Brand Name: da Vinci S recall (Intuitive Surgical, Inc.)

Medical deviceClass IIOngoing

Summary: Intuitive Surgical, Inc. is recalling Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endo. Reason: Due to increased complaints for broken/frayed grip cables for reusable instruments.

What is being recalled

Brand Name: da Vinci S, Si Monopolar Curved Scissors Instrument Product Name: da Vinci S and Si Monopolar Curved Scissors Instrument Model/Catalog Number: 420179 Software Version: N/A Product Description: EndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient). Component: N/A

Why

Due to increased complaints for broken/frayed grip cables for reusable instruments.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - U.S. Nationwide distribution in the states of CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MO, MS, NC, NJ, NY, OH, PA, TX, VA and WI. The countries of Argentina, Brazil, Canada, Chile, China, Cyprus, Czech Republic, Ecuador, France, Germany, Greece, Guadeloupe, India, Indonesia, Italy, Japan, Kuwait, Lebanon, Malaysia, Mexico, Pakistan, Philippines, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, South Korea, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam. (states named: California, Florida, Georgia, Iowa, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Missouri, Mississippi, North Carolina, New Jersey, New York, Ohio, Pennsylvania, Texas, Virginia, Wisconsin)

How to tell if yours is affected

Lot Code: 420179: Monopolar Curved Scissors (Hot Shears) GTIN: 00886874111505 FDA Medical Device Number: D088920 Material Number: 420179-10 Batch / System Numbers: M10101102 SH0935 M10101201 SH0935 M10101202 SH1279 M10101203 SH0059 M10101204 USG294 M10101206 USG294 M10101206 SH1378 M10101214 USG022 M10101214 SH0035 M10101214 SH0495 M10110104 SH1279 M10110111 SH1378 M10110112 USG294 M10110113 SH0075 M10110121 SG704 M10110121 SH1279 M10110124 SH0495 M10110126 USG721 M10110126 SH1391 M10110127 USG721 M10110127 SH1279 M10110127 SH2144 M10110202 USG294 M10110202 SH0495 M10110203 SH2144 M10110208 SG704 M10110210 SH1378 M10110216 SH1368 M10110216 USG294 M10110216 SH1391 M10110217 SH0495 M10110217 SH1279 M10110222 SH1279 M10110228 SH0547 M10110228 USG721 M10110228 USG022 M10110228 SH1378 M10110228 SH1279 M10110301 SH1279 M10110303 SH0495 M10110304 USG294 M10110304 SH0729 M10110304 SH0761 M10110307 USG294 M10110308 USG294 M10110308 SH0035 M10110308 SH0495 M10110308 SH1279 M10110311 USG294 M10110311 SH2144 M10110315 USG294 M10110318 USG294 M10110318 SH0935 M10110318 SH1279 M10110323 USG022 M10110324 SH0059 M10110324 USG022 M10110324 SH0495 M10110328 USG294 M10110328 SH0935 M10110328 SH1279 M10110329 USG294 M10110329 SH0935 M10110331 SH0935 M10110405 SH1378 M10110413 USG294 M10110413 SH1279 M10110414 SG704 M10110414 SH0495 M10110420 SH1279 M10110425 SH0035 M10110426 SH0059 M10110428 USG022 M10110429 SG2026 M10110429 SH0035 M10110504 USG294 M10110512 USG294 M10110512 USG022 M10110512 SH2144 M10110517 USG294 M10110517 SH0495 M10110518 SH0035 M10110524 USG294 M10110524 SH1279 M10110525 USG294 M10110526 USG294 M10110526 SH1378 M10110602 SH1279 M10110603 SH1378 M10110614 SH1279 M10110616 SH1368 M10110617 SH1182 M10110617 SH0495 M10110621 SH2144 M10110622 SH0935 M10110623 SH0729 M10110624 SH1378 M10110629 USG294 M10110630 SH1279 M10110630 SH2144 M10110701 SH1378 M10110705 SH0035 M10110711 USG294 M10110712 SH1378 M10110712 SH0495 M10110715 SH1279 M10110715 SH2144 M10110718 SH0761 M10110721 SH0935 M10110728 SH0729 M10110728 SH0761 M10110729 USG721 M10110801 SH1279 M10110802 SH0935 M10110803 USG294 M10110803 USG022 M10110803 SH2144 M10110805 SG704 M10110805 SH0495 M10110808 SH0035 M10110809 USG721 M10110810 SH1279 M10110811 USG022 M10110811 SH0729 M10110812 USG721 M10110812 SH0729 M10110815 SH0935 M10110815 SH1279 M10110816 SG704 M10110816 USG721 M10110816 SH0729 M10110819 USG294 M10110822 SG704 M10110822 USG294 M10110823 SH0729 M10110824 SH0059 M10110824 SH0729 M10110824 SH0495 M10110825 USG721 M10110826 USG721 M10110829 SH1279 M10110830 USG022 M10110902 SG704 M10110902 SH1279 M10110902 SH0761 M10110906 USG022 M10110908 USG721 M10110908 SH1378 M10110908 SH1279 M10110909 SH0495 M10110913 USG721 M10110922 SH2144 M10110923 USG294 M10110923 SH0729 M10110930 SH0935 M10110930 SH0495 M10111003 SG704 M10111004 USG294 M10111005 USG294 M10111006 SH0729 M10111007 SH1368 M10111007 SH0035 M10111007 SH2144 M10111010 SH0935 M10111011 SH0495 M10111011 SH1279 M10111013 SH1378 M10111015 USG022 M10111017 SH1015 M10111018 SH1368 M10111018 SH1279 M10111019 USG294 M10111020 SH0547 M10111020 SH1279 M10111021 USG022 M10111027 SG704 M10111027 SH1378 M10111028 SH0059 M10111028 USG294 M10111028 SH2144 M10111031 SH1015 M10111104 SH1182 M10111107 SH0059 M10111107 SH0729 M10111107 SH0761 M10111108 SH1279 M10111108 SH2144 M10111109 SH0495 M10111111 SH0935 M10111114 USG294 M10111114 SH0729 M10111114 SH1279 M10111115 SG704 M10111115 USG294 M10111115 USG022 M10111116 SH0825 M10111117 SH0729 M10111117 SH1391 M10111117 SH0761 M10111118 SH0935 M10111122 SH0495 M10111122 SH0761 M10111123 SH1390 M10111123 USG294 M10111123 SH0035 M10111124 USG294 M10111129 SH0935 M10111130 SH1378 M10111201 SH1368 M10111202 USG294 M10111202 SH0729 M10111208 USG294 M10111208 SH1279 M10111212 SH1378 M10111213 SG704 M10111213 SH1279 M10111214 USG294 M10111214 SH0729 M10111214 SH1015 M10111216 SH0495 M10111221 USG294 M10111221 SH1378 M10120104 SH1182 M10120104 USG022 M10120104 SH1279 M10120105 USG294 M10120109 USG022 M10120111 SH2144 M10120112 USG721 M10120113 SH0495 M10120117 SH0495 M10120117 SH1279 M10120127 USG294 M10120130 SG704 M10120130 SH1279 M10120131 SH0729 M10120201 SH1390 M10120201 SH0729 M10120203 SH0935 M10120208 SH0729 M10120208 SH1279 M10120208 SH2144 M10120209 SH1378 M10120213 SH0495 M10120214 SH1279 M10120216 SG704 M10120216 SH0495 M10120217 SH1182 M10120217 SH0761 M10120221 SH0729 M10120222 SH1391 M10120224 SH1390 M10120224 USG294 M10120224 SH0729 M10120228 SH1279 M10120301 SH1015 M10120305 SH1279 M10120308 SH0729 M10120309 SH0935 M10120312 USG022 M10120312 SH0935 M10120312 SH1015 M10120313 SH0825 M10120315 SH1279 M10120315 SH2144 M10120316 SH1390 M10120316 SH0729 M10120316 SH1279 M10120317 SH1276 M10120320 SG704 M10120320 SH1368 M10120320 SH1378 M10120320 SH0729 M10120321 SG2026 M10120326 SH1076 M10120327 SH0935 M10120330 SH0059 M10120330 USG294 M10120402 SH1279 M10120403 SH0729 M10120403 SH0495 M10120404 SH1279 M10120409 SH2144 M10120411 SG2026 M10120411 SH2144 M10120412 SH0935 M10120412 SH0761 M10120413 SH1276 M10120416 SH1276 M10120423 USG294 M10120424 USG294 M10120424 SH0035 M10120425 SH1015 M10120425 SH1279 M10120426 SH1390 M10120501 SG704 M10120501 USG721 M10120503 USG022 M10120503 SH0729 M10120503 SH2144 M10120503 SH1276 M10120507 SH1279 M10120515 SH1182 M10120515 SG704 M10120515 USG721 M10120516 SH2144 M10120518 SH1368 M10120521 SH0761 M10120522 USG294 