Gelsoft Plus recall (Vascutek, Ltd.)
Medical deviceClass IIOngoing
Summary: Vascutek, Ltd. is recalling Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE,. Reason: Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.
What is being recalled
Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE, 636007PE, 631512PE, 636012PE, 636014PE, 632211P50E, 633016PE, 631407PE, 631809PE, 632010PE, 631608P50E, 632520PE, 632412PE, 631510PE, 635022PE, 631224PE, 632412P50E, 631507PE, 633014PE, 632010P50E, 635020PE, 631809P50E. Gelweave, REF: 736038E, 736036E, 736034E, 736032ST/10DE, 736032E, 736030ST/10DE, 736030E, 736028ST/10DE, 736028E, 736026ST/10DE, 736026E, 736024ST/10DE, 736024E, 736022ST/10DE, 736022E, 736020E, 736018E, 736016E, 736014E, 736012E, 736010E, 736008E, 736006E, 735034ST/8AE, 735034/8RME, 735034/10SEE, 735034/10RME, 735032STE, 735032ST/8AE, 735032/8SE, 735032/8RME, 735032/10SEE, 735032/10RME, 735030STE, 735030ST/8AE, 735030/8SE, 735030/8RME, 735030/10SEE, 735030/10RME, 735028STE, 735028ST/8AE, 735028E, 735028/8SE, 735028/8RME, 735028/10RME, 735026STE, 735026ST/8AE, 735026E, 735026/8SE, 735026/8RME, 735026/10RME, 735024STE, 735024ST/8AE, 735024E, 735024/8SE, 735024/8RME, 735024/10SEE, 735024/10RME, 735022STE, 735022E, 735022/8SE, 735022/8RME, 735020E, 735020/8RME, 735018E, 734038E, 734036E, 734034E, 734034CX4RMSE, 734034/8E, 734034/10E, 734032E, 734032CX4RMSE, 734032/8E, 734032/10E, 734030E, 734030CX4RMSE, 734030/8E, 734030/10E, 734028CX4RMSE, 734028/8E, 734028/10E, 734026CX4RMSE, 734026/8E, 734026/10E, 734024CX4RMSE, 734024/8E, 734024/10E, 734022CX4RMSE, 734022/8E, 734022/10E, 734020CX4RMSE, 734020/8E, 734020/10E, 734018/8E, 73341088E, 73341088/8RME, 73341088/8E, 73341088/10RME, 73341088/10E, 73321488/8ARME, 7332148/10RMFE, 7332148/10RMEE, 733212810/10STDE, 7332128/10RMFE, 7332128/10RMEE, 73321088E, 73321088/8SDE, 73321088/8RME, 73321088/8E, 73321088/10RME, 73321088/10E, 7330888/8E, 7330668/6E, 733038E, 733036E, 733034E, 733032E, 733030E, 733028E, 733026E, 733024E, 733024/6AE, 733022E, 733022/6AE, 733020E, 733020/6AE, 733018E, 733018/6AE, 733016E, 733014E, 73301488/8ARME, 7330148/10RMFE, 7330148/10RMEE, 73301412/10ETE, 733012E, 733012810/10STDE, 7330128/10RMFE, 7330128/10RMEE, 733010E, 73301088E, 73301088/8SDE, 73301088/8RME, 73301088/8E, 73301088/10RME, 73301088/10E, 733010108/8S4E, 733010108/8RME, 733010/10RME, 733008E, 733006E, 7328888/8E, 7328668/6E, 73281488/8ARME, 7328148/10RMEE, 73281412/10ETE, 73281288/8E, 732812810/10STDE, 7328128/10RMFE, 7328128/10RMEE, 73281088E, 73281088/8SDE, 73281088/8RME, 73281088/8E, 73281088/10RME, 73281088/10E, 732810108/8S4E, 7326888/8E, 7326668/6E, 73261488/8ARME, 7326148/10RMEE, 732612810/10STDE, 7326128/10RMFE, 7326128/10RMEE, 73261088E, 73261088/8SDE, 73261088/8RME, 73261088/8E, 