See RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling See RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Mo. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
See RES for full list. Medline Convenience Kits: 1) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069G; 2) CHP COCHLEAR SUPPLEMENT CDS, Model Number: CDS983069I; 3) HEAD & NECK CDS-LF, Model Number: CDS983782C; 4) BAPTIST FLOYD BRONCHOSCOPY, Model Number: DYKE1456D; 5) BRONCHOSCOPY SETUP KIT, Model Number: DYKE1955; 6) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156; 7) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156A; 8) KIT ENT HEAD AND NECK GENERIC, Model Number: DYKMBNDL156B;
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
1) CDS983069G, UDI-DI: 10193489923926(each), 40193489923927(case), Lot Number: 21HBO921; 2) CDS983069G, UDI-DI: 10193489923926(each), 40193489923927(case), Lot Number: 21IBD399; 3) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22FBX804; 4) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22GBE731; 5) CDS983069I, UDI-DI: 10195327071868(each), 40195327071869(case), Lot Number: 22JBW315; 6) CDS983782C, UDI-DI: 10193489592047(each), 40193489592048(case), Lot Number: 21DKA756; 7) DYKE1456D, UDI-DI: 10195327639938(each), 40195327639939(case), Lot Number: 24CBK289; 8) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 24ALA266; 9) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 24DLA606; 10) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25CLB002; 11) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25DLB145; 12) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25FLA904; 13) DYKE1955, UDI-DI: 10195327539832(each), 40195327539833(case), Lot Number: 25HLA345; 14) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21ABK323; 15) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21BBG062; 16) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21CBD340; 17) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21DBJ855; 18) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21FBO537; 19) DYKMBNDL156, UDI-DI: 10193489478242(each), 40193489478243(case), Lot Number: 21FBS478; 20) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21GBE737; 21) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21HBF494; 22) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21IBP624; 23) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21JBA009; 24) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 21JBB014; 25) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22ABG015; 26) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22BBA231; 27) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22CBX230; 28) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22DBL428; 29) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22EBS812; 30) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22FBT038; 31) DYKMBNDL156A, UDI-DI: 10193489951912(each), 40193489951913(case), Lot Number: 22GBH498; 32) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 22IBU647; 33) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 22KBH786; 34) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23ABQ109; 35) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23DBM060; 36) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23EBS366; 37) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23FBG212; 38) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23HBT877; 39) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23HBV381; 40) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23JBQ487; 41) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 23JBT911; 42) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24ABL113; 43) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24DBV609; 44) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24EBF125; 45) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24GBD477; 46) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24GBR935; 47) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24HBU706; 48) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 24KBJ157; 49) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25ABK769; 50) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25BBM402; 51) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25DBQ008; 52) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25EBH975; 53) DYKMBNDL156B, UDI-DI: 10195327240363(each), 40195327240364(case), Lot Number: 25GBX739; 54) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21ABH824; 55) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21BBP836; 56) DYKMBNDL174, UDI-DI: 10193489478334(each), 40193489478335(case), Lot Number: 21FBQ554; 57) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 21LBY299; 58) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 22BBS429; 59) DYKMBNDL174A, UDI-DI: 10193489951882(each), 40193489951883(case), Lot Number: 22EBO543; 60) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 22KBK464; 61) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 22LBI002; 62) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 23ABH675; 63) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 23JBN238; 64) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24CBA389; 65) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24DBB691; 66) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 24EBP260; 67) DYKMBNDL174B, UDI-DI: 10195327240370(each), 40195327240371(case), Lot Number: 25LBM606; 68) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21ABP337; 69) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21BBG851; 70) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBF521; 71) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBH559; 72) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBM511; 73) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBS041; 74) DYKMBNDL196, UDI-DI: 10193489479393(each), 40193489479394(case), Lot Number: 21CBT411; 75) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 22JLB001; 76) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 22KLA452; 77) DYKMBNDL1A, UDI-DI: 10195327245832(each), 40195327245833(case), Lot Number: 23BLA884; 78) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23CLA769; 79) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23DLA495; 80) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23HLA624; 81) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23ILA756; 82) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23JLA058; 83) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23KLA620; 84) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 23LLA422; 85) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24FLA349; 86) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24GLA055; 87) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24HLA924; 88) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24HLB156; 89) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24ILA560; 90) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 24JLA055; 91) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25BLA837; 92) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25CLA312; 93) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25DLA251; 94) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25FLA791; 95) DYKMBNDL1B, UDI-DI: 10195327352738(each), 40195327352739(case), Lot Number: 25FLA924; 96) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA179; 97) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA232; 98) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25HLA843; 99) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25ILA035; 100) DYKMBNDL1C, UDI-DI: 10198459448591(each), 40198459448592(case), Lot Number: 25ILA386; 101) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21ABJ862; 102) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21ABJ985; 103) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21BBL111; 104) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21CBS031; 105) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21DBK216; 106) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21EBM706; 107) DYKMBNDL32, UDI-DI: 10193489478723(each), 40193489478724(case), Lot Number: 21FBT798; 108) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 21IBD784; 109) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 21JBJ743; 110) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22EBO633; 111) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22EBQ053; 112) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22FBE193; 113) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22GBF522; 114) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22GBQ602; 115) DYKMBNDL32A, UDI-DI: 10193489952018(each), 40193489952019(case), Lot Number: 22HBP925; 116) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 22IBL167; 117) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 22JBW957; 118) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23BBS356; 119) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23DBP706; 120) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23EBB247; 121) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23EBT500; 122) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23FBP266; 123) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23HBH678; 124) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23HBX217; 125) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23IBA611; 126) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 23LBP166; 127) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24ABH039; 128) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24ABK306; 129) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24CBR483; 130) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24DBU451; 131) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24EBU213; 132) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24GBC830; 133) DYKMBNDL32B, UDI-DI: 10195327240325(each), 40195327240326(case), Lot Number: 24GBP188; 134) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22JBG017; 135) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22JBQ673; 136) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 22LBN128; 137) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 23LBF184; 138) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24BBT350; 139) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24CBD159; 140) DYKMBNDL66A, UDI-DI: 10195327240349(each), 40195327240340(case), Lot Number: 24DBD514; 141) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24DBR155; 142) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24FBG015; 143) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24HBH734; 144) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24IBO509; 145) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24KBT017; 146) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 24LBT153; 147) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25ABH117; 148) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25CBR258; 149) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25JBF754; 150) DYKMBNDL66B, UDI-DI: 10198459000607(each), 40198459000608(case), Lot Number: 25LBI754; 151) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22JBD610; 152) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22JBW066; 153) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 22LBR829; 154) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23ABK217; 155) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23CBJ801; 156) DYKMBNDL88A, UDI-DI: 10195327240400(each), 40195327240401(case), Lot Number: 23CBN843; 157) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24BBJ842; 158) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24BBL266; 159) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24DBS395; 160) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24EBF126; 161) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24EBU598; 162) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24GBN944; 163) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24GBU190; 164) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24JBL602; 165) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 24JBX567; 166) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25ABH118; 167) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25ABS392; 168) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25CBI877; 169) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25EBK536; 170) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25EBU506; 171) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25FBT988; 172) DYKMBNDL88B, UDI-DI: 10195327633691(each), 40195327633692(case), Lot Number: 25HBT644; 173) DYKMBNDL88C, UDI-DI: 10198459584886(each), 40198459584887(case), Lot Number: 25LBP011; 174) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24DBR908; 175) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24GBH580; 176) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24HBJ497; 177) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24HBN645; 178) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24IBQ080; 179) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 24KBA441; 180) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 25GBG516; 181) DYKMBNDL96C, UDI-DI: 10198459000782(each), 40198459000783(case), Lot Number: 25HBP413; 182) DYNJ01831I, UDI-DI: 10889942572502(each), 40889942572503(case), Lot Number: 21EMA909; 183) DYNJ0387711F, UDI-DI: 10198459005893(each), 40198459005894(case), Lot Number: 24HMA304; 184) DYNJ0855090P, UDI-DI: 10198459605826(each), 40198459605827(case), Lot Number: 25LMA083; 185) DYNJ17005I, UDI-DI: 10193489241860(each), 40193489241861(case), Lot Number: 23HBV662; 186) DYNJ26780I, UDI-DI: 10193489238822(each), 40193489238823(case), Lot Number: 23GBP197; 187) DYNJ27320V, UDI-DI: 10195327104719(each), 40195327104710(case), Lot Number: 22LBQ654; 188) DYNJ30604R, UDI-DI: 10195327514198(each), 40195327514199(case), Lot Number: 23LBU616; 189) DYNJ32060C, UDI-DI: 10884389637265(each), 40884389637266(case), Lot Number: 22LBS348; 190) DYNJ32809I, UDI-DI: 10193489448795(each), 40193489448796(case), Lot Number: 22IBP091; 191) DYNJ33585F, UDI-DI: 10195327469528(each), 40195327469529(case), Lot Number: 23JDA407; 192) DYNJ33586D, UDI-DI: 10195327223366(each), 40195327223367(case), Lot Number: 23JDA459; 193) DYNJ41169, UDI-DI: 10884389906699(each), 40884389906690(case), Lot Number: 22LBR515; 194) DYNJ45175B, UDI-DI: 10889942485215(each), 40889942485216(case), Lot Number: 24HLA032; 195) DYNJ47481C, UDI-DI: 10889942544028(each), 40889942544029(case), Lot Number: 21BBP023; 196) DYNJ47882F, UDI-DI: 10195327319472(each), 40195327319473(case), Lot Number: 23HME434; 197) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23CBR177; 198) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23CBT206; 199) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23GBA277; 200) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23HBJ541; 201) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23JBA900; 202) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23LBO079; 203) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 23LBS232; 204) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 24ABR658; 205) DYNJ48401B, UDI-DI: 10195327173098(each), 40195327173099(case), Lot Number: 24DBG199; 206) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25ABD063; 207) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25BBT722; 208) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25CBM527; 209) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25EBQ061; 210) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25FBS319; 211) DYNJ48401C, UDI-DI: 10198459175749(each), 40198459175740(case), Lot Number: 25HBA711; 212) DYNJ51934A, UDI-DI: 10193489584776(each), 40193489584777(case), Lot Number: 23HMG725; 213) DYNJ52916A, UDI-DI: 10195327311735(each), 40195327311736(case), Lot Number: 24GMK024; 214) DYNJ54721F, UDI-DI: 10195327021290(each), 40195327021291(case), Lot Number: 23EBG645; 215) DYNJ54757B, UDI-DI: 10195327021085(each), 40195327021086(case), Lot Number: 23FBQ117; 216) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 24IBV894; 217) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 24LBN788; 218) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25BBM674; 219) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25DBB366; 220) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25EBB513; 221) DYNJ54964A, UDI-DI: 10198459033483(each), 40198459033484(case), Lot Number: 25EBV663; 222) DYNJ54964B, UDI-DI: 10198459428814(each), 40198459428815(case), Lot Number: 25HBR203; 223) DYNJ54964B, UDI-DI: 10198459428814(each), 40198459428815(case), Lot Number: 25KBH303; 224) DYNJ55350G, UDI-DI: 10195327173753(each), 40195327173754(case), Lot Number: 24CBT148; 225) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 21IBQ696; 226) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 21IBQ699; 227) DYNJ56840B, UDI-DI: 10193489416077(each), 40193489416078(case), Lot Number: 22GBJ777; 228) DYNJ57200C, UDI-DI: 10198459306754(each), 40198459306755(case), Lot Number: 25DMF322; 229) DYNJ57200C, UDI-DI: 10198459306754(each), 40198459306755(case), Lot Number: 25EMA242; 230) DYNJ57688F, UDI-DI: 10195327034610(each), 40195327034611(case), Lot Number: 22HMH769; 231) DYNJ61286C, UDI-DI: 10195327416263(each), 40195327416264(case), Lot Number: 24BBT395; 232) DYNJ61393A, UDI-DI: 10193489444179(each), 40193489444170(case), Lot Number: 23CMB523; 233) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 23HBC470; 234) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 23LBP950; 235) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 24BBI869; 236) DYNJ61935D, UDI-DI: 10195327453336(each), 40195327453337(case), Lot Number: 