M10120524 USG294 M10120524 SH0729 M10120528 USG294 M10120528 SH1279 M10120529 SH0935 M10120529 SH1276 M10120604 SH0495 M10120606 SH1378 M10120607 SH0547 M10120607 SH0825 M10120612 USG294 M10120612 SH0035 M10120612 SH1279 M10120612 SH0761 M10120614 SH1390 M10120614 SH1368 M10120615 SH0729 M10120622 SG704 M10120622 SH1378 M10120622 SH1015 M10120626 USG294 M10120626 SH1279 M10120627 SH2144 M10120629 USG022 M10120629 SH1378 M10120629 SH0935 M10120629 SH0495 M10120629 SH0761 M10120702 SH0729 M10120702 SH1015 M10120705 SH0059 M10120705 SH1368 M10120705 USG294 M10120705 SH0495 M10120705 SH0761 M10120706 SH0935 M10120706 SH1391 M10120709 SH1390 M10120711 USG294 M10120711 SH1076 M10120713 USG294 M10120713 SH0729 M10120713 SH1279 M10120713 SH2144 M10120713 SH0761 M10120719 USG294 M10120719 SH1279 M10120723 SG704 M10120726 USG721 M10120730 SH0729 M10120730 SH0035 M10120730 SH0761 M10120731 SH1385 M10120801 SH2144 M10120802 SH0059 M10120807 SH1378 M10120808 SH1076 M10120809 SH0729 M10120809 SH1276 M10120810 USG294 M10120813 USG721 M10120815 SH1378 M10120821 SH0935 M10120822 SH1378 M10120822 SH0935 M10120823 USG022 M10120823 SH0729 M10120823 SH0761 M10120827 USG294 M10120828 SH1279 M10120903 SG704 M10120903 SH1276 M10120904 SH0495 M10120906 USG022 M10120911 SH0825 M10120919 SH1276 M10120920 SH0729 M10120920 SH0035 M10120925 SH0495 M10120925 SH1015 M10120928 SH0729 M10121001 SH0059 M10121002 SH1015 M10121004 USG294 M10121004 SH0729 M10121004 SH0495 M10121004 SH1385 M10121006 SH1391 M10121008 SH1279 M10121008 SH0761 M10121010 USG294 M10121012 USG294 M10121015 SH1378 M10121016 SH0495 M10121017 SH2144 M10121019 SH1368 M10121019 SH0825 M10121019 SH0935 M10121023 SH0729 M10121025 USG022 M10121029 SH0729 M10121029 SH1391 M10121029 SH0761 M10121031 USG294 M10121031 SH0935 M10121102 SH1279 M10121106 USG294 M10121106 USG721 M10121112 USG721 M10121113 USG689 M10121114 SG704 M10121114 USG294 M10121115 SH1390 M10121115 USG721 M10121115 SH1391 M10121121 USG294 M10121123 SH0059 M10121123 SH1390 M10121123 SH0495 M10121126 SH0935 M10121128 SH0059 M10121128 SH0729 M10121129 SH1276 M10121204 USG022 M10121205 SH2144 M10121206 SH0059 M10121215 SH1378 M10121215 SH1391 M10121215 SH0495 M10121218 USG294 M10121219 SH0547 M10130102 SH0547 M10130102 SH0059 M10130102 SH1279 M10130104 USG294 M10130108 SH1390 M10130108 USG022 M10130108 SH1279 M10130109 SH1368 M10130110 SH1368 M10130110 SH0935 M10130114 SH0729 M10130115 USG294 M10130116 SH1570 M10130116 SH0935 M10130117 SH0059 M10130117 SH0825 M10130117 SH0729 M10130117 SH0495 M10130118 USG294 M10130121 SH1390 M10130122 USG721 M10130122 SH1015 M10130123 SH1182 M10130124 SH1378 M10130124 SH0761 M10130205 USG022 M10130207 SH1391 M10130207 SH1015 M10130208 SG704 M10130211 USG689 M10130212 SH0729 M10130212 SH0761 M10130213 SH0547 M10130214 SH0729 M10130214 SH0495 M10130215 SH1182 M10130215 SH1378 M10130222 SH1390 M10130227 SH0495 M10130228 SH1622 M10130304 SH0825 M10130304 SH0495 M10130305 SH1570 M10130306 SH0761 M10130306 SH1276 M10130307 USG294 M10130308 SG2026 M10130308 SH1378 M10130308 SH1279 M10130313 USG294 M10130313 SH0935 M10130315 SH2144 M10130315 SH0761 M10130318 SH0059 M10130319 SH1279 M10130319 SH2144 M10130322 SH1279 M10130402 USG022 M10130404 USG022 M10130408 SH1378 M10130408 SH0729 M10130409 SH0059 M10130409 SH0761 M11110225 SG704 M11110323 SH1368 M11110517 SH0729 M11110517 SH1279 M11110517 SH2144 M11110525 SH0761 M11110526 SG704 M11110531 SH0495 M11110602 USG294 M11110602 SH0729 M11110623 SH1368 M11110623 SH1279 M11110625 SH0547 M11110625 USG294 M11110630 SG704 M11110630 USG294 M11110701 USG721 M11110701 SH0495 M11110701 SH0761 M11110808 SH0495 M11110808 SH0761 M11110809 USG294 M11110809 SH1391 M11110809 SH2144 M11110810 SH0761 M11110825 SH0059 M11110825 SH0761 M11110826 SH2144 M11110929 SG2026 M11110929 SH1279 M11111026 SH0729 M11111027 USG294 M11111028 SH0935 M11111028 SH0495 M11111117 USG022 M11111117 SH0729 M11111117 SH0761 M11111124 SH1015 M11111129 USG721 M11111130 USG294 M11111130 SH2144 M11120103 SH0059 M11120103 SH0729 M11120116 USG294 M11120116 SH0495 M11120126 SH0035 M11120127 SH1279 M11120130 USG294 M11120131 SH1182 M11120208 USG294 M11120209 SH0761 M11120210 SH0935 M11120213 SH0729 M11120214 SG2026 M11120330 USG294 M11120402 SH1391 M11120403 USG022 M11120404 SH1279 M11120517 USG294 M11120517 SH1279 M11120522 SH1368 M11120522 SH1276 M11120607 USG294 M11120607 USG022 M11120608 USG022 M11120717 USG294 M11120801 SH1015 M11120813 SH1368 M11120814 SH0729 M11120814 SH1276 M11120816 SH0729 M11120821 USG294 M11120906 SH1378 M11120919 USG689 M11120928 SH0935 M11120928 SH1276 M11121004 USG294 M11121004 SH1015 M11121004 SH1279 M11121010 USG294 M11121011 SH1182 M11121015 SH1378 M11121016 SH2144 M11121114 SH1557 M11121114 SH1378 M11121114 SH0935 M11121114 SH1279 M11121114 SH1276 M11121115 SH1076 M11121116 SH1279 M11121129 USG294 M11121129 SH0761 M11121130 SH0059 M11121203 USG294 M11121205 SH0035 M11121206 SH1604 M11121206 SH1385 M11121211 SG704 M11121211 SH0935 M11121211 SH2144 M11121214 SH1622 M11121218 SH0495 M11121219 SH1279 M11130112 USG294 M11130121 SH1276 M11130124 USG294 M11130124 SH0729 M11130124 SH1279 M11130207 USG721 M11130211 SH1378 M11130213 USG294 M11130214 SH2144 M11130225 SH0729 M11130226 USG721 M11130227 SH0059 M11130227 USG294 M11130228 SH0729 M11130301 SG704 M11130306 SH0059 M11130307 SH1378 M11130315 SH1378 M11130315 SH1391 M11130315 SH1015 M11130322 SH0729 M11130322 SH1276 M11130402 SH1279 M11130404 SH1015 M12110228 SH1279 M12120614 USG294 M12120614 SH1378 M12120801 SH1378 M12120816 SH1279 M12120906 SH1279 M12121011 SH1276 M12121105 SH0495 M12121207 SH1368 M12121211 SH0729 M12130109 SH0495 M12130111 SG704 M12130213 SH1279 M12130221 USG721 M12130226 SH0729 M12130226 SH0035 M12130320 USG294 M12130320 SH0935 M12130320 SH2144 M13120801 SH1390 M13120801 SH0495 M13121206 SH1378 M13130110 USG294 M13130110 SH0761 M13130111 USG721 M13130111 SH0935 M13130112 SH0035 M20111026 SH0935 M20111118 USG294 M20111206 SH0935 M20111206 SH1279 M20120116 USG294 M20120116 SH2144 M20120125 SH1391 M20120125 SH1076 M20120306 SH0495 M20120307 SH0935 M20120307 SH1279 M20120308 SH0035 M20120314 USG294 M20120314 SH0495 M20120514 USG721 M20120514 SH0825 M20120514 SH0729 M20120514 SH1276 S12110708 USG294 Material Number: 420179-12 Batch / System Numbers: M10130601 SH1390 M10130603 SH0825 M10130603 SH0495 M10130605 USG022 M10130605 SH2144 M10130606 SH1570 M10130611 SH1390 M10130612 SH2144 M10130612 SH1276 M10130614 SH0935 M10130619 SG704 M10130620 SH1378 M10130621 SH0729 M10130622 SH1276 M10130703 SH1182 M10130703 SH0035 M10130704 SH2144 M10130710 SH1378 M10130713 SH0729 M10130713 SH1385 M10130713 SH0761 M10130716 SH0761 M10130716 SH1276 M10130717 SG2026 M10130717 SH0935 M10130717 SH1279 M10130718 SH0729 M10130722 USG294 M10130722 SH0761 M10130725 USG022 M10130725 SH1279 M10130726 SH1557 M10130726 USG294 M10130726 SH1391 M10130731 