Medline Kits: 1) CATH LAB CDS recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline Kits: 1) CATH LAB CDS, Model Number: CDS840228F; 2) IR PACK, Model Number: DYNDH2104; 3) CARDIAC CATH PACK-LF, Model Number: DYNJT2164M; 4) ANGIOGRAPHY TRAY, Model Number: DYNJT3078; 5) CATH PACK, Model Number: DYNJT4190; 6) ANGIO VASCULAR PACK, Model Number: DYNJT5717; 7) HEART CATH P. Reason: The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
What is being recalled
Medline Kits: 1) CATH LAB CDS, Model Number: CDS840228F; 2) IR PACK, Model Number: DYNDH2104; 3) CARDIAC CATH PACK-LF, Model Number: DYNJT2164M; 4) ANGIOGRAPHY TRAY, Model Number: DYNJT3078; 5) CATH PACK, Model Number: DYNJT4190; 6) ANGIO VASCULAR PACK, Model Number: DYNJT5717; 7) HEART CATH PACK, Model Number: DYNJ0381139R; 8) CARDIAC CATH PACK, Model Number: DYNJ04529L; 9) CUSTOM RADIOLOGY PACK, Model Number: DYNJ07955L; 10) STVZ CARDIAC CATH PACK, Model Number: DYNJ24619J; 11) CATH TRAY #2, Model Number: DYNJ24706D; 12) CATH LAB PACK, Model Number: DYNJ24802Q; 13) CATH PACK, Model Number: DYNJ30268K; 14) HEART CATHETERIZATION PACK, Model Number: DYNJ30927R; 15) HEART CATH TRAY PACK-LF, Model Number: DYNJ31944D; 16) CATH PACK-LF, Model Number: DYNJ33061D; 17) CATH PACK, Model Number: DYNJ34638K; 18) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ38120C; 19) CATH PACK, Model Number: DYNJ40904G; 20) CARDIAC CATH-LF, Model Number: DYNJ41057D; 21) CATH LAB PACK, Model Number: DYNJ41172; 22) ANGIOGRAPHY PACK, Model Number: DYNJ42134G; 23) CATH LAB DRAPE PACK, Model Number: DYNJ42455B; 24) ANGIOGRAPHY PACK, Model Number: DYNJ42543B; 25) CATH LAB PACK, Model Number: DYNJ42900D; 26) ANGIOGRAPHY PACK NO SYR, Model Number: DYNJ43009K; 27) CATH PACK (WITH MANIFOLD), Model Number: DYNJ44515L; 28) ANGIO IR DRAPE PACK 319696, Model Number: DYNJ44593I; 29) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ44850J; 30) CATH LAB PACK, Model Number: DYNJ45296; 31) ANGIO PACK, Model Number: DYNJ50739J; 32) CARDIAC CATH PACK, Model Number: DYNJ51103F; 33) IR ANGIO BODY, Model Number: DYNJ51765D; 34) INTERVENTIONAL TRAY, Model Number: DYNJ52501C; 35) CARDIAC CATH IMF 86040- LF, Model Number: DYNJ53947J; 36) CARDIAC CATH PACK, Model Number: DYNJ54101I; 37) ANGIO KIT, Model Number: DYNJ55686C; 38) CATH LAB PACK, Model Number: DYNJ56141B; 39) ANGIO NEURO PACK, Model Number: DYNJ57685D; 40) CATH PACK, Model Number: DYNJ57729F; 41) CATH PACK, Model Number: DYNJ59632F; 42) CATH LAB PACK WITH NAMIC, Model Number: DYNJ60853D; 43) ANGIO PACK ALEX, Model Number: DYNJ63539C; 44) ANGIOGRAPHY CV RAD PACK, Model Number: DYNJ64936B; 45) CATH LAB PACK, Model Number: DYNJ65185D; 46) CATH PACK, Model Number: DYNJ65619A; 47) CATH I W SPLIT DRAPE PACK, Model Number: DYNJ66182; 48) CATH PACK SHD, Model Number: DYNJ68384B; 49) CATH LAB PACK, Model Number: DYNJ69094B; 50) CATH LAB ACCESSORIES, Model Number: DYNJ69213B; 51) REYES CATH PACK, Model Number: DYNJ69345A; 52) TMH ANGIOGRAPHY TRAY, Model Number: DYNJ80054A; 53) ANGIOGRAPHY TRAY, Model Number: DYNJ80156; 54) CATH PACK, Model Number: DYNJ80516C; 55) CATH PACK, Model Number: DYNJ80730; 56) ANGIO DRAPE PACK 2 WINDOW, Model Number: DYNJ81066B; 57) OR ANGIOGRAM PACK, Model Number: DYNJ81435F; 58) HEART CATH PACK, Model Number: DYNJ81527B; 59) HEART CATH PACK, Model Number: DYNJ83764D; 60) CATH LAB PACK, Model Number: DYNJ84048A; 61) ANGIO TRAY NO LIDO, Model Number: DYNJ85557B; 62) CARDIAC CATH PACK, Model Number: DYNJ85746B; 63) ANGIOGRAPHY PACK, Model Number: DYNJ86313; 64) ANGIOGRAPHY DRAPE PACK, Model Number: DYNJ86787; 65) CATH PACK, Model Number: DYNJ87098A; 66) CATH LAB TAVR PACK, Model Number: DYNJ88532; 67) SMT LEFT HEART CATH, Model Number: DYNJ89274; 68) VASCULAR CATH PACK, Model Number: DYNJ89352; 69) CATH LAB PACK, Model Number: DYNJ89805; 70) CARDIAC CATH MERCY WASHINGTON, Model Number: DYNJ89934; 71) ANGIO, Model Number: DYNJ900278G; 72) TAVR ICC VER. F, Model Number: DYNJ904401F; 73) TAVR, Model Number: DYNJ908646C; 74) LHK-UPMC CHILDRENS MAIN OR 2, Model Number: NAM0009; 75) KIT ANGIO CSTM-EVERGREEN, Model Number: NAM0020; 76) GP-CARDIAC CATH PACK-LF, Model Number: PHS397170009D; 77) LEFT HEART KIT - MARION, Model Number: VASC1073A; 78) RHK - VAMC RICHMOND, Model Number: VASC1185B; 79) LHK-MEADOWVIEW REG, Model Number: VASC1263; 80) LHK - BSC 3V ON S10RA, Model Number: VASC1269; 81) LHK - BSC 3V ON S10MLL, Model Number: VASC1270; 82
Why
The 510(k) regulatory clearances for certain Control Syringes, Guidewires, and/or High-Pressure Tubing have been rescinded.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
US Nationwide distribution.
How to tell if yours is affected