24BBK459; 237) DYNJ62093B, UDI-DI: 10193489969351(each), 40193489969352(case), Lot Number: 23ABE410; 238) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24ABK239; 239) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24DBV691; 240) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24EBS530; 241) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24GBM556; 242) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24HBO119; 243) DYNJ64024B, UDI-DI: 10195327556617(each), 40195327556618(case), Lot Number: 24KBS794; 244) DYNJ66509A, UDI-DI: 10193489462708(each), 40193489462709(case), Lot Number: 24HMA316; 245) DYNJ67208, UDI-DI: 10193489391671(each), 40193489391672(case), Lot Number: 20XBA769; 246) DYNJ67328B, UDI-DI: 10195327271503(each), 40195327271504(case), Lot Number: 23HMD036; 247) DYNJ69746B, UDI-DI: 10195327475222(each), 40195327475223(case), Lot Number: 24JLA247; 248) DYNJ81280A, UDI-DI: 10195327491284(each), 40195327491285(case), Lot Number: 23JME498; 249) DYNJ81280A, UDI-DI: 10195327491284(each), 40195327491285(case), Lot Number: 25LME467; 250) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24CBF775; 251) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24CBI987; 252) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24DBR594; 253) DYNJ84258A, UDI-DI: 10195327650742(each), 40195327650743(case), Lot Number: 24HBK264; 254) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 24IBD271; 255) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 24IBP146; 256) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25CBA913; 257) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25DBO402; 258) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25EBG049; 259) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25FBL492; 260) DYNJ84258B, UDI-DI: 10198459117985(each), 40198459117986(case), Lot Number: 25HBI876; 261) DYNJ85078A, UDI-DI: 10195327532543(each), 40195327532544(case), Lot Number: 24HMA736; 262) DYNJ85340, UDI-DI: 10195327491239(each), 40195327491230(case), Lot Number: 23KMC008; 263) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24DBN706; 264) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24GBQ754; 265) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24GBQ919; 266) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24IBC003; 267) DYNJ86194A, UDI-DI: 10195327682293(each), 40195327682294(case), Lot Number: 24IBR498; 268) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24CBG395; 269) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24HBV552; 270) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 24IBL260; 271) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 25CBG746; 272) DYNJ86235, UDI-DI: 10195327606534(each), 40195327606535(case), Lot Number: 25LMD475; 273) DYNJ87378, UDI-DI: 10198459034787(each), 40198459034788(case), Lot Number: 24GMG790; 274) DYNJ87378, UDI-DI: 10198459034787(each), 40198459034788(case), Lot Number: 24HMA738; 275) DYNJ87378A, UDI-DI: 10198459186394(each), 40198459186395(case), Lot Number: 24KMI285; 276) DYNJ87378A, UDI-DI: 10198459186394(each), 40198459186395(case), Lot Number: 25EMB836; 277) DYNJ87378C, UDI-DI: 10198459390647(each), 40198459390648(case), Lot Number: 25GME299; 278) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25GMJ757; 279) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25IMA065; 280) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25IMC112; 281) DYNJ87378D, UDI-DI: 10198459467257(each), 40198459467258(case), Lot Number: 25JMI380; 282) DYNJ87451, UDI-DI: 10198459040283(each), 40198459040284(case), Lot Number: 24HBH452; 283) DYNJ88498, UDI-DI: 10198459166181(each), 40198459166182(case), Lot Number: 24KMC641; 284) DYNJ88498, UDI-DI: 10198459166181(each), 40198459166182(case), Lot Number: 24KMF610; 285) DYNJ901306D, UDI-DI: 10889942564736(each), 40889942564737(case), Lot Number: 23IMB807; 286) DYNJ902993F, UDI-DI: 10193489490817(each), 40193489490818(case), Lot Number: 23IMA480; 287) DYNJ902993F, UDI-DI: 10193489490817(each), 40193489490818(case), Lot Number: 23IMD509; 288) DYNJ905866, UDI-DI: 10193489547184(each), 40193489547185(case), Lot Number: 23CBW319; 289) DYNJ906292B, UDI-DI: 10193489877588(each), 40193489877589(case), Lot Number: 24BBV224; 290) DYNJ906292B, UDI-DI: 10193489877588(each), 40193489877589(case), Lot Number: 24CBI703; 291) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21GBG669; 292) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21GBR282; 293) DYNJ906841A, UDI-DI: 10193489392197(each), 40193489392198(case), Lot Number: 21HBN905; 294) DYNJ907739A, UDI-DI: 10193489976991(each), 40193489976992(case), Lot Number: 23GBW213; 295) DYNJ907739A, UDI-DI: 10193489976991(each), 40193489976992(case), Lot Number: 23IBK879; 296) DYNJ907739A, UDI-DI: 10193489976991(each), 40193489976992(case), Lot Number: 24CBA406; 297) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25ABE814; 298) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25ABP522; 299) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25CBE654; 300) DYNJ907739B, UDI-DI: 10195327671778(each), 40195327671779(case), Lot Number: 25JBR114; 301) DYNJ908953, UDI-DI: 10195327215293(each), 40195327215294(case), Lot Number: 23KME667; 302) DYNJ911062, UDI-DI: 10198459248139(each), 40198459248130(case), Lot Number: 25BBT142; 303) DYNJ911062, UDI-DI: 10198459248139(each), 40198459248130(case), Lot Number: 25FBI843; 304) DYNJ911062, UDI-DI: 10198459248139(each), 40198459248130(case), Lot Number: 25GBI875; 305) DYNJ911623, UDI-DI: 10198459441097(each), 40198459441098(case), Lot Number: 25HBH612; 306) DYNJT5739, UDI-DI: 10198459499661(each), 40198459499662(case), Lot Number: 25IMF120; 307) PHS393087007C, UDI-DI: 10193489222777(each), 40193489222778(case), Lot Number: 21ADA448
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2097-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | May 13, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 11575 units |