USG294 M10130731 SH0495 M10130801 SH1182 M10130802 SH0495 M10130805 SH0059 M10130805 SH0729 M10130806 USG294 M10130806 SH1391 M10130812 SH1276 M10130814 SH0035 M10130815 SH1378 M10130815 SH0729 M10130815 SH2144 M10130816 SH1570 M10130816 USG721 M10131105 USG294 M10131111 SH0935 M10131111 SH0729 M10131115 USG294 M10131115 SH1279 M10131119 USG022 M10131120 USG294 M10131120 SH1378 M10131120 SH1279 M10131121 SG704 M10131121 USG022 M10131122 SH0935 M10131122 SH0761 M10131123 SH0547 M10131123 SH1837 M10131123 SH1390 M10131128 SH0825 M10131128 SH1378 M10131202 SH0761 M10131202 SH1276 M10131204 SH1604 M10131206 SH1182 M10131206 SH0547 M10131206 SH2144 M10131207 SH0729 M10131207 SH1279 M10131207 SH1276 M10131211 USG294 M10131212 SH0059 M10131212 USG022 M10131212 SH0729 M10131212 SH1279 M10131214 SG704 M10131216 SH0729 M10131217 SH1570 M10131217 SH1557 M10131217 SH0729 M10131217 SH1279 M10131217 SH2144 M10131218 SH1368 M10140108 SH0761 M10140113 SH1378 M10140115 SH1390 M10140115 SH0935 M10140116 SH1557 M10140116 USG721 M10140116 SH1622 M10140117 USG721 M10140120 USG294 M10140120 USG022 M10140123 USG294 M10140123 SH1015 M10140212 SH0547 M10140212 SH1368 M10140212 USG294 M10140212 SH0761 M10140213 SG704 M10140213 SH0729 M10140213 SH0495 M10140213 SH1279 M10140217 SH1276 M10140306 SH1570 M10140328 SH1378 M10140328 SH1391 M10140401 SH1557 M10140401 SH0729 M10140402 SH1385 M10140402 SH1279 M10140416 SH1997 M10140416 SH1391 M11130527 SH1279 M11130603 SH0059 M11130603 SH0729 M11130606 USG721 M11130606 SH1385 M11130607 SH1837 M11130607 SH1279 M11130611 SH0059 M11130613 USG721 M11130613 SH1391 M11130614 SH1570 M11130614 SH1837 M11130618 SH0059 M11130618 SH1390 M11130622 SH1390 M11130622 SH1279 M11130622 SH1276 M11130703 SH1390 M11130708 USG022 M11130709 USG294 M11130710 SH1837 M11130710 SH0935 M11130710 SH1279 M11130711 SH0495 M11130712 USG294 M11130713 SH1368 M11130713 SH0729 M11130716 SH0059 M11130716 SH0729 M11130716 SH2144 M11130718 USG689 M11130718 USG294 M11130722 SH0825 M11130723 SH1378 M11130724 USG294 M11130726 SG704 M11130726 SH1557 M11130726 SH0761 M11130730 SH1390 M11130805 SH1368 M11130805 SG2026 M11130809 USG689 M11130813 SG704 M11130813 USG294 M11130814 SH1837 M11130814 USG022 M11130815 SH0729 M11130815 SH1015 M11130816 USG294 M11130816 SH0729 M11130816 SH1279 M11131111 SH0495 M11131111 USG518 M11131202 USG294 M11131202 SH1378 M11131204 SH0935 M11131205 SH1391 M11131205 SH1279 M11131207 USG721 M11131207 SH0935 M11131207 SH0729 M11131207 SH0495 M11131207 SH2144 M11140108 USG294 M11140108 SH0495 M11140116 SH1378 M11140116 SH1279 M11140211 SH1557 M11140211 SH1378 M11140228 USG721 M11140228 SH1279 M12130531 SH0761 M12130603 SH0729 M12130608 SH0761 M12130610 SG704 M12130611 SH0059 M12130613 USG022 M12130613 SH1391 M12130615 USG518 M12130620 SH0935 M12130620 SH1385 M12130621 SH2144 M12130624 USG294 M12130624 SH1015 M12130702 SH0729 M12130704 SH0495 M12130705 SH1279 M12130708 SH1385 M12130709 SH1276 M12130710 SH1076 M12130710 SH0495 M12130712 USG721 M12130712 SH1385 M12130715 SH0547 M12130715 USG022 M12130717 SH0729 M12130717 SH1391 M12130722 SH0035 M12130726 SG704 M12130726 SH1279 M12130729 SH0729 M12130801 SH1182 M12130802 USG294 M12130802 SH2144 M12130805 SH1557 M12130807 USG294 M12130807 USG022 M12130808 SH1279 M12130809 SH1378 M12130809 SH0761 M12131123 USG689 M13130601 USG294 M13130605 SH0935 M13130605 SH1276 M13130606 SH0035 M13130608 SH0059 M13130613 SH0935 M13130613 SH0035 M13130615 SH1279 M13130617 USG294 M13130620 USG721 M13130621 SH1368 M13130621 SH1378 M13130622 SH1015 M13130624 SH1368 M13130624 SH0761 M13130626 SH0729 M13130702 USG294 M13130703 SH1279 M13130703 SH2144 M13130708 SH1279 M13130712 SH2144 M13130712 SH1276 M13130715 USG721 M13130715 SH1378 M13130715 SH0729 M13130716 USG022 M13130716 SH1076 M13130722 SH1570 M13130722 USG518 M13130724 SH1279 M13130726 USG294 M13130726 SH0825 M13130726 SH0935 M13130731 USG294 M13130731 SH0729 M13130807 SH1378 M13130808 USG294 M13131123 SH1557 M13131123 SH0729 M14130606 SH1391 M14130606 USG518 M14130607 USG294 M14130610 USG294 M14130610 SH0729 M14130613 SH0035 M14130614 SH1622 M14130617 USG294 M14130617 SH0761 M14130620 USG294 M14130620 USG721 M14130621 SH0825 M14130626 SH1570 M14130626 SH1391 M14130702 SH1279 M14130703 USG294 M14130724 SH1015 M14130724 SH1276 M14130725 SH1622 M14130730 USG294 N10130926 SH0495 N10131217 SH1015 N10131217 SH2144 N10140106 SH1015 N10140109 SH1385 N10140110 SH1378 N10140120 SH0729 N10140120 SH1279 N10140120 SH0761 N10140123 SH1378 N10140130 SH0825 N10140130 SH1378 N10140130 SH0935 N10140204 SH1837 N10140204 USG721 N10140204 SH0729 N10140207 SH1276 N10140210 SH0059 N10140210 SH1557 N10140210 USG294 N10140210 SH0729 N10140210 SH2125 N10140211 SH0935 N10140214 SG704 N10140214 USG294 N10140214 SH0495 N10140217 USG294 N10140221 SH0495 N10140224 SH0729 N10140313 USG294 N10140313 SH1378 N10140313 SH0729 N10140313 SH0495 N10140314 SH1557 N10140325 SH0761 N10140401 SH1390 N10140401 SH2186 N10140403 SH1182 N10140403 USG294 N10140408 SH2144 N10140410 SH1378 N10140410 SH0495 N10140410 SH1015 N10140414 USG294 N10140422 USG294 N10140422 SH0761 N10140423 USG294 N10140424 SH1378 N10140425 SH1390 N10140425 USG022 N10140428 SH1015 N10140428 SH1279 N10140428 SH2144 N10140501 USG294 N10140501 SH1391 N10140501 SH2125 N10140501 USG518 N10140507 SH1557 N10140507 SH0729 N10140508 SH0729 N10140512 SH1378 N10140512 SH0935 N10140512 SH1279 N10140513 USG518 N10140514 USG721 N10140519 SH1182 N10140519 USG022 N10140519 SH0495 N10140520 USG721 N10140520 SH0729 N10140523 SH1557 N10140523 SH2144 N10140527 USG721 N10140527 SH1279 N10140530 USG721 N11140204 SH0059 N11140204 USG294 N11140204 SH0729 N11140204 SH0035 N11140314 SH0547 N11140314 USG294 N11140314 USG022 N11140321 USG022 N11140423 SH1378 N11140423 SH2125 N11140428 SH0935 N11140428 SH0761 N11140501 SG704 N11140501 SH0035 N11140501 SH2125 N11140508 SH0059 N11140508 SH0761 N11140509 SH0729 N11140509 SH2144 N11140513 SH0547 N11140514 SH1378 N11140515 SH1279 N11140515 SH0761 N11140516 SH1622 N11140519 SH0547 N11140519 SH1837 N11140521 USG294 N11140526 USG689 N11140526 SH0825 N11140527 SH0729 N11140527 SH1279 N11140529 SH1378 N11140530 SH1076 N12140321 SH1182 N12140325 SH2144 N12140515 SH1378 N13140325 SH0729 N13140424 USG689 S10130603 SH1279 S11130530 SH0761 S11130603 USG294 S11130603 SH1378 Material Number: 420179-14 Batch / System Numbers: N10140620 SH1182 N11140730 SH1182 N13140808 SH1182 N11140904 SH1182 N11150109 SH1182 N11150228 SH1182 N12140620 SH1570 N11140923 SH1570 N11141111 SH1570 N11150228 SH1570 N11140630 SG704 N10140704 SG704 N10140721 SG704 N10140729 SG704 N10140811 SG704 N13140808 SG704 N11140814 SG704 N10140829 SG704 N10140917 SG704 N10140926 SG704 N10141217 SG704 N12141217 SG704 N13150302 SG704 N11140703 SH0547 N10140909 SH0547 N13140909 SH0547 N11141213 SH0547 N10150212 SH0547 N10140617 SH0059 N12140704 SH0059 N10140718 