Azurion 7 M12 recall (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.)
Medical deviceClass IIOngoing
Summary: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. is recalling Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;. Reason: Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
What is being recalled
Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;
Why
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen; (states named: Georgia)
How to tell if yours is affected
(1) System Code: 722078; UDI: 00884838085251; System Serial Number: 302, 585, 410, 753, 523, 293, 840, 291, 536, 537, 870, 613, 598, 679, 224, 829, 249, 546, 478, 479, 513, 278, 437, 438, 456, 566, 833, 488, 327, 550, 571, 544, 545, 465, 376, 31, 921, 700, 165, 445, 539, 654, 656, 657, 658, 436, 1007, 862, 786, 308, 396, 576, 121, 239, 383, 491, 125, 500, 645, 379, 399, 803, 415, 296, 136, 890, 626, 697, 559, 532, 538, 185, 166, 257, 258, 684, 359, 143, 661, 824, 649, 493, 720, 520, 521, 556, 557, 558, 187, 902, 269, 317, 522, 611, 677, 709, 535, 170, 723, 251, 298, 22, 305, 725, 728, 524, 290, 441, 384, 480, 481, 364, 294, 906, 240, 256, 603, 289, 719, 163, 215, 271, 788, 412, 627, 141, 644, 292, 844, 876, 748, 483, 497, 309, 232, 425, 635, 612, 797, 247, 733, 162, 517, 888, 923, 204, 205, 206, 195, 197, 218, 444, 986, 561, 503, 689, 499, 135, 589, 38, 977, 738, 476, 380, 453, 24, 690, 356, 343, 344, 174, 898, 482, 378, 764, 693, 694, 770, 771, 772, 773, 439, 191, 938, 980, 981, 982, 307, 628, 235, 139, 178, 179, 265, 266, 511, 526, 746, 264, 287, 361, 362, 1003, 348, 897, 845, 960, 1036, 967, 374, 375, 564, 142, 469, 934, 565, 835, 836, 929, 975, 681, 1069, 1070, 1071, 1072, 893, 1061, 382, 324, 604, 713, 114, 849, 326, 659, 346, 347, 26, 621, 553, 325, 648, 1058, 1068, 339, 340, 414, 687, 688, 237, 958, 751, 881, 698, 610, 20, 846, 625, 424, 800, 838, 590, 150, 762, 485, 726, 727, 1039, 847, 959, 962, 865, 180, 371, 769, 570, 572, 1031, 792, 702, 196, 28, 29, 683, 541, 213, 320, 1025, 943, 651, 722, 242, 36, 431, 672, 177, 578, 912, 355, 504, 699, 270, 306, 190, 366, 52, 398, 470, 562, 489, 147, 160, 743, 152, 759, 630, 729, 151, 334, 815, 429, 449, 757, 506, 703, 149, 818, 647, 419, 328, 584, 223, 192, 779, 468, 222, 200, 134, 262, 855, 284, 918, 919, 277, 976, 13, 15, 502, 817, 795, 498, 574, 577, 254, 255, 448, 27, 631, 632, 599, 370, 216, 451, 428, 467, 450, 102, 662, 189, 633, 512, 175, 796, 459, 460, 718, 462, 464, 891, 768, 263, 944, 669, 112, 23, 452, 338, 367, 440, 490, 181, 145, 250, 705, 49, 50, 51, 617, 336, 226, 761, 220, 221, 1037, 799, 839, 354, 153, 154, 957, 471, 357, 130, 132, 133, 261, 695, 158, 507, 854, 477, 785, 789, 404, 877, 806, 422, 194, 801, 802, 140, 608, 878, 952, 10, 123, 319, 686, 805, 397, 127, 128, 253, 594, 988, 300, 301, 321, 332, 35, 17, 