 1647) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23JBS811; 1648) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 25BBN761; 1649) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 24DBJ635; 1650) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 23KBI250; 1651) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 23JBK792; 1652) DYNJ80879D, UDI-DI: 10195327428976(each), 40195327428977(case), Lot Number: 23LBM359; 1653) DYNJ80879D, UDI-DI: 10195327428976(each), 40195327428977(case), Lot Number: 23JBW728; 1654) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23FBR818; 1655) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23EBN355; 1656) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23CBG415; 1657) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23CBF631; 1658) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 22KBI778; 1659) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 22DBA249; 1660) DYNJ81014A, UDI-DI: 10195327495848(each), 40195327495849(case), Lot Number: 23JBH396; 1661) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 24DBH527; 1662) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23KBS890; 1663) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23JBV322; 1664) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23IBB042; 1665) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23FBV730; 1666) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 23CBD541; 1667) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 23ABU291; 1668) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22JBM301; 1669) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22HBW957; 1670) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22GBZ212; 1671) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22EBG823; 1672) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23HME214; 1673) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23FMF387; 1674) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23EME110; 1675) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23CMI084; 1676) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22HMB593; 1677) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22HMB592; 1678) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22EME413; 1679) DYNJ81905, UDI-DI: 10195327154998(each), 40195327154999(case), Lot Number: 22GBA833; 1680) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25LME657; 1681) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25JMH386; 1682) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25IMD543; 1683) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25GMF869; 1684) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25EMA396; 1685) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25CMH914; 1686) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25BMB151; 1687) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25AMH197; 1688) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24KMG661; 1689) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24IMI184; 1690) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24HMH952; 1691) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24EMG933; 1692) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMK399A; 1693) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMK399; 1694) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMJ076; 1695) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24BME257; 1696) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24AME898; 1697) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23LMB675; 1698) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23KMB630; 1699) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23JBU264; 1700) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23IBO987; 1701) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23GBS218; 1702) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23DBP302; 1703) DYNJ81921, UDI-DI: 10195327155544(each), 40195327155545(case), Lot Number: 22HBD171; 1704) DYNJ82742, UDI-DI: 10195327207588(each), 40195327207589(case), Lot Number: 24AMI098; 1705) DYNJ82784, UDI-DI: 10195327210267(each), 40195327210268(case), Lot Number: 22IBT259; 1706) DYNJ82785, UDI-DI: 10195327210274(each), 40195327210275(case), Lot Number: 22JBG012; 1707) DYNJ83008, UDI-DI: 10195327226237(each), 40195327226238(case), Lot Number: 23GBT347; 1708) DYNJ83181, UDI-DI: 10195327240080(each), 40195327240081(case), Lot Number: 22LBD470; 1709) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23IBL690; 1710) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23GBF023; 1711) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23FBC442; 1712) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23EBC088; 1713) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 22KBI755; 1714) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 24EBE829; 1715) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23LBG771; 1716) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23IBO006; 1717) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23GBM802; 1718) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23BBS888; 1719) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 22KBR110; 1720) DYNJ83323, UDI-DI: 10195327260293(each), 40195327260294(case), Lot Number: 23IBB023; 1721) DYNJ83401, UDI-DI: 10195327264314(each), 40195327264315(case), Lot Number: 23CDC415; 1722) DYNJ83401, UDI-DI: 10195327264314(each), 40195327264315(case), Lot Number: 23ADA122; 1723) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23KDB925; 1724) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23EDA014; 1725) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23DDB319; 1726) DYNJ83683, UDI-DI: 10195327302498(each), 40195327302499(case), Lot Number: 23DBE880; 1727) DYNJ83948, UDI-DI: 10195327318154(each), 40195327318155(case), Lot Number: 23KBT912; 1728) DYNJ83999, UDI-DI: 10195327325428(each), 40195327325429(case), Lot Number: 23DLB004; 1729) DYNJ84473, UDI-DI: 10195327368975(each), 40195327368976(case), Lot Number: 23ILA147; 1730) DYNJ84648A, UDI-DI: 10195327449940(each), 40195327449941(case), Lot Number: 23HBQ821; 1731) DYNJ84937B, UDI-DI: 10198459089510(each), 40198459089511(case), Lot Number: 24KMB597; 1732) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25KBF640; 1733) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25IBQ365; 1734) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25GBK279; 1735) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25ABM165; 1736) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 24DBT813; 1737) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 23IBH150; 1738) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25KBL958; 1739) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25HBA955; 1740) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25FBH505; 1741) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25EBK149; 1742) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 24KBB114; 1743) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 23IBK215; 1744) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25JBM584; 1745) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25IBP658; 1746) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25DBL816; 1747) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25ABM160; 1748) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 24LBA279; 1749) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 24DBO515; 1750) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 23IBK219; 1751) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25LBM028; 1752) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25KBJ312; 1753) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25GBU913; 1754) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25EBH468; 1755) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24JBE493; 1756) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24HBS004; 1757) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24GBC603; 1758) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24DBA418; 1759) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 23IBH149; 1760) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25LBQ771; 1761) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25JBJ844; 1762) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25GBR600; 1763) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25FBN620; 1764) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25DBD521; 1765) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25ABS847; 1766) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24KBM440; 1767) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24HBP882; 1768) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24GBJ336; 1769) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24GBA885; 1770) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24FBE223; 1771) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24DBE599; 1772) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24ABW604; 1773) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24ABU575; 1774) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 23IBL567; 1775) DYNJ85246, UDI-DI: 10195327482794(each), 40195327482795(case), Lot Number: 24JBA875; 1776) DYNJ85246, UDI-DI: 10195327482794(each), 40195327482795(case), Lot Number: 23KBS394; 1777) DYNJ85902, UDI-DI: 10195327570224(each), 40195327570225(case), Lot Number: 24ABW109; 1778) DYNJ85904, UDI-DI: 10195327569204(each), 40195327569205(case), Lot Number: 24ABW111; 1779) DYNJ86358, UDI-DI: 10195327623920(each), 40195327623921(case), Lot Number: 24DBT962; 1780) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25EBR951; 1781) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), L ot Number: 25DBI319; 1782) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25CBL656; 1783) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25CBD157; 1784) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25BBG889; 1785) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 24LBD035; 1786) DYNJ87523, UDI-DI: 10198459040511(each), 40198459040512(case), Lot Number: 24HBC472; 1787) DYNJ87530, UDI-DI: 10198459040207(each), 40198459040208(case), Lot Number: 24HBO727; 1788) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 25CBT519; 1789) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 25CBS110; 1790) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 24HBB599; 1791) DYNJ87922, UDI-DI: 10198459078668(each), 40198459078669(case), Lot Number: 24JBC382; 1792) DYNJ87922, UDI-DI: 10198459078668(each), 40198459078669(case), Lot Number: 24HBH778; 1793) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25LBI670; 1794) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25JBR829; 1795) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25GBX075; 