Medline medical procedure kits recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645B; 3. ENT PK 1008847, Medline Kit Number/SKU DYNJ42319C; 4. ENDO SINUS PACK, Medline Kit Number/SKU D. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

What is being recalled

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645B; 3. ENT PK 1008847, Medline Kit Number/SKU DYNJ42319C; 4. ENDO SINUS PACK, Medline Kit Number/SKU DYNJ66449; 5. ENT, Medline Kit Number/SKU DYNJ900282I; 6. SEPTO/ENDO, Medline Kit Number/SKU DYNJ902831D; 7. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ906630.

Why

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 23DBM600; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 22JBM081; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 22ABJ346; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21KBO873; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21KBO877; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21JBX107; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21BBE213; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21BBB024; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 22AMD386; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21LMC822; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21JMF409; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21IMC184; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21GMB555; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21EMD626; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21CMC668; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21AMB081; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23FMD067; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23EMB306; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23CMD967; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22KMB384; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22OBB494; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22BBT747; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22ABA711; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21LBG900; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21KBD492; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21EBS624; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21CBY739; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21ABL492; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22KBD211; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22JBX995; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22JBL500; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 24CBB115; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 22ABB463; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 22ABB461; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 21KBT464; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 21KBP591; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23EBS895; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23CBU007; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23ABE854; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22LBE910; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22JBP701; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22FBV579; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22BBZ713; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22BBZ714; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21LBD683; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21IBM004; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21FBJ190; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21EBE932; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21DBA613; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21CBI323; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22FBM378; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22EBE987; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22OBI162; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21KBF463; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21IBU411; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21GBJ251; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21DBU812.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2024-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 19, 2026
ReportedMay 6, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity2153 kits