73261088/10RME, 73261088/10E, 732610108/8S4E, 732610/10RME, 732528E, 732526E, 732524E, 732522E, 732520E, 732518E, 7324888/8E, 7324668/6E, 73241488/8ARME, 7324148/10RMEE, 732412E, 732412810/10STDE, 7324128/10RMEE, 732412/6E, 73241088E, 73241088/8SDE, 73241088/8RME, 73241088/8E, 73241088/10RME, 73241088/10E, 732410108/8S4E, 73221488/8ARME, 732214/10/10CRME, 732212810/10STDE, 732212/10/8CRME, 732212/10/8ARME, 732211E, 73221088/8RME, 73221088/8E, 73221088/10E, 732210108/8S4E, 7320888/8E, 7320668/6E, 732038E, 732034E, 732034CX4RME, 732034/8E, 732034/10E, 732032E, 732032CX4RME, 732032/8E, 732032/10E, 732030E, 732030CX4RME, 732030/8E, 732030/10E, 732028CX4RME, 732028/8E, 732028/10E, 732026CX4RME, 732026/8E, 732026/10E, 732024CX4RME, 732024/8E, 732024/10E, 732022/10E, 732014/8X2E, 732014/8X2AE, 732014/10X2E, 732014/10X2AE, 732012810/10STDE, 732012/8X2E, 732012/8X2AE, 732012/10X2E, 732010E, 73201088/8RME, 73201088/8E, 732010108/8S4E, 732010/6E, 731809E, 731809/6E, 731608XL55E, 731608E, 731608/8/10RME, 731608/6E, 731608/10/10RME, 731516E, 731514E, 731512E, 731510E, 731508E, 731506E, 7314108/10RME, 731407XL55E,
Why
Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution including in the states of FL, PA, TX, NC, NY, CA, NJ, OH, MI, DC, CO, VA, GA, IL, OR, ND, VT, MS, MA, IN, WA, AZ, MN, MD, IA, OK, MO, AL, WI, LA, AR, TN, CT, ME, NE, SC, KS, NV, KY, RI, WV, UT, MT, ID, AK, NH, DE, NM, SD, HI, WY. (states named: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, District of Columbia, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Wisconsin, West Virginia, Wyoming)
How to tell if yours is affected
All Gelsoft Plus devices (any item number ending 'E-B' e.g. 631206PE-B); REF, Item Number, UDI-DI(GTIN): 635024PE, 635024PE-B, 5037881166636; 631514PE, 631514PE-B, 5037881166858; 631222PE, 631222PE-B, 5037881166896; 633008PE, 633008PE-B, 5037881166698; 633010PE, 633010PE-B, 5037881166711; 633007PE, 633007PE-B, 5037881166681; 636010PE, 636010PE-B, 5037881166520; 632522PE, 632522PE-B, 5037881166780; 633012PE, 633012PE-B, 5037881166728; 635018PE, 635018PE-B, 5037881166605; 631516PE, 631516PE-B, 5037881166865; 636016PE, 636016PE-B, 5037881166551; 631220PE, 631220PE-B, 5037881166889; 631608PE, 631608PE-B, 5037881166933; 631206PE, 631206PE-B, 5037881166919; 631407P50E, 631407P50E-B, 5037881167008; 631508PE, 631508PE-B, 5037881166827; 636008PE, 636008PE-B, 5037881166513; 636006PE, 636006PE-B, 5037881166063; 632518PE, 632518PE-B, 5037881166759; 634008PE, 634008PE-B, 5037881166667; 634006PE, 634006PE-B, 5037881166643; 631218PE, 631218PE-B, 5037881166872; 632211PE, 632211PE-B, 5037881166971; 634007PE, 634007PE-B, 5037881166650; 633006PE, 633006PE-B, 5037881166674; 631506PE, 631506PE-B, 5037881166803; 632524PE, 632524PE-B, 5037881166797; 