1796) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25GBG555; 1797) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25CBC626; 1798) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 24LBK468; 1799) DYNJ88343, UDI-DI: 10198459146541(each), 40198459146542(case), Lot Number: 25FBN319; 1800) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25JBW661; 1801) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25CBS371; 1802) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25BBO562; 1803) DYNJ900312F, UDI-DI: 10195327698010(each), 40195327698011(case), Lot Number: 24LMH420; 1804) DYNJ900473F, UDI-DI: 10195327448141(each), 40195327448142(case), Lot Number: 23IMD804; 1805) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25GBJ998; 1806) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25EBN039; 1807) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25BBU433; 1808) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25BBM255; 1809) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25ABA922; 1810) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24LBH537; 1811) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24IBP307; 1812) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24HBR222; 1813) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24EBI630; 1814) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24DBR216; 1815) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22FDB319; 1816) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22EDB848; 1817) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22DDC460; 1818) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22DDA354; 1819) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21LDB775; 1820) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21LDA249; 1821) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21JDB219; 1822) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21HDB766; 1823) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21GDA613; 1824) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21FDA216; 1825) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21DDB318; 1826) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21BDB150; 1827) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25HDA524; 1828) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23IDA598; 1829) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23GDC145; 1830) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23FDB393; 1831) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23EDB016; 1832) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23CDC445; 1833) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23BDB808; 1834) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23ADB159; 1835) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 22KDA621; 1836) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 22IDB505; 1837) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22EDA396; 1838) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22DDB174; 1839) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22ADA779; 1840) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21KDB540; 1841) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21JDB654; 1842) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21GDA051; 1843) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21FDB070; 1844) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21BDB155; 1845) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23JDA516; 1846) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23GDA643; 1847) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23EDC178; 1848) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23EDA531; 1849) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23CDB246; 1850) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23BDA103; 1851) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22LDA717; 1852) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22KDA430; 1853) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22IDA935; 1854) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22GDB557; 1855) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22GDB004; 1856) DYNJ901293C, UDI-DI: 10193489210934(each), 40193489210935(case), Lot Number: 21BBQ036; 1857) DYNJ901293C, UDI-DI: 10193489210934(each), 40193489210935(case), Lot Number: 20LBO414; 1858) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 22GBA672; 1859) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21LBQ931; 1860) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21LBB250; 1861) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21KBR503; 1862) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21JBK076; 1863) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21HBI249; 1864) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21GBF747; 1865) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21DBP543; 1866) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23EBF123; 1867) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23EBD058; 1868) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23DBE024; 1869) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23BBU922; 1870) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23BBF811; 1871) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBK712; 1872) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBK367; 1873) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBB795; 1874) DYNJ901293G, UDI-DI: 10195327405267(each), 40195327405268(case), Lot Number: 23KBX231; 1875) DYNJ901293G, UDI-DI: 10195327405267(each), 40195327405268(case), Lot Number: 23JBM583; 1876) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24KBB223; 1877) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24IBM824; 1878) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24HBK597; 1879) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24EBO749; 1880) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24CBM164; 1881) DYNJ901293J, UDI-DI: 10198459210525(each), 40198459210526(case), Lot Number: 25ABB962; 1882) DYNJ901294D, UDI-DI: 10193489201055(each), 40193489201056(case), Lot Number: 21CBM996; 1883) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 24ABO438; 1884) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23KBW620; 1885) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23JBN868; 1886) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23IBI898; 1887) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23HBJ256; 1888) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23FBO393; 1889) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23DBS344; 1890) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23BBF812; 1891) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 22JBI903; 1892) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24KBP826; 1893) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24IBB759; 1894) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24FBL612; 1895) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24EBM171; 1896) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24DBE522; 1897) DYNJ901294I, UDI-DI: 10198459210471(each), 40198459210472(case), Lot Number: 25FBR246; 1898) DYNJ901294I, UDI-DI: 10198459210471(each), 40198459210472(case), Lot Number: 25ABO543; 1899) DYNJ901312G, UDI-DI: 10195327462543(each), 40195327462544(case), Lot Number: 23IMA558; 1900) DYNJ901364G, UDI-DI: 10193489656060(each), 40193489656061(case), Lot Number: 24JLA825; 1901) DYNJ901364I, UDI-DI: 10198459224874(each), 40198459224875(case), Lot Number: 25ILA425; 1902) DYNJ901485J, UDI-DI: 10195327284411(each), 40195327284412(case), Lot Number: 23HMI283; 1903) DYNJ901486F, UDI-DI: 10195327284428(each), 40195327284429(case), Lot Number: 23HMH007; 1904) DYNJ901515C, UDI-DI: 10193489732436(each), 40193489732437(case), Lot Number: 21CBV915; 1905) DYNJ901752M, UDI-DI: 10193489604580(each), 40193489604581(case), Lot Number: 21ADB750; 1906) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 25ABN022; 1907) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24LBN814; 1908) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24HBU924; 1909) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24GBC321; 1910) DYNJ901883M, UDI-DI: 10198459288807(each), 40198459288808(case), Lot Number: 25DBR925; 1911) DYNJ902019J, UDI-DI: 10195327394424(each), 40195327394425(case), Lot Number: 23IBP535; 1912) DYNJ902019J, UDI-DI: 10195327394424(each), 40195327394425(case), Lot Number: 23IBP386; 1913) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 25DBG926; 1914) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21GBE547; 1915) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21DBC079; 1916) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21BBQ998; 1917) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22OBA319; 1918) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22FBG142; 1919) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22DBM174; 1920) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22ABT877; 1921) DYNJ902247G, UDI-DI: 10195327189587(each), 40195327189588(case), Lot Number: 22KBS876; 1922) DYNJ902247G, UDI-DI: 10195327189587(each), 40195327189588(case), Lot Number: 22IBN886; 1923) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24JBS825; 1924) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24IBB754; 1925) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24DBB200; 1926) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24BBJ347; 1927) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23LBN371; 1928) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23JBX532; 1929) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23GBI471; 1930) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23FBF178; 1931) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23ABP539; 1932) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25FBL542; 1933) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25BBM532; 1934) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25ABC841; 1935) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 24LBP686; 1936) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25KBE472; 1937) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25JBO058; 1938) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25JBD155; 1939) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22OBM158; 1940) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22JBE981; 1941) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22GBF980; 1942) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22FBW219; 1943) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22DBL831; 1944) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22ABH935; 1945) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22ABH933; 1946) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 21LBC621; 1947) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 