SH0059 N11140820 SH0059 N11140916 SH0059 N10141119 SH0059 N10141125 SH0059 N11150218 SH0059 N11140729 SH1997 N12140620 SH1557 N10140701 SH1557 N10140711 SH1557 N11140724 SH1557 N10140812 SH1557 N11140822 SH1557 N10140828 SH1557 N10140909 SH1557 N10141022 SH1557 N10141111 SH1557 N13141124 SH1557 N11141209 SH1557 N11141213 SH1557 N14141223 SH1557 N10150106 SH1557 N10150115 SH1557 N10150128 SH1557 N11150213 SH1557 N11150219 SH1557 N10150303 SH1557 N10150109 SH1964 N11150218 SH1964 N11140623 SH1837 N10141010 SH1837 N12140612 SH1390 N10140701 SH1390 N10140910 SH1390 N10141206 SH1390 N10150210 SH1390 N13150302 SH1390 N12140612 USG689 N11140701 USG689 N11141001 USG689 N11140806 SH1368 N10141116 SH1368 N12141213 SH1368 N11140610 USG294 N10140620 USG294 N10140701 USG294 N11140701 USG294 N10140711 USG294 N10140718 USG294 N11140730 USG294 N11140731 USG294 N11140820 USG294 N10140829 USG294 N13141023 USG294 N14141023 USG294 N11141104 USG294 N10141201 USG294 N11141210 USG294 N12141213 USG294 N12141217 USG294 N12150121 USG294 N11150123 USG294 N12150212 USG294 N10150219 USG294 N11150228 USG294 N13141125 USG294 N12140710 USG721 N10140806 USG721 N10140814 USG721 N11140829 USG721 N11140901 USG721 N10140903 USG721 N10141029 USG721 N11141103 USG721 N10141211 USG721 N14141223 USG721 N11150211 USG721 N10140701 USG022 N10140703 USG022 N10140806 USG022 N10140814 USG022 N11140917 USG022 N11141211 USG022 N10141213 USG022 N10141218 USG022 N11150115 USG022 N11150209 USG022 N11150211 USG022 N11141110 SH2186 N11141206 SH2186 N12141208 SH2186 N10141218 SH2186 N10140619 SH0825 N11140806 SH0825 N11140903 SH0825 N10141022 SH0825 N14141218 SH0825 N10140620 SH1378 N13140623 SH1378 N10140630 SH1378 N12140704 SH1378 N11140724 SH1378 N10140729 SH1378 N10140909 SH1378 N10140905 SH1378 N11141022 SH1378 N13141023 SH1378 N11141104 SH1378 N10141127 SH1378 N12141129 SH1378 N10141202 SH1378 N11141208 SH1378 N13141218 SH1378 N10141222 SH1378 N11150116 SH1378 N11150219 SH1378 N11150225 SH1378 N10140701 SH0935 N12140710 SH0935 N10140813 SH0935 N13140808 SH0935 N11140909 SH0935 N10141028 SH0935 N10141104 SH0935 N11141208 SH0935 N12141210 SH0935 N11150126 SH0935 N11150204 SH0935 N12150212 SH0935 N12140623 SH0729 N12140704 SH0729 N10140721 SH0729 N10140730 SH0729 N11140729 SH0729 N10140815 SH0729 N10140820 SH0729 N10140904 SH0729 N11140904 SH0729 N10140909 SH0729 N12140910 SH0729 N13141022 SH0729 N11141105 SH0729 N10141127 SH0729 N12141202 SH0729 N11150119 SH0729 N10150109 SH0729 N11150123 SH0729 N11150211 SH0729 N10150224 SH0729 N13150225 SH0729 N10140630 SH0035 N11140801 SH0035 N10141001 SH0035 N11141201 SH0035 N10150218 SH0035 N10140617 SH1391 N13140808 SH1391 N11141129 SH1391 N12150216 SH1391 N13150224 SH1391 N10140607 SH2125 N12141125 SH2125 N10140711 SH1622 N12150121 SH1622 N12140910 SH1076 N11140611 SH1604 N12141023 SH1604 N12141211 SH1604 N10140617 SH0495 N10140703 SH0495 N11140707 SH0495 N10140722 SH0495 N11140812 SH0495 N11140904 SH0495 N10140910 SH0495 N10141103 SH0495 N10141124 SH0495 N11141208 SH0495 N11141217 SH0495 N10141222 SH0495 N11150203 SH0495 N10150225 SH0495 N13150302 SH0495 N10140607 SH1015 N11140612 SH1015 N10140711 SH1015 N11140724 SH1015 N11140926 SH1015 N10141001 SH1015 N10141022 SH1015 N10141106 SH1015 N12141116 SH1015 N10150209 SH1015 N12150216 SH1015 N10150216 SH1015 N13150302 SH1015 N10140619 SH1385 N10140729 SH1385 N12140808 SH1385 N11140917 SH1385 N10141202 SH1385 N10150219 SH1385 N13150302 SH1385 N11140612 SH1279 N10140613 SH1279 N10140620 SH1279 N11140630 SH1279 N11140711 SH1279 N10140718 SH1279 N10140820 SH1279 N13140909 SH1279 N11140926 SH1279 N10141010 SH1279 N10141104 SH1279 N10141110 SH1279 N13141116 SH1279 N12141208 SH1279 N12141211 SH1279 N12141217 SH1279 N10150109 SH1279 N11150218 SH1279 N10150228 SH1279 N11150303 SH1279 N11140610 SH2144 N11140611 SH2144 N10140904 SH2144 N10140903 SH2144 N11140909 SH2144 N10141021 SH2144 N12141116 SH2144 N12141124 SH2144 N11141201 SH2144 N12141222 SH2144 N10150203 SH2144 N10150226 SH2144 N12140630 SH0761 N10140722 SH0761 N11140801 SH0761 N11140923 SH0761 N11141103 SH0761 N11141201 SH0761 N12141213 SH0761 N11141217 SH0761 N12150121 SH0761 N11150211 SH0761 N11140630 SH1276 N11140711 SH1276 N10140806 SH1276 N11140822 SH1276 N10141110 USG518 N10141116 USG518 N12141125 USG518 N10150217 USG518 Material Number: 420179-15 Batch / System Numbers: N11150625 SH1182 N10150810 SH1182 N11150810 SH1182 N10150911 SH1182 N10150916 SH1182 N11150922 SH1182 N10151117 SH1182 N12151202 SH1182 N10160107 SH1182 N10160129 SH1182 N11160309 SH1182 N10160620 SH1182 N12160708 SH1182 N10160715 SH1182 N10161014 SH1182 N11161116 SH1182 N11170106 SH1182 N12170110 SH1182 N10170126 SH1182 N11170217 SH1182 N10170309 SH1182 N11170526 SH1182 N10170626 SH1182 N11170706 SH1182 N11170627 SH1182 N10170830 SH1182 N10171128 SH1182 N10180125 SH1182 N10180222 SH1182 N10180326 SH1182 N10180316 SH1182 N10180424 SH1182 N10150605 SH1570 N10150603 SH1570 N11150825 SH1570 N10151015 SH1570 N10151202 SH1570 N11151216 SH1570 N10160112 SH1570 N10160422 SH1570 N14160725 SH1570 N12160712 SH1570 N10161221 SH1570 N12170309 SH1570 N11170427 SH1570 N10170630 SH1570 N11170609 SH1570 N10180208 SH1570 N10180405 SH1570 N11180220 SH1570 N11150311 SG704 N10150416 SG704 N11150505 SG704 N11150609 SG704 N11150730 SG704 N10150729 SG704 N10150801 SG704 N10151005 SG704 N10151202 SG704 N12151208 SG704 N10151222 SG704 N10160119 SG704 N10160220 SG704 N13160305 SG704 N11160504 SG704 N11160609 SG704 N13160801 SG704 N11160912 SG704 N10161013 SG704 N11161102 SG704 N11170103 SG704 N11170217 SG704 N10170223 SG704 N10170421 SG704 N11170427 SG704 N10170515 SG704 N11170526 SG704 N10170626 SG704 N10170714 SG704 N10170728 SG704 N10171002 SG704 N10171115 SG704 N10171204 SG704 N10171207 SG704 N10180104 SG704 N10180125 SG704 N10180202 SG704 N11180123 SG704 N10180321 SG704 N10180306 SG704 N10180618 SG704 N13150304 SH0547 N12150511 SH0547 N11150519 SH0547 N10150624 SH0547 N10150921 SH0547 N10150923 SH0547 N11151210 SH0547 N13151208 SH0547 N10160210 SH0547 N10160414 SH0547 N10160519 SH0547 N11160614 SH0547 N11160811 SH0547 N11150312 SH0059 N11150401 SH0059 N12150319 SH0059 N10150507 SH0059 N10150519 SH0059 N10150601 SH0059 N10150602 SH0059 N10150623 SH0059 N10150731 SH0059 N10150903 SH0059 N10150929 SH0059 N11151030 SH0059 N10151209 SH0059 N12151202 SH0059 N10151221 SH0059 N11151210 SH0059 N10160718 SH0059 N12160815 SH0059 N13160823 SH0059 N10161020 SH0059 N10161104 SH0059 N10161116 SH0059 N11161117 SH0059 N10170425 SH0059 N10170426 SH0059 N10170509 SH0059 N10170610 SH0059 N10170704 SH0059 N10170717 SH0059 N10170718 SH0059 N10170720 SH0059 N10170926 SH0059 N10171109 SH0059 N10171205 SH0059 N10171215 SH0059 N10171214 SH0059 N10180202 SH0059 N10180509 SH0059 N11180511 SH0059 N11150312 SH1557 N10150330 SH1557 N10150417 SH1557 N11150429 SH1557 N10150526 SH1557 N11150528 SH1557 N10150622 SH1557 N10150803 SH1557 