Philips Ingenia 3.0T CX with MR Elastography recall (Philips North America)
Medical deviceClass IIOngoing
Summary: Philips North America is recalling Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.. Reason: The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
What is being recalled
Philips Ingenia 3.0T CX with MR Elastography (MRE). 1. Model Number (REF): 781271. 2. Model Number (REF): 782105. 510(k) Numbers: K173079, K183063, K193215, K213583, K223458, K251397, K251808.
Why
The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Canada, Chile, Cuba, Cyprus, Denmark, Ecuador, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Kenya, Lebanon, Lithuania, Malta, Mexico, Myanmar, Netherlands, Pakistan, Palestine, Peru, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Kingdom.
How to tell if yours is affected
1. Model Number (REF): 781271. UDI: (01)00884838068452(21)78017, (01)00884838068452(21)78013, (01)00884838068452(21)78014, (01)00884838068452(21)78034, (01)00884838068452(21)78058, (01)00884838068452(21)78073, (01)00884838068452(21)78075, (01)00884838068452(21)78092, (01)00884838068452(21)78112, (01)00884838068452(21)78123, (01)00884838068452(21)78201, (01)00884838068452(21)78202, (01)00884838068452(21)78215, (01)00884838068452(21)78217, (01)00884838068452(21)78239, (01)00884838068452(21)78247, (01)00884838068452(21)78260,(01)00884838068452(21)78286, (01)00884838068452(21)78287, (01)00884838068452(21)78026. Serial Numbers: 78017, 78013, 78014, 78034, 78058, 78073, 78075, 78092, 78112, 78123, 78201, 78202, 78215, 78217, 78239, 78247, 78260, 78286, 78287, 78026. 2. Model Number (REF): 782105. UDI: (01)00884838098312(21)78532, (01)00884838098312(21)78576. Serial Numbers: 78532, 78576.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-1959-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | April 14, 2026 |
| Reported | May 6, 2026 |
| Company | Philips North America |
| Company location | Cambridge, MA, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 22 units |
More recalls from Philips North America · all medical device recalls · recalls from April 2026