21IBC819; 1948) DYNJ902308D, UDI-DI: 10195327276737(each), 40195327276738(case), Lot Number: 22LBF699; 1949) DYNJ902308F, UDI-DI: 10195327326937(each), 40195327326938(case), Lot Number: 23DBJ927; 1950) DYNJ902308F, UDI-DI: 10195327326937(each), 40195327326938(case), Lot Number: 23DBF023; 1951) DYNJ902308G, UDI-DI: 10195327412173(each), 40195327412174(case), Lot Number: 23FBH153; 1952) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24DBU297; 1953) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24BBG964; 1954) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24ABR435; 1955) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23LBT893; 1956) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23KBT585; 1957) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23IBJ191; 1958) DYNJ902310D, UDI-DI: 10193489315257(each), 40193489315258(case), Lot Number: 21ADB455; 1959) DYNJ902310D, UDI-DI: 10193489315257(each), 40193489315258(case), Lot Number: 21ADB189; 1960) DYNJ902321M, UDI-DI: 10193489863956(each), 40193489863957(case), Lot Number: 21DKA251; 1961) DYNJ902352D, UDI-DI: 10889942804177(each), 40889942804178(case), Lot Number: 21BDA474; 1962) DYNJ902362I, UDI-DI: 10195327510497(each), 40195327510498(case), Lot Number: 24BDA089; 1963) DYNJ902362I, UDI-DI: 10195327510497(each), 40195327510498(case), Lot Number: 24ADB157; 1964) DYNJ902366A, UDI-DI: 10193489616255(each), 40193489616256(case), Lot Number: 21CBM977; 1965) DYNJ902438J, UDI-DI: 10193489761450(each), 40193489761451(case), Lot Number: 21CDB481; 1966) DYNJ902438J, UDI-DI: 10193489761450(each), 40193489761451(case), Lot Number: 21ADB962; 1967) DYNJ902473L, UDI-DI: 10195327261436(each), 40195327261437(case), Lot Number: 23KLA717; 1968) DYNJ902473L, UDI-DI: 10195327261436(each), 40195327261437(case), Lot Number: 23KLA364; 1969) DYNJ902567I, UDI-DI: 10195327388577(each), 40195327388578(case), Lot Number: 23HMG997; 1970) DYNJ902665O, UDI-DI: 10195327421380(each), 40195327421381(case), Lot Number: 23IMI035; 1971) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22LMB722; 1972) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22IMH146; 1973) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22FMB393; 1974) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22EMD118; 1975) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22CMD603; 1976) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 21GMF509; 1977) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 21FMB735; 1978) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25FME478; 1979) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25CMD002; 1980) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25AME975; 1981) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24JMA066; 1982) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24IMF329; 1983) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24HMC925; 1984) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24EMH532; 1985) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24EMC395; 1986) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24CMJ425; 1987) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24CMG427; 1988) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23LMG552; 1989) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23JME975; 1990) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23JMA960; 1991) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23HME545; 1992) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23GMC011; 1993) DYNJ902916I, UDI-DI: 10195327356989(each), 40195327356980(case), Lot Number: 23JDB783; 1994) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23LBV796; 1995) DYNJ903053D, UDI-DI: 10193489712568(each), 40193489712569(case), Lot Number: 23AMA115; 1996) DYNJ903175C, UDI-DI: 10195327435912(each), 40195327435913(case), Lot Number: 23LBT799; 1997) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23KBM048; 1998) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25DBP020; 1999) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25CBU462; 2000) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25ABA361; 2001) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24JBY080; 2002) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24CBD528; 2003) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24ABV428; 2004) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25KBT140; 2005) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25JBP549; 2006) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25JBM545; 2007) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23EBR769; 2008) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23EBL640; 2009) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23CBU917; 2010) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23CBG054; 2011) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22IBP136; 2012) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22IBO773; 2013) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22GBE114; 2014) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22DBR197; 2015) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21LBT148; 2016) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21KBR505; 2017) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21IBP909; 2018) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21IBP908; 2019) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21GBG879; 2020) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21FBE274; 2021) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21DBO218; 2022) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 24BBD430; 2023) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 24ABF212; 2024) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23LBV018; 2025) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23LBN776; 2026) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24KBG005; 2027) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24DBO100; 2028) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24CBU456; 2029) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 23EBW544; 2030) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 23CBQ348; 2031) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22IBA639; 2032) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22HBH661; 2033) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22FBW884; 2034) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21JBQ513; 2035) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21IBK686; 2036) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21EBD387; 2037) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21ABI686; 2038) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 23CBN882; 2039) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22NBF998; 2040) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22IBO536; 2041) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22GBS690; 2042) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22EBA227; 2043) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21IBK141; 2044) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21FBM616; 2045) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21FBI614; 2046) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21DBM942; 2047) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21CBN613; 2048) DYNJ904021Q, UDI-DI: 10198459140440(each), 40198459140441(case), Lot Number: 24JDC025; 2049) DYNJ904073D, UDI-DI: 10195327597191(each), 40195327597192(case), Lot Number: 24LMF733; 2050) DYNJ904082J, UDI-DI: 10195327062002(each), 40195327062003(case), Lot Number: 22GBH238; 2051) DYNJ904083C, UDI-DI: 10193489282832(each), 40193489282833(case), Lot Number: 22KDA657; 2052) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23EBS007; 2053) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23DBL273; 2054) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23DBD010; 2055) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23CBK134; 2056) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23ABS790; 2057) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23ABS351; 2058) DYNJ904202D, UDI-DI: 10193489219807(each), 40193489219808(case), Lot Number: 21BDB502; 2059) DYNJ904202D, UDI-DI: 10193489219807(each), 40193489219808(case), Lot Number: 21BDA737; 2060) DYNJ904232K, UDI-DI: 10193489834277(each), 40193489834278(case), Lot Number: 22FDA699; 2061) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21HBQ139; 2062) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21GBI161; 2063) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21DBU882; 2064) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 24ABI434; 2065) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23LBP282; 2066) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23LBJ439; 2067) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23JBW627; 2068) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23IBA527; 2069) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23HBO357; 2070) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23GBM309; 2071) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23GBL919; 2072) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23EBM510; 2073) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23CBD237; 2074) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23BBL710; 2075) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23ABC466; 2076) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22KBK520; 2077) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22HBS687; 2078) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO762; 2079) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO761; 2080) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO760; 2081) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22EBG657; 2082) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22DBI520; 2083) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22DBH466; 2084) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22CBD647; 2085) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22BBD028; 2086) DYNJ904256D, UDI-DI: 10195327173814(each), 40195327173815(case), Lot Number: 24HBJ793; 2087) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21JBU284; 2088) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21GBJ327; 2089) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21EBN308; 2090) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21EBC430; 2091) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21DBF301; 2092) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 23CBI402; 2093) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 22IBD724; 2094) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 22DBJ147; 2095) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24GBM228; 2096) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24FBD344; 