636007PE, 636007PE-B, 5037881166506; 631512PE,631512PE-B, 5037881166841; 636012PE, 636012PE-B, 5037881166537; 636014PE, 636014PE-B, 5037881166544; 632211P50E, 632211P50E-B, 5037881167046; 633016PE, 633016PE-B, 5037881166742; 631407PE, 631407PE-B, 5037881166926; 631809PE, 631809PE-B, 5037881166940; 632010PE, 632010PE-B, 5037881166964; 631608P50E, 631608P50E-B, 5037881167015; 632520PE, 632520PE-B, 5037881166773; 632412PE, 632412PE-B, 5037881166988; 631510PE, 631510PE-B, 5037881166834; 635022PE, 635022PE-B, 5037881166629; 631224PE, 631224PE-B, 5037881166902; 632412P50E, 632412P50E-B, 5037881167053; 631507PE, 631507PE-B, 5037881166810; 633014PE, 633014PE-B, 5037881166735; 632010P50E, 632010P50E-B, 5037881167039; 635020PE, 635020PE-B, 5037881166612; 631809P50E, 631809P50E-B, 5037881167022. All Gelweave devices (any item number ending E-B e.g. 7320128/10RMEE-B); REF, Item Number, UDI-DI(GTIN): 736038E, 736038E-B, 5037881169415; 736036E, 736036E-B, 5037881169408; 736034E, 736034E-B, 5037881169392; 736032ST/10DE, 736032ST/10DE-B, 5037881172620; 736032E, 736032E-B, 5037881169385; 736030ST/10DE, 736030ST/10DE-B, 5037881172637; 736030E , 736030E-B, 5037881169378; 736028ST/10DE, 736028ST/10DE-B, 5037881172644; 736028E, 736028E-B, 5037881169361; 736026ST/10DE, 736026ST/10DE-B, 5037881172651; 736026E, 736026E-B, 5037881169354; 736024ST/10DE, 736024ST/10DE-B, 5037881172668; 736024E, 736024E-B, 5037881169347; 736022ST/10DE, 736022ST/10DE-B, 5037881172675; 736022E, 736022E-B, 5037881169330; 736020E, 736020E-B, 5037881169323; 736018E, 736018E-B, 5037881169316; 736016E, 736016E-B, 5037881114057; 736014E, 736014E-B, 5037881114040; 736012E, 736012E-B, 5037881168548; 736010E, 736010E-B, 5037881168517; 736008E, 736008E-B, 5037881168494; 736006E, 736006E-B, 5037881166056; 735034ST/8AE, 735034ST/8AE-B, 5037881173061; 735034/8RME, 735034/8RME-B, 5037881170039; 735034/10SEE, 735034/10SEE-B, 5037881171883; 735034/10RME, 735034/10RME-B, 5037881169958; 735032STE, 735032STE-B, 5037881175102; 735032ST/8AE, 735032ST/8AE-B, 5037881173078; 735032/8SE, 735032/8SE-B, 5037881172095; 735032/8RME, 735032/8RME-B, 5037881170046; 735032/10SEE, 735032/10SEE-B, 5037881171890; 735032/10RME, 735032/10RME-B, 5037881169965; 735030STE, 735030STE-B, 5037881175126; 735030ST/8AE, 735030ST/8AE-B, 5037881173085; 735030/8SE, 735030/8SE-B, 5037881172101; 735030/8RME, 735030/8RME-B, 5037881170053; 735030/10SEE, 735030/10SEE-B, 5037881171906; 735030/10RME, 735030/10RME-B, 5037881169972; 735028STE, 735028STE-B, 5037881175133; 735028ST/8AE, 735028ST/8AE-B, 5037881173092; 735028E, 735028E-B 5037881168876; 735028/8SE, 735028/8SE-B, 5037881172118; 735028/8RME, 735028/8RME-B, 5037881170060; 735028/10RME, 735028/10RME-B, 5037881169989; 735026STE, 735026STE-B, 5037881175140; 735026ST/8AE, 