341, 563, 314, 316, 496, 641, 953, 30, 950, 331, 652, 1005, 349, 236, 979, 831, 501, 841, 842, 905, 682, 843, 830, 323, 315, 680, 365, 282, 283, 602, 203, 202, 40, 673, 674, 675, 676, 609, 295, 368, 369, 615, 618, 310, 199, 455, 668, 352, 193, 280, 667, 884, 567, 183, 749, 692, 32, 241, 935, 407, 408, 505, 138, 157, 752, 461, 463, 423, 554, 252, 169, 426, 650, 351, 442, 750, 968, 207, 208, 209, 210, 211, 622, 623, 629, 129, 529, 144, 400, 401, 411, 259, 457, 579, 926, 597, 899, 322, 978, 173, 418, 637, 670, 909, 601, 37, 1033, 1035, 227, 1009, 1008, 989, 984, 987, 907, 1081, 1004, 1086, 1084, 1085, 1048, 1059, 1034, 1080, 1015, 875, 1026, 1077, 716, 717, 1045, 940, 942, 945, 946, 947, 948, 1052, 971, 1062, 1063, 994, 701, 993, 620, 1075, 1046, 970, 985, 798, 1074, 856, 710, 1011, 1064, 1065, 1066, 1067, 1000, 1056, 992, 663, 1057, 964, 965, 1043, 735, 736, 819, 820, 272, 1049, 1073, 1079, 712, 714, 732, 774, 381, 233, 638, 318, 593, 624, 775, 776, 21, 267, 268, 487, 421, 760, 542, 614, 864, 148, 837, 458, 596, 377, 281, 784, 991, 528, 42, 44, 45, 41, 43, 466, 704, 219, 217, 260, 103, 430, 342, 126, 184, 778, 547, 586, 879, 159, 363, 896, 642, 472, 39, 172, 782, 961, 33, 643, 581, 552, 525, 311, 312, 510, 857, 731, 432, 433, 636, 954, 447, 333, 358, 540, 619, 492, 527, 600, 816, 434, 892, 990, 124, 122, 146, 201, 509, 495, 515, 353, 303, 304, 186, 286, 548, 518, 486, 474, 475, 1047, 853, 243, 244, 335, 549, 46, 616, 972, 533, 777, 182, 1032, 1051, 850, 696, 345, 575, 939, 869, 874, 871, 1010, 1029, 1030, 18, 551, 1013, 917, 543, 516, 12, 416, 155, 446, 685, 715, 225, 1082, 913, 582, 908, 19, 435, 16, 11, 167, 664, 901, 508, 530, 198; (2) System Code: 722223; UDI: 00884838099241; System Serial Number: 312, 324, 332, 585, 85, 533, 189, 479, 1038, 47, 1146, 352, 391, 328, 681, 1052, 1130, 388, 581, 855, 905, 570, 77, 448, 495, 571, 171, 1051, 439, 126, 628, 36, 1088, 580, 1179, 1241, 1065, 799, 577, 1060, 1107, 1114, 323, 252, 626, 261, 207, 329, 1001, 801, 560, 561, 205, 379, 440, 204, 192, 197, 177, 37, 340, 1097, 282, 357, 1015, 214, 44, 206, 210, 336, 1231, 343, 365, 390, 1067, 1106, 1144, 186, 187, 631, 87, 373, 514, 599, 601, 602, 69, 638, 728, 937, 944, 945, 951, 852, 256, 679, 596, 67, 932, 503, 432, 131, 301, 456, 704, 659, 242, 746, 1161, 1090, 1265, 455, 457, 221, 454, 829, 816, 318, 229, 230, 262, 308, 637, 124, 198, 433, 883, 889, 566, 1084, 880, 665, 188, 367, 52, 371, 892, 127, 105, 572, 104, 334, 627, 458, 1113, 463, 487, 470, 96, 277, 616, 1140, 211, 850, 223, 1242, 872, 877, 97, 1211, 161, 1132, 386, 1123, 208, 1061, 361, 258, 597, 727, 1008, 1296, 180, 1181, 835, 668, 377, 882, 565, 1103, 196, 130, 1185, 1150, 886, 887, 452, 453, 269, 888, 1188, 567, 966, 449, 435, 441, 669, 351, 508, 750, 164, 139, 355, 619, 1236, 133, 89, 370, 598, 714, 119, 1306, 