Medline medical procedure kits recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU CDS982389J; 3. VP SHUNT ADULT PACK-LF, Medline Kit Number/SKU DYNJ0117497G; 4. PICC LINE PACK, Medline Kit Number. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

What is being recalled

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU CDS982389J; 3. VP SHUNT ADULT PACK-LF, Medline Kit Number/SKU DYNJ0117497G; 4. PICC LINE PACK, Medline Kit Number/SKU DYNJ04592K; 5. VP SHUNT PACK-LF, Medline Kit Number/SKU DYNJ0750993L; 6. PEDI VP SHUNT ADD A PACK-LF, Medline Kit Number/SKU DYNJ0753933F; 7. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861B; 8. A-V FISTULA PACK GH-LF, Medline Kit Number/SKU DYNJ39861D; 9. EVD VP SHUNT DBS PACK 319705, Medline Kit Number/SKU DYNJ44603C; 10. BURR HOLE SHUNT PACK, Medline Kit Number/SKU DYNJ56819G; 11. CHRISTUS CHILDRENS VP SHUNT PK, Medline Kit Number/SKU DYNJ61174A; 12. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213C; 13. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213F; 14. NASAL ENDO PACK, Medline Kit Number/SKU DYNJ84501; 15. RO VP SHUNT PACK, Medline Kit Number/SKU PHS392834010D.