2097) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24BBD690; 2098) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23LBM413; 2099) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23KBG974; 2100) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23IBU342; 2101) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23GBM378; 2102) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23EBL642; 2103) DYNJ904458G, UDI-DI: 10193489286366(each), 40193489286367(case), Lot Number: 21CDB169; 2104) DYNJ904458G, UDI-DI: 10193489286366(each), 40193489286367(case), Lot Number: 21ADB432; 2105) DYNJ904498I, UDI-DI: 10195327105334(each), 40195327105335(case), Lot Number: 22DBT224; 2106) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24IBE071; 2107) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24HBF630; 2108) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24GBD132; 2109) DYNJ904681F, UDI-DI: 10193489803815(each), 40193489803816(case), Lot Number: 21BBB752; 2110) DYNJ904777C, UDI-DI: 10195327349349(each), 40195327349340(case), Lot Number: 23FBC935; 2111) DYNJ904777C, UDI-DI: 10195327349349(each), 40195327349340(case), Lot Number: 23EBU756; 2112) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 23JBV743; 2113) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22GBA676; 2114) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22CBZ135; 2115) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22ABW913; 2116) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21KBZ698; 2117) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21GBH416; 2118) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBL904; 2119) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBL903; 2120) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBC377; 2121) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBA001; 2122) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21CBM824; 2123) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21BBK007; 2124) DYNJ904784A, UDI-DI: 10195327190422(each), 40195327190423(case), Lot Number: 22IBA582; 2125) DYNJ904784D, UDI-DI: 10195327350864(each), 40195327350865(case), Lot Number: 23EBL415; 2126) DYNJ904784F, UDI-DI: 10195327431266(each), 40195327431267(case), Lot Number: 23HBU587; 2127) DYNJ904784F, UDI-DI: 10195327431266(each), 40195327431267(case), Lot Number: 23HBK149; 2128) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25KBI026; 2129) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25KBE198; 2130) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25GBI837; 2131) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25EBQ792; 2132) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25CBU452 ; 2133) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 24BBR413; 2134) DYNJ904843J, UDI-DI: 10193489928938(each), 40193489928939(case), Lot Number: 23KBK541; 2135) DYNJ904843J, UDI-DI: 10193489928938(each), 40193489928939(case), Lot Number: 23JBV222; 2136) DYNJ904995D, UDI-DI: 10193489735307(each), 40193489735308(case), Lot Number: 21ADB018; 2137) DYNJ905067C, UDI-DI: 10195327098681(each), 40195327098682(case), Lot Number: 24AMF906; 2138) DYNJ905304A, UDI-DI: 10193489633634(each), 40193489633635(case), Lot Number: 23FLA163; 2139) DYNJ905334A, UDI-DI: 10889942885923(each), 40889942885924(case), Lot Number: 23IBT639; 2140) DYNJ905350B, UDI-DI: 10195327039264(each), 40195327039265(case), Lot Number: 23JLA091; 2141) DYNJ905687C, UDI-DI: 10195327361624(each), 40195327361625(case), Lot Number: 24BMH064; 2142) DYNJ905750B, UDI-DI: 10193489246858(each), 40193489246859(case), Lot Number: 21ADB693; 2143) DYNJ905779, UDI-DI: 10889942971824(each), 40889942971825(case), Lot Number: 21ABT001; 2144) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25DBQ990; 2145) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25CBQ416; 2146) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25BBR363; 2147) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24JBG311; 2148) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24GBI990; 2149) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABV903; 2150) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABK092; 2151) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABG424; 2152) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23JBI205; 2153) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23FBT546; 2154) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23FBD281; 2155) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 23DBG524; 2156) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 23ABN481; 2157) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 22GBM712; 2158) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 22ABF205; 2159) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21LBV265; 2160) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21KBK295; 2161) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21HBQ156; 2162) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21GBM188; 2163) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21EBO481; 2164) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24GBL459; 2165) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24EBU549; 2166) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24BBV164; 2167) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24BBA227; 2168) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23LBV082; 2169) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23LBA464; 2170) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23KBK365; 2171) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23GBF834; 2172) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23EBV687; 2173) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23DBO515; 2174) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23BBG666; 2175) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23ABG370; 2176) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22LBT258; 2177) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22JBC736; 2178) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBG539; 2179) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBF669; 2180) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBE604; 2181) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBA089; 2182) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21LBV306; 2183) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21JBZ318; 2184) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21HBQ064; 2185) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21GBO014; 2186) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21CBJ290; 2187) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 24JBX485; 2188) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 24EBT891; 2189) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 23LBK530; 2190) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 23LBB395; 2191) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23LDB587; 2192) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23LDA751; 2193) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 24ADA142; 2194) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23LDA290; 2195) DYNJ905986I, UDI-DI: 10198459387814(each), 40198459387815(case), Lot Number: 25HDB205; 2196) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 24ADA155; 2197) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24EDC137; 2198) DYNJ905987G, UDI-DI: 10198459505225(each), 40198459505226(case), Lot Number: 25LMC401; 2199) DYNJ905987G, UDI-DI: 10198459505225(each), 40198459505226(case), Lot Number: 25IMG396; 2200) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 24ADA098; 2201) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23LDB456; 2202) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23LDA254; 2203) DYNJ906027F, UDI-DI: 10195327419578(each), 40195327419579(case), Lot Number: 23IBI649; 2204) DYNJ906061F, UDI-DI: 10193489279146(each), 40193489279147(case), Lot Number: 21IDA821; 2205) DYNJ906065F, UDI-DI: 10193489861075(each), 40193489861076(case), Lot Number: 21IDB654; 2206) DYNJ906065K, UDI-DI: 10198459051463(each), 40198459051464(case), Lot Number: 24IDB713; 2207) DYNJ906066G, UDI-DI: 10193489861082(each), 40193489861083(case), Lot Number: 22DDB504; 2208) DYNJ906118, UDI-DI: 10193489668568(each), 40193489668569(case), Lot Number: 21BDB916; 2209) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23GBE754; 2210) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23FBN758; 2211) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23EBL829; 2212) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23DBH806; 2213) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23BBE573; 2214) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 22KBM272; 2215) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 22KBA603; 2216) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 21IBU423; 2217) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 23JBS718; 2218) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 21JBP608; 2219) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 23IBO924; 2220) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 23ABC574; 2221) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 22JBU059; 2222) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 22FBF437; 2223) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 21GBM697; 2224) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 21CBB037; 2225) DYNJ906369A, UDI-DI: 10193489286991(each), 40193489286992(case), Lot Number: 21CBZ745; 2226) DYNJ906385F, UDI-DI: 10195327094751(each), 40195327094752(case), Lot Number: 23LBK526; 2227) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22JBM029; 2228) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22HBS418; 2229) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22HBD336; 2230) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22GBX091; 2231) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22EBD169; 2232) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22EBC265; 2233) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24IBC264; 2234) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24HBP008; 2235) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24GBV890; 2236) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24DBU228; 2237) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24ABU196; 2238) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23JBR108; 2239) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23HBV241; 2240) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23HBF888; 2241) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23GBM188; 2242) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23EBV656; 2243) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23BBJ904; 2244) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25FBM372; 