Medline medical procedure kits recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medline Kit Number/SKU DYNJ47568B; 3. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555B; 4. UROLOGY ROBOTIC, Medline K. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

What is being recalled

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medline Kit Number/SKU DYNJ47568B; 3. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555B; 4. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555C; 5. UROLOGY ROBOTIC, Medline Kit Number/SKU DYNJ55555D.

Why

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide distribution.

How to tell if yours is affected

Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 22EMB486; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 22CMI117; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 22CMB218; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 22CMB219; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21LMA406; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21LMA805; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21JMF667; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21IMF752; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21HMG136; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21HMB278; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21GMD367; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21CKA503; Medline Kit Number/SKU DYNJ0751220P: UDI/DI each 10193489815122, UDI/DI case 40193489815123, Lot Number: 21AKA471; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23JBT331; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23IBL993; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23GBN118; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23FBQ324; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23EBO916; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23CBI861; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23BBE825; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23ABD998; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22LBA177; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22IBJ001; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22FBY332; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22FBX502; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22EBE351; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 22ABO417; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21WBA306; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21KBG927; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21GBW059; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21GBW057; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21EBP877; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 21EBP891; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 26ABT067; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 25LBO620; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 25LBC010; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24KBR338; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24HBS302; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24HBR212; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24FBM377; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24EBC753; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24BBN421; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 24BBG081; Medline Kit Number/SKU DYNJ47568B: UDI/DI each 10193489846058, UDI/DI case 40193489846059, Lot Number: 23LBC938; Medline Kit Number/SKU DYNJ55555B: UDI/DI each 10195327040284, UDI/DI case 40195327040285, Lot Number: 22ABW137; Medline Kit Number/SKU DYNJ55555B: UDI/DI each 10195327040284, UDI/DI case 40195327040285, Lot Number: 22ABD653; Medline Kit Number/SKU DYNJ55555B: UDI/DI each 10195327040284, UDI/DI case 40195327040285, Lot Number: 21LBN774; Medline Kit Number/SKU DYNJ55555C: UDI/DI each 10195327154738, UDI/DI case 40195327154739, Lot Number: 22IBD321; Medline Kit Number/SKU DYNJ55555C: UDI/DI each 10195327154738, UDI/DI case 40195327154739, Lot Number: 22HBB560; Medline Kit Number/SKU DYNJ55555C: UDI/DI each 10195327154738, UDI/DI case 40195327154739, Lot Number: 22GBG766; Medline Kit Number/SKU DYNJ55555C: UDI/DI each 10195327154738, UDI/DI case 40195327154739, Lot Number: 22DBU479; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23KBQ339; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23JBF048; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23EBX527; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23DBG858; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23CBX731; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 23CBF699; Medline Kit Number/SKU DYNJ55555D: UDI/DI each 10195327276041, UDI/DI case 40195327276042, Lot Number: 22LBJ888.

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2034-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedMarch 19, 2026
ReportedMay 6, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity4324 kits