1) CDS840228F, UDI-DI: 10889942778546(each), 40889942778547(case), Lot Number: 25ALA420; 2) CDS840228F, UDI-DI: 10889942778546(each), 40889942778547(case), Lot Number: 25CLA774; 3) CDS840228F, UDI-DI: 10889942778546(each), 40889942778547(case), Lot Number: 24LLA038; 4) DYNDH2104, UDI-DI: 10198459273964(each), 40198459273965(case), Lot Number: 25DBR609; 5) DYNJT2164M, UDI-DI: 10198459003165(each), 40198459003166(case), Lot Number: 24JBT885; 6) DYNJT3078, UDI-DI: 10198459305146(each), 40198459305147(case), Lot Number: 25DBN722; 7) DYNJT4190, UDI-DI: 10198459428234(each), 40198459428235(case), Lot Number: 25HBA387; 8) DYNJT5717, UDI-DI: 10198459496684(each), 40198459496685(case), Lot Number: 26ALA778; 9) DYNJ0381139R, UDI-DI: 10198459276019(each), 40198459276010(case), Lot Number: 25JLA029; 10) DYNJ04529L, UDI-DI: 10198459391460(each), 40198459391461(case), Lot Number: 25JLA312; 11) DYNJ07955L, UDI-DI: 10889942628896(each), 40889942628897(case), Lot Number: 24HBM316; 12) DYNJ07955L, UDI-DI: 10889942628896(each), 40889942628897(case), Lot Number: 23HBX895; 13) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 23DBP876; 14) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 23FBG933; 15) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 24GBE563; 16) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 24HBT787; 17) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 24KBC564; 18) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 24LBL445; 19) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 25DBF355; 20) DYNJ24619J, UDI-DI: 10195327295639(each), 40195327295630(case), Lot Number: 25EBG106; 21) DYNJ24706D, UDI-DI: 10889942297122(each), 40889942297123(case), Lot Number: 25BBB510; 22) DYNJ24802Q, UDI-DI: 10198459091476(each), 40198459091477(case), Lot Number: 25DBJ212; 23) DYNJ30268K, UDI-DI: 10198459132636(each), 40198459132637(case), Lot Number: 25EMH059; 24) DYNJ30268K, UDI-DI: 10198459132636(each), 40198459132637(case), Lot Number: 25FMA010; 25) DYNJ30268K, UDI-DI: 10198459132636(each), 40198459132637(case), Lot Number: 25GME803; 26) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 24JME497; 27) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25AME060; 28) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25CMA483; 29) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25CMI507; 30) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25DMB230; 31) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25DMF455; 32) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25EMB900; 33) DYNJ30927R, UDI-DI: 10195327539160(each), 40195327539161(case), Lot Number: 25EMI646; 34) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 23JBA288; 35) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 23JBB352; 36) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 23LBC983; 37) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24ABR897; 38) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24EBQ350; 39) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24FBM413; 40) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24GBA097; 41) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24HBU538; 42) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24IBM289; 43) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24JBJ568; 44) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24LBL092; 45) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 25CBF566; 46) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 25DBG717; 47) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 24ABR608; 48) DYNJ31944D, UDI-DI: 10193489440171(each), 40193489440172(case), Lot Number: 25EBA519; 49) DYNJ33061D, UDI-DI: 10193489841428(each), 40193489841429(case), Lot Number: 25BBJ181; 50) DYNJ34638K, UDI-DI: 10195327630614(each), 40195327630615(case), Lot Number: 25LLA963; 51) DYNJ38120C, UDI-DI: 10193489979398(each), 40193489979399(case), Lot Number: 23JBL270; 52) DYNJ38120C, UDI-DI: 10193489979398(each), 40193489979399(case), Lot Number: 23HBW572; 53) DYNJ40904G, UDI-DI: 10195327650117(each), 40195327650118(case), Lot Number: 25ABD202; 54) DYNJ40904G, UDI-DI: 10195327650117(each), 40195327650118(case), Lot Number: 25CBK291; 55) DYNJ40904G, UDI-DI: 10195327650117(each), 40195327650118(case), Lot Number: 25EBB772; 56) DYNJ40904G, UDI-DI: 10195327650117(each), 40195327650118(case), Lot Number: 25EBS985; 57) DYNJ41057D, UDI-DI: 10889942341764(each), 40889942341765(case), Lot Number: 24FMB990; 58) DYNJ41172, UDI-DI: 10884389926567(each), 40884389926568(case), Lot Number: 25DBJ237; 59) DYNJ42134G, UDI-DI: 10195327644581(each), 40195327644582(case), Lot Number: 24JBF696; 60) DYNJ42134G, UDI-DI: 10195327644581(each), 40195327644582(case), Lot Number: 24KBD845; 61) DYNJ42134G, UDI-DI: 10195327644581(each), 40195327644582(case), Lot Number: 24IBS009; 62) DYNJ42455B, UDI-DI: 10198459198441(each), 40198459198442(case), Lot Number: 25JBU218; 63) DYNJ42543B, UDI-DI: 10198459026157(each), 40198459026158(case), Lot Number: 25GLB081; 64) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21CBA559; 65) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21DBB663; 66) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21EME084; 67) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21GMD914; 68) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21IMC485; 69) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21JMD380; 70) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21LMA418; 71) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 21LMG019; 72) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22GMA059; 73) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22HMC848; 74) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22IMG049; 75) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22LMD909; 76) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22LMD910; 77) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22LMD911; 78) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23DMD899; 79) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23DMH723; 80) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23GMI663; 81) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23IMA328; 82) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23IMH816; 83) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23LMB744; 84) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24BMC690; 85) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24CME643; 86) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24DMF294; 87) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24DMH722; 88) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24FMD725; 89) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24GMH186; 90) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24HMB859; 91) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24JMC165; 92) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24KMA538; 93) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24KMD828; 94) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 25AMI402; 95) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 25BMC463; 96) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 25CMA551; 97) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 22DMD535; 98) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 23FMB301; 99) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 24GMA164; 100) DYNJ42900D, UDI-DI: 10193489203028(each), 40193489203029(case), Lot Number: 25CME925; 101) DYNJ43009K, UDI-DI: 10198459243974(each), 40198459243975(case), Lot Number: 25BMH143; 102) DYNJ43009K, UDI-DI: 10198459243974(each), 40198459243975(case), Lot Number: 25DMH529; 103) DYNJ43009K, UDI-DI: 10198459243974(each), 40198459243975(case), Lot Number: 25DMK344; 104) DYNJ43009K, UDI-DI: 10198459243974(each), 40198459243975(case), Lot Number: 25GMG243; 105) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 24JBB567; 106) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25CBO965; 107) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25CBT443; 108) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25DBL293; 109) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25EBK886; 110) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25GBL828; 111) DYNJ44515L, UDI-DI: 10198459054518(each), 40198459054519(case), Lot Number: 25KMF800; 112) DYNJ44593I, UDI-DI: 10198459280221(each), 40198459280222(case), Lot Number: 25GBN239; 113) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23IBB504; 114) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23JBJ153; 115) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23KBP581; 116) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 25ABE986; 117) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 25BBO499; 118) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 25CBQ262; 119) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 25DBT178; 120) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 25GBK671; 121) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21ABF853; 122) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21BBR426; 123) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21DBG025; 124) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21EBA979; 125) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21GBJ604; 126) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21IBN506; 127) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 21KBI794; 128) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22ABH556; 129) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22BBE828; 130) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22EBL005; 131) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22GBK923; 132) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22GBK924; 133) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22IBS866; 134) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22JBW730; 135) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22KBS657; 136) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22OBI203; 137) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23BBC590; 138) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23CBS434; 139) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23FBA280; 140) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23GBE356; 141) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23HBK673; 142) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23JBT743; 143) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23JBU445; 144) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24ABT870; 145) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24CBC657; 146) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24CBS511; 147) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24EBP127; 148) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24GBQ822; 149) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24HBH343; 150) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24JBO868; 151) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24KBS706; 152) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24LBE189; 153) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 25ABP082; 154) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 25CBK764; 155) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 25DBH257; 156) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 25EBM039; 157) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 22CBP589; 158) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 23EBC276; 159) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 24FBR295; 160) DYNJ45296, UDI-DI: 10888277200784(each), 40888277200785(case), Lot Number: 25GBO282; 161) DYNJ50739J, UDI-DI: 10198459421785(each), 40198459421786(case), Lot Number: 25GBJ678; 162) DYNJ51103F, UDI-DI: 10193489985344(each), 40193489985345(case), Lot Number: 25KDA536; 163) DYNJ51765D, UDI-DI: 10198459308048(each), 40198459308049(case), Lot Number: 25GME161; 164) DYNJ52501C, UDI-DI: 10193489820461(each), 40193489820462(case), Lot Number: 25GBN389; 165) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 24ELA396; 166) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 24ELA650; 167) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 24HLB182; 168) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25ALA448; 169) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25ALA801; 170) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25BLA350; 171) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25BLA406; 172) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25DLB012; 173) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25ELA058; 174) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25FLA819; 175) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25GLA626; 176) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 24GLA955; 177) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25HLA565; 178) DYNJ53947J, UDI-DI: 10198459010545(each), 40198459010546(case), Lot Number: 25KLA585; 179) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 23HBG218; 180) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 23HBZ255; 181) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 23JBC069; 182) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 23KBK087; 183) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 23LBG529; 184) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24DBR595; 185) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24FBM938; 186) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24HBG268; 187) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24IBL499; 188) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24JBR909; 189) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24KBU615; 190) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24LBT555; 191) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 25BBT208; 192) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 25CBI708; 193) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 25EBK966; 194) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 25FBC618; 195) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 24GBS514; 196) DYNJ54101I, UDI-DI: 10195327394677(each), 40195327394678(case), Lot Number: 25GBP762; 197) DYNJ55686C, UDI-DI: 10195327566470(each), 40195327566471(case), Lot Number: 25CBK952; 198) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 24IBH768; 199) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 24JBL744; 200) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 24JBT214; 201) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 24LBL456; 202) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25BBH102; 203) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25DBJ497; 204) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25FBB113; 205) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25HBB616; 206) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25HBI901; 207) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 24HBK984; 208) DYNJ56141B, UDI-DI: 10198459069536(each), 40198459069537(case), Lot Number: 25JBW313; 209) DYNJ57685D, UDI-DI: 10198459346552(each), 40198459346553(case), Lot Number: 25GMB889; 210) DYNJ57729F, UDI-DI: 10198459395673(each), 40198459395674(case), Lot Number: 25HBB613; 211) DYNJ59632F, UDI-DI: 10198459230097(each), 40198459230098(case), Lot Number: 25LLA739; 212) DYNJ60853D, UDI-DI: 10195327023966(each), 40195327023967(case), Lot Number: 24DBR134; 213) DYNJ63539C, UDI-DI: 10198459323201(each), 40198459323202(case), Lot Number: 25JBS069; 214) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 22HBY850; 215) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 22LBQ523; 216) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23ABK408; 217) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23BBP118; 218) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23EBI424; 219) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23FBF158; 220) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23GBM124; 221) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23HBW852; 222) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23JBA954; 223) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23KBI356; 224) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24BBO109; 225) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24CBT861; 226) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24EBG430; 227) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24GBN951; 228) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24GBV433; 229) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24HBK331; 230) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24KBI485; 231) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24LBD301; 232) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 25ABL627; 233) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 25BBJ788; 234) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 25CBG888; 235) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 25FBB380; 236) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 23DBP201; 237) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 24FBA798; 238) DYNJ64936B, UDI-DI: 10195327223205(each), 40195327223206(case), Lot Number: 25GBJ737; 239) DYNJ65185D, UDI-DI: 10198459269509(each), 40198459269500(case), Lot Number: 25CBA241; 240) DYNJ65619A, UDI-DI: 10195327427825(each), 40195327427826(case), Lot Number: 25BMH948; 241) DYNJ66182, UDI-DI: 10193489320985(each), 40193489320986(case), Lot Number: 24DMA244; 242) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24EBG456; 243) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24FBN898; 244) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24GBS573; 245) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24IBL917; 246) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24LBM711; 247) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 25CBU912; 248) DYNJ68384B, UDI-DI: 10195327341329(each), 40195327341320(case), Lot Number: 24BBU798; 249) DYNJ69094B, UDI-DI: 10195327567804(each), 40195327567805(case), Lot Number: 25HBT584; 250) DYNJ69094B, UDI-DI: 10195327567804(each), 40195327567805(case), Lot Number: 25JBK598; 251) DYNJ69094B, UDI-DI: 10195327567804(each), 40195327567805(case), Lot Number: 25HBR477; 252) DYNJ69213B, UDI-DI: 10198459001109(each), 40198459001100(case), Lot Number: 25GBJ711; 253) DYNJ69345A, UDI-DI: 10195327227753(each), 40195327227754(case), Lot Number: 24CBK985; 254) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24ABS572; 255) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24EBS233; 256) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24FBT638; 257) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24GBS648; 258) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24IBU381; 259) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24JBQ403; 260) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24KBG054; 261) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24LBJ083; 262) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 25BBK791; 263) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 25DBD994; 264) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 24CBU370; 265) DYNJ80054A, UDI-DI: 10195327516536(each), 40195327516537(case), Lot Number: 25EBG388; 266) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22ABY003; 267) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22DBN383; 268) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22EBG312; 269) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22JBT107; 270) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22LBJ502; 271) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 22OBI374; 272) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 23CBV611; 273) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 23EBG614; 274) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 23FBT410; 275) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 23JBV583; 276) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24ABU892; 277) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24ABV456; 278) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24FBG657; 279) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24GBN004; 280) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24HBD223; 281) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24IBC153; 282) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 24KBH026; 283) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 25EBG390; 284) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 23IBG062; 285) DYNJ80156, UDI-DI: 10195327012533(each), 40195327012534(case), Lot Number: 25GBJ833; 286) DYNJ80516C, UDI-DI: 10198459627002(each), 40198459627003(case), Lot Number: 25LBF566; 287) DYNJ80730, UDI-DI: 10195327076313(each), 40195327076314(case), Lot Number: 25JBW545; 288) DYNJ81066B, UDI-DI: 10198459019609(each), 40198459019600(case), Lot Number: 24IBV032; 