2245) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25DBJ963; 2246) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25DBH514; 2247) DYNJ906529B, UDI-DI: 10195327162085(each), 40195327162086(case), Lot Number: 23KMB560; 2248) DYNJ906529B, UDI-DI: 10195327162085(each), 40195327162086(case), Lot Number: 23HMF718; 2249) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21CDC011; 2250) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21ADA953; 2251) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21ADA649; 2252) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 22CBP273; 2253) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21LBU091; 2254) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21IBA121; 2255) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21GBQ428; 2256) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21GBG878; 2257) DYNJ906891A, UDI-DI: 10193489390773(each), 40193489390774(case), Lot Number: 21HBK630; 2258) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22OBA312; 2259) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22JBL535; 2260) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22IBK691; 2261) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22GBF054; 2262) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22EBO622; 2263) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22DBK648; 2264) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 21IBI423; 2265) DYNJ906925A, UDI-DI: 10193489410709(each), 40193489410700(case), Lot Number: 21BBP039; 2266) DYNJ906932A, UDI-DI: 10193489411812(each), 40193489411813(case), Lot Number: 21CBE124; 2267) DYNJ906932A, UDI-DI: 10193489411812(each), 40193489411813(case), Lot Number: 21ABE629; 2268) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21FBF204; 2269) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21EBC594; 2270) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21DBP476; 2271) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23IBW104; 2272) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23FBU054; 2273) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23FBU011; 2274) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23DBE078; 2275) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23CBW421; 2276) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23CBG971; 2277) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22JBR766; 2278) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22IBA642; 2279) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22HBW697; 2280) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22GBL768; 2281) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22GBG496; 2282) DYNJ906934A, UDI-DI: 10193489411942(each), 40193489411943(case), Lot Number: 21CBE125; 2283) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 24ABK789; 2284) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23LBQ687; 2285) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23JBQ233; 2286) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23IBI732; 2287) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23FBU055; 2288) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23FBU012; 2289) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23EBD628; 2290) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23BBU919; 2291) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23BBB416; 2292) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23ABG140; 2293) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22OBD787; 2294) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22NBF781; 2295) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22LBF478; 2296) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22JBF979; 2297) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22GBG314; 2298) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22EBV617; 2299) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21LBO447; 2300) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21KBD951; 2301) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21IBJ189; 2302) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21HBO551; 2303) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21GBN206; 2304) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23EBR992; 2305) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23EBQ838; 2306) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBP814; 2307) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBL573; 2308) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBK693; 2309) DYNJ906986A, UDI-DI: 10193489448481(each), 40193489448482(case), Lot Number: 22EBR476; 2310) DYNJ907145A, UDI-DI: 10193489411614(each), 40193489411615(case), Lot Number: 21DBA527; 2311) DYNJ907169A, UDI-DI: 10193489431865(each), 40193489431866(case), Lot Number: 22GBF388; 2312) DYNJ907169A, UDI-DI: 10193489431865(each), 40193489431866(case), Lot Number: 22DBN969; 2313) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24ABD872; 2314) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 23JBW513; 2315) DYNJ907366A, UDI-DI: 10195327474669(each), 40195327474660(case), Lot Number: 23ILA762; 2316) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22IMA627; 2317) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22CMB538; 2318) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22AMG437; 2319) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21IMG285; 2320) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21FME071; 2321) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21CBU592; 2322) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21ABB064; 2323) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23GMC399; 2324) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23FMG424; 2325) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23EMF510; 2326) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 24AMG458; 2327) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23HMC702; 2328) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23DMA497; 2329) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23BMI414; 2330) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 25CMI159; 2331) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 25BMF925; 2332) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 24HMB473; 2333) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 23AMF670; 2334) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22IME107; 2335) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22HMC207; 2336) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22GMA663; 2337) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22EMA492; 2338) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22CMG917; 2339) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22BMC147; 2340) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21LMG466; 2341) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21GBD502; 2342) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21CBO856; 2343) DYNJ907394B, UDI-DI: 10195327359225(each), 40195327359226(case), Lot Number: 23EMD892; 2344) DYNJ907394B, UDI-DI: 10195327359225(each), 40195327359226(case), Lot Number: 23DMH233; 2345) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 23CMD411; 2346) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 23AMF672; 2347) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22LMI031; 2348) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22KMA556; 2349) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22EMG662; 2350) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22DME010; 2351) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22DMD242; 2352) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22CMG107; 2353) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22BMA354; 2354) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22AMG438; 2355) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21LME763; 2356) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21JMG857; 2357) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21HMH140; 2358) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21FBO458; 2359) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21DBU998; 2360) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21DBK092; 2361) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21BBM403; 2362) DYNJ907402A, UDI-DI: 10195327324544(each), 40195327324545(case), Lot Number: 23FMI433; 2363) DYNJ907402A, UDI-DI: 10195327324544(each), 40195327324545(case), Lot Number: 23FMB745; 2364) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMH367; 2365) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMG263; 2366) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMF091; 2367) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23AMA389; 2368) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22LMC940; 2369) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22IMB402; 2370) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22HME124; 2371) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22FMI039; 2372) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22CMD161; 2373) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AME949; 2374) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AMD448; 2375) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AMC615; 2376) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21LMF667; 2377) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21LMA712; 2378) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21JMA294; 2379) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21HMC149; 2380) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21FBC533; 2381) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21EBC422; 2382) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21DBQ728; 2383) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21BBM502; 2384) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21BBM404; 2385) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23GMH562; 2386) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23GMA340; 2387) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23EMC989; 2388) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23LMD520; 2389) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23FMG426; 2390) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23CMD940; 2391) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 25FME364; 2392) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 25BMI451; 2393) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 24KMF521; 2394) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 24FMF042; 2395) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22JMI412; 2396) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22IMD821; 2397) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22HMC453; 2398) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22EMG725; 