N10150909 SH1557 N11150825 SH1557 N10151009 SH1557 N11151012 SH1557 N12151118 SH1557 N11160116 SH1557 N10160215 SH1557 N10160225 SH1557 N10160408 SH1557 N11160504 SH1557 N10160516 SH1557 N11160606 SH1557 N10160623 SH1557 N11160614 SH1557 N11160715 SH1557 N11160808 SH1557 N13160823 SH1557 N11160930 SH1557 N12160930 SH1557 N10161128 SH1557 N11161202 SH1557 N10161219 SH1557 N10170201 SH1557 N11170120 SH1557 N10170225 SH1557 N12170404 SH1557 N11170421 SH1557 N10170509 SH1557 N11170526 SH1557 N10170606 SH1557 N10170706 SH1557 N10170718 SH1557 N10170810 SH1557 N10170815 SH1557 N10171011 SH1557 N10171023 SH1557 N10171117 SH1557 N10180123 SH1557 N10180202 SH1557 N10180216 SH1557 N10180302 SH1557 N11180413 SH1557 N10180423 SH1557 N10180306 SH1557 N10180418 SH1557 N10150427 SH1964 N11150429 SH1964 N10150622 SH1964 N11150626 SH1964 N11150728 SH1964 N10150814 SH1964 N10151014 SH1964 N11150928 SH1964 N10151102 SH1964 N10160106 SH1964 N10160303 SH1964 N10160321 SH1964 N10160505 SH1964 N10160705 SH1964 N11160809 SH1964 N10161028 SH1964 N11161107 SH1964 N11161125 SH1964 N10170309 SH1964 N10170404 SH1964 N10170519 SH1964 N10170526 SH1964 N11170620 SH1964 N10170707 SH1964 N11170906 SH1964 N10171101 SH1964 N10180119 SH1964 N10180201 SH1964 N10180209 SH1964 N10180122 SH1964 N10180322 SH1964 N10180706 SH1964 N11150622 SH1837 N12150725 SH1837 N10150804 SH1837 N10150714 SH1837 N10160105 SH1837 N12160831 SH1837 N12161014 SH1837 N10161128 SH1837 N10170913 SH1837 N10171020 SH1837 N10180219 SH1837 N11180511 SH1837 N10180301 SH1822 N13150318 SH1390 N10150518 SH1390 N10150624 SH1390 N10150623 SH1390 N10151009 SH1390 N10151028 SH1390 N12151202 SH1390 N11160225 SH1390 N12160305 SH1390 N11160419 SH1390 N11160428 SH1390 N10160628 SH1390 N11160725 SH1390 N11160714 SH1390 N13160830 SH1390 N13160815 SH1390 N11161010 SH1390 N11161028 SH1390 N10161124 SH1390 N11161223 SH1390 N11170421 SH1390 N10170505 SH1390 N10170612 SH1390 N10170707 SH1390 N10171027 SH1390 N10171101 SH1390 N10180112 SH1390 N10180215 SH1390 N10180302 SH1390 N10180306 SH1390 N10150325 USG689 N10150630 USG689 N12150525 USG689 N10150729 USG689 N10150814 USG689 N10150904 USG689 N10160415 USG689 N10160512 USG689 N11160901 USG689 N10161003 USG689 N10170202 USG689 N10170209 USG689 N12170304 USG689 N11171221 USG689 N10171214 USG689 N10180326 USG689 N10180503 USG689 N10150422 SH1368 N11150708 SH1368 N10150730 SH1368 N11150922 SH1368 N10160106 SH1368 N11160213 SH1368 N10160317 SH1368 N11160418 SH1368 N10170116 SH1368 N11170517 SH1368 N11171220 SH1368 N10171214 SH1368 N10180413 SH1368 N10180504 SH1368 N11150306 USG294 N10150324 USG294 N10150326 USG294 N11150407 USG294 N13150318 USG294 N11150416 USG294 N10150424 USG294 N11150505 USG294 N11150519 USG294 N10150528 USG294 N11150529 USG294 N10150616 USG294 N12150609 USG294 N10150708 USG294 N10150727 USG294 N10150804 USG294 N10150810 USG294 N10150824 USG294 N10150902 USG294 N10150909 USG294 N10150916 USG294 N11150824 USG294 N10151005 USG294 N11151013 USG294 N12150820 USG294 N11151014 USG294 N11151030 USG294 N10151203 USG294 N12151006 USG294 N11151208 USG294 N10151216 USG294 N10160104 USG294 N12160116 USG294 N11160128 USG294 N10160210 USG294 N10160213 USG294 N10160301 USG294 N10160307 USG294 N11160312 USG294 N10160321 USG294 N11160330 USG294 N10160422 USG294 N10160505 USG294 N10160512 USG294 N10160517 USG294 N10160613 USG294 N10160623 USG294 N10160630 USG294 N10160714 USG294 N13160803 USG294 N11160809 USG294 N11160815 USG294 N11160818 USG294 N11160906 USG294 N14160815 USG294 N10160930 USG294 N14160913 USG294 N10161010 USG294 N10161021 USG294 N10161014 USG294 N11161124 USG294 N10160915 USG294 N11170103 USG294 N10170117 USG294 N11170126 USG294 N11170120 USG294 N11170309 USG294 N11170327 USG294 N10170330 USG294 N11170427 USG294 N10170428 USG294 N10170427 USG294 N10170511 USG294 N10170607 USG294 N11170606 USG294 N10170610 USG294 N10170620 USG294 N12170610 USG294 N10170707 USG294 N10170720 USG294 N10170803 USG294 N10170810 USG294 N10170901 USG294 N10170907 USG294 N10170929 USG294 N10171030 USG294 N10171128 USG294 N10171207 USG294 N10171219 USG294 N10180119 USG294 N10180130 USG294 N11180123 USG294 N10180209 USG294 N10180314 USG294 N10180326 USG294 N10180413 USG294 N10180424 USG294 N10180316 USG294 N11180423 USG294 N10180306 USG294 N10180510 USG294 N10150304 USG721 N11150312 USG721 N10150325 USG721 N10150602 USG721 N10150608 USG721 N11150622 USG721 N11150625 USG721 N12150725 USG721 N11150814 USG721 N10151019 USG721 N10150725 USG721 N11151110 USG721 N11151205 USG721 N12151202 USG721 N10151210 USG721 N11160513 USG721 N10160601 USG721 N10160606 USG721 N10160609 USG721 N10161020 USG721 N10161026 USG721 N11161007 USG721 N11161020 USG721 N10161128 USG721 N11160921 USG721 N11161214 USG721 N10170105 USG721 N11170213 USG721 N10170228 USG721 N10170224 USG721 N10170426 USG721 N10170512 USG721 N10170519 USG721 N10170620 USG721 N11170627 USG721 N10170720 USG721 N10170726 USG721 N10170810 USG721 N10170817 USG721 N10170815 USG721 N10171025 USG721 N10171026 USG721 N10171124 USG721 N10171206 USG721 N10180104 USG721 N10180131 USG721 N10180207 USG721 N10180301 USG721 N10180216 USG721 N11180103 USG721 N10180220 USG721 N10150428 USG022 N13150427 USG022 N10150603 USG022 N10150622 USG022 N10150623 USG022 N12150806 USG022 N10150923 USG022 N10151118 USG022 N13151204 USG022 N10160206 USG022 N11160310 USG022 N10160316 USG022 N10160419 USG022 N10160531 USG022 N11160615 USG022 N11160725 USG022 N10160830 USG022 N11160927 USG022 N10161028 USG022 N11161107 USG022 N10161122 USG022 N11170120 USG022 N11170403 USG022 N10170504 USG022 N10170515 USG022 N11160811 USG022 N10170626 USG022 N11170706 USG022 N10170720 USG022 N11170814 USG022 N10171017 USG022 N10171101 USG022 N11171221 USG022 N10171214 USG022 N10180504 USG022 N10180322 USG022 N10150605 SH2186 N10150624 SH2186 N10151106 SH2186 N12151202 SH2186 N11160513 SH2186 N10160705 SH2186 N12170225 SH2186 N10170404 SH2186 N10170530 SH2186 N10171023 SH2186 N10171128 SH2186 N10180426 SH2186 N12150511 SH0825 N11150820 SH0825 N11151205 SH0825 N11160120 SH0825 N10160303 SH0825 N14160725 SH0825 N10160809 SH0825 N14160815 SH0825 N11160923 SH0825 N10161031 SH0825 N10170311 SH0825 N10170526 SH0825 N10170630 SH0825 N10171215 SH0825 N10180129 SH0825 N10180328 SH0825 N10150312 SH1378 N10150318 SH1378 N10150506 SH1378 N11150525 SH1378 N10150529 SH1378 N11150529 SH1378 N10150624 SH1378 N10150623 SH1378 N10150908 SH1378 N10150915 SH1378 N10150929 SH1378 N10151016 SH1378 N10150714 SH1378 N12151006 SH1378 N12151202 SH1378 N10160212 SH1378 N10160301 SH1378 N10160308 SH1378 N12160305 SH1378 N10160328 SH1378 N10160421 SH1378 N10160512 SH1378 N11160504 SH1378 N10160518 SH1378 N11160628 SH1378 N11160714 SH1378 N10160718 SH1378 N11160823 SH1378 N13160815 SH1378 N10161013 SH1378 N10161104 SH1378 N10161215 SH1378 N11170103 SH1378 N10170208 SH1378 N10170209 SH1378 N10170213 SH1378 N11170223 SH1378 