Philips Respironics Trilogy EV300 recall (Philips Respironics, Inc.)

Medical deviceClass IOngoing

Summary: Philips Respironics, Inc. is recalling Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.. Reason: In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.

What is being recalled

Philips Respironics Trilogy EV300, Software Version 1.05.15.00. Continuous home-use ventilator device.

Why

In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.

How serious is it

Class I is the most serious level: FDA considers there is a reasonable probability that using or being exposed to this product will cause serious health problems or death.

Where it was sold

Worldwide Distribution. US Nationwide, Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe. (states named: Georgia)

How to tell if yours is affected

All Model No.; All UDI; All Serial No.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1648-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass I
StatusOngoing
Recall startedMarch 2, 2026
ReportedApril 8, 2026
CompanyPhilips Respironics, Inc.
Company locationMurrysville, PA, United States
Initiated byVoluntary: Firm initiated
Quantity113,717 units