2399) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22DMG429; 2400) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22BME206; 2401) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22AMG111; 2402) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21JMA677; 2403) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21HMF279; 2404) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21GMB733; 2405) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21EBN820; 2406) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21EBE793; 2407) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21CBS019; 2408) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21CBF261; 2409) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 23BMG352; 2410) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22LMD329; 2411) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22KME421; 2412) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22JMG905; 2413) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22GME949; 2414) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22DME804; 2415) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22BME152; 2416) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22BMD527; 2417) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22AME098; 2418) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22AMD977; 2419) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21KMC798; 2420) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21JMB209; 2421) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21GBG470; 2422) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21GBD601; 2423) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21FBK863; 2424) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21FBH069; 2425) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21EBQ749; 2426) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21CBR937; 2427) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23HMA290; 2428) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23FME339; 2429) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23EMK083; 2430) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23DMI890; 2431) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23DMF108; 2432) DYNJ907544, UDI-DI: 10193489497298(each), 40193489497299(case), Lot Number: 21DBM325; 2433) DYNJ907683C, UDI-DI: 10195327536480(each), 40195327536481(case), Lot Number: 24ALA185; 2434) DYNJ907683D, UDI-DI: 10195327641849(each), 40195327641840(case), Lot Number: 24DLA934; 2435) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22DBO377; 2436) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22DBI117; 2437) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22CBD802; 2438) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24EBC033; 2439) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24DBB570; 2440) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24CBI490; 2441) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23KBG210; 2442) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23JBW637; 2443) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23JBK535; 2444) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24KBJ963; 2445) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24IBS859; 2446) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24HBP511; 2447) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24GBP839; 2448) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24EBT841; 2449) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24DBL919; 2450) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24CBR860; 2451) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23LBV060; 2452) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23KBE321; 2453) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23JBV731; 2454) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23JBN816; 2455) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23HBI650; 2456) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23FBD177; 2457) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23EBJ494; 2458) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23BBN968; 2459) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23ABE629; 2460) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22LBP490; 2461) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22HBS664; 2462) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22HBG360; 2463) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25KBT033; 2464) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25IBS487; 2465) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25HBJ943; 2466) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25GBE390; 2467) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25EBF761; 2468) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25CBE672; 2469) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25CBA960; 2470) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24EBA383; 2471) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24CBD690; 2472) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24ABS275; 2473) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24ABD885; 2474) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 23LBF416; 2475) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 23IBV289; 2476) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 25ABF310; 2477) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24KBD526; 2478) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24IBH056; 2479) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24HBM062; 2480) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23HBY687; 2481) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23DBI178; 2482) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23BBG123; 2483) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23ABE898; 2484) DYNJ908091A, UDI-DI: 10195327015343(each), 401953270 15344(case), Lot Number: 22LBE941; 2485) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 22KBK450; 2486) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 22JBO771; 2487) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23KBP882; 2488) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23IBA498; 2489) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23HBA413; 2490) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23GBC887; 2491) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23CBG930; 2492) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23CBE990; 2493) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23BBG422; 2494) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23ABL473; 2495) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22OBM193; 2496) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBW398; 2497) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBO246; 2498) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBH049; 2499) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22JBF721; 2500) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22IBT662; 2501) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22IBK801; 2502) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22FBM782; 2503) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22DBG490; 2504) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22CBG743; 2505) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 21LBI471; 2506) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 21KBL408; 2507) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22GLB112; 2508) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22GLB042; 2509) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLB062; 2510) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA782; 2511) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA293; 2512) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA116; 2513) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22DLA667; 2514) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22CLB151; 2515) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22ALA136; 2516) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 21LLA171; 2517) DYNJ908232, UDI-DI: 10195327013523(each), 40195327013524(case), Lot Number: 22ADB298; 2518) DYNJ908232, UDI-DI: 10195327013523(each), 40195327013524(case), Lot Number: 21KDB711; 2519) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25LBR103; 2520) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25IBF012; 2521) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25GBU152; 2522) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25FBE954; 2523) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25CBE077; 2524) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24LBO070; 2525) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24LBN862; 2526) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24HBR104; 2527) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24GBN485; 2528) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24EBN304; 2529) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24DBV190; 2530) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24DBK634; 2531) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23HBN967; 2532) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23HBK245; 2533) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23GBS766; 2534) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23CBG248; 2535) DYNJ908305, UDI-DI: 10195327030308(each), 40195327030309(case), Lot Number: 22HBY634; 2536) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22IBF959; 2537) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22IBE835; 2538) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22HBT451; 2539) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22HBA389; 2540) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22BBC124; 2541) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 24BBV260; 2542) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23LBT932; 2543) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23LBS778; 2544) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23KBR860; 2545) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23EBE317; 2546) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24JBS437; 2547) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24IBM825; 2548) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24HBV288; 2549) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24HBI970; 2550) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24FBS775; 2551) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25DBL478; 2552) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25CBD533; 2553) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 24LBU118; 2554) DYNJ908515B, UDI-DI: 10195327527112(each), 40195327527113(case), Lot Number: 24BMJ733; 2555) DYNJ908515B, UDI-DI: 