Why

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23JBW804; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23JBX015; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23JBD094; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23IBQ501; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23FBO564; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23EBW686; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 23DBO524; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22GBU281; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22EBR456; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22EBM213; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22EBN031; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22DBP331; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22MBA145; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 22ABM907; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21KBE066; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21JBE527; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21HBH244; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21HBE520; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21HBE521; Medline Kit Number/SKU CDS840193V: UDI/DI each 10193489822540, UDI/DI case 40193489822541, Lot Number : 21FBC630; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 22IBU058; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 22HBH306; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 22GBQ912; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 22DBT185; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 22DBI063; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 21IBM934; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 21GBN290; Medline Kit Number/SKU CDS982389J: UDI/DI each 10193489242553, UDI/DI case 40193489242554, Lot Number : 21DBG187; Medline Kit Number/SKU DYNJ0117497G: UDI/DI each 10193489896541, UDI/DI case 40193489896542, Lot Number : 21IBP593; Medline Kit Number/SKU DYNJ0117497G: UDI/DI each 10193489896541, UDI/DI case 40193489896542, Lot Number : 21HBA351; Medline Kit Number/SKU DYNJ0117497G: UDI/DI each 10193489896541, UDI/DI case 40193489896542, Lot Number : 21GBE684; Medline Kit Number/SKU DYNJ0117497G: UDI/DI each 10193489896541, UDI/DI case 40193489896542, Lot Number : 21GBA846; Medline Kit Number/SKU DYNJ0117497G: UDI/DI each 10193489896541, UDI/DI case 40193489896542, Lot Number : 21FBI072; Medline Kit Number/SKU DYNJ04592K: UDI/DI each 10193489215205, UDI/DI case 40193489215206, Lot Number : 23JLA230; Medline Kit Number/SKU DYNJ0750993L: UDI/DI each 10193489201390, UDI/DI case 40193489201391, Lot Number : 22FBH340; Medline Kit Number/SKU DYNJ0750993L: UDI/DI each 10193489201390, UDI/DI case 40193489201391, Lot Number : 21CKA561; Medline Kit Number/SKU DYNJ0750993L: UDI/DI each 10193489201390, UDI/DI case 40193489201391, Lot Number : 21BKA591; Medline Kit Number/SKU DYNJ0750993L: UDI/DI each 10193489201390, UDI/DI case 40193489201391, Lot Number : 21AKB149; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 23KBE105; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 23JBH870; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 23IBS256; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 23HBW099; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 23EBB384; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 22JBP897; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 21ABI491; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 21ABG881; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 21ABH208; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 21ABF236; Medline Kit Number/SKU DYNJ0753933F: UDI/DI each 10193489283273, UDI/DI case 40193489283274, Lot Number : 21ABB836; Medline Kit Number/SKU DYNJ39861B: UDI/DI each 10193489477511, UDI/DI case 40193489477512, Lot Number : 21KMH013; Medline Kit Number/SKU DYNJ39861B: UDI/DI each 10193489477511, UDI/DI case 40193489477512, Lot Number : 21JME705; Medline Kit Number/SKU DYNJ39861B: UDI/DI each 10193489477511, UDI/DI case 40193489477512, Lot Number : 21JMD611; Medline Kit Number/SKU DYNJ39861B: UDI/DI each 10193489477511, UDI/DI case 40193489477512, Lot Number : 21IME971; Medline Kit Number/SKU DYNJ39861D: UDI/DI each 10198459433559, UDI/DI case 40198459433550, Lot Number : 26BME412; Medline Kit Number/SKU DYNJ39861D: UDI/DI each 10198459433559, UDI/DI case 40198459433550, Lot Number : 26AMJ494; Medline Kit Number/SKU DYNJ39861D: UDI/DI each 10198459433559, UDI/DI case 40198459433550, Lot Number : 25KMG940; Medline Kit Number/SKU DYNJ39861D: UDI/DI each 10198459433559, UDI/DI case 40198459433550, Lot Number : 25GMH433; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23KBN958; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23JBH478; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23HBT146; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23GBG624; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23CBC001; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23ABU575; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 23ABL787; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22LBM663; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22LBJ926; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22JBG520; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22IBR340; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22EBU759; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22DBR774; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22BBZ813; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 22BBZ792; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21KBG387; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21IBR350; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21GBV108; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21FBJ559; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21DBE602; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 21CBJ874; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 26BBI620; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 26BBA242; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25JBD488; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25HBM158; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25GBE204; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25EBR912; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25BBU542; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 25ABW421; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24KBO504; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24JBM902; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24IBK100; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24GBE540; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24FBK645; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24DBS967; Medline Kit Number/SKU DYNJ44603C: UDI/DI each 10193489767032, UDI/DI case 40193489767033, Lot Number : 24BBG984; Medline Kit Number/SKU DYNJ56819G: UDI/DI each 10195327073749, UDI/DI case 40195327073740, Lot Number : 22HBB961; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21KLA936; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21JLA574; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21HLA401; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21GLB120; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21GLA452; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21GLA295; Medline Kit Number/SKU DYNJ61174A: UDI/DI each 10193489860740, UDI/DI case 40193489860741, Lot Number : 21CLA652; Medline Kit Number/SKU DYNJ68213C: UDI/DI each 10195327511999, UDI/DI case 40195327511990, Lot Number : 23IMI021; Medline Kit Number/SKU DYNJ68213C: UDI/DI each 10195327511999, UDI/DI case 40195327511990, Lot Number : 24HMA726; Medline Kit Number/SKU DYNJ68213C: UDI/DI each 10195327511999, UDI/DI case 40195327511990, Lot Number : 24GMI771; Medline Kit Number/SKU DYNJ68213C: UDI/DI each 10195327511999, UDI/DI case 40195327511990, Lot Number : 24DMB264; Medline Kit Number/SKU DYNJ68213C: UDI/DI each 10195327511999, UDI/DI case 40195327511990, Lot Number : 23KMJ671; Medline Kit Number/SKU DYNJ68213F: UDI/DI each 10198459501173, UDI/DI case 40198459501174, Lot Number : 25IMD349; Medline Kit Number/SKU DYNJ84501: UDI/DI each 10195327371845, UDI/DI case 40195327371846, Lot Number : 23GBI579; Medline Kit Number/SKU DYNJ84501: UDI/DI each 10195327371845, UDI/DI case 40195327371846, Lot Number : 24ABL755; Medline Kit Number/SKU DYNJ84501: UDI/DI each 10195327371845, UDI/DI case 40195327371846, Lot Number : 23LBC527; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 22CDC077; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 22ADC035; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 22ADB161; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 22ADA979; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 21LDA084; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 21HDB299; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 21DDC112; Medline Kit Number/SKU PHS392834010D: UDI/DI each 10193489222432, UDI/DI case 40193489222433, Lot Number : 21ADC003.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2017-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 19, 2026
ReportedMay 6, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity3588 kits