N10170316 SH1378 N11170403 SH1378 N10170421 SH1378 N11170503 SH1378 N10170606 SH1378 N10170610 SH1378 N10170713 SH1378 N10170802 SH1378 N10170814 SH1378 N10170831 SH1378 N10170926 SH1378 N10171020 SH1378 N10171027 SH1378 N10171114 SH1378 N10171129 SH1378 N10180102 SH1378 N10180113 SH1378 N10180206 SH1378 N10180213 SH1378 N10180305 SH1378 N11180125 SH1378 N11180103 SH1378 N10180328 SH1378 N10180306 SH1378 N10180418 SH1378 N10180509 SH1378 N11180511 SH1378 N10180702 SH1378 N10150306 SH0935 N10150408 SH0935 N12150511 SH0935 N11150605 SH0935 N10150622 SH0935 N12150609 SH0935 N10150923 SH0935 N11150922 SH0935 N10151014 SH0935 N10151026 SH0935 N12151006 SH0935 N10151214 SH0935 N10151216 SH0935 N10151222 SH0935 N11160310 SH0935 N12160309 SH0935 N11160606 SH0935 N10160617 SH0935 N11160714 SH0935 N10160718 SH0935 N10161003 SH0935 N11161004 SH0935 N10161020 SH0935 N11161026 SH0935 N11161124 SH0935 N10170110 SH0935 N12170208 SH0935 N10170301 SH0935 N11170410 SH0935 N10170510 SH0935 N10170606 SH0935 N10170615 SH0935 N10170623 SH0935 N10170707 SH0935 N10170901 SH0935 N11170906 SH0935 N10171026 SH0935 N10171113 SH0935 N10171208 SH0935 N10180111 SH0935 N10180130 SH0935 N10180216 SH0935 N10180326 SH0935 N10180405 SH0935 N11180413 SH0935 N10180306 SH0935 N10150326 SH0729 N11150326 SH0729 N12150311 SH0729 N10150424 SH0729 N10150514 SH0729 N10150611 SH0729 N10150619 SH0729 N10150622 SH0729 N12150722 SH0729 N10150730 SH0729 N10150731 SH0729 N10150824 SH0729 N11150911 SH0729 N11150824 SH0729 N10150929 SH0729 N10151009 SH0729 N11150924 SH0729 N10151014 SH0729 N10150725 SH0729 N11151020 SH0729 N10151026 SH0729 N10151209 SH0729 N11151205 SH0729 N10160107 SH0729 N10160128 SH0729 N13160128 SH0729 N10160215 SH0729 N10160301 SH0729 N11160310 SH0729 N11160317 SH0729 N10160322 SH0729 N10160426 SH0729 N11160506 SH0729 N10160607 SH0729 N10160615 SH0729 N10160620 SH0729 N11160624 SH0729 N12160815 SH0729 N11160811 SH0729 N13160823 SH0729 N10161005 SH0729 N12161014 SH0729 N10161122 SH0729 N10170111 SH0729 N11170324 SH0729 N12170329 SH0729 N11170417 SH0729 N10170505 SH0729 N11170512 SH0729 N10170707 SH0729 N10170712 SH0729 N10170803 SH0729 N10170926 SH0729 N10171023 SH0729 N10171207 SH0729 N10171206 SH0729 N10171228 SH0729 N10180102 SH0729 N10180111 SH0729 N10180122 SH0729 N10180618 SH0729 N10150330 SH0035 N13150427 SH0035 N11150722 SH0035 N11150904 SH0035 N12151204 SH0035 N12151208 SH0035 N10160412 SH0035 N11160513 SH0035 N10160803 SH0035 N10161010 SH0035 N10161013 SH0035 N10170203 SH0035 N11170405 SH0035 N10170610 SH0035 N10170803 SH0035 N10171109 SH0035 N10180111 SH0035 N10180301 SH0035 N10150901 SH1391 N13150304 SH2125 N11150818 SH2125 N10160706 SH2125 N11160815 SH2125 N11161012 SH2125 N11161208 SH2125 N10170620 SH2125 N10170901 SH2125 N10171109 SH2125 N10180103 SH2125 N10180221 SH2125 N10180122 SH2125 N10180418 SH2125 N10150417 SH1622 N10150804 SH1622 N11160105 SH1622 N10160226 SH1622 N11160923 SH1622 N10170316 SH1622 N10171002 SH1622 N10171027 SH1622 N10180129 SH1622 N10160322 SH1076 N10160803 SH1076 N11161013 SH1076 N10170106 SH1076 N11170607 SH1076 N10171025 SH1076 N10171109 SH1076 N10171204 SH1076 N10180207 SH1076 N11180220 SH1076 N12151204 SH1604 N11160614 SH1604 N10161004 SH1604 N10170202 SH1604 N10170322 SH1604 N10170408 SH1604 N12170606 SH1604 N10171030 SH1604 N10171219 SH1604 N10180221 SH1604 N10180305 SH1604 N10180424 SH1604 N10180511 SH1604 N10180418 SH1604 N11150401 SH0495 N10150413 SH0495 N10150506 SH0495 N11150506 SH0495 N10150611 SH0495 N11150609 SH0495 N10150623 SH0495 N11150706 SH0495 N10150804 SH0495 N10150813 SH0495 N10150929 SH0495 N10151118 SH0495 N11151208 SH0495 N11160823 SH0495 N10161114 SH0495 N10170209 SH0495 N11170317 SH0495 N11170526 SH0495 N11170606 SH0495 N11170706 SH0495 N10170714 SH0495 N10180413 SH0495 N10150304 SH1015 N12150311 SH1015 N11150416 SH1015 N10150506 SH1015 N11150509 SH1015 N10150528 SH1015 N11150814 SH1015 N10150921 SH1015 N10151012 SH1015 N10151109 SH1015 N10151110 SH1015 N11151103 SH1015 N10151106 SH1015 N10160106 SH1015 N10160118 SH1015 N12160128 SH1015 N10160224 SH1015 N10160307 SH1015 N12160305 SH1015 N10160420 SH1015 N10160504 SH1015 N10160630 SH1015 N11160708 SH1015 N12160712 SH1015 N10160801 SH1015 N10160802 SH1015 N10160816 SH1015 N11160815 SH1015 N11160831 SH1015 N10161017 SH1015 N10161110 SH1015 N10161104 SH1015 N10161107 SH1015 N10161116 SH1015 N10161219 SH1015 N11170126 SH1015 N11170309 SH1015 N10170322 SH1015 N11170425 SH1015 N11170427 SH1015 N11170517 SH1015 N11170526 SH1015 N10170607 SH1015 N10170724 SH1015 N10170816 SH1015 N10170824 SH1015 N10171102 SH1015 N10171128 SH1015 N10180424 SH1015 N10180227 SH1015 N10180316 SH1015 N10180328 SH1015 N10150521 SH1385 N11150609 SH1385 N11151013 SH1385 N11151207 SH1385 N11151210 SH1385 N11160419 SH1385 N10160706 SH1385 N11160906 SH1385 N14160913 SH1385 N11161007 SH1385 N11161124 SH1385 N11170111 SH1385 N11170606 SH1385 N10170704 SH1385 N10170710 SH1385 N11170609 SH1385 N10171206 SH1385 N11171220 SH1385 N10180206 SH1385 N10180413 SH1385 N10180405 SH1385 N10180412 SH1385 N11180216 SH1385 N10150311 SH1279 N10150313 SH1279 N11150429 SH1279 N11150507 SH1279 N10150519 SH1279 N10150528 SH1279 N11150529 SH1279 N11150626 SH1279 N11150722 SH1279 N10150722 SH1279 N10150805 SH1279 N10150901 SH1279 N10150902 SH1279 N11150820 SH1279 N10150915 SH1279 N10151015 SH1279 N10151016 SH1279 N10151117 SH1279 N12151118 SH1279 N10160106 SH1279 N12160116 SH1279 N12160130 SH1279 N13160219 SH1279 N10160308 SH1279 N10160412 SH1279 N10160420 SH1279 N11160418 SH1279 N10160516 SH1279 N10160523 SH1279 N10160627 SH1279 N11160712 SH1279 N11160802 SH1279 N10160901 SH1279 N12160831 SH1279 N11161003 SH1279 N11161007 SH1279 N11160719 SH1279 N11161130 SH1279 N10170103 SH1279 N10170201 SH1279 N10170324 SH1279 N11170330 SH1279 N10170425 SH1279 N10170526 SH1279 N10170608 SH1279 N11170607 SH1279 N11170620 SH1279 N10170717 SH1279 N10170724 SH1279 N10170815 SH1279 N10170929 SH1279 N10171030 SH1279 N10171102 SH1279 N10180103 SH1279 N10180119 SH1279 N10180131 SH1279 N10180202 SH1279 N10180215 SH1279 N10180326 SH1279 N11180413 SH1279 N10180328 SH1279 N10180220 SH1279 N11150306 SH2144 N10150407 SH2144 N10150413 SH2144 N10150521 SH2144 N10150602 SH2144 N11150708 SH2144 N11150722 SH2144 N10150911 SH2144 N11150724 SH2144 N11151012 SH2144 N10151202 SH2144 N10151216 SH2144 N13160128 SH2144 N10160206 SH2144 N10160222 SH2144 N10160303 SH2144 N10160322 SH2144 N10160504 SH2144 N11160509 SH2144 N10160714 SH2144 N10160718 SH2144 N13160912 SH2144 N11160928 SH2144 N12160930 SH2144 N10161107 SH2144 N11161125 SH2144 N10170103 SH2144 N10170202 SH2144 N11170223 SH2144 N11170403 SH2144 N10170421 SH2144 N10170505 SH2144 N10170515 SH2144 N10170626 SH2144 N10170824 SH2144 N10170926 SH2144 N10171115 SH2144 N10171221 SH2144 N10180119 SH2144 N10180221 SH2144 N11180125 SH2144 N11180216 SH2144 N10180220 SH2144 N10180706 SH2144 N13150304 SH0761 