735026ST/8AE-B, 5037881173108; 735026E, 735026E-B, 5037881168869; 735026/8SE, 735026/8SE-B, 5037881172125; 735026/8RME, 735026/8RME-B, 5037881170077; 735026/10RME, 735026/10RME-B, 5037881169996; 735024STE, 735024STE-B, 5037881175164; 735024ST/8AE, 735024ST/8AE-B, 5037881173115; 735024E, 735024E-B, 5037881168845; 735024/8SE, 735024/8SE-B, 5037881172132; 735024/8RME, 735024/8RME-B, 5037881170084; 735024/10SEE, 735024/10SEE-B, 5037881171937; 735024/10RME, 735024/10RME-B, 5037881170008; 735022STE, 735022STE-B, 5037881175188; 735022E, 735022E-B, 5037881168838; 735022/8SE, 735022/8SE-B, 5037881172149; 735022/8RME, 735022/8RME-B, 5037881170091; 735020E, 735020E-B, 5037881168821; 735020/8RME, 735020/8RME-B, 5037881170107; 735018E, 735018E-B, 5037881168814; 734038E, 734038E-B, 5037881169309; 734036E, 734036E-B, 5037881169293; 734034E, 734034E-B, 5037881169286; 734034CX4RMSE, 734034CX4RMSE-B, 5037881170794; 734034/8E, 734034/8E-B, 5037881172460; 734034/10E, 734034/10E-B, 5037881172224; 734032E, 734032E-B, 5037881169279; 734032CX4RMSE, 734032CX4RMSE-B, 5037881170800; 734032/8E, 734032/8E-B, 5037881172477; 734032/10E, 734032/10E-B, 5037881172231; 734030E, 734030E-B, 5037881169262; 734030CX4RMSE, 734030CX4RMSE-B, 5037881170817; 734030/8E, 734030/8E-B, 5037881172484; 734030/10E, 734030/10E-B, 5037881172262; 734028CX4RMSE, 734028CX4RMSE-B, 5037881170824; 734028/8E, 734028/8E-B, 5037881172491; 734028/10E, 734028/10E-B, 5037881172286; 734026CX4RMSE, 734026CX4RMSE-B, 5037881170831; 734026/8E, 734026/8E-B, 5037881172507; 734026/10E, 734026/10E-B, 5037881172309; 734024CX4RMSE, 734024CX4RMSE-B, 5037881170848; 734024/8E, 734024/8E-B, 5037881172514; 734024/10E, 734024/10E-B, 5037881172323; 734022CX4RMSE, 734022CX4RMSE-B, 5037881170855; 734022/8E, 734022/8E-B, 5037881172521; 734022/10E, 734022/10E-B, 5037881172347; 734020CX4RMSE, 734020CX4RMSE-B, 5037881170862; 734020/8E, 734020/8E-B, 5037881172538; 734020/10E, 734020/10E-B, 5037881172354; 734018/8E, 734018/8E-B, 5037881172545; 73341088E, 73341088E-B, 5037881171562; 73341088/8RME, 73341088/8RME-B, 5037881169873; 73341088/8E, 73341088/8E-B, 5037881171371; 73341088/10RME, 73341088/10RME-B, 5037881169774; 73341088/10E, 73341088/10E-B, 5037881171296; 73321488/8ARME, 73321488/8ARME-B, 5037881171043; 7332148/10RMFE, 7332148/10RMFE-B, 5037881174136; 7332148/10RMEE, 7332148/10RMEE-B, 5037881174334; 733212810/10STDE, 733212810/10STDE-B, 5037881172552; 7332128/10RMFE, 7332128/10RMFE-B, 5037881174266; 7332128/10RMEE, 7332128/10RMEE-B, 5037881174433; 73321088E, 73321088E-B, 5037881171579; 73321088/8SDE, 73321088/8SDE-B, 5037881172699; 73321088/8RME, 73321088/8RME-B, 5037881169880; 73321088/8E, 73321088/8E-B, 5037881171388; 73321088/10RME, 73321088/10RME-B, 5037881169781; 73321088/10E, 73321088/10E-B, 