GE Healthcare Centricity Universal Viewer Zero Footprint Client recall (GE Healthcare)

Medical deviceClass IIOngoing

Summary: GE Healthcare is recalling GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological. Reason: Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,.

What is being recalled

GE Healthcare Centricity Universal Viewer Zero Footprint Client, Model Numbers: 2110344-039, 2110344-046, 2110344-047; System, Image Processing, Radiological

Why

Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Belarus, Brazil, Canada, China, France, Georgia, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Lebanon, Malaysia, Netherlands, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkiye, United Arab Emirates, United Kingdom, Vietnam. (states named: Georgia)

How to tell if yours is affected

1) 2110344-039, UDI/DI 00840682102988, System ID Numbers: 105906-2-Centricity Universal Viewer Zero Footprint / ZFP-02312614, 4491-1-Centricity Universal Viewer Zero Footprint / ZFP-22180897, 10908-1-Centricity Universal Viewer Zero Footprint / ZFP-25199248, 10908-7-Centricity Universal Viewer Zero Footprint / ZFP-22203146, 10908-7-Centricity Universal Viewer Zero Footprint / ZFP-25188736, 4263-3-Centricity Universal Viewer Zero Footprint / ZFP-00106869, 4263-3-Centricity Universal Viewer Zero Footprint / ZFP-00106870, 4263-5-Centricity Universal Viewer Zero Footprint / ZFP-01300798, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25222931, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248683, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248688, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248690, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248695, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248713, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-25248716, 3064-15-Centricity Universal Viewer Zero Footprint / ZFP-26608643, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26580437, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26694204, 10458-9-Centricity Universal Viewer Zero Footprint / ZFP-26694229, 105197-2-Centricity Universal Viewer Zero Footprint / ZFP-20813950, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210052, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210057, 4642-6-Centricity Universal Viewer Zero Footprint / ZFP-25210107, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-20900298, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-21180523, 103829-6-Centricity Universal Viewer Zero Footprint / ZFP-21180525, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26390526, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26694221, 3212-5-Centricity Universal Viewer Zero Footprint / ZFP-26694222, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-24463042, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-25239139, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-25239140, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-26714200, 10360-3-Centricity Universal Viewer Zero Footprint / ZFP-26714202, 3471-36-Centricity Universal Viewer Zero Footprint / ZFP-24399434, 3471-36-Centricity Universal Viewer Zero Footprint / ZFP-26295812, 1198-18-Centricity Universal Viewer Zero Footprint / ZFP-02348419, 1198-18-Centricity Universal Viewer Zero Footprint / ZFP-18990433, 1198-78-Centricity Universal Viewer Zero Footprint / ZFP-22223452, 100581-1-Centricity Universal Viewer Zero Footprint / ZFP-25168130, 8746-1-Centricity Universal Viewer Zero Footprint / ZFP-23900206, 8746-1-Centricity Universal Viewer Zero Footprint / ZFP-21318104, 1207-14-Centricity Universal Viewer Zero Footprint / ZFP-24836084, 2123-19-Centricity Universal Viewer Zero Footprint / ZFP-24836053, 102901-1-Centricity Universal Viewer Zero Footprint / ZFP-23251238, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466083, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-26132903, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-26132895, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466076, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466077, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466078, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466079, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-24466088, 3922-18-Centricity Universal Viewer Zero Footprint / ZFP-00098228, 103696-33-Centricity