N10150413 SH0761 N11150509 SH0761 N11150521 SH0761 N10150617 SH0761 N10150619 SH0761 N11150622 SH0761 N11150601 SH0761 N12150730 SH0761 N11150814 SH0761 N10150917 SH0761 N10151012 SH0761 N10151013 SH0761 N11151012 SH0761 N10151104 SH0761 N11151027 SH0761 N12151117 SH0761 N10151221 SH0761 N10160225 SH0761 N10160307 SH0761 N11160321 SH0761 N11160330 SH0761 N11160418 SH0761 N10160609 SH0761 N11160620 SH0761 N10160705 SH0761 N11160823 SH0761 N12160831 SH0761 N13160912 SH0761 N11160922 SH0761 N10161027 SH0761 N10161110 SH0761 N11161110 SH0761 N11160719 SH0761 N10170202 SH0761 N11170126 SH0761 N10170213 SH0761 N10170210 SH0761 N11170304 SH0761 N11170309 SH0761 N10170324 SH0761 N10170405 SH0761 N10170424 SH0761 N10170504 SH0761 N10170606 SH0761 N10170706 SH0761 N10170704 SH0761 N10170712 SH0761 N10170913 SH0761 N10170929 SH0761 N10171025 SH0761 N10171113 SH0761 N10171205 SH0761 N11171221 SH0761 N10180112 SH0761 N10180130 SH0761 N10180206 SH0761 N10180226 SH0761 N10180223 SH0761 N10180412 SH0761 N10180511 SH0761 N10180418 SH0761 N10180509 SH0761 N10180702 SH0761 N10150304 SH1276 N11150312 SH1276 N10151215 SH1276 N10150318 USG518 N13150427 USG518 N11160105 USG518 N11160312 USG518 N10170105 USG518 N11170103 RSH0152 N11170424 RSH0152 N11170706 RSH0152 N10171214 RSH0152 N10180122 RSH0152 Material Numbers: 420179-18 Batch / System Numbers: N12180517 SH1182 N11180518 SH1182 N11180518 SH1570 N10180516 SH0059 N10180521 SH0059 N10180409 SH0059 N10180411 SH1557 N11180605 SH1557 N10180613 SH1557 N10180615 SH1557 N10180409 SH1557 N10171213 SH1557 N10180604 SH1964 N10180615 SH1964 N10180409 SH1964 N10180416 SH1964 N11180518 SH1822 N11180605 SH1390 N12180517 USG689 N10180507 USG689 N11180524 USG294 N11180228 USG721 N11180517 USG721 N12180517 USG721 N10180507 USG721 N11180323 USG022 N11180517 USG022 N10180409 USG022 N10180529 SH2186 N11180110 SH0825 N11180525 SH0825 N11180228 SH1378 N10180430 SH1378 N11180517 SH1378 N11180524 SH1378 N10180416 SH0935 N10180523 SH0729 N10180521 SH0729 N10180525 SH0729 N10180521 SH2125 N10180619 SH2125 N11180518 SH1622 N10180516 SH1076 N10180615 SH1076 N10180605 SH0495 N10180430 SH1015 N10180409 SH1015 N10180416 SH1015 N10180507 SH1015 N11180613 SH1385 N10171123 SH1385 N12180517 SH1279 N10180521 SH1279 N10180613 SH1279 N11180524 SH1279 N11180601 SH2144 N11180613 SH2144 N11180517 SH0761 N10180619 SH0761 Material Number: 420179-22 Batch / System Numbers: N10180926 SH1182 N10181205 SH1182 N11181102 SH1182 N10190110 SH1182 N10190131 SH1182 N10190617 SH1182 N10190415 SH1182 N14190701 SH1182 N11190812 SH1182 N14190729 SH1182 N12200302 SH1182 N10200803 SH1182 N11200810 SH1182 N11181009 SG704 N11181019 SG704 N10181024 SG704 N10181129 SG704 N10181203 SG704 N11180927 SG704 N10190603 SH0059 N11200810 SH1997 N10180920 SH1557 N10181005 SH1557 N10181015 SH1557 N10181101 SH1557 N11181210 SH1557 N11181023 SH1557 N11181030 SH1557 N11181122 SH1557 N10190102 SH1557 N10190109 SH1557 N10190124 SH1557 N10190205 SH1557 N10190315 SH1557 N10190415 SH1557 N11190708 SH1557 N12190715 SH1557 N11190722 SH1557 N12190729 SH1557 N10190729 SH1557 N10190805 SH1557 N10190812 SH1557 N11191028 SH1557 N10191209 SH1557 N11200127 SH1557 N10200221 SH1557 N10200317 SH1557 N10200406 SH1557 N10180926 SH1557 N10200218 SH1557 N10200803 SH1557 N11200817 SH1557 N10200525 SH1557 N11180921 SH1964 N11180924 SH1964 N11181010 SH1964 N10181113 SH1964 N10181102 SH1964 N11181029 SH1964 N11190114 SH1964 N10190201 SH1964 N10190306 SH1964 N12190603 SH1964 N11190701 SH1964 N10190708 SH1964 N11190708 SH1964 N10190722 SH1964 N12190722 SH1964 N10190808 SH1964 N10190902 SH1964 N13190805 SH1964 N13200127 SH1964 N10200221 SH1964 N12200127 SH1964 N10200406 SH1964 N11200810 SH1964 N11200309 SH1964 N10190206 SH1837 N11190715 SH1837 N10200218 SH1837 N10200622 SH1837 N10200525 SH1837 N10180920 SH1822 N10190121 SH1822 N11190715 SH1822 N11191028 SH1822 N11200810 SH1822 N10180918 SH1390 N11180919 SH1390 N10181217 SH1390 N10181102 SH1390 N10180927 SH1390 N11181030 SH1390 N14190701 SH1390 N10190722 SH1390 N13190805 SH1390 N11200323 SH1390 N10200309 SH1390 N11200309 SH1390 N11200817 SH1390 N10181122 USG689 N10181023 USG689 N11181030 USG689 N10190201 USG689 N11190812 USG689 N10181204 SH1368 N11191104 SH1368 N12200302 SH1368 N10180921 USG294 N11181024 USG294 N10181211 USG294 N10190610 USG294 N11190722 USG294 N10200323 USG294 N10200803 USG294 N11180920 USG721 N11180926 USG721 N10181003 USG721 N10181024 USG721 N11181211 USG721 N10190109 USG721 N10190111 USG721 N11190114 USG721 N11190327 USG721 N12190603 USG721 N10190415 USG721 N11190701 USG721 N12190708 USG721 N14190729 USG721 N10191007 USG721 N11191104 USG721 N11200127 USG721 N10200317 USG721 N11200323 USG721 N11200309 USG721 N10180919 USG022 N10181212 USG022 N10181217 USG022 N11181102 USG022 N10190104 USG022 N10190109 USG022 N10190603 USG022 N10190701 USG022 N11190617 USG022 N10190805 USG022 N10200217 USG022 N10200317 USG022 N14200127 USG022 N11200309 USG022 N10180926 SH2186 N10181017 SH2186 N10180925 SH2186 N10190225 SH2186 N10190315 SH2186 N11190327 SH2186 N10190722 SH2186 N11190805 SH2186 N11200817 SH2186 N10190201 SH0825 N11190715 SH0825 N12181023 SH1378 N10180928 SH1378 N10181102 SH1378 N11181023 SH1378 N10181207 SH1378 N11181022 SH1378 N11181101 SH1378 N10190107 SH1378 N10190111 SH1378 N10190225 SH1378 N10190306 SH1378 N11190327 SH1378 N10190617 SH1378 N10190708 SH1378 N11190709 SH1378 N12190722 SH1378 N13190805 SH1378 N11191104 SH1378 N11180918 SH0935 N11180921 SH0935 N11180925 SH0935 N10181122 SH0935 N10181114 SH0935 N10190104 SH0935 N10190124 SH0935 N10190131 SH0935 N10190214 SH2180 N12190603 SH2180 N10200221 SH2180 N12200127 SH2180 N10181005 SH0729 N10181203 SH0729 N11181101 SH0729 N10190104 SH0729 N10190315 SH0729 N12190708 SH0729 N11190715 SH0729 N11181101 SH1391 N10181211 SH1391 N11180921 SH2125 N10181218 SH2125 N10190124 SH2125 N10190312 SH2125 N10190603 SH2125 N12190708 SH2125 N10181101 SH1622 N10190114 SH1622 N10180919 SH1076 N12181023 SH1076 N11181210 SH1076 N11181029 SH1076 N10190124 SH1076 N10190701 SH1076 N12190722 SH1076 N11200810 SH1076 N11200309 SH1076 N11181008 SH1604 N10181205 SH1604 N13190708 SH1604 N11191028 SH1604 N10191209 SH1604 N12200302 SH1604 N10200317 SH1604 N10200406 SH1604 N10200601 SH1604 N12181011 SH0495 N11181010 SH0495 N10181204 SH0495 N10181212 SH0495 N10190114 SH0495 N10190314 SH0495 N11190327 SH0495 N13190708 SH0495 N11190805 SH0495 N12191014 SH0495 N10200803 SH0495 N10181113 SH1015 N11181114 SH1015 N12181023 SH1015 N10181012 SH1015 N11181101 SH1015 N10190124 SH1015 N11190124 SH1015 N10190306 SH1015 N10190315 SH1015 N10190408 SH1015 N11190722 SH1015 N11190812 SH1015 N14200127 SH1015 N10200221 SH1015 N11200323 SH1015 N10200406 SH1015 N11180925 SH1385 N11181009 SH1385 N10181015 SH1385 N10181113 SH1385 N11181114 SH1385 N10181217 SH1385 N11181011 SH1385 N10181221 SH1385 N10181218 SH1385 N10190115 SH1385 N10190205 SH1385 