Oxoid Agglutinating Sera recall (Remel Europe Ltd.)

Medical deviceClass IIOngoing

Summary: Remel Europe Ltd. is recalling Oxoid Agglutinating Sera, Salmonella 9-0 R30957301. Reason: Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.

What is being recalled

Oxoid Agglutinating Sera, Salmonella 9-0 R30957301

Why

Affected lots are not performing to IFU criteria showing cross reactions with 2-0 within 60 seconds. Cross-reactivity may result in the misidentification of a Salmonella 9-O serotype as a 2-O serotype.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution to the states of AL, CA, FL, IA, IL, IN, MI, MN, MO, NC, NE, OH, SD, TX, WA. (states named: Alabama, California, Florida, Iowa, Illinois, Indiana, Michigan, Minnesota, Missouri, North Carolina, Nebraska, Ohio, South Dakota, Texas, Washington)

How to tell if yours is affected

UDI-DI 05056080500270 lots 6172600, 6172741, 6171393, 6150168, 6107767, 3766399, 3766400, 3780497, 3719592, 3709358, 6266130, 6253410, 6260215, 6222350, 6233859, 6233858

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1923-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 25, 2026
ReportedApril 29, 2026
CompanyRemel Europe Ltd.
Company locationDartford, N/A, United Kingdom
Initiated byVoluntary: Firm initiated
Quantity117 US; 950 OUS

More recalls from Remel Europe Ltd. · all medical device recalls · recalls from March 2026

Official FDA record