1307, 871, 697, 698, 1035, 1036, 116, 154, 100, 912, 99, 349, 387, 136, 775, 903, 867, 680, 527, 1227, 629, 1264, 834, 1226, 155, 837, 17, 18, 804, 235, 249, 477, 1025, 1177, 1153, 1210, 167, 1290, 1293, 227, 965, 999, 46, 1131, 141, 1282, 1005, 1129, 1115, 237, 348, 287, 1325, 135, 414, 416, 417, 132, 814, 836, 1245, 356, 809, 810, 303, 1214, 1099, 813, 1167, 1232, 1248, 322, 587, 591, 812, 875, 904, 756, 516, 917, 1016, 10, 1256, 537, 582, 694, 1055, 576, 134, 247, 961, 963, 1222, 1217, 1272, 584, 7, 923, 933, 709, 193, 128, 393, 434, 615, 1165, 854, 550, 657, 1283, 1196, 1258, 407, 1212, 612, 926, 353, 398, 555, 848, 931, 213, 364, 909, 910, 918, 916, 265, 1243, 142, 168, 518, 476, 671, 683, 302, 863, 864, 1033, 894, 25, 1295, 482, 919, 693, 553, 1191, 868, 114, 191, 409, 106, 506, 232, 446, 403, 1125, 823, 670, 803, 1284, 1253, 1254, 579, 289, 703, 705, 311, 776, 1104, 719, 480, 113, 609, 822, 1095, 1137, 1042, 1043, 1078, 519, 521, 520, 522, 523, 524, 525, 526, 858, 362, 375, 747, 272, 81, 661, 723, 821, 862, 140, 761, 63, 109, 1003, 1251, 1041, 1271, 998, 316, 317, 78, 978, 202, 1276, 824, 915, 711, 108, 185, 754, 928, 846, 847, 893, 1262, 1263, 769, 755, 906, 313, 314, 238, 239, 1034, 430, 299, 1247, 363, 40, 574, 486, 354, 710, 279, 280, 281, 845, 960, 306, 992, 989, 618, 225, 226, 173, 212, 8, 9, 157, 165, 344, 1147, 431, 590, 1083, 492, 28, 274, 753, 424, 425, 817, 820, 1237, 818, 819, 1207, 1301, 405, 974, 975, 976, 635, 924, 217, 381, 395, 111, 691, 815, 436, 437, 438, 35, 830, 938, 331, 1010, 1011, 79, 1267, 295, 65, 234, 1246, 796, 337, 1075, 270, 468, 376, 450, 451, 49, 505, 650, 1059, 1259, 1050, 465, 68, 636, 20, 383, 245, 246, 1261, 222, 374, 930, 793, 794, 795, 849, 712, 861, 1291, 325, 907, 908, 859, 605, 61, 544, 1289, 1285, 851, 1201, 921, 589, 1100, 273, 294, 800, 806, 333, 4, 118, 1195, 1197, 1135, 429, 264, 720, 707, 606, 43, 392, 418, 759, 1280, 179, 445, 215, 556, 1027, 21, 442, 791, 807, 76, 922, 1275, 427, 257, 266, 1268, 1336, 228, 460, 554, 194, 954, 956, 174, 1151, 684, 558, 559, 563, 564, 724, 1077, 702, 1092, 48, 779, 422, 583, 573, 413, 995, 541, 600, 183, 6, 1190, 699, 700, 701, 664, 876, 419, 666, 12, 243, 1017, 1037, 283, 1286, 30, 502, 790, 22, 778, 1117, 1239, 540, 169, 485, 26, 787, 507, 396, 382, 176, 986, 685, 682, 1141, 721, 91, 378, 773, 640, 641, 1021, 805, 304, 443, 562, 115, 259, 985, 159, 160, 994, 1186, 170, 82, 199, 271, 988, 481, 828, 860, 278, 219, 461, 549, 802, 66, 298, 552, 285, 497, 962, 534, 1073, 613, 1250, 1270, 1335, 611, 1122, 1124, 692, 766, 996, 997, 652, 900, 614, 517, 1172, 990, 991, 1176, 1310, 957, 780, 781, 1304, 686, 687, 688, 689, 1287, 1288, 1332, 1171, 634, 1198, 979, 1334, 843, 651, 678, 1317, 782, 987, 798, 825, 797, 1302, 1128, 604, 1321, 1322, 935, 838, 1218, 1225, 811, 1328, 762, 758, 