Distal Humeral Plates with Angular Stability with the below descriptions and... recall (I.T.S. GmbH)
Medical deviceClass IIOngoing
Summary: I.T.S. GmbH is recalling Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N,. Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
What is being recalled
Distal Humeral Plates with Angular Stability with the below descriptions and corresponding article numbers. 1. Cortical Screw, Locking; Article Numbers: 37351-10-N, 37351-12-N, 37351-14-N, 37351-16-N, 37351-18-N, 37351-20-N, 37351-22-N, 37351-24-N, 37351-26-N, 37351-28-N, 37351-30-N, 37351-32-N, 37351-34-N, 37351-36-N, 37351-38-N, 37351-40-N, 37351-42-N, 37351-44-N, 37351-46-N, 37351-48-N, 37351-50-N, 37351-55-N, 37351-60-N, 37351-65-N, 37351-70-N, 37351-75-N, 37351-80-N, 37351-85-N. 2. Humeral Plate, Distal; Article Numbers: 21241-5, 21241-7, 21242-7, 21243-10, 21244-9.
Why
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W. (states named: Alabama, Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, Nebraska, New Jersey, New Mexico, Nevada, Ohio, Oklahoma, Pennsylvania, South Carolina, Texas, West Virginia)
How to tell if yours is affected
1. Article Number (Lot Numbers): 37351-10-N (18/240314, 679/122205), 37351-12-N (18/200330, 18/943977, 26/F65544, 679/032401, 679/052417, 679/122261), 37351-14-N (18/201331, 18/511504, 18/944978, 26/F65545, 679/032402, 679/032514, 679/052244, 679/052418, 679/052516, 679/072520), 37351-16-N (18/202332, 18/521505, 18/945979, 26/F65163, 679/032423, 679/032504, 679/052201, 679/062466, 679/062504, 679/072521, 679/122251), 37351-18-N (18/203333, 18/531506, 18/946980, 18/971215, 26/F65546, 679/022329, 679/022516, 679/032430, 679/062206, 679/062505, 679/072522), 37351-20-N (18/231334, 18/541507, 18/952981, 26/F65164, 26/F65547, 679/032431, 679/062230, 679/072501, 679/122202), 37351-22-N (18/233335, 18/953982, 26/F65548, 679/062233, 679/092111, 679/122210), 37351-24-N (18/238336, 18/956983, 26/F65549, 679/042415, 679/072203, 679/112242), 37351-26-N (18/204337, 18/947984, 26/F65166, 26/F65550, 679/022330, 679/042416, 679/062207), 37351-28-N (18/234338, 18/954985, 26/F65551, 679/022331, 679/042423, 679/072204), 37351-30-N (18/205339, 18/948986, 26/F65552, 679/042424, 679/072205, 679/122209), 37351-32-N (18/235340, 18/955987, 26/F65553, 679/042425, 679/062234), 37351-34-N (18/206341, 18/491108, 26/F65554, 679/012306, 679/042426, 679/062208), 37351-36-N (18/239342, 18/571109, 18/571503, 26/F65555, 679/022324, 679/052401, 679/062209), 37351-38-N (18/230679, 18/501110, 18/862696, 679/052402, 679/122252), 37351-40-N (147/15088L, 18/207343, 18/506128, 18/511111, 679/052411), 37351-42-N (18/581112, 679/032515), 37351-44-N (18/591113, 679/022511), 37351-46-N (18/601114), 37351-48-N (18/611115, 679/032517), 37351-50-N (18/513248, 18/621116, 18/912703, 679/032518, 679/062508), 37351-55-N (18/652407, 679/032519), 37351-60-N (18/662408), 37351-65-N (679/022512, 679/052236), 37351-70-N (18/631117, 679/052237), 37351-75-N (18/641118, 679/052238), 37351-80-N (679/052239, 679/072406), 37351-85-N (679/052240). 2. Article Number (Lot Numbers): 21241-5 (770/201905), 21241-7 (770/201906), 21242-7 (770/201901), 21243-10 (770/201901), 21244-9 (770/201905).
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1602-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 23, 2026 |
| Reported | April 1, 2026 |
| Company | I.T.S. GmbH |
| Company location | Lasnitzhohe, Austria |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 12,059 units |
More recalls from I.T.S. GmbH · all medical device recalls · recalls from January 2026