cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample... recall (Roche Diagnostics Operations, Inc.)

Medical deviceClass IIOngoing

Summary: Roche Diagnostics Operations, Inc. is recalling cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 0810559619. Reason: Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

What is being recalled

cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unit, material number 08464502001 cobas pro SSU, material number 09205632001 Affected Assays: (Product Name/ Catalog Number/ Application Code/ Application Short Name) 1. Cystatin C Gen.2 08105596190 20510 CYSC2 2. Ferritin Gen.4 08057648190 20571 FER4X 3. Lipoprotein (a) Gen.2 08106126190 20861 LPA2X 4. Lipoprotein (a) molarity 08106126160 20864 LPA2-X 5. Vancomycin Gen.3 08058849190 21211 VANC3O 6. Kappa Free Light Chains Partner Channel 08896640190 21421 KFLCX 7. Lambda Free Light Chains Partner Channel 08896631190 21430 LFLC 8. fCAL turbo Partner Channel 08910367190 21490 FCAL

Why

Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

US Nationwide distribution.

How to tell if yours is affected

cobas pro sample supply unit, UDI-DI 07613336158852 cobas pro SSU, UDI-DI 07613336179499 all software versions previous to 03-02

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1476-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 20, 2026
ReportedMarch 11, 2026
CompanyRoche Diagnostics Operations, Inc.
Company locationIndianapolis, IN, United States
Initiated byVoluntary: Firm initiated
Quantity1261 total