289) DYNJ81066B, UDI-DI: 10198459019609(each), 40198459019600(case), Lot Number: 24KBL790; 290) DYNJ81066B, UDI-DI: 10198459019609(each), 40198459019600(case), Lot Number: 24HBX059; 291) DYNJ81435F, UDI-DI: 10198459434754(each), 40198459434755(case), Lot Number: 25IBP952; 292) DYNJ81527B, UDI-DI: 10198459084935(each), 40198459084936(case), Lot Number: 25CBS792; 293) DYNJ81527B, UDI-DI: 10198459084935(each), 40198459084936(case), Lot Number: 25EBR150; 294) DYNJ83764D, UDI-DI: 10198459067891(each), 40198459067892(case), Lot Number: 25IBU004; 295) DYNJ83764D, UDI-DI: 10198459067891(each), 40198459067892(case), Lot Number: 25IBA135; 296) DYNJ84048A, UDI-DI: 10195327350567(each), 40195327350568(case), Lot Number: 25EBG981; 297) DYNJ85557B, UDI-DI: 10198459140099(each), 40198459140090(case), Lot Number: 25CBH200; 298) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 24KBG717; 299) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 24LBN809; 300) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 25BBA684; 301) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 25CBS896; 302) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 25EBB234; 303) DYNJ85746B, UDI-DI: 10198459187292(each), 40198459187293(case), Lot Number: 25GBM051; 304) DYNJ86313, UDI-DI: 10195327620011(each), 40195327620012(case), Lot Number: 24JLA184; 305) DYNJ86313, UDI-DI: 10195327620011(each), 40195327620012(case), Lot Number: 24ILA638; 306) DYNJ86787, UDI-DI: 10195327675165(each), 40195327675166(case), Lot Number: 25CMD965; 307) DYNJ86787, UDI-DI: 10195327675165(each), 40195327675166(case), Lot Number: 24EMF718; 308) DYNJ87098A, UDI-DI: 10198459273599(each), 40198459273590(case), Lot Number: 25BBT130; 309) DYNJ87098A, UDI-DI: 10198459273599(each), 40198459273590(case), Lot Number: 25CBR012; 310) DYNJ87098A, UDI-DI: 10198459273599(each), 40198459273590(case), Lot Number: 25DBE588; 311) DYNJ88532, UDI-DI: 10198459172366(each), 40198459172367(case), Lot Number: 25BMG293; 312) DYNJ89274, UDI-DI: 10198459258954(each), 40198459258955(case), Lot Number: 25BBT742; 313) DYNJ89352, UDI-DI: 10198459269615(each), 40198459269616(case), Lot Number: 25GBU445; 314) DYNJ89805, UDI-DI: 10198459290756(each), 40198459290757(case), Lot Number: 25CBF895; 315) DYNJ89934, UDI-DI: 10198459305238(each), 40198459305239(case), Lot Number: 25DMH974; 316) DYNJ89934, UDI-DI: 10198459305238(each), 40198459305239(case), Lot Number: 25EMJ512; 317) DYNJ89934, UDI-DI: 10198459305238(each), 40198459305239(case), Lot Number: 25GMG106; 318) DYNJ900278G, UDI-DI: 10195327554101(each), 40195327554102(case), Lot Number: 25LLA045; 319) DYNJ904401F, UDI-DI: 10198459566684(each), 40198459566685(case), Lot Number: 25JDB197; 320) DYNJ908646C, UDI-DI: 10198459208898(each), 40198459208899(case), Lot Number: 25IBL114; 321) NAM0009, UDI-DI: 10195327270742(each), 40195327270743(case), Lot Number: 25EMA601; 322) NAM0009, UDI-DI: 10195327270742(each), 40195327270743(case), Lot Number: 24AMF168; 323) NAM0020, UDI-DI: 10195327403126(each), 40195327403127(case), Lot Number: 25BBS852; 324) PHS397170009D, UDI-DI: 10198459506642(each), 40198459506643(case), Lot Number: 25KDA544; 325) VASC1073A, UDI-DI: 10889942781881(each), 40889942781882(case), Lot Number: 25CBI989; 326) VASC1185B, UDI-DI: 10195327243395(each), 40195327243396(case), Lot Number: 25FBB593; 327) VASC1185B, UDI-DI: 10195327243395(each), 40195327243396(case), Lot Number: 25EBK520; 328) VASC1263, UDI-DI: 10195327000080(each), 40195327000081(case), Lot Number: 25GBM923; 329) VASC1269, UDI-DI: 10193489962048(each), 40193489962049(case), Lot Number: 25LBM049; 330) VASC1270, UDI-DI: 10193489962055(each), 40193489962056(case), Lot Number: 22EBP401; 331) VASC1270, UDI-DI: 10193489962055(each), 40193489962056(case), Lot Number: 22EBB968; 332) VASC1272, UDI-DI: 10193489962987(each), 40193489962988(case), Lot Number: 25DBN048; 333) VASC1273, UDI-DI: 10193489962994(each), 40193489962995(case), Lot Number: 24LBE294; 334) VASC1273, UDI-DI: 10193489962994(each), 40193489962995(case), Lot Number: 24KBU582; 335) VASC1274, UDI-DI: 10193489963007(each), 40193489963008(case), Lot Number: 24JBN014; 336) VASC1274, UDI-DI: 10193489963007(each), 40193489963008(case), Lot Number: 24KBP576; 337) VASC1274, UDI-DI: 10193489963007(each), 40193489963008(case), Lot Number: 23HBE103; 338) VASC1283, UDI-DI: 10193489979794(each), 40193489979795(case), Lot Number: 22LME362; 339) VASC1288, UDI-DI: 10193489996487(each), 40193489996488(case), Lot Number: 23KBX706; 340) VASC1295, UDI-DI: 10195327019440(each), 40195327019441(case), Lot Number: 24HBW862; 341) VASC1295, UDI-DI: 10195327019440(each), 40195327019441(case), Lot Number: 24JBE244; 342) VASC1295, UDI-DI: 10195327019440(each), 40195327019441(case), Lot Number: 24GBS118; 343) VASC1307A, UDI-DI: 10198459281211(each), 40198459281212(case), Lot Number: 25BBS813; 344) VASC1311, UDI-DI: 10195327092900(each), 40195327092901(case), Lot Number: 25CBG058; 345) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 22HBI793; 346) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 22HBJ759; 347) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 22LBB103; 348) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 23CBR430; 349) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 23FBE694; 350) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 23KBX573; 351) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 24DBH866; 352) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 24HBK999; 353) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 24JBR627; 354) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25ABS944; 355) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25BBD432; 356) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25DBC056; 357) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25DBD636; 358) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25DBQ283; 359) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 23JBI686; 360) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25DBS918; 361) VASC1321A, UDI-DI: 10195327223380(each), 40195327223381(case), Lot Number: 25EBV367; 362) VASC1325, UDI-DI: 10195327127466(each), 40195327127467(case), Lot Number: 24IBR613; 363) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 22FBG046; 364) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 22IBK994; 365) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23BBH558; 366) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23BBI789; 367) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23CBM020; 368) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23FBP237; 369) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23HBA351; 370) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23IBF804; 371) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 23KBJ680; 372) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24ABC821; 373) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24BBI096; 374) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24DBW081; 375) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24GBI591; 376) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24HBS193; 377) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24JBC754; 378) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24KBR290; 379) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24LBF931; 380) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 25ABN410; 381) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 25ABS437; 382) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 25EBS376; 383) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 22KBE304; 384) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 24CBJ019; 385) VASC1340, UDI-DI: 10195327160197(each), 40195327160198(case), Lot Number: 25GBJ908; 386) VASC1366, UDI-DI: 10195327242701(each), 40195327242702(case), Lot Number: 24LBE764; 387) VASC1395, UDI-DI: 10195327336707(each), 40195327336708(case), Lot Number: 23IBH009; 388) VASC1395, UDI-DI: 10195327336707(each), 40195327336708(case), Lot Number: 23EBG672; 389) VASC1406A, UDI-DI: 10195327563288(each), 40195327563289(case), Lot Number: 25GBJ756; 390) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 23HBT638; 391) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 24LBL698; 392) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 25ABE332; 393) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 25CBU962; 394) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 25EBA222; 395) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 25EBH282; 396) VASC1416, UDI-DI: 10195327409609(each), 40195327409600(case), Lot Number: 25EBP333; 397) VASC1424, UDI-DI: 10195327417475(each), 40195327417476(case), Lot Number: 24LBQ573; 398) VASC1433, UDI-DI: 10195327436964(each), 40195327436965(case), Lot Number: 23JBP992; 399) VASC1445, UDI-DI: 10195327469122(each), 40195327469123(case), Lot Number: 25HBV179; 400) VASC1456, UDI-DI: 10195327501228(each), 40195327501229(case), Lot Number: 24LBQ327; 401) VASC1458A, UDI-DI: 10198459030444(each), 40198459030445(case), Lot Number: 25GBN426; 402) VASC1472, UDI-DI: 10195327541347(each), 40195327541348(case), Lot Number: 25EBB614; 403) VASC1474, UDI-DI: 10195327544058(each), 40195327544059(case), Lot Number: 25DBU129; 404) VASC1474, UDI-DI: 10195327544058(each), 40195327544059(case), Lot Number: 25EBJ096; 405) VASC1474, UDI-DI: 10195327544058(each), 40195327544059(case), Lot Number: 25CBU804; 406) VASC1476, UDI-DI: 10195327555450(each), 40195327555451(case), Lot Number: 24CBL307; 407) VASC1486, UDI-DI: 10195327581916(each), 40195327581917(case), Lot Number: 25EBG340; 408) VASC1486, UDI-DI: 10195327581916(each), 40195327581917(case), Lot Number: 25EBB574; 409) VASC1489A, UDI-DI: 10198459155024(each), 40198459155025(case), Lot Number: 24JBX902; 410) VASC1489A, UDI-DI: 10198459155024(each), 40198459155025(case), Lot Number: 25ABK840; 411) VASC1489A, UDI-DI: 10198459155024(each), 40198459155025(case), Lot Number: 25BBN319; 412) VASC1505, UDI-DI: 10195327673826(each), 40195327673827(case), Lot Number: 25HBV570; 413) VASC1516A, UDI-DI: 10198459150708(each), 40198459150709(case), Lot Number: 25CBO552; 414) VASC1516A, UDI-DI: 10198459150708(each), 40198459150709(case), Lot Number: 25DBL094; 415) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 