Medline Convenience Kits: 1) ARTHROSCOPY CDS recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
Medline Convenience Kits: 1) ARTHROSCOPY CDS, Model Number: CDS984609B; 2) ARTHROSCOPY TRAYS, Model Number: DYNJ22501G; 3) ARTHROSCOPY PACK, Model Number: DYNJ32715F; 4) CHS ARTHROSCOPY PACK, Model Number: DYNJ50834C; 5) ARTHROSCOPY TRAY-LF, Model Number: DYNJ51069
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide.
How to tell if yours is affected
1) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBL738; 2) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 23BBJ847; 3) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBF936; 4) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22IBE795; 5) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22FBP788; 6) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22BBD031; 7) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 22ABJ831; 8) CDS984609B, UDI-DI: 10889942719372(each), 40889942719373(case), Lot Number: 21JBJ732; 9) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21CBZ902; 10) DYNJ22501G, UDI-DI: 10889942234110(each), 40889942234111(case), Lot Number: 21ABK153; 11) DYNJ32715F, UDI-DI: 10889942530038(each), 40889942530039(case), Lot Number: 21ADB529; 12) DYNJ50834C, UDI-DI: 10889942568178(each), 40889942568179(case), Lot Number: 21BDA710; 13) DYNJ51069, UDI-DI: 10889942118410(each), 40889942118411(case), Lot Number: 21BDB116
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1422-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | February 25, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 175 units |