COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number... recall (Cook Incorporated)

Medical deviceClass IIOngoing

Summary: Cook Incorporated is recalling COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number G44806; Reference P. Reason: Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

What is being recalled

COOK MEDICAL Spectrum¿ Central Venous Tray: Reference Part Number C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, Order Number G44800; Reference Part Number C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, Order Number G44809; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, Order Number G44806; Reference Part Number C-UTLMYJ-701J-RSC-ABRM-HC-IHI-FST-A-RD, Order Number G44807.

Why

Products from the affected device lots were labeled with expiration dates that exceed the true shelf life.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448001(17)280512(10)NS16597252, Lot Number NS16597252; C-UQLMYJ-1001J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448001(17)280513(10)NS16600633, Lot Number NS16600633; C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, UDI (01)00827002448094(17)280327(10)16530788, Lot Number 16530788; C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, UDI (01)00827002448094(17)280401(10)16537255, Lot Number 16537255; C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, UDI (01)00827002448094(17)280417(10)16562522, Lot Number 16562522; C-UTLMYJ-701J-ABRM-HC-IHI-FST-A-RD, UDI (01)00827002448094(17)280602(10)16625687, Lot Number 16625687; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280416(10)NS16523791X, Lot Number NS16523791X; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280514(10)NS16523837X, Lot Number NS16523837X; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280603(10)NS16523837XX, Lot Number NS16523837XX; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280324(10)NS16523838, Lot Number NS16523838; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280324(10)NS16523839, Lot Number NS16523839; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280324(10)NS16523840, Lot Number NS16523840; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588043, Lot Number NS16588043; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588056, Lot Number NS16588056; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588057, Lot Number NS16588057; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588058, Lot Number NS16588058; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588059, Lot Number NS16588059; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588060, Lot Number NS16588060; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588061, Lot Number NS16588061; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588062, Lot Number NS16588062; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280505(10)NS16588063, Lot Number NS16588063; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280507(10)NS16590737, Lot Number NS16590737; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280507(10)NS16590738, Lot Number NS16590738; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280507(10)NS16590739, Lot Number NS16590739; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596195, Lot Number NS16596195; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596196, Lot Number NS16596196; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596197, Lot Number NS16596197; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596198, Lot Number NS16596198; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596199, Lot Number NS16596199; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596200, Lot Number NS16596200; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596201, Lot Number NS16596201; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596202, Lot Number NS16596202; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16596430, Lot Number NS16596430; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597248, Lot Number NS16597248; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597249, Lot Number NS16597249; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597269, Lot Number NS16597269; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597270, Lot Number NS16597270; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597271, Lot Number NS16597271; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597272, Lot Number NS16597272; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597273, Lot Number NS16597273; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597274, Lot Number NS16597274; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597275, Lot Number NS16597275; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597276, Lot Number NS16597276; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280512(10)NS16597277, Lot Number NS16597277; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280513(10)NS16600640, Lot Number NS16600640; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280513(10)NS16600641, Lot Number NS16600641; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280513(10)NS16600642, Lot Number NS16600642; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280513(10)NS16600653, Lot Number NS16600653; C-UTLMYJ-701J-RSC-ABRM-HC-FST-A-RD, UDI (01)00827002448063(17)280513(10)NS16600654, Lot Number NS16600654; C-UTLMYJ-701J-RSC-ABRM-HC-IHI-FST-A-RD, UDI (01)00827002448070(17)280602(10)16625668, Lot Number 16625668.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1789-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 5, 2026
ReportedApril 15, 2026
CompanyCook Incorporated
Company locationBloomington, IN, United States
Initiated byVoluntary: Firm initiated
Quantity2348 units