10195327527112(each), 40195327527113(case), Lot Number: 24BMF113; 2556) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 24ABV074; 2557) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23GBB423; 2558) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23DBD850; 2559) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23BBU902; 2560) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23ABM615; 2561) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23ABC148; 2562) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 22GBD311; 2563) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 22DBG283; 2564) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23JBY557; 2565) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23IBF496; 2566) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23HBO429; 2567) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23GBI824; 2568) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23EBT798; 2569) DYNJ908644D, UDI-DI: 10195327404307(each), 40195327404308(case), Lot Number: 23HBF890; 2570) DYNJ908644D, UDI-DI: 10195327404307(each), 40195327404308(case), Lot Number: 23GBG730; 2571) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24HBH035; 2572) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24GBC844; 2573) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24EBD242; 2574) DYNJ908703B, UDI-DI: 10198459091599(each), 40198459091590(case), Lot Number: 24JBW648; 2575) DYNJ908703C, UDI-DI: 10198459229633(each), 40198459229634(case), Lot Number: 25ABV138; 2576) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25GBO748; 2577) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25EBS805; 2578) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25DBE997; 2579) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25CBC210; 2580) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24JBF640; 2581) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24IBS872; 2582) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24GBT134; 2583) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24EBJ708; 2584) DYNJ908766A, UDI-DI: 10195327267834(each), 40195327267835(case), Lot Number: 25GLB098; 2585) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 24BBK238; 2586) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23KBW175; 2587) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23KBM022; 2588) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23JBY247; 2589) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23GBU940; 2590) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23GBU776; 2591) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23EBJ221; 2592) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23ABS838; 2593) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23ABC149; 2594) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 22LBJ720; 2595) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 22IBO061; 2596) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25JBQ642; 2597) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25HBR514; 2598) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25CBP163; 2599) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25CBG036; 2600) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25BBS821; 2601) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25ABF284; 2602) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 24LBD495; 2603) DYNJ908861, UDI-DI: 10195327195281(each), 40195327195282(case), Lot Number: 23GBH518; 2604) DYNJ909027B, UDI-DI: 10195327617301(each), 40195327617302(case), Lot Number: 25ALA334; 2605) DYNJ909027B, UDI-DI: 10195327617301(each), 40195327617302(case), Lot Number: 24KLA186; 2606) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23DMF511; 2607) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23CME124; 2608) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23BME735; 2609) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMI315; 2610) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMF978; 2611) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMB428; 2612) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 22LMB692; 2613) DYNJ909142D, UDI-DI: 10195327696115(each), 40195327696116(case), Lot Number: 24JMG326; 2614) DYNJ909142D, UDI-DI: 10195327696115(each), 40195327696116(case), Lot Number: 24IMH986; 2615) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25CMF877; 2616) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25BMI753; 2617) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25AMI497; 2618) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25AMA338; 2619) DYNJ909142G, UDI-DI: 10198459354496(each), 40198459354497(case), Lot Number: 25HMH554; 2620) DYNJ909142G, UDI-DI: 10198459354496(each), 40198459354497(case), Lot Number: 25FME290; 2621) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23GBG727; 2622) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23FBB682; 2623) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23DBG095; 2624) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23FMD557; 2625) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23EMF267; 2626) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23EME409; 2627) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24IMA563; 2628) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24HMB850; 2629) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24GMH511; 2630) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24CMG434; 2631) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24BMD766; 2632) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24AMA487; 2633) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 23LMA802; 2634) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 23JME303; 2635) DYNJ909505B, UDI-DI: 10198459124440(each), 40198459124441(case), Lot Number: 24JMI945; 2636) DYNJ909505C, UDI-DI: 10198459173561(each), 40198459173562(case), Lot Number: 25DMH508; 2637) DYNJ909505D, UDI-DI: 10198459594229(each), 40198459594220(case), Lot Number: 25LMB637; 2638) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 24ILA570; 2639) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 23GLA881; 2640) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 23GLA007; 2641) DYNJ909510A, UDI-DI: 10198459124457(each), 40198459124458(case), Lot Number: 24JLB326; 2642) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25ELA490; 2643) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25DLA665; 2644) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25DLA162; 2645) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25GBC518; 2646) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25CBO176; 2647) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25ABK779; 2648) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24LBH562; 2649) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24JBW034; 2650) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24JBI640; 2651) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24HBD480; 2652) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24GBI979; 2653) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24CBB520; 2654) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25HLA515; 2655) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25FLA990; 2656) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25ELA790; 2657) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25CLB033; 2658) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25BLA450; 2659) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24LLA402; 2660) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24KLB020; 2661) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24HLA301; 2662) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24DLA572; 2663) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24DLA175; 2664) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24ALB062; 2665) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25IME399; 2666) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25HMF682; 2667) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25CMH632; 2668) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25BME883; 2669) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25AMI296; 2670) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24LMA925; 2671) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24JMF321; 2672) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24DMA086; 2673) DYNJ909627, UDI-DI: 10195327415198(each), 40195327415199(case), Lot Number: 23IBL905; 2674) DYNJ909630, UDI-DI: 10195327416843(each), 40195327416844(case), Lot Number: 23IBK293; 2675) DYNJ909630, UDI-DI: 10195327416843(each), 40195327416844(case), Lot Number: 23GBR009; 2676) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24DBB572; 2677) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24CBP841; 2678) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24CBK434; 2679) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23LBC689; 2680) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23JBV722; 2681) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23JBO869; 2682) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23IBR224; 2683) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24CBI514; 2684) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24BBL043; 2685) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24ABH462; 2686) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 23LBN237; 2687) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 23HBV261; 2688) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25IBD880; 2689) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25HBA067; 2690) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25GBO024; 2691) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25EBL872; 2692) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25DBF495; 2693) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25ABF309; 2694) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24HBM093; 2695) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24GBP217; 2696) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24DBR221; 2697) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABS346; 2698) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABC094; 2699) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABC089; 2700) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24JBN452; 2701) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24GBM375; 2702) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24FBH757; 2703) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24EBB383; 2704) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24DBM437; 2705) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24DBI478; 2706) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24CBO933; 