Medline medical procedure kits recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Numb. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

What is being recalled

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ39461J; 3. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060I; 4. PK-MAD,SPINE, Medline Kit Number/SKU DYNJ58060J; 5. NEURO BASIN KIT, Medline Kit Number/SKU DYNJ80200B; 6. NEURO-SPINE BASIN PACK, Medline Kit Number/SKU DYNJ83019; 7. ENT PACK, Medline Kit Number/SKU DYNJ86328; 8. EDMOND SPINE PACK, Medline Kit Number/SKU DYNJ89263; 9. ENT PACK, Medline Kit Number/SKU DYNJ89675; 10. ENT PACK, Medline Kit Number/SKU DYNJ89675A; 11. ENT/PAIN PACK, Medline Kit Number/SKU DYNJT7493.

Why

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25IMD003; Medline Kit Number/SKU DYNJ0532910L: UDI/DI each 10198459503870, UDI/DI case 40198459503871, Lot Number: 25HMH643; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22GBS127; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 22FBB472; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25FBE752; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25EBA260; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25DBG176; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25CBD819; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 25BBA593; Medline Kit Number/SKU DYNJ39461J: UDI/DI each 10195327112721, UDI/DI case 40195327112722, Lot Number: 24KBJ163; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23KMI415; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMK023; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23JMB211; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23IMC591; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23HMG291; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23GMG136; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23FME285; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23EME227; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 23CMD903; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMI915; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24KMH419; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24JMG571; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24HMF066; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24FMG130; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24CMI441; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24BMD229; Medline Kit Number/SKU DYNJ58060I: UDI/DI each 10195327326159, UDI/DI case 40195327326150, Lot Number: 24AMH016; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26BMB944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 26AMJ197; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25KMC587; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25JMD275; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25IMC944; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25GMF811; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25FME706; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25EMA516; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 25CME296; Medline Kit Number/SKU DYNJ58060J: UDI/DI each 10198459210662, UDI/DI case 40198459210663, Lot Number: 24LMD726; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 22HBH553; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 26BBF722; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25JBB835; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBO944; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBL998; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBH356; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25EBF751; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25DBB435; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 25BBC547; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24KBR868; Medline Kit Number/SKU DYNJ80200B: UDI/DI each 10195327213978, UDI/DI case 40195327213979, Lot Number: 24JBV969; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBB952; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 23BBC125; Medline Kit Number/SKU DYNJ83019: UDI/DI each 10195327228842, UDI/DI case 40195327228843, Lot Number: 22JBR211; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25LBS635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25KBE004; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25HBR489; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25FBL484; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25DBD392; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 25ABD635; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24KBB462; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24JBE518; Medline Kit Number/SKU DYNJ86328: UDI/DI each 10195327622381, UDI/DI case 40195327622382, Lot Number: 24GBE890; Medline Kit Number/SKU DYNJ89263: UDI/DI each 10198459253348, UDI/DI case 40198459253349, Lot Number: 25BBK629; Medline Kit Number/SKU DYNJ89675: UDI/DI each 10198459281693, UDI/DI case 40198459281694, Lot Number: 25CME067; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMG054; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 26AMA771; Medline Kit Number/SKU DYNJ89675A: UDI/DI each 10198459417436, UDI/DI case 40198459417437, Lot Number: 25GME903; Medline Kit Number/SKU DYNJT7493: UDI/DI each 10198459626258, UDI/DI case 40198459626259, Lot Number: 26ABA254.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2026-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 19, 2026
ReportedMay 6, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity9805 kits