5037881171302; 7330888/8E, 7330888/8E-B, 5037881170657; 7330668/6E, 7330668/6E-B, 5037881170718; 733038E, 733038E-B, 5037881169163; 733036E , 733036E-B, 5037881169156; 733034E, 733034E-B, 5037881169149; 733032E, 733032E-B, 5037881169132; 733030E, 733030E-B, 5037881169118; 733028E, 733028E-B, 5037881169101; 733026E, 733026E-B, 5037881169095; 733024E, 733024E-B, 5037881169088; 733024/6AE, 733024/6AE-B, 5037881171845; 733022E, 733022E-B, 5037881169071; 733022/6AE, 733022/6AE-B, 5037881171852; 733020E, 733020E-B, 5037881169064; 733020/6AE, 733020/6AE-B, 5037881171869; 733018E, 733018E-B, 5037881169057; 733018/6AE, 733018/6AE-B, 5037881171876; 733016E, 733016E-B, 5037881168715; 733014E, 733014E-B, 5037881168708; 73301488/8ARME, 73301488/8ARME-B, 5037881171050; 7330148/10RMFE, 7330148/10RMFE-B, 5037881174150; 7330148/10RMEE, 7330148/10RMEE-B, 5037881174358; 73301412/10ETE, 73301412/10ETE-B, 5037881170206; 733012E, 733012E-B, 5037881168692; 733012810/10STDE, 733012810/10STDE-B, 5037881172569; 7330128/10RMFE, 7330128/10RMFE-B, 5037881174273; 7330128/10RMEE, 7330128/10RMEE-B, 5037881174440; 733010E, 733010E-B, 5037881168678; 73301088E, 73301088E-B, 5037881171586; 73301088/8SDE, 73301088/8SDE-B, 5037881172705; 73301088/8RME, 73301088/8RME-B, 5037881169897; 73301088/8E, 73301088/8E-B, 5037881171395; 73301088/10RME, 73301088/10RME-B, 5037881169811; 73301088/10E, 73301088/10E-B, 5037881171319; 733010108/8S4E, 733010108/8S4E-B, 5037881170589; 733010108/8RME, 733010108/8RME-B, 5037881169576; 733010/10RME, 733010/10RME-B, 5037881169729; 733008E, 733008E-B, 5037881168654; 733006E, 733006E-B, 5037881168616; 7328888/8E, 7328888/8E-B, 5037881170664; 7328668/6E, 7328668/6E-B, 5037881170725; 73281488/8ARME, 73281488/8ARME-B, 5037881171067; 7328148/10RMEE, 7328148/10RMEE-B, 5037881174365; 73281412/10ETE, 73281412/10ETE-B, 5037881170213; 73281288/8E, 73281288/8E-B, 5037881171289; 732812810/10STDE, 732812810/10STDE-B, 5037881172576; 7328128/10RMFE, 7328128/10RMFE-B, 5037881174280; 7328128/10RMEE, 7328128/10RMEE-B, 5037881174457; 73281088E, 73281088E-B, 5037881171593; 73281088/8SDE, 73281088/8SDE-B, 5037881172712; 73281088/8RME, 73281088/8RME-B, 5037881169903; 73281088/8E, 73281088/8E-B, 5037881171401; 73281088/10RME, 73281088/10RME-B, 5037881169828; 73281088/10E, 73281088/10E-B, 5037881171326; 732810108/8S4E, 732810108/8S4E-B, 5037881170596; 7326888/8E, 7326888/8E-B, 5037881170671; 7326668/6E, 7326668/6E-B, 5037881170732; 73261488/8ARME, 73261488/8ARME-B, 5037881171074; 7326148/10RMEE, 7326148/10RMEE-B, 5037881174372; 732612810/10STDE, 732612810/10STDE-B, 5037881172583; 7326128/10RMFE, 7326128/10RMFE-B, 5037881174297; 7326128/10RMEE, 7326128/10RMEE-B, 5037881174464; 73261088E, 73261088E-B, 5037881171609; 73261088/8SDE, 73261088/8SDE-B, 5037881172729; 