Atellica CH Urine Albumin recall (Siemens Healthcare Diagnostics, Inc.)
Medical deviceClass IIOngoing
Summary: Siemens Healthcare Diagnostics, Inc. is recalling Atellica CH Urine Albumin (UAlb). Material Number: 11537225. Reason: Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
What is being recalled
Atellica CH Urine Albumin (UAlb). Material Number: 11537225
Why
Falsely depressed UAlb patient results may occur. Affected samples with any urine albumin concentration above the UAlb measuring interval >40.0 mg/dL may report values between 6.2 mg/dL and 40.0 mg/dL.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Ireland, Israel, Italy, Korea, Latvia, Malaysia, Netherlands, New Zealand, Norway, Portugal, Romania, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, United¿Kingdom.
How to tell if yours is affected
Material Number: 11537225. UDI-DI: 00630414611099. Lot Numbers: All lot numbers.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1484-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 20, 2026 |
| Reported | March 11, 2026 |
| Company | Siemens Healthcare Diagnostics, Inc. |
| Company location | Tarrytown, NY, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 4,885 units |