1057, 1330, 763, 764, 765, 767, 953, 644, 1169, 1228, 421, 655, 662, 873, 874, 92, 1133, 1136, 632, 792, 218, 784, 760, 535, 542, 543, 895, 426, 1160, 117, 195, 255, 42, 839, 708, 536, 254, 1108, 41, 914, 359, 360, 408, 1019, 642, 90, 643, 286, 288, 290, 72, 74, 80, 1184, 955, 1063, 423, 706, 1252, 291, 483, 1007, 241, 625, 656, 751, 774, 630, 531, 939, 941, 1024, 478, 1143, 645, 646, 647, 648, 786, 660, 2, 840, 24, 674, 358, 1079, 110, 649, 102, 315, 1109, 1235, 38, 673, 1066, 925, 940, 934, 201, 1229, 121, 122, 1189, 902, 1116, 123, 34, 1206, 250, 64, 172, 369, 175, 1298, 1056, 490, 1013, 1014, 1026, 832, 578, 473, 138, 462, 841, 842, 1159, 808, 305, 530, 1074, 474, 475, 515, 620, 768, 857, 913, 103, 27, 603, 1158, 623, 968, 983, 1009, 993, 972, 1199, 1230, 1257, 267, 268, 1209, 293, 639, 1202, 182, 1040, 156, 158, 224, 163, 1157, 389, 1216, 510, 958, 1039, 1081, 444, 1192, 1238, 1045, 83, 491, 420, 1233, 969, 967, 952, 1023, 512, 569, 610, 1030, 551, 1012, 410, 1044, 330, 896, 1178, 251, 406, 404, 1004, 86, 397, 33, 748, 547, 1000, 399, 400, 366, 1018, 513, 345, 380, 94, 621, 95, 13, 428, 672, 1200, 977, 927, 1119, 1120, 1029, 190, 1208, 879, 901, 1224, 296, 297, 263, 342, 539, 726, 494, 1260, 319, 320, 586, 300, 276, 493, 624, 1111, 292, 368, 777, 84, 788, 789, 184, 338, 467, 107, 166, 1193, 898, 622, 1062, 1082, 1053, 200, 509, 532, 653, 1096, 695, 592, 884, 633, 327, 749, 394, 1215, 943, 971; (3) System Code: 722233; UDI: 00884838116764; System Serial Number: 64, 122, 169, 170, 267, 96, 144, 212, 153, 259, 118, 214, 41, 59, 124, 147, 103, 108, 111, 172, 127, 192, 218, 219, 250, 123, 37, 38, 183, 273, 304, 17, 137, 76, 300, 12, 201, 238, 82, 288, 295, 297, 298, 8, 141, 26, 27, 235, 68, 14, 138, 139, 117, 279, 58, 240, 28, 7, 280, 281, 31, 165, 166, 62, 100, 95, 29, 200, 203, 204, 205, 120, 310, 285, 292, 161, 181, 110, 112, 173, 160, 51, 75, 202, 86, 197, 176, 223, 211, 136, 227, 81, 16, 24, 184, 129, 156, 222, 253, 30, 284, 134, 143, 233, 162, 71, 40, 48, 101, 180, 105, 106, 133, 57, 61, 186, 189, 89, 90, 91, 145, 44, 231, 268, 312, 85, 221, 193, 194, 83, 84, 87, 88, 19, 5, 6, 18, 178, 60, 257, 55, 239, 243, 130, 234, 9, 32, 159, 217, 260, 56, 20, 225, 291, 50, 11, 72, 107, 35, 46, 47, 74, 43, 104, 78, 290, 272, 249, 77, 80, 174, 163, 97, 102, 146, 148, 149, 109, 15, 13, 39, 206, 67;
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1677-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | March 3, 2026 |
| Reported | April 8, 2026 |
| Company | PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. |
| Company location | Best, Netherlands |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 1994 units (360 US, 1634 OUS) |
More recalls from PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · all medical device recalls · recalls from March 2026