Medline medical procedure kits recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DY. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
What is being recalled
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572B; 5. MICROSURGERY PACK, Medline Kit Number/SKU DYNJ67284A; 6. BASIC EYE PACK, Medline Kit Number/SKU DYNJ81411A; 7. SMR SINUS PACK, Medline Kit Number/SKU DYNJ84478; 8. FESS/SEPTO PACK, Medline Kit Number/SKU DYNJ85025; 9. PLASTIC EYE PACK, Medline Kit Number/SKU DYNJ89030; 10. NEURO, Medline Kit Number/SKU DYNJ902935K; 11. OCULAR, Medline Kit Number/SKU DYNJ909065; 12. OCULAR, Medline Kit Number/SKU DYNJ909065A.
Why
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution.
How to tell if yours is affected
Medline Kit Number/SKU DYNJ48725A: UDI/DI each 10193489853896, UDI/DI case 40193489853897, Lot Number: 24IMG757; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22FME241; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22DMC340; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21JMC829; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21IMC497; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21HMD139; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21GME747; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21FMA440; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21DBU602; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21CBJ936; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23IBC227; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23FBA458; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23DBC239; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23ABI841; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBP260; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBO842; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBN801; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22GBL794; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24HBB998; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24DBQ664; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 24CBQ260; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 26ABC645; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25LBB014; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25JBM975; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25HBN493; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25GBM010; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25FBJ206; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25CBN002; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25ABF979; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 25ABA086; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 24JBE583; Medline Kit Number/SKU DYNJ64572B: UDI/DI each 10198459131752, UDI/DI case 40198459131753, Lot Number: 24JBG667; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25KBQ749; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25JBF743; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 25IBA573; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24IBO985; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24IBB916; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24GBY148; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 24EBO714; Medline Kit Number/SKU DYNJ67284A: UDI/DI each 10195327575731, UDI/DI case 40195327575732, Lot Number: 23LBH320; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23KBO017; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23JBA275; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23GBK387; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23CBD302; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 26ABF562; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25GBF664; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25EBB838; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 25BBQ021; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24LBB454; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24JBE139; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24GBI010; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24FBR965; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24DBQ797; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 24BBH448; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23LBL498; Medline Kit Number/SKU DYNJ81411A: UDI/DI each 10195327320836, UDI/DI case 40195327320837, Lot Number: 23KBX879; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23IBE948; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23FBB413; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 24ABN367; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 24ABA089; Medline Kit Number/SKU DYNJ84478: UDI/DI each 10195327370107, UDI/DI case 40195327370108, Lot Number: 23KBU302; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25LBL874; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25HBD880; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 25EBI919; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24LBG184; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24LBF578; Medline Kit Number/SKU DYNJ85025: UDI/DI each 10195327447564, UDI/DI case 40195327447565, Lot Number: 24JBS764; Medline Kit Number/SKU DYNJ89030: UDI/DI each 10198459229725, UDI/DI case 40198459229726, Lot Number: 25JBH439; Medline Kit Number/SKU DYNJ89030: UDI/DI each 10198459229725, UDI/DI case 40198459229726, Lot Number: 25ABP922; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23IBW005; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23IBG214; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23FBP153; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23FBN621; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23EBW204; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23EBE141; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24DBP594; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24BBB058; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 24BBB100; Medline Kit Number/SKU DYNJ902935K: UDI/DI each 10195327370442, UDI/DI case 40195327370443, Lot Number: 23LBE522; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 23CMB048; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 23CMA495; Medline Kit Number/SKU DYNJ909065: UDI/DI each 10195327260019, UDI/DI case 40195327260010, Lot Number: 22LMD749; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 23HMI877; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 23HMH413; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25GMJ496; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25DMD357; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 25AMJ130; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24IME711; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24HME507; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24DMF364; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24CMI687; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24CMG984; Medline Kit Number/SKU DYNJ909065A: UDI/DI each 10195327460648, UDI/DI case 40195327460649, Lot Number: 24BMI425.
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2037-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | March 19, 2026 |
| Reported | May 6, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 3290 kits |
More recalls from Medline Industries, LP · all medical device recalls · recalls from March 2026