Medline Convenience Kits: 1) ANGIO PROCEDURE PACK recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4A; . Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What is being recalled

Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AB; 2) OR ARTERIOGRAM, Model Number:CDS985094F; 3) INTRAVASOR CATH KIT, Model Number:DYK1060193I; 4) KIT SURG ONC LAP DIAG/CV CATH, Model Number:DYKMBNDL153; 5) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4A; 6) KIT ACES INSERT PERCUTANEOUS T, Model Number:DYKMBNDL4B; 7) VEIN PROCEDURE KIT, Model Number:DYNDA2076C; 8) CARDIAC CATH PACK-LF, Model Number:DYNJ0150605AC; 9) CARDIAC CATH PACK-LF, Model Number:DYNJ0368386D; 10) CATH LAB PACK-LF, Model Number:DYNJ0648473Y; 11) NBMC CATH PACK-LF, Model Number:DYNJ0803367I; 12) RADIOLOGY PACK-LF, Model Number:DYNJ0928124C; 13) ANGIO PACK-LF, Model Number:DYNJ0954970I; 14) SPECIALS TRAY RADIOLOGY, Model Number:DYNJ16803D; 15) CARDIAC CATH PACK, Model Number:DYNJ19946M; 16) CATH LAB PACK, Model Number:DYNJ30955I; 17) CATH LAB PACK, Model Number:DYNJ31773K; 18) CATH PACK-LF, Model Number:DYNJ33061D; 19) DB CCL OAKWOOD PACK-LF, Model Number:DYNJ33620D; 20) CARDIAC CATH PACK MRH-LF, Model Number:DYNJ35300M; 21) CATH LAB PK, Model Number:DYNJ35643C; 22) CARDIAC CATH PACK-LF, Model Number:DYNJ39066B; 23) ANGIOGRAPHY PACK, Model Number:DYNJ42134G; 24) CATH PACK-LF, Model Number:DYNJ43101C; 25) OR ANGIO PACK-LF, Model Number:DYNJ43415D; 26) OR ANGIO PACK-LF, Model Number:DYNJ43415F; 27) ANGIOPLASTY PACK, Model Number:DYNJ44700J; 28) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ44850J; 29) RADIOLOGICAL PACK, Model Number:DYNJ44889O; 30) CARDIAC CATH PACK, Model Number:DYNJ47370B; 31) ANGIOGRAPHY PACK, Model Number:DYNJ53129D; 32) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJ56772G; 33) DR. MOHINDRA PACK, Model Number:DYNJ57602G; 34) NE ENDO VASCULAR PACK, Model Number:DYNJ57940B; 35) CARDIAC CATH PACK - Q1, Model Number:DYNJ58048C; 36) OR ANGIO, Model Number:DYNJ59931B; 37) LPCH PERRY CATH PACK, Model Number:DYNJ67378C; 38) HOSPITAL REGIONAL DE CONCEPCIO, Model Number:DYNJ67918A; 39) CATH LAB PACK, Model Number:DYNJ68633; 40) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197B; 41) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197C; 42) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197D; 43) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197F; 44) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197G; 45) STEWARD CATH LAB PACK GSS, Model Number:DYNJ81197I; 46) BASIC CARDIAC CATH PACK, Model Number:DYNJ81593B; 47) CATH LAB PACK, Model Number:DYNJ83073; 48) UMC EL PASO CATH PACK, Model Number:DYNJ83089; 49) DR. MOHINDRA PACK, Model Number:DYNJ83819A; 50) VZ VAS/ANG PACK, Model Number:DYNJ83981; 51) WCPN CARDIOLOGY PACK, Model Number:DYNJ84539; 52) CATH LAB, Model Number:DYNJ85511; 53) CATH LAB, Model Number:DYNJ85511A; 54) ANGIOGRAPHY DRAPE PACK, Model Number:DYNJT1520K; 55) CVC BUNDLE KIT, Model Number:ECVC6225A; 56) CVC 3L 7F 20CM VANTEX, Model Number:ECVC7265A; 57) ANGIOGRAPHY TRAY, Model Number:MNS9130

Why

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

1) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24IBQ637; 2) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24KBF279; 3) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 24LBF132; 4) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25CBJ807; 5) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25DBO471; 6) 00-401824AB, UDI-DI: 10198459089169(each), 40198459089160(case), Lot Number: 25GBK506; 7) CDS985094F, UDI-DI: 10195327657642(each), 40195327657643(case), Lot Number: 25AMG933; 8) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 22KBQ113; 9) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23ABU376; 10) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23DBD193; 11) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23GBT819; 12) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23HBQ238; 13) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23IBR569; 14) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23KBH770; 15) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 23LBO849; 16) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24ABW181; 17) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24EBM208; 18) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24GBA239; 19) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24IBE619; 20) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24JBA719; 21) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24KBJ312; 22) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 24LBM892; 23) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25BBK679; 24) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25CBB540; 25) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25DBA780; 26) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25FBR530; 27) DYK1060193I, UDI-DI: 10884389871737(each), 40884389871738(case), Lot Number: 25GBQ911; 28) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21ABI976; 29) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21BBG516; 30) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 21LBJ748; 31) DYKMBNDL153, UDI-DI: 10193489478440(each), 40193489478441(case), Lot Number: 22ABO093; 32) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21GLA287; 33) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21GLA930; 34) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 21KLA100; 35) DYKMBNDL4A, UDI-DI: 10193489904512(each), 40193489904513(case), Lot Number: 22ALA279; 36) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 22HLB260; 37) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23BLA583; 38) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23GLA176; 39) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23GLA710; 40) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23HLA316; 41) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23ILA268; 42) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23JLA979; 43) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 23KLA402; 44) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 24CLA777; 45) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 24FLA816; 46) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25ALA014; 47) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25ELA738; 48) DYKMBNDL4B, UDI-DI: 10195327200589(each), 40195327200580(case), Lot Number: 25HLA757; 49) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24HMD928; 50) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24KMC111; 51) DYNDA2076C, UDI-DI: 10198459001208(each), 40198459001209(case), Lot Number: 24LMB205; 52) DYNJ0150605AC, UDI-DI: 10198459106422(each), 40198459106423(case), Lot Number: 24IMB736; 53) DYNJ0368386D, UDI-DI: 10889942060245(each), 40889942060246(case), Lot Number: 23BMB501; 54) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24GBT979; 55) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24HBQ060; 56) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24IBS786; 57) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 24LBE715; 58) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25ABH209; 59) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25BBT175; 60) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25CBO084; 61) DYNJ0648473Y, UDI-DI: 10195327647834(each), 40195327647835(case), Lot Number: 25EBR586; 62) DYNJ0803367I, UDI-DI: 10195327461027(each), 40195327461028(case), Lot Number: 23IMC122; 63) DYNJ0803367I, UDI-DI: 10195327461027(each), 40195327461028(case), Lot Number: 23JMA137; 64) DYNJ0928124C, UDI-DI: 10193489484427(each), 40193489484428(case), Lot Number: 23JLB288; 65) DYNJ0954970I, UDI-DI: 10198459297571(each), 40198459297572(case), Lot Number: 25FMJ515; 66) DYNJ16803D, UDI-DI: 10195327503710(each), 40195327503711(case), Lot Number: 23JMC203; 67) DYNJ19946M, UDI-DI: 10193489907933(each), 40193489907934(case), Lot Number: 23HBO857; 68) DYNJ30955I, UDI-DI: 10195327428051(each), 40195327428052(case), Lot Number: 23GBN513; 69) DYNJ31773K, UDI-DI: 10195327245856(each), 40195327245857(case), Lot Number: 23ABB938; 70) DYNJ33061D, UDI-DI: 10193489841428(each), 40193489841429(case), Lot Number: 21CBI315; 71) DYNJ33620D, UDI-DI: 10889942677832(each), 40889942677833(case), Lot Number: 21ADA464; 72) DYNJ35300M, UDI-DI: 10195327101169(each), 40195327101160(case), Lot Number: 22JMB302; 73) DYNJ35643C, UDI-DI: 10195327111946(each), 40195327111947(case), Lot Number: 23CBI301; 74) DYNJ39066B, UDI-DI: 10193489937268(each), 40193489937269(case), Lot Number: 23HMG433; 75) DYNJ42134G, UDI-DI: 10195327644581(each), 40195327644582(case), Lot Number: 24IBS009; 76) DYNJ43101C, UDI-DI: 10888277737327(each), 40888277737328(case), Lot Number: 21ADA535; 77) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 23IBP533; 78) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 24ABI910; 79) DYNJ43415D, UDI-DI: 10195327512873(each), 40195327512874(case), Lot Number: 24CBG256; 80) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24GBV382; 81) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24HBM254; 82) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24IBU874; 83) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24KBR090; 84) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 24LBQ306; 85) DYNJ43415F, UDI-DI: 10195327674731(each), 40195327674732(case), Lot Number: 25EBF775; 86) DYNJ44700J, UDI-DI: 10195327277413(each), 40195327277414(case), Lot Number: 25JDA679; 87) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 21KBO116; 88) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22ABI539; 89) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22ABK883; 90) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22CBV346; 91) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22EBP030; 92) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22FBN577; 93) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22HBM547; 94) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22IBT510; 95) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22JBC588; 96) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22JBY064; 97) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22LBG760; 98) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 22OBA581; 99) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23ABR589; 100) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23BBD736; 101) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23DBI845; 102) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23EBG893; 103) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23FBF258; 104) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23GBP204; 105) DYNJ44850J, UDI-DI: 10193489976342(each), 40193489976343(case), Lot Number: 23IBB504; 106) DYNJ44889O, UDI-DI: 10193489979411(each), 40193489979412(case), Lot Number: 22AME756; 107) DYNJ44889O, UDI-DI: 10193489979411(each), 40193489979412(case), Lot Number: 22CMA383; 108) DYNJ47370B, UDI-DI: 10889942975723(each), 40889942975724(case), Lot Number: 22LBP706; 109) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22GMI134; 110) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22IMG735; 111) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22JME805; 112) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22KMC442; 113) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 22LMD047; 114) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23BMC700; 115) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23CMB279; 116) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23DMF448; 117) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23EME724; 118) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23FME772; 119) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23GMI002; 120) DYNJ53129D, UDI-DI: 10195327107222(each), 40195327107223(case), Lot Number: 23HMH791; 121) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22AMD580; 122) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22BMG182; 123) DYNJ56772G, UDI-DI: 10195327054069(each), 40195327054060(case), Lot Number: 22CMH914; 124) DYNJ57602G, UDI-DI: 10195327338169(each), 40195327338160(case), Lot Number: 23BBK044; 125) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23GBE283; 126) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23GBP181; 127) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23JBB692; 128) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23JBX727; 129) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 23KBB465; 130) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABP912; 131) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABS524; 132) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24ABT394; 133) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24CBC189; 134) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24CBO358; 135) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24DBF723; 136) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24EBH891; 137) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24FBH066; 138) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24GBA907; 139) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24GBQ359; 140) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24JBB314; 141) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24JBC107; 142) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24KBN562; 143) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 24LBJ716; 144) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25BBA430; 145) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25DBH293; 146) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25EBL462; 147) DYNJ57940B, UDI-DI: 10195327450762(each), 40195327450763(case), Lot Number: 25FBK782; 148) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23FMF145; 149) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 22LMA967; 150) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23AMI769; 151) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23BME160; 152) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23DMI350; 153) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23FME284; 154) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23GMF491; 155) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23IMD765; 156) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23KMH102; 157) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 23LMD057; 158) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24CMF744; 159) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24CMG318; 160) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24DMC509; 161) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24DMG251; 162) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24EMH576; 163) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24GMH576; 164) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24HMF813; 165) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24JMJ426; 166) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 24KMD474; 167) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25AMG240; 168) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25CMD739; 169) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25DMK093; 170) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25FMC822; 171) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25GMF770; 172) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25HMD080; 173) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25JME465; 174) DYNJ58048C, UDI-DI: 10195327285098(each), 40195327285099(case), Lot Number: 25KMD960; 175) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22BBX590; 176) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22DBM388; 177) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22DBS611; 178) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22FBU003; 179) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22HBI355; 180) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22KBH299; 181) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 22LBA917; 182) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23ABL095; 183) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23CBU118; 184) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23EBN753; 185) DYNJ59931B, UDI-DI: 10195327106546(each), 40195327106547(case), Lot Number: 23GBC772; 186) DYNJ67378C, UDI-DI: 10195327697501(each), 40195327697502(case), Lot Number: 24HME465; 187) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 21LBP804; 188) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22CBZ026; 189) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22CBZ027; 190) DYNJ67918A, UDI-DI: 10195327042776(each), 40195327042777(case), Lot Number: 22OBG779; 191) DYNJ68633, UDI-DI: 10193489837667(each), 40193489837668(case), Lot Number: 21BBQ520; 192) DYNJ81197B, UDI-DI: 10195327427818(each), 40195327427819(case), Lot Number: 23FBA862; 193) DYNJ81197B, UDI-DI: 10195327427818(each), 40195327427819(case), Lot Number: 23GBC445; 194) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23GBJ036; 195) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23HBZ547; 196) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23IBT017; 197) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23JBF780; 198) DYNJ81197C, UDI-DI: 10195327460839(each), 40195327460830(case), Lot Number: 23JBR165; 199) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24ABW070; 200) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24DBF346; 201) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24EBU925; 202) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24GBA095; 203) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24HBK118; 204) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24IBL386; 205) DYNJ81197D, UDI-DI: 10195327590093(each), 40195327590094(case), Lot Number: 24JBE696; 206) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 24LBU027; 207) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25CBG703; 208) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25DBT193; 209) DYNJ81197F, UDI-DI: 10198459225000(each), 40198459225001(case), Lot Number: 25EBR556; 210) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25FBT968; 211) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25HBT944; 212) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25LBF577; 213) DYNJ81197G, UDI-DI: 10198459369179(each), 40198459369170(case), Lot Number: 25KBE898; 214) DYNJ81197I, UDI-DI: 10198459640056(each), 40198459640057(case), Lot Number: 26ABG934; 215) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23EBT723; 216) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23GBO156; 217) DYNJ81593B, UDI-DI: 10195327405298(each), 40195327405299(case), Lot Number: 23IBJ841; 218) DYNJ83073, UDI-DI: 10195327234645(each), 40195327234646(case), Lot Number: 23DMI662; 219) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 22JMI325; 220) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23BMB766; 221) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23DMA913; 222) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23DMD375; 223) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23EMF492; 224) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23FMC409; 225) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23GMD133; 226) DYNJ83089, UDI-DI: 10195327230838(each), 40195327230839(case), Lot Number: 23IMB693; 227) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23CBS129; 228) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23EBU525; 229) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23FBJ743; 230) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23FBV261; 231) DYNJ83819A, UDI-DI: 10195327338176(each), 40195327338177(case), Lot Number: 23HBX967; 232) DYNJ83981, UDI-DI: 10195327324124(each), 40195327324125(case), Lot Number: 23CMI086; 233) DYNJ84539, UDI-DI: 10195327375706(each), 40195327375707(case), Lot Number: 23FMH066; 234) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 23JBP557; 235) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 23JBR927; 236) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24ABR587; 237) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24BBN470; 238) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24CBV309; 239) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24DBS240; 240) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24EBN545; 241) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24FBB518; 242) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24GBK818; 243) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24HBG260; 244) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24IBI836; 245) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24KBR659; 246) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 24LBK715; 247) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25BBE395; 248) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25BBP468; 249) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25CBR179; 250) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25EBB233; 251) DYNJ85511, UDI-DI: 10195327511302(each), 40195327511303(case), Lot Number: 25EBR907; 252) DYNJ85511A, UDI-DI: 10198459413735(each), 40198459413736(case), Lot Number: 25GBN456; 253) DYNJ85511A, UDI-DI: 10198459413735(each), 40198459413736(case), Lot Number: 25HBM139; 254) DYNJT1520K, UDI-DI: 10195327436261(each), 40195327436262(case), Lot Number: 23JLA175; 255) ECVC6225A, UDI-DI: 10653160333418(each), 653160333411(case), Lot Number: 2025110690; 256) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2023071090; 257) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2024021590; 258) ECVC7265A, UDI-DI: 10653160345145(each), 653160345148(case), Lot Number: 2024042390; 259) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22GBC976; 260) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22HBO815; 261) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22JBU760; 262) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 22LBK966; 263) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23ABR642; 264) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23CBI570; 265) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23GBG640; 266) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23JBN131; 267) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 23LBC300; 268) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24ABY158; 269) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24DBC089; 270) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24FBF351; 271) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24FBT379; 272) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24IBF858; 273) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 24JBW207; 274) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25ABS583; 275) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25DBF439; 276) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25EBA128; 277) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25FBI982; 278) MNS9130, UDI-DI: 10653160281214(each), 40653160281215(case), Lot Number: 25HBP222

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2092-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 7, 2026
ReportedMay 13, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity39341 units