Universal Viewer Zero Footprint / ZFP-20098431, 103696-4-Centricity Universal Viewer Zero Footprint / ZFP-20036715, 4578-1-Centricity Universal Viewer Zero Footprint / ZFP-25014571, 100468-1-Centricity Universal Viewer Zero Footprint / ZFP-24480189, 2440-1-Centricity Universal Viewer Zero Footprint / ZFP-22206970, 613493-Centricity Universal Viewer Zero Footprint / ZFP-26356112, 672721-Centricity Universal Viewer Zero Footprint / ZFP-25151288, 313741-Centricity Universal Viewer Zero Footprint / ZFP-26548614, 101223-6-Centricity Universal Viewer Zero Footprint / ZFP-20569855, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575531, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575532, 101223-22-Centricity Universal Viewer Zero Footprint / ZFP-20575543, 101223-16-Centricity Universal Viewer Zero Footprint / ZFP-20575533, 101223-16-Centricity Universal Viewer Zero Footprint / ZFP-20575535, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02283242, 283039-Centricity Universal Viewer Zero Footprint / ZFP-02170574, 283039-Centricity Universal Viewer Zero Footprint / ZFP-01819807, 104894-1-Centricity Universal Viewer Zero Footprint / ZFP-01225123, 104894-1-Centricity Universal Viewer Zero Footprint / ZFP-01225124, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21180575, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21200874, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265320, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265321, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265322, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21200873, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265323, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01225125, 659003-Centricity Universal Viewer Zero Footprint / ZFP-01265343, 659003-Centricity Universal Viewer Zero Footprint / ZFP-21207211, 1539601-Centricity Universal Viewer Zero Footprint / ZFP-26379461, A41676-Centricity Universal Viewer Zero Footprint / ZFP-21214076, 246573-Centricity Universal Viewer Zero Footprint / ZFP-26617853, M40455-Centricity Universal Viewer Zero Footprint / ZFP-00044391, C41917-Centricity Universal Viewer Zero Footprint / ZFP-00117236, D41917-Centricity Universal Viewer Zero Footprint / ZFP-00117239, M42001-Centricity Universal Viewer Zero Footprint / ZFP-02394845, 16921400-Centricity Universal Viewer Zero Footprint / ZFP-25194303, 509982-Centricity Universal Viewer Zero Footprint / ZFP-26608620, 509982-Centricity Universal Viewer Zero Footprint / ZFP-19670323, HC0220-Centricity Universal Viewer Zero Footprint / ZFP-01286629, HC0220-Centricity Universal Viewer Zero Footprint / ZFP-26660354, HC2720-Centricity Universal Viewer Zero Footprint / ZFP-00163499, UC2847-Centricity Universal Viewer Zero Footprint / ZFP-00260806, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-01461578, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703904, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703931, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-26703990, 600791-Centricity Universal Viewer Zero Footprint / ZFP-24944069, 600791-Centricity Universal Viewer Zero Footprint / ZFP-26859675, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-24467504, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-02286748, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624821, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624822, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624823, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624824, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624826, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624828, UC1989-Centricity Universal Viewer Zero Footprint / ZFP-26624829, 351738-Centricity Universal Viewer Zero Footprint / ZFP-26304641, 434487-Centricity Universal Viewer Zero Footprint / ZFP-22169581, 722453-Centricity Universal Viewer Zero Footprint / ZFP-25287444, 560496-Centricity Universal Viewer Zero Footprint / ZFP-26748573, HCNE01-Centricity Universal Viewer Zero Footprint / ZFP-20108177, 1257845X - 434487-Centricity Universal Viewer Zero Footprint / ZFP-22203508, 2010846X - 601860-Centricity Universal Viewer Zero Footprint / ZFP-24336091, HC0565-Centricity Universal Viewer Zero Footprint / ZFP-26608610, 4625913X - HC0565-Centricity Universal Viewer Zero Footprint / ZFP-25017402, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-00001140, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666351, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666352, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666354, X9001282-Centricity Universal Viewer Zero Footprint / ZFP-26666356, 4790503-Centricity Universal Viewer Zero Footprint / ZFP-26599556, 12711614-Centricity Universal Viewer Zero Footprint / ZFP-26449902, 81077-Centricity Universal Viewer Zero Footprint / ZFP-25304950, 81077-Centricity Universal Viewer Zero Footprint / ZFP-25329396, 591292-Centricity Universal Viewer Zero Footprint / ZFP-21198479, B53686-Centricity Universal Viewer Zero Footprint / ZFP-01271448, D58105-Centricity Universal Viewer Zero Footprint / ZFP-25204625, 16872968-Centricity Universal Viewer Zero Footprint / ZFP-24855625, 16872968-Centricity Universal Viewer Zero Footprint / ZFP-26929832, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24448019, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-26640843, 16999320-Centricity Universal Viewer Zero Footprint / ZFP-24833747, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24434112, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24439350, 553884-Centricity Universal Viewer Zero Footprint / ZFP-25203487, 553884-Centricity Universal Viewer Zero Footprint / ZFP-26906289, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24838806, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-24305476, 11388609-Centricity Universal Viewer Zero Footprint / ZFP-26127359, JP-283699-Centricity Universal Viewer Zero Footprint / ZFP-20978978, JP-260056-Centricity Universal Viewer Zero Footprint / ZFP-20587825, JP-330087-Centricity Universal Viewer Zero Footprint / ZFP-24931880, 851228-Centricity Universal Viewer Zero Footprint / ZFP-26136995, KW1018-Centricity Universal Viewer Zero Footprint / ZFP-26917029, KW1018-Centricity Universal Viewer Zero Footprint / ZFP-26917028, 60020-Centricity Universal Viewer Zero Footprint / ZFP-01819295, 17026623-Centricity Universal Viewer Zero Footprint / ZFP-25248193, 4510134-Centricity Universal Viewer Zero Footprint / ZFP-24525506, 17191257-Centricity Universal Viewer Zero Footprint / ZFP-26628542, 17066964-Centricity Universal Viewer Zero Footprint / ZFP-24484202, 1886799-Centricity Universal Viewer Zero Footprint / ZFP-25169256, 16951628-Centricity Universal Viewer Zero Footprint / ZFP-26453548, SA1232-Centricity Universal Viewer Zero Footprint / ZFP-25227096, SA1233-Centricity Universal Viewer Zero Footprint / ZFP-00066573, SA1013-Centricity Universal Viewer Zero Footprint / ZFP-22879437, SA1013-Centricity Universal Viewer Zero Footprint / ZFP-24281567, 17111683-Centricity Universal Viewer Zero Footprint / ZFP-21733830, 82667-Centricity Universal Viewer Zero Footprint / ZFP-25194300, 17021465-Centricity Universal Viewer Zero Footprint / ZFP-24487930, 747887-Centricity Universal Viewer Zero Footprint / ZFP-25210479, MPX431-Centricity Universal Viewer Zero Footprint / ZFP-24465830, MEX255-Centricity Universal Viewer Zero Footprint / ZFP-26802528, GED649-Centricity Universal Viewer Zero Footprint / ZFP-19666476, 947379-Centricity Universal Viewer Zero Footprint / ZFP-24846063, O051CA01-Centricity Universal Viewer Zero Footprint / ZFP-20618153, 228493-Centricity Universal Viewer Zero Footprint / ZFP-24821472, 407237-Centricity Universal Viewer Zero Footprint / ZFP-20997395, 730140-Centricity Universal Viewer Zero Footprint / ZFP-00038193, 0853550198-Centricity Universal Viewer Zero Footprint / ZFP-02093675, 6051-Centricity Universal Viewer Zero Footprint / ZFP-00116181, 6002-Centricity Universal Viewer Zero Footprint / ZFP-00158571, 486961-Centricity Universal Viewer Zero Footprint / ZFP-25022095, AE1564-Centricity Universal Viewer Zero Footprint / ZFP-00555894, AE1564-Centricity Universal Viewer Zero Footprint / ZFP-26560468, RWEAE-Centricity Universal Viewer Zero Footprint / ZFP-00165095, 243119-Centricity Universal Viewer Zero Footprint / ZFP-21322536, 243119-Centricity Universal Viewer Zero Footprint / ZFP-26631095, 370259-Centricity Universal Viewer Zero Footprint / ZFP-25309376, 370259-Centricity Universal Viewer Zero Footprint / ZFP-26784611, 17053632-Centricity Universal Viewer Zero Footprint / ZFP-24829942, 6345592-Centricity Universal Viewer Zero Footprint / ZFP-24825373, 17152365-Centricity Universal Viewer Zero Footprint / ZFP-26632357, 17040488-Centricity Universal Viewer Zero Footprint / ZFP-26404455. 