73261088/8RME, 73261088/8RME-B, 5037881169910; 73261088/8E, 73261088/8E-B, 5037881171418; 73261088/10RME, 73261088/10RME-B, 5037881169835; 73261088/10E, 73261088/10E-B, 5037881171333; 732610108/8S4E, 732610108/8S4E-B, 5037881170602; 732610/10RME, 732610/10RME-B, 5037881169743; 732528E, 732528E-B, 5037881168982; 732526E, 732526E-B, 5037881168975; 732524E, 732524E-B, 5037881168968; 732522E, 732522E-B, 5037881168951; 732520E, 732520E-B, 5037881168944; 732518E, 732518E-B, 5037881168937; 7324888/8E, 7324888/8E-B, 5037881170688; 7324668/6E, 7324668/6E-B, 5037881170749; 73241488/8ARME, 73241488/8ARME-B, 5037881171081; 7324148/10RMEE, 7324148/10RMEE-B, 5037881174389; 732412E, 732412E-B, 5037881174907; 732412810/10STDE, 732412810/10STDE-B, 5037881172590; 7324128/10RMEE, 7324128/10RMEE-B, 5037881174495; 732412/6E, 732412/6E-B, 5037881174983; 73241088E, 73241088E-B, 5037881171616; 73241088/8SDE, 73241088/8SDE-B, 5037881172736; 73241088/8RME, 73241088/8RME-B, 5037881169927; 73241088/8E, 73241088/8E-B, 5037881171425; 73241088/10RME, 73241088/10RME-B, 5037881169842; 73241088/10E, 73241088/10E-B, 5037881171340; 732410108/8S4E, 732410108/8S4E-B, 5037881170619; 73221488/8ARME, 73221488/8ARME-B, 5037881171098; 732214/10/10CRME, 732214/10/10CRME-B, 5037881171661; 732212810/10STDE, 732212810/10STDE-B, 5037881172606; 732212/10/8CRME, 732212/10/8CRME-B, 5037881171678; 732212/10/8ARME, 732212/10/8ARME-B, 5037881171692; 732211E, 732211E-B, 5037881174839; 73221088/8RME, 73221088/8RME-B, 5037881169934; 73221088/8E, 73221088/8E-B, 5037881171432; 73221088/10E, 73221088/10E-B, 5037881171357; 732210108/8S4E, 732210108/8S4E-B, 5037881170626; 7320888/8E, 7320888/8E-B, 5037881170701; 7320668/6E, 7320668/6E-B, 5037881170763; 732038E, 732038E-B, 5037881169255; 732034E, 732034E-B, 5037881169248; 732034CX4RME, 732034CX4RME-B, 5037881170879; 732034/8E, 732034/8E-B, 5037881172392; 732034/10E, 732034/10E-B, 5037881172156; 732032E, 732032E-B, 5037881169231; 732032CX4RME, 732032CX4RME-B, 5037881170886; 732032/8E, 732032/8E-B, 5037881172408; 732032/10E, 732032/10E-B, 5037881172163; 732030E, 732030E-B, 5037881169224; 732030CX4RME, 732030CX4RME-B, 5037881170893; 732030/8E, 732030/8E-B, 5037881172415; 732030/10E, 732030/10E-B, 5037881172170; 732028CX4RME, 732028CX4RME-B, 5037881170909; 732028/8E, 732028/8E-B, 5037881172422; 732028/10E, 732028/10E-B, 5037881172187; 732026CX4RME, 732026CX4RME-B, 5037881170916; 732026/8E, 732026/8E-B, 5037881172439; 732026/10E, 732026/10E-B, 5037881172194; 732024CX4RME, 732024CX4RME-B, 5037881170923; 732024/8E, 732024/8E-B, 5037881172446; 732024/10E, 732024/10E-B, 5037881172200; 732022/10E, 732022/10E-B, 5037881172217; 732014/8X2E, 732014/8X2E-B, 5037881171739; 732014/8X2AE, 732014/8X2AE-B, 5037881176277; 732014/10X2E, 732014/10X2E-B, 