Recall Ledger

Heraeus recall (Heraeus Medical GmbH (Dental Division))

Medical deviceClass IIIOngoing

Summary: Heraeus Medical GmbH (Dental Division) is recalling Heraeus, PALAMIX duo. Material Number: 66057897.. Reason: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What is being recalled

Heraeus, PALAMIX duo. Material Number: 66057897.

Why

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

How serious is it

Class III is the least serious level: using or being exposed to this product is not likely to cause health problems, but it breaks FDA rules.

Where it was sold

Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.; (states named: Puerto Rico)

How to tell if yours is affected

Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-1998-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass III
StatusOngoing
Recall startedMarch 25, 2026
ReportedMay 6, 2026
CompanyHeraeus Medical GmbH (Dental Division)
Company locationWehrheim, Germany
Initiated byVoluntary: Firm initiated
Quantity1,125 units

More recalls from Heraeus Medical GmbH (Dental Division) · all medical device recalls · recalls from March 2026

Official FDA record