24GBL420; 416) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 24HBX102; 417) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 24IBV647; 418) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 24JBP816; 419) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 24JBS223; 420) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 25CBB663; 421) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 25DBD834; 422) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 25EBF084; 423) VASC1528, UDI-DI: 10198459021107(each), 40198459021108(case), Lot Number: 25HBO061; 424) VASC1603, UDI-DI: 10198459142260(each), 40198459142261(case), Lot Number: 24KMD745; 425) VASC1604, UDI-DI: 10198459142321(each), 40198459142322(case), Lot Number: 25DBL601; 426) VASC1608, UDI-DI: 10198459143519(each), 40198459143510(case), Lot Number: 25CBS980; 427) VASC1625, UDI-DI: 10198459161698(each), 40198459161699(case), Lot Number: 25BBT179; 428) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 24KBD713; 429) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25ABP307; 430) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25DBB874; 431) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25DBK489; 432) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25DBS122; 433) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25EBK745; 434) VASC1628, UDI-DI: 10198459167492(each), 40198459167493(case), Lot Number: 25GBJ696; 435) VASC1633, UDI-DI: 10198459180309(each), 40198459180300(case), Lot Number: 24LBC090; 436) VASC1633, UDI-DI: 10198459180309(each), 40198459180300(case), Lot Number: 25EBB507; 437) VASC1633, UDI-DI: 10198459180309(each), 40198459180300(case), Lot Number: 25EBV466; 438) VASC1633, UDI-DI: 10198459180309(each), 40198459180300(case), Lot Number: 25GBJ765; 439) VASC1655, UDI-DI: 10198459218491(each), 40198459218492(case), Lot Number: 25ABG238; 440) VASC1655, UDI-DI: 10198459218491(each), 40198459218492(case), Lot Number: 25CBA276; 441) VASC1655, UDI-DI: 10198459218491(each), 40198459218492(case), Lot Number: 25GBS583; 442) VASC1673, UDI-DI: 10198459246944(each), 40198459246945(case), Lot Number: 25BBD501; 443) VASC1673, UDI-DI: 10198459246944(each), 40198459246945(case), Lot Number: 25HBM036; 444) VASC1678, UDI-DI: 10198459241468(each), 40198459241469(case), Lot Number: 25BBJ631; 445) VASC1678, UDI-DI: 10198459241468(each), 40198459241469(case), Lot Number: 25CBJ145; 446) VASC1678, UDI-DI: 10198459241468(each), 40198459241469(case), Lot Number: 25DBI494; 447) VASC1679, UDI-DI: 10198459242151(each), 40198459242152(case), Lot Number: 25CBD675; 448) VASC1679, UDI-DI: 10198459242151(each), 40198459242152(case), Lot Number: 25CBI195; 449) VASC1679, UDI-DI: 10198459242151(each), 40198459242152(case), Lot Number: 25CBJ039; 450) VASC1679, UDI-DI: 10198459242151(each), 40198459242152(case), Lot Number: 25DBT853; 451) VASC1679, UDI-DI: 10198459242151(each), 40198459242152(case), Lot Number: 25GBS586; 452) VASC1681, UDI-DI: 10198459243837(each), 40198459243838(case), Lot Number: 25CBA812; 453) VASC1681, UDI-DI: 10198459243837(each), 40198459243838(case), Lot Number: 25CBG063; 454) VASC1681, UDI-DI: 10198459243837(each), 40198459243838(case), Lot Number: 25CBK100; 455) VASC1681, UDI-DI: 10198459243837(each), 40198459243838(case), Lot Number: 25DBF823; 456) VASC1707, UDI-DI: 10198459269349(each), 40198459269340(case), Lot Number: 25CBQ372; 457) VASC1707, UDI-DI: 10198459269349(each), 40198459269340(case), Lot Number: 25CBS015; 458) VASC1717, UDI-DI: 10198459285387(each), 40198459285388(case), Lot Number: 25EBC181; 459) VASC1747, UDI-DI: 10198459329388(each), 40198459329389(case), Lot Number: 25EMG054; 460) VASC1751, UDI-DI: 10198459337222(each), 40198459337223(case), Lot Number: 25EMH057; 461) VASC1757, UDI-DI: 10198459347832(each), 40198459347833(case), Lot Number: 25FBA376; 462) VASC1787, UDI-DI: 10198459386015(each), 40198459386016(case), Lot Number: 25HBL457; 463) VASC1802, UDI-DI: 10198459401626(each), 40198459401627(case), Lot Number: 25IBF016; 464) 60033123, UDI-DI: 10193489042436(each), 40193489042437(case), Lot Number: 25DBC934; 465) 60033123, UDI-DI: 10193489042436(each), 40193489042437(case), Lot Number: 25DBT579; 466) 6010108121, UDI-DI: 10193489104929(each), 40193489104920(case), Lot Number: 25CBC919; 467) 6010108121, UDI-DI: 10193489104929(each), 40193489104920(case), Lot Number: 24JBD975; 468) 601832314, UDI-DI: 10193489095173(each), 40193489095174(case), Lot Number: 25DBC058; 469) 601832314, UDI-DI: 10193489095173(each), 40193489095174(case), Lot Number: 25DBS920; 470) 601832314, UDI-DI: 10193489095173(each), 40193489095174(case), Lot Number: 25EBB562; 471) 601832314, UDI-DI: 10193489095173(each), 40193489095174(case), Lot Number: 25EBJ329; 472) 60232091, UDI-DI: 10193489053258(each), 40193489053259(case), Lot Number: 25GBS552
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1835-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | February 25, 2026 |
| Reported | April 29, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 160883 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from February 2026