See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS recall (Medline Industries, LP)

Medical deviceClass IIOngoing

Summary: Medline Industries, LP is recalling See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABD. Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What is being recalled

See RES for complete list. Medline Convenience Kits: 1) PLASTIC BIN W/MISC KITS, Model Number: ACC010499; 2) BREAST-HERNIA-PORT CDS-LF, Model Number: CDS984853J; 3) KIT UROLOGY URETHROPLASTY, Model Number: DYKMBNDL117C; 4) KIT PLASTIC INSERT TISSUE EXPA, Model Number: DYKMBNDL158; 5) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170; 6) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170A; 7) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL170B; 8) KIT PLASTIC ABDOMINOPLASTY, Model Number: DYKMBNDL182;

Why

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

How serious is it

Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.

Where it was sold

Worldwide - US Nationwide distribution.

How to tell if yours is affected

1) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21DDA661; 2) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21FDA888; 3) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21GDB740; 4) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21HDB434; 5) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 21JDB530; 6) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23HDB786; 7) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23IDB182; 8) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 23LDC086; 9) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24IDA403; 10) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 24JDA650; 11) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25ADB705; 12) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDA228; 13) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25BDB702; 14) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25FDA013; 15) ACC010499, UDI-DI: 10193489845709(each), 40193489845700(case), Lot Number: 25GDB671; 16) CDS984853J, UDI-DI: 10195327694180(each), 40195327694181(case), Lot Number: 24DDC169; 17) CDS984853J, UDI-DI: 10195327694180(each), 40195327694181(case), Lot Number: 24EDA131; 18) DYKMBNDL117C, UDI-DI: 10195327371104(each), 40195327371105(case), Lot Number: 23DLA834; 19) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21ABI978; 20) DYKMBNDL158, UDI-DI: 10193489478266(each), 40193489478267(case), Lot Number: 21CBE548; 21) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 20LBY951; 22) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21ABJ202; 23) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21BBH637; 24) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21DBN049; 25) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 21IBV640; 26) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22EBA143; 27) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22HBN456; 28) DYKMBNDL170, UDI-DI: 10193489477450(each), 40193489477451(case), Lot Number: 22IBC515; 29) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 22JBM366; 30) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23GBR888; 31) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBB877; 32) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23HBG507; 33) DYKMBNDL170A, UDI-DI: 10195327240813(each), 40195327240814(case), Lot Number: 23IBB851; 34) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBD539; 35) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24FBL546; 36) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 24IBB931; 37) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBA340; 38) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25BBN216; 39) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBK034; 40) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25DBS731; 41) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25EBM448; 42) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25GBA895; 43) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25HBP137; 44) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25KBR870; 45) DYKMBNDL170B, UDI-DI: 10195327633783(each), 40195327633784(case), Lot Number: 25LBM160; 46) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABI990; 47) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21ABX025; 48) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 21DBC569; 49) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22BBN306; 50) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22EBO665; 51) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 22FBC879; 52) DYKMBNDL182, UDI-DI: 10193489485370(each), 40193489485371(case), Lot Number: 23LBI220; 53) DYKMBNDL182A, UDI-DI: 10195327633806(each), 40195327633807(case), Lot Number: 24BBN507; 54) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24DBR913; 55) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBL678; 56) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24IBQ469; 57) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 24KBS891; 58) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25ABL533; 59) DYKMBNDL182B, UDI-DI: 10198459000478(each), 40198459000479(case), Lot Number: 25FBO935; 60) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21ABI992; 61) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21CBG305; 62) DYKMBNDL183, UDI-DI: 10193489478495(each), 40193489478496(case), Lot Number: 21JBJ252; 63) DYKMBNDL183A, UDI-DI: 10195327246594(each), 40195327246595(case), Lot Number: 23BBI487; 64) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBJ591; 65) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24DBT970; 66) DYKMBNDL183B, UDI-DI: 10195327633813(each), 40195327633814(case), Lot Number: 24GBC829; 67) DYKMBNDL34A, UDI-DI: 10193489872033(each), 40193489872034(case), Lot Number: 21DBQ512; 68) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21ABJ002; 69) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21CBD335; 70) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21GBR352; 71) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 21LBJ742; 72) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22BBU074; 73) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22GBM861; 74) DYKMBNDL39, UDI-DI: 10193489477764(each), 40193489477765(case), Lot Number: 22IBC434; 75) DYKMBNDL39A, UDI-DI: 10195327241308(each), 40195327241309(case), Lot Number: 23ABH677; 76) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 24JBO020; 77) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25GBW751; 78) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25HBK505; 79) DYKMBNDL39C, UDI-DI: 10198459112461(each), 40198459112462(case), Lot Number: 25HBQ024; 80) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21ABL308; 81) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21BBM839; 82) DYKMBNDL46, UDI-DI: 10193489478761(each), 40193489478762(case), Lot Number: 21GBR353; 83) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 20LBY967; 84) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABJ589; 85) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABM089; 86) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21ABO607; 87) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21BBH629; 88) DYKMBNDL54, UDI-DI: 10193489478822(each), 40193489478823(case), Lot Number: 21CBG791; 89) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABJ013; 90) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21ABL345; 91) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21DBC571; 92) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21GBR324; 93) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBK340; 94) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBM983; 95) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21HBO272; 96) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21IBV702; 97) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 21JBS176; 98) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBN457; 99) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22HBQ100; 100) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22IBR235; 101) DYKMBNDL69, UDI-DI: 10193489477726(each), 40193489477727(case), Lot Number: 22JBC409; 102) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC618; 103) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBC626; 104) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBJ150; 105) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22JBQ852; 106) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 22KBO876; 107) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23FBP246; 108) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23GBL960; 109) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23JBJ695; 110) DYKMBNDL69A, UDI-DI: 10195327241155(each), 40195327241156(case), Lot Number: 23LBA275; 111) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24EBR715; 112) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24GBF481; 113) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24HBM099; 114) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 24KBT018; 115) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25FBH462; 116) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25GBL984; 117) DYKMBNDL69B, UDI-DI: 10195327633653(each), 40195327633654(case), Lot Number: 25KBP763; 118) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 20XBB257; 119) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABJ216; 120) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21ABX017; 121) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21BBK531; 122) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21DBC549; 123) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBF907; 124) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21FBR398; 125) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21HBK310; 126) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBB097; 127) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21IBV643; 128) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21JBJ751; 129) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21KBY757; 130) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBO572; 131) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 21LBY278; 132) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABG018; 133) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22ABL251; 134) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22BBL848; 135) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22CBX537; 136) DYKMBNDL78, UDI-DI: 10193489478044(each), 40193489478045(case), Lot Number: 22EBS816; 137) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ019; 138) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBJ143; 139) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22JBQ856; 140) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 22KBO921; 141) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBO451; 142) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23GBP831; 143) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23JBN862; 144) DYKMBNDL78A, UDI-DI: 10195327241179(each), 40195327241170(case), Lot Number: 23LBE774; 145) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBF892; 146) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24CBR114; 147) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBE953; 148) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24EBU597; 149) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24HBD157; 150) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 24LBL035; 151) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBE752; 152) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBO685; 153) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25BBU089; 154) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25CBG915; 155) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBE325; 156) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25EBM007; 157) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25FBP527; 158) DYKMBNDL78B, UDI-DI: 10195327633684(each), 40195327633685(case), Lot Number: 25HBS118; 159) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABM088; 160) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21ABU822; 161) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21CBM835; 162) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21EBH614; 163) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBF538; 164) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21GBR338; 165) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21IBT011; 166) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21JBH111; 167) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBD628; 168) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 21LBH705; 169) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBF307; 170) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22FBT013; 171) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22HBO907; 172) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22IBC435; 173) DYKMBNDL84, UDI-DI: 10193489477801(each), 40193489477802(case), Lot Number: 22OBH455; 174) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24IBE175; 175) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24JBE336; 176) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBI135; 177) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 24LBS883; 178) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25DBO560; 179) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25FBB154; 180) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25GBF486; 181) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25HBG553; 182) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBP864; 183) DYKMBNDL84C, UDI-DI: 10198459112454(each), 40198459112455(case), Lot Number: 25LBQ326; 184) DYNJ0785310G, UDI-DI: 10193489268966(each), 40193489268967(case), Lot Number: 23KMH075; 185) DYNJ0854650J, UDI-DI: 10193489832426(each), 40193489832427(case), Lot Number: 24IMF274; 186) DYNJ36469I, UDI-DI: 10195327513931(each), 40195327513932(case), Lot Number: 24BBU341; 187) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21ADA533; 188) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21BDB779; 189) DYNJ38988J, UDI-DI: 10193489315233(each), 