5037881171722; 732014/10X2AE, 732014/10X2AE-B, 5037881175355; 732012810/10STDE, 732012810/10STDE-B, 5037881172613; 732012/8X2E, 732012/8X2E-B, 5037881171753; 732012/8X2AE, 732012/8X2AE-B, 5037881175324; 732012/10X2E, 732012/10X2E-B, 5037881171746; 732010E, 732010E-B, 5037881174822; 73201088/8RME, 73201088/8RME-B, 5037881169941; 73201088/8E, 73201088/8E-B, 5037881171449; 732010108/8S4E, 732010108/8S4E-B, 5037881170633; 732010/6E, 732010/6E-B, 5037881174976; 731809E, 731809E-B, 5037881174792; 731809/6E, 731809/6E-B, 5037881174945; 731608XL55E, 731608XL55E-B, 5037881175348; 731608E, 731608E-B, 5037881174747; 731608/8/10RME, 731608/8/10RME-B, 5037881175195; 731608/6E, 731608/6E-B, 5037881174921; 731608/10/10RME, 731608/10/10RME-B, 5037881175201; 731516E, 731516E-B, 5037881168807; 731514E, 731514E-B, 5037881168791; 731512E, 731512E-B, 5037881168784; 731510E, 731510E-B, 5037881168760; 731508E, 731508E-B, 5037881168746; 731506E, 731506E-B, 5037881168722; 7314108/10RME, 7314108/10RME-B, 5037881176468; 731407XL55E, 731407XL55E-B, 5037881175362; 731407E, 731407E-B, 5037881174686; 731407/7E, 731407/7E-B, 5037881175119; 731407/10/10RME, 731407/10/10RME-B, 5037881175218; 731228E, 731228E-B, 5037881169040; 731226E, 731226E-B, 5037881169033; 731224E, 731224E-B, 5037881169026; 731222E, 731222E-B, 5037881169019; 731220E, 731220E-B, 5037881169002; 731218E, 731218E-B, 5037881168999; 7312108/10RME, 7312108/10RME-B, 5037881176444; 731210/8X2E, 731210/8X2E-B, 5037881171456; 731206XL55E, 731206XL55E-B, 5037881175379; 731206E, 731206E-B, 5037881174631; 731206/10/10RME, 731206/10/10RME-B, 5037881175225; 731038E, 731038E-B, 5037881169217; 731036E, 731036E-B, 5037881169200; 731034E, 731034E-B, 5037881169194; 731032E, 731032E-B, 5037881169187; 731030E, 731030E-B, 5037881169170; 731010/8X2E, 731010/8X2E-B, 5037881171463; 730034ADPE, 730034ADPE-B, 5037881173986; 730032ADPSLE, 730032ADPSLE-B, 5037881173603; 730032ADPE, 730032ADPE-B, 5037881173993; 730030ADPSLE, 730030ADPSLE-B, 5037881173634; 730030ADPE, 730030ADPE-B, 5037881174013; 730028ADPSLE, 730028ADPSLE-B, 5037881173641; 730028ADPE, 730028ADPE-B, 5037881174037; 730026ADPE, 730026ADPE-B, 5037881174044; 730024ADPE, 730024ADPE-B, 5037881174051; 730022ADPE, 730022ADPE-B, 5037881174068; 730020ADPE, 730020ADPE-B, 5037881174082; 730018ADPE, 730018ADPE-B, 5037881174099; 730016ADPE, 730016ADPE-B, 5037881174112.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1492-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | December 18, 2025 |
| Reported | March 11, 2026 |
| Company | Vascutek, Ltd. |
| Company location | Inchinnan, N/A, United Kingdom |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 37430 |
More recalls from Vascutek, Ltd. · all medical device recalls · recalls from December 2025