N11190617 SH1385 N14190701 SH1385 N11190708 SH1385 N12190708 SH1385 N14190708 SH1385 N10190729 SH1385 N13190805 SH1385 N10191007 SH1385 N10191021 SH1385 N11191104 SH1385 N11180919 SH1279 N10180921 SH1279 N10181015 SH1279 N10181203 SH1279 N10181121 SH1279 N10181012 SH1279 N10181123 SH1279 N10190102 SH1279 N11181029 SH1279 N10190312 SH1279 N10190610 SH1279 N11190701 SH1279 N14190708 SH1279 N11190709 SH1279 N10190730 SH1279 N10191021 SH1279 N10200221 SH1279 N11200217 SH1279 N12200302 SH1279 N10200803 SH1279 N10200810 SH1279 N11180920 SH2144 N10180926 SH2144 N11181024 SH2144 N11180927 SH2144 N10181102 SH2144 N11181030 SH2144 N10190115 SH2144 N10190227 SH2144 N12190603 SH2144 N12190701 SH2144 N11190715 SH2144 N10200323 SH2144 N10180919 SH0761 N11181010 SH0761 N10181017 SH0761 N10181101 SH0761 N10181114 SH0761 N10181102 SH0761 N10180927 SH0761 N10181023 SH0761 N11181031 SH0761 N11181122 SH0761 N10190116 SH0761 N10190513 SH0761 N14190708 SH0761 N10190709 SH0761 N10190805 SH0761 N11191104 SH0761 N10191209 SH0761 N11200217 SH0761 N10200317 SH0761 N10200218 SH0761 N10200601 SH0761 N10200810 SH0761 N10200622 SH0761 N10181217 USG518 N11180920 RSH0152 N11191104 RSH0152 Material Number: 420179-23 Batch / System Numbers: N13190522 SH1182 N10191209 SH1182 N10200831 SH1182 N11201026 SH1182 N10201130 SH1182 N11201207 SH1182 N11210316 SH1182 N10210405 SH1182 N11191007 SH1997 N11191014 SH1997 N11191209 SH1997 N11201026 SH1997 N11201130 SH1997 N10210830 SH1997 N10210913 SH1997 K11220828 SH1997 K11231123 SH1997 K10240104 SH1997 K10240125 SH1997 N11191007 SH1557 N10191007 SH1557 N11191111 SH1557 N10200107 SH1557 N13200120 SH1557 N10200831 SH1557 N11201026 SH1557 N10201117 SH1557 N12210202 SH1557 N10210222 SH1557 N12210316 SH1557 N10210405 SH1557 N11210419 SH1557 N10210503 SH1557 N10210614 SH1557 N10210726 SH1557 N10210628 SH1557 N11210927 SH1557 N10211011 SH1557 K10220104 SH1557 N10191014 SH1964 N11200107 SH1964 N10200420 SH1964 N12200420 SH1964 N11200727 SH1964 N11200921 SH1964 N10201123 SH1964 N10201214 SH1964 N12201214 SH1964 N12210202 SH1964 N12210316 SH1964 N11210316 SH1964 N10210405 SH1964 N11210419 SH1964 N10210510 SH1964 N11210503 SH1964 N10210503 SH1964 N10210614 SH1964 N10210726 SH1964 N10210628 SH1964 N10210830 SH1964 N10210913 SH1964 N10211011 SH1964 N11210927 SH1964 K10220208 SH1964 K11220425 SH1964 K10220605 SH1964 K10220807 SH1964 K10220918 SH1964 N10191209 SH1837 N10201123 SH1837 K10220403 SH1837 K10240208 SH1837 N10200114 SH1822 N10201214 SH1822 N10210215 SH1822 N10210222 SH1822 N11190610 SH1390 N10190827 SH1390 N10200420 SH1390 N11200727 SH1390 N11201207 SH1390 N12210316 SH1390 N11210503 SH1390 N10210614 SH1390 N10210726 SH1390 N11210927 SH1390 N10210920 SH1390 K10220131 SH1390 N10211206 SH1390 N11210913 SH1390 K11220522 SH1390 K10220605 SH1390 K10220807 SH1390 K11220828 SH1390 K10230209 SH1390 N10190909 USG689 N10191014 USG689 N13200210 USG689 K10220605 USG689 K10220807 USG689 N11191028 SH1368 N10200824 SH1368 K10240104 SH1368 K10240125 SH1368 N11191007 USG294 N11200120 USG294 N11200210 USG294 N10200420 USG294 N11200921 USG294 N10201026 USG294 N10201123 USG294 N11201207 USG294 N10210112 USG294 N10210118 USG294 N10210222 USG294 N10210405 USG294 N10210412 USG294 N11210503 USG294 N10210802 USG294 N10210830 USG294 K10220131 USG294 K10220208 USG294 K10220403 USG294 K11220425 USG294 K11220522 USG294 K10220605 USG294 K10220918 USG294 K11220828 USG294 K10230209 USG294 K10240104 USG294 K10240125 USG294 N10190826 USG721 N10191007 USG721 N10190827 USG721 N10190917 USG721 N11200107 USG721 N10200120 USG721 N11200120 USG721 N10200703 USG721 N10200330 USG721 N10200831 USG721 N12200831 USG721 N10201130 USG721 N11210621 USG721 N10191014 USG022 N10200120 USG022 N10200420 USG022 N12200420 USG022 N11200921 USG022 N11201207 USG022 N10210111 USG022 N10210118 USG022 N10210215 USG022 N10210316 USG022 N10210412 USG022 N11210621 USG022 N10210405 USG022 N10210726 USG022 N11210823 USG022 N11210927 USG022 N11210913 USG022 N10210712 USG022 K10220104 USG022 K10220208 USG022 K11220425 USG022 K11220522 USG022 K10220605 USG022 K10220807 USG022 K10220918 USG022 N11210316 USG022 K11220828 USG022 K10230209 USG022 N10190826 SH2186 N11191209 SH2186 N12200420 SH2186 N10200914 SH2186 N10201026 SH2186 N10200420 SH2186 N12201214 SH2186 N10210118 SH2186 N11210419 SH2186 N10210503 SH2186 N12210524 SH2186 N10210726 SH2186 N10210913 SH2186 N10191209 SH0825 N11201026 SH0825 N10200810 SH0825 N10210726 SH0825 N10191104 SH1378 N11200113 SH1378 N11191209 SH1378 N13200120 SH1378 N10200703 SH1378 N10200330 SH1378 N11201214 SH1378 N10210112 SH1378 N10210118 SH1378 N12210202 SH1378 N12200831 SH1378 N13210222 SH1378 N11210316 SH1378 N10210405 SH1378 N11210412 SH1378 N10210503 SH1378 N10210830 SH1378 N10200420 SH0729 N11200727 SH0729 N11200921 SH0729 N11201214 SH0729 N11210215 SH0729 N13200120 SH0035 N10191209 SH1391 N10201019 SH1391 N11200921 SH1391 N10190827 SH2125 N13190522 SH2125 N10190909 SH1622 N11191014 SH1076 N10191209 SH1076 N10210112 SH1076 N10190909 SH1076 N10200914 SH1076 N11210412 SH1076 K10220918 SH1076 N10190909 SH1604 N10200703 SH1604 N10200907 SH1604 N11200120 SH0495 N10200810 SH0495 N11191104 SH1015 N11200107 SH1015 N13200120 SH1015 N10200120 SH1015 N13200210 SH1015 N11200713 SH1015 N10200831 SH1015 N11201026 SH1015 K10220124 SH1015 N10190909 SH1385 N14190522 SH1385 N10190827 SH1385 N11200107 SH1385 N10200204 SH1385 N10190827 SH1279 N10200114 SH1279 N11200120 SH1279 N11200210 SH1279 N12200420 SH1279 N10200831 SH1279 N10200420 SH1279 N10210111 SH1279 N11210111 SH1279 N10210614 SH1279 N10210726 SH1279 N10210802 SH1279 N10210913 SH1279 N10211011 SH1279 N10211206 SH1279 K10220104 SH1279 K10220208 SH1279 K10220403 SH1279 K11220425 SH1279 K11220522 SH1279 K10220605 SH1279 K10220807 SH1279 N10200107 SH2144 N11200107 SH2144 N10200713 SH2144 N10210111 SH2144 N10210118 SH2144 N10210412 SH2144 N10210628 SH2144 N10210802 SH2144 N10210830 SH2144 K11220425 SH2144 K10230209 SH2144 K11231123 SH2144 N11190819 SH0761 N11191028 SH0761 N11200107 SH0761 N12200120 SH0761 N10200114 SH0761 N11200210 SH0761 N11201214 SH0761 N12210202 SH0761 N10201130 SH0761 N12200831 SH0761 N10210405 SH0761 N10210503 SH0761 N10210614 SH0761 N10210726 SH0761 N10210830 SH0761 N10210628 SH0761 N10210913 SH0761 N11210913 SH0761 N10211206 SH0761 K10220208 SH0761 K11220425 SH0761 N10210222 SH0761 K11220522 SH0761 K10220605 SH0761 K10220807 SH0761 K11220828 SH0761 K10230209 SH0761 N10211011 SH0761 N12200420 RSH0152 N10210510 RSH0152 K10230209 RSH0152 K10220208 SH1513 K10220605 SH1513 K10220807 SH1513 K10220918 SH1513 K10230209 SH1513 K10240125 SH1513

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1696-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedDecember 9, 2025
ReportedApril 8, 2026
CompanyIntuitive Surgical, Inc.
Company locationSunnyvale, CA, United States
Initiated byVoluntary: Firm initiated
Quantity86904