40193489315234(case), Lot Number: 21CDB475; 190) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22GMF684; 191) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22HMH034; 192) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 22IMF713; 193) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23AME665; 194) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23HMA900; 195) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23IMG016; 196) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23KMA468; 197) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 23LMB124; 198) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24AMI035; 199) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24CMD088; 200) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24DMJ046; 201) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24GMA471; 202) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24IME867; 203) DYNJ39831F, UDI-DI: 10195327176396(each), 40195327176397(case), Lot Number: 24KMD041; 204) DYNJ44852K, UDI-DI: 10195327127589(each), 40195327127580(case), Lot Number: 22CMI118; 205) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC374; 206) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23AMC530; 207) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23CMF343; 208) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23DMD096; 209) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23FMH424; 210) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 23HME689; 211) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24AMB315; 212) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24CMC758; 213) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24GMI186; 214) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 24KME600; 215) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25CME690; 216) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25EME929; 217) DYNJ44852M, UDI-DI: 10195327325220(each), 40195327325221(case), Lot Number: 25HMH591; 218) DYNJ50957G, UDI-DI: 10193489448863(each), 40193489448864(case), Lot Number: 23GBV065; 219) DYNJ55358G, UDI-DI: 10195327173500(each), 40195327173501(case), Lot Number: 23FBN323; 220) DYNJ56737A, UDI-DI: 10193489395594(each), 40193489395595(case), Lot Number: 23CME388; 221) DYNJ58334D, UDI-DI: 10195327515225(each), 40195327515226(case), Lot Number: 24ABU081; 222) DYNJ60210A, UDI-DI: 10193489985108(each), 40193489985109(case), Lot Number: 21KBP457; 223) DYNJ64927F, UDI-DI: 10195327143091(each), 40195327143092(case), Lot Number: 23FBV335; 224) DYNJ68844A, UDI-DI: 10195327216504(each), 40195327216505(case), Lot Number: 23HMB858; 225) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 21LBP376; 226) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBD661; 227) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23BBP453; 228) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23CBG820; 229) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23HBY228; 230) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23JBU455; 231) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 23KBN973; 232) DYNJ69351, UDI-DI: 10193489934212(each), 40193489934213(case), Lot Number: 24BBL537; 233) DYNJ82998B, UDI-DI: 10195327377861(each), 40195327377862(case), Lot Number: 24AMB126; 234) DYNJ83788, UDI-DI: 10195327305833(each), 40195327305834(case), Lot Number: 23CBO212; 235) DYNJ83884B, UDI-DI: 10198459234620(each), 40198459234621(case), Lot Number: 25GBO120; 236) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24EBR736; 237) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24FBL440; 238) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 24JBV233; 239) DYNJ83950A, UDI-DI: 10198459018510(each), 40198459018511(case), Lot Number: 25ABF915; 240) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25ABT952; 241) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25CBF253; 242) DYNJ83950B, UDI-DI: 10198459257827(each), 40198459257828(case), Lot Number: 25DBT195; 243) DYNJ83950C, UDI-DI: 10198459363948(each), 40198459363949(case), Lot Number: 25HBM610; 244) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 24CBQ297; 245) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25ABJ747; 246) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25BBD646; 247) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25DBG170; 248) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25GBD859; 249) DYNJ86165, UDI-DI: 10195327599324(each), 40195327599325(case), Lot Number: 25HBS096; 250) DYNJ86816, UDI-DI: 10195327679903(each), 40195327679904(case), Lot Number: 24FMB609; 251) DYNJ87444, UDI-DI: 10198459040122(each), 40198459040123(case), Lot Number: 24GBQ988; 252) DYNJ902005I, UDI-DI: 10195327347710(each), 40195327347711(case), Lot Number: 24DLA844; 253) DYNJ902558G, UDI-DI: 10193489728859(each), 40193489728850(case), Lot Number: 23IMC207; 254) DYNJ902558G, UDI-DI: 10193489728859(each), 40193489728850(case), Lot Number: 23HMG989; 255) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22GBT713; 256) DYNJ903555D, UDI-DI: 10195327084615(each), 40195327084616(case), Lot Number: 22HBF672; 257) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23FBN475; 258) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBG876; 259) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP518; 260) DYNJ903555F, UDI-DI: 10195327224585(each), 40195327224586(case), Lot Number: 23GBP519; 261) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 23DBS608; 262) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 23HBQ462; 263) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24HBK601; 264) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24IBQ524; 265) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24JBQ211; 266) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24JBU528; 267) DYNJ903849C, UDI-DI: 10195327333355(each), 40195327333356(case), Lot Number: 24KBB289; 268) DYNJ907847A, UDI-DI: 10193489937589(each), 40193489937580(case), Lot Number: 23IMF989; 269) DYNJ907874A, UDI-DI: 10195327241247(each), 40195327241248(case), Lot Number: 23FLA249; 270) DYNJ910457A, UDI-DI: 10198459262272(each), 40198459262273(case), Lot Number: 25GMJ928; 271) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24GME383; 272) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24IMF741; 273) DYNJ910462, UDI-DI: 10195327695637(each), 40195327695638(case), Lot Number: 24JMD608; 274) DYNJT3283, UDI-DI: 10198459323645(each), 40198459323646(case), Lot Number: 25JBI499; 275) DYNJT3417, UDI-DI: 10198459324888(each), 40198459324889(case), Lot Number: 25EBN203; 276) DYNJVB91096B, UDI-DI: 10195327257156(each), 40195327257157(case), Lot Number: 24IDA577; 277) PHS806973G, UDI-DI: 10193489926996(each), 40193489926997(case), Lot Number: 23EBQ702

What to do

Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.

Recall facts

Recall numberZ-2099-2026
CategoryMedical device
AgencyU.S. Food and Drug Administration (FDA)
SeverityClass II
StatusOngoing
Recall startedJanuary 7, 2026
ReportedMay 13, 2026
CompanyMedline Industries, LP
Company locationNorthfield, IL, United States
Initiated byVoluntary: Firm initiated
Quantity5497 units