GEM Premier 5000 recall (Instrumentation Laboratory)
Medical deviceClass IIOngoing
Summary: Instrumentation Laboratory is recalling GEM Premier 5000; Part No. 00055445010 & 00055445011.. Reason: Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
What is being recalled
GEM Premier 5000; Part No. 00055445010 & 00055445011.
Why
Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jordan, Japan, South Korea, Kazakhstan, Lebanon, Lithuania, Luxembourg, Latvia, Macau, Mexico, Malaysia, Netherlands, New Zealand, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, San Marino, South Africa, Thailand, Tunisia, Turkey, Uruguay, Uzbekistan & Vietnam. (states named: Georgia)
How to tell if yours is affected
Part number: 00055445010; UDI: 08426950807810; Lots No.: 251023N, 251024J, 251114K, 251117A, 251117AC, 251117AE, 251117F, 251117I, 251117J, 251117Y, 251118A, 251119A, 251120U, 251121Q, 251121S, 251121T, 251121W, 251121X, 251124K, 251124O, 251124U, 251125I, 251126C, 251126D, 251202H, 251203L, 251203M, 251204R, 251204S, 251205K, 251205L, 251208J, 251208L, 251215D, 251215W, 251215X, 251216S, 251216V, 251217W, 251217X, 251218R, 251218Y, 251219B, 251219C, 251219E, 251219H, 251219I, 251219P, 251219Q, 251223B, 251223D, 251223F, 251223K, 251229AL, 251229D, 251229E, 251229F, 251229S, 251229U, 251230R, 251231B, 251231D, 251231E, 251231H, 251231M. Part number: 00055445011 UDI: 08426950807827 Kits distributed in US: 96 Lots No.: 251117L, 251118O, 251124K, 251125I, 251201X, 251202G, 251203M, 251204O, 251223K, 251229W, 251230A.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1543-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 20, 2026 |
| Reported | March 18, 2026 |
| Company | Instrumentation Laboratory |
| Company location | Bedford, MA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1,989 units |