Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1 recall (GE Medical Systems, LLC)

Medical deviceClass IIOngoing

Summary: GE Medical Systems, LLC is recalling Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 5826659-029, 5826659-032, 5826659-033, 5826659-028, 5826659-034, 5826659-030, 5826659-019, 5826659-022. Reason: There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

What is being recalled

Centricity Universal Viewer Software Versions 7.0 through 7.0 Sp2.0.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 5826659-027, 5826659-029, 5826659-032, 5826659-033, 5826659-028, 5826659-034, 5826659-030, 5826659-019, 5826659-022, 5826659-026, 5826659-023, 5826659-031, 5826659-025, 5826659-002, 5826659-021, 5826659-037, 5826659-014, 5826659-016, 5826659-013

Why

There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

UDI Numbers: (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P001.58258320011 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)008406821 45794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P00465825832001 (01)00840682145794(10)70P00465825832001 (01)00840682145794(10)70P00465825832001 (01)00840682145794(10)70P00465825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P0105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0041.58258320011 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P00445825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00465825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P00485825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P00425825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0041.58258320011 (01)00840682145794(10)70P0041.58258320011 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004.58258320011 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P15825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004.58258320011 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P00475825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P00495825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P2005825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P0115825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1025825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P1045825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P1035825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P004105825832001 (01)00840682145794(10)70P2015825832001 (01)00840682145794(10)70P004105825832001

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1569-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 30, 2026
ReportedMarch 25, 2026
CompanyGE Medical Systems, LLC
Company locationWaukesha, WI, United States
Initiated byVoluntary: Firm initiated
Quantity1568 units