LUX-DX II recall (Boston Scientific Corporation)
Medical deviceClass IIOngoing
Summary: Boston Scientific Corporation is recalling LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5. Reason: For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
What is being recalled
LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Why
For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide distribution - US Nationwide and the country of Belgium.
How to tell if yours is affected
LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2048-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | March 30, 2026 |
| Reported | May 13, 2026 |
| Company | Boston Scientific Corporation |
| Company location | Saint Paul, MN, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 2 units |