Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1 recall (GE Medical Systems, LLC)
Medical deviceClass IIOngoing
Summary: GE Medical Systems, LLC is recalling Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-72. Reason: There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
What is being recalled
Centricity Universal Viewer Software Versions 6.0 through 6.0 Sp10.4.1, a device that displays medical images (including mammograms) and data from various imaging sources, Model Numbers 2088026-406, 2100169-001, 2088026-985, 2088026-971, 2088026-614, 2088026-713, 2088026-994, 2100491-006, 2088026-723, 2088026-308, 2104867-044, 2100491-012, 2088026-962, 2088026-969, 2102676-001, 2088026-936, 2088026-960, K1220JWED, 2088026-806, 2088026-506, 2088026-026, 2088026-970, 2088026-943, 2088026-990, 2088026-940, 2088026-959, 2088026-967, 2088026-989, 2104867-045, 2088026-993, 2088026-306, 2102675-010, 2088026-948, 2088026-957, 2088026-973, 2088026-946, 2088026-715, 2088026-906
Why
There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, which could allow an unauthorized individual to potentially impact system availability and/or manipulate data.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
UDI Numbers: (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)60P90111.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)60SP90182094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP102094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90122094097001G (01)00840682103800(10)60SP90112094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60SP90162094097001G (01)00840682103800(10)60P90110.2094097001H (01)00840682103800(10)60P90110.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP4.0.22094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP90162094097001G (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60SP90172094097001H (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60P1031.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.22094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP5.0.1.12094097001D (01)00840682103800(10)6.0SP5.0.12094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60SP90142094097001G (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90152094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP102094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60SP10.12094097001G (01)00840682103800(10)60SP10.12094097001G (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP90132094097001G (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP82094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P103.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP72094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60SP5.0.32094097001D (01)00840682103800(10)6.0SP9012094097001F (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7.12094097001H (01)00840682103800(10)6.0SP7.12094097001E (01)00840682103800(10)60SP5.0.42094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.22094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP7022094097001D (01)00840682103800(10)6.0SP7022094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)SP32094097001C (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60SP5.0.4.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)6.0SP62094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP5.0.32094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)6.0SP50412094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P102.2094097001H (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P104.2094097001H (01)00840682103800(10)60P901122094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP3.12094097001D (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P104120 94097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)60P10412094097001 (01)00840682103800(10)6.0SP52094097001D (01)00840682103800(10)6.0SP92094097001D (01)00840682103800(10)60P10412094097001
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1570-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 30, 2026 |
| Reported | March 25, 2026 |
| Company | GE Medical Systems, LLC |
| Company location | Waukesha, WI, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 797 units |