2) 2110344-046, UDI/DI 00840682102988, System ID Numbers: 6579-11-Centricity Universal Viewer Zero Footprint / ZFP-26923415, 10458-1-Centricity Universal Viewer Zero Footprint / ZFP-26898336, 3886-4-Centricity Universal Viewer Zero Footprint / ZFP-22180910, 3886-4-Centricity Universal Viewer Zero Footprint / ZFP-22180906, 9094-1-Centricity Universal Viewer Zero Footprint / ZFP-26705438, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839308, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839310, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839311, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839312, 2457-1-Centricity Universal Viewer Zero Footprint / ZFP-26839313, 4326-2-Centricity Universal Viewer Zero Footprint / ZFP-26715184, 10190-9-Centricity Universal Viewer Zero Footprint / ZFP-26850722, 4084-1-Centricity Universal Viewer Zero Footprint / ZFP-26747422, 103696-30-Centricity Universal Viewer Zero Footprint / ZFP-21237479, 4204-1-Centricity Universal Viewer Zero Footprint / ZFP-26884848, 4204-1-Centricity Universal Viewer Zero Footprint / ZFP-26884853, 17094888-Centricity Universal Viewer Zero Footprint / ZFP-26918064, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26747414, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26769258, 268882-Centricity Universal Viewer Zero Footprint / ZFP-26769261, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388499, 104991-4-Centricity Universal Viewer Zero Footprint / ZFP-21350494, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388493, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388494, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388495, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388497, 104991-6-Centricity Universal Viewer Zero Footprint / ZFP-24388498, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388490, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388491, 104991-5-Centricity Universal Viewer Zero Footprint / ZFP-24388492, A91901-Centricity Universal Viewer Zero Footprint / ZFP-24484469, 145790 - HC0399-Centricity Universal Viewer Zero Footprint / ZFP-24374380, HC0014-Centricity Universal Viewer Zero Footprint / ZFP-21359919, 255473-Centricity Universal Viewer Zero Footprint / ZFP-26865306, JP-280250-Centricity Universal Viewer Zero Footprint / ZFP-26671908, JP-283699-Centricity Universal Viewer Zero Footprint / ZFP-26689509, JP-280250-Centricity Universal Viewer Zero Footprint / ZFP-02360544, JP-272375-Centricity Universal Viewer Zero Footprint / ZFP-26918260, 347891-Centricity Universal Viewer Zero Footprint / ZFP-23297397, 347891-Centricity Universal Viewer Zero Footprint / ZFP-26880666, 17195358-Centricity Universal Viewer Zero Footprint / ZFP-23986377, RU1086-Centricity Universal Viewer Zero Footprint / ZFP-26799525, RU1282-Centricity Universal Viewer Zero Footprint / ZFP-26800953, 17121345-Centricity Universal Viewer Zero Footprint / ZFP-26796880, RU8902-Centricity Universal Viewer Zero Footprint / ZFP-17775055, SA2591-Centricity Universal Viewer Zero Footprint / ZFP-26880655, 521920-Centricity Universal Viewer Zero Footprint / ZFP-26912759, 659749-Centricity Universal Viewer Zero Footprint / ZFP-26824375, ZA2239-Centricity Universal Viewer Zero Footprint / ZFP-26805562, 16959088-Centricity Universal Viewer Zero Footprint / ZFP-26777223, GEX184-Centricity Universal Viewer Zero Footprint / ZFP-26368616, 242833-Centricity Universal Viewer Zero Footprint / ZFP-26825072, 80182-Centricity Universal Viewer Zero Footprint / ZFP-26876144, 16882165-Centricity Universal Viewer Zero Footprint / ZFP-26673881, 16880927-Centricity Universal Viewer Zero Footprint / ZFP-26818090. 3) 2110344-047, UDI/DI 00840682102988, System ID Numbers: Local DB site#000330-Centricity Universal Viewer Zero Footprint / ZFP-22138067, Local DB site#000330-Centricity Universal Viewer Zero Footprint / ZFP-22138066, JP-341108-Centricity Universal Viewer Zero Footprint / ZFP-25012607, JP-220062-Centricity Universal Viewer Zero Footprint / ZFP-24475044.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1475-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 16, 2026
ReportedMarch 11, 2026
CompanyGE Healthcare
Company locationChicago, IL, United States
Initiated byVoluntary: Firm initiated
Quantity250 units