Tandem pumps are battery-operated infusion pumps capable of both basal and bolus... recall (Tandem Diabetes Care, Inc.)

Medical deviceClass IIOngoing

Summary: Tandem Diabetes Care, Inc. is recalling Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pump. Reason: The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.

What is being recalled

Tandem pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient).

Why

The Czech language user guide contained multiple translation errors. The most significant error was that it incorrectly instructs users to verify that the infusion set is connected to the body, when it should instead direct users to confirm that the infusion set is not connected to the body. Following this instruction could result in unintended insulin delivery, potentially leading to a hypoglycemic event.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

International distribution in the countries of Czech Republic, Slovakia.

How to tell if yours is affected

Version 7.8.1/ UDI: (01)008500066132625

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1636-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 30, 2026
ReportedApril 1, 2026
CompanyTandem Diabetes Care, Inc.
Company locationSan Diego, CA, United States
Initiated byVoluntary: Firm initiated
Quantity1,380 units