2707) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24CBG292; 2708) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 23KBM218; 2709) DYNJ909710A, UDI-DI: 10198459327674(each), 40198459327675(case), Lot Number: 25GBE494; 2710) DYNJ909710A, UDI-DI: 10198459327674(each), 40198459327675(case), Lot Number: 25FBE958; 2711) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25KBO393; 2712) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25HBM092; 2713) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25GBR226; 2714) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 24AMC915; 2715) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23KMF164; 2716) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23IMD312; 2717) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23HMH032; 2718) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24CMG034; 2719) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24CME898; 2720) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24BMK233; 2721) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24BMH621; 2722) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24AMI270; 2723) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 23LMA320; 2724) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 23KME642; 2725) DYNJ909750B, UDI-DI: 10195327543525(each), 40195327543526(case), Lot Number: 23LBG699; 2726) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24EBC579; 2727) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24CBQ758; 2728) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24BBP450; 2729) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 25GMA576; 2730) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 25EMG060; 2731) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24JMD236; 2732) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24EMH866; 2733) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24BMF839; 2734) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24AMD056; 2735) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 23LMB636; 2736) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24GMF289; 2737) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24AMA700; 2738) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25GBT129; 2739) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25FBB523; 2740) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25DBS303; 2741) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25CBB461; 2742) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24LBR437; 2743) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24LBA993; 2744) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24KBK896; 2745) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24KBF862; 2746) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24CBP577; 2747) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25JBW820; 2748) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25JBH697; 2749) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25FBP016; 2750) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25EBE470; 2751) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25CBT525; 2752) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24FBL642; 2753) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24CBN438; 2754) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25LBS897; 2755) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25KBC349; 2756) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25GBO042; 2757) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25FBF532; 2758) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25DBI472; 2759) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25DBB387; 2760) DYNJ910326, UDI-DI: 10195327628154(each), 40195327628155(case), Lot Number: 24DBM380; 2761) DYNJ910326B, UDI-DI: 10198459075001(each), 40198459075002(case), Lot Number: 24KBQ357; 2762) DYNJ910326B, UDI-DI: 10198459075001(each), 40198459075002(case), Lot Number: 24HBL975; 2763) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25DBC255; 2764) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25CBT588; 2765) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25CBT572; 2766) DYNJ910358, UDI-DI: 10195327648015(each), 40195327648016(case), Lot Number: 24DDC162; 2767) DYNJ910416, UDI-DI: 10195327680954(each), 40195327680955(case), Lot Number: 24EBA717; 2768) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25IMG659; 2769) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25GMB870; 2770) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25DMI838; 2771) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25DMB038; 2772) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25BMD967; 2773) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25AMA256; 2774) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24LMH172; 2775) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24KMH139; 2776) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24KMD003; 2777) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24GMG658; 2778) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24FMD548; 2779) DYNJ910431A, UDI-DI: 10198459588051(each), 40198459588052(case), Lot Number: 25LMG122; 2780) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25BBD954; 2781) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25BBD805; 2782) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 24FBP464; 2783) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 25BBC824; 2784) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 24LBH399; 2785) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 24JBR694; 2786) DYNJ910578C, UDI-DI: 10198459272653(each), 40198459272654(case), Lot Number: 25DBB506; 2787) DYNJ910578C, UDI-DI: 10198459272653(each), 40198459272654(case), Lot Number: 25CBG904; 2788) DYNJ910678A, UDI-DI: 10198459336737(each), 40198459336738(case), Lot Number: 25GBN594; 2789) DYNJ910678A, UDI-DI: 10198459336737(each), 40198459336738(case), Lot Number: 25FBG095; 2790) DYNJ910732C, UDI-DI: 10198459494062(each), 40198459494063(case), Lot Number: 25LME962; 2791) DYNJ910732C, UDI-DI: 10198459494062(each), 40198459494063(case), Lot Number: 25JML588; 2792) DYNJ910733C, UDI-DI: 10198459494079(each), 40198459494070(case), Lot Number: 25LME692; 2793) DYNJ910733C, UDI-DI: 10198459494079(each), 40198459494070(case), Lot Number: 25JBB540; 2794) DYNJ910864, UDI-DI: 10198459161308(each), 40198459161309(case), Lot Number: 24KBB660; 2795) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25LMA455; 2796) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25HME905; 2797) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25EMF898; 2798) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25IBJ122; 2799) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25GBM729; 2800) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25FBD829; 2801) DYNJ911026B, UDI-DI: 10198459530531(each), 40198459530532(case), Lot Number: 25KBC353; 2802) DYNJ911032, UDI-DI: 10198459234460(each), 40198459234461(case), Lot Number: 25BBA642; 2803) DYNJ911032A, UDI-DI: 10198459338229(each), 40198459338220(case), Lot Number: 25KBM944; 2804) DYNJ911032A, UDI-DI: 10198459338229(each), 40198459338220(case), Lot Number: 25GBN709; 2805) DYNJ911355, UDI-DI: 10198459326288(each), 40198459326289(case), Lot Number: 25EBV320; 2806) DYNJ911356, UDI-DI: 10198459326509(each), 40198459326500(case), Lot Number: 25KBT505; 2807) DYNJ911356, UDI-DI: 10198459326509(each), 40198459326500(case), Lot Number: 25JBA531; 2808) DYNJ911644, UDI-DI: 10198459502200(each), 40198459502201(case), Lot Number: 25IBP089; 2809) DYNJ911838, UDI-DI: 10198459552786(each), 40198459552787(case), Lot Number: 25LBM762; 2810) DYNJ911838, UDI-DI: 10198459552786(each), 40198459552787(case), Lot Number: 25KBE192; 2811) DYNJ911956, UDI-DI: 10198459587726(each), 40198459587727(case), Lot Number: 25LBF532; 2812) DYNJ911956, UDI-DI: 10198459587726(each), 40198459587727(case), Lot Number: 25LBA968; 2813) DYNJ9682800V, UDI-DI: 10195327257248(each), 40195327257249(case), Lot Number: 22KDA037; 2814) DYNJC9306L, UDI-DI: 10195327478148(each), 40195327478149(case), Lot Number: 23IMD838; 2815) DYNJG901000A, UDI-DI: 10195327677077(each), 40195327677078(case), Lot Number: 25DLA778; 2816) DYNJS3013, UDI-DI: 10884389201572(each), 40884389201573(case), Lot Number: 22DBO761; 2817) DYNJT2144R, UDI-DI: 10889942399192(each), 40889942399193(case), Lot Number: 20LMF883; 2818) DYNJT3352, UDI-DI: 10198459324123(each), 40198459324124(case), Lot Number: 25EBJ794; 2819) DYNJT3397, UDI-DI: 10198459324918(each), 40198459324919(case), Lot Number: 25FBN276; 2820) DYNJT3411, UDI-DI: 10198459325052(each), 40198459325053(case), Lot Number: 25FBN247; 2821) DYNJT7166, UDI-DI: 10198459602740(each), 40198459602741(case), Lot Number: 25KMI656; 2822) DYNJV0056AA, UDI-DI: 10198459489112(each), 40198459489113(case), Lot Number: 25LMC937; 2823) DYNJV0154AB, UDI-DI: 10193489336191(each), 40193489336192(case), Lot Number: 21BBM420; 2824) DYNJVB91045, UDI-DI: 10193489284478(each), 40193489284479(case), Lot Number: 21BDB808; 2825) DYNJVB91128D, UDI-DI: 10195327367916(each), 40195327367917(case), Lot Number: 23JLA241; 2826) HM280B, UDI-DI: 10195327481179(each), 40195327481170(case), Lot Number: 23JBT647; 2827) OPS767M, UDI-DI: 10193489552966(each), 40193489552967(case), Lot Number: 21HDA963; 2828) OPS767M, UDI-DI: 10193489552966(each), 40193489552967(case), Lot Number: 21BDA327; 2829) PHS389811I, UDI-DI: 10193489333657(each), 40193489333658(case), Lot Number: 22AME448; 2830) PHS389913T, UDI-DI: 10193489333671(each), 40193489333672(case), Lot Number: 20LMF727; 2831) PHS389913U, UDI-DI: 10193489453287(each), 40193489453288(case), Lot Number: 21BMC221; 2832) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22HMF957; 2833) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22HMF956; 2834) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22EMB748; 2835) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22DMC685; 2836) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22AMD873; 2837) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21KME416; 2838) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21JMB398; 2839) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21HMB975; 2840) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21FMB052; 2841) PHS393071004C, UDI-DI: 10193489222623(each), 40193489222624(case), Lot Number: 21ADA094; 2842) PHS393205012C, UDI-DI: 10193489222852(each), 40193489222853(case), Lot Number: 21ADA066; 2843) PHS518775J, UDI-DI: 10195327034627(each), 40195327034628(case), Lot Number: 22HMG648; 2844) PHS807386J, UDI-DI: 10193489927078(each), 40193489927079(case), Lot Number: 23HBM378
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1421-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | February 25, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 117907 units |