See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS recall (Medline Industries, LP)
Medical deviceClass IIOngoing
Summary: Medline Industries, LP is recalling See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, . Reason: Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What is being recalled
See RES for complete list. Medline Convenience Kits: 1) SPINAL CDS, Model Number: CDS940087AG; 2) SPINAL CDS, Model Number: CDS940087AI; 3) TOTAL HIP CDS-LF, Model Number: CDS980832Q; 4) TOTAL HIP CDS-LF, Model Number: CDS980832R; 5) TOTAL HIP CDS-LF, Model Number: CDS980832S; 6) TOTAL KNEE CDS-LF, Model Number: CDS980833Q; 7) TOTAL KNEE CDS-LF, Model Number: CDS980833R; 8) TOTAL HIP PROCEDURE, Model Number: CDS980865V; 9) TOTAL HIP PROCEDURE, Model Number: CDS980865W;
Why
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
How serious is it
Class II is the middle level: using or being exposed to this product may cause temporary or medically reversible health problems, and the chance of serious harm is remote.
Where it was sold
Worldwide - US Nationwide distribution.
How to tell if yours is affected
1) CDS940087AG, UDI-DI: 10193489847260(each), 40193489847261(case), Lot Number: 21EBC900; 2) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21GBI183; 3) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21IBJ260; 4) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21JBF936; 5) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 21KBH163; 6) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22BBY060; 7) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBA274; 8) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBF695; 9) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22EBG548; 10) CDS940087AI, UDI-DI: 10193489933758(each), 40193489933759(case), Lot Number: 22GBE442; 11) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21BBL223; 12) CDS980832Q, UDI-DI: 10193489788655(each), 40193489788656(case), Lot Number: 21FBS254; 13) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21GBR895; 14) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBU197; 15) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21IBV695; 16) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBB625; 17) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21KBQ650; 18) CDS980832R, UDI-DI: 10193489900972(each), 40193489900973(case), Lot Number: 21LBO219; 19) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK858; 20) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBK859; 21) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBL549; 22) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22GBM598; 23) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22HBS531; 24) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22JBH355; 25) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBI093; 26) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 22OBK666; 27) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBJ844; 28) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23BBK758; 29) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23CBG614; 30) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23DBR887; 31) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23EBW653; 32) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23GBP400; 33) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23IBF278; 34) CDS980832S, UDI-DI: 10195327064495(each), 40195327064496(case), Lot Number: 23JBI054; 35) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBB808; 36) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21GBV804; 37) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21JBO077; 38) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21KBF393; 39) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBC616; 40) CDS980833Q, UDI-DI: 10193489900989(each), 40193489900980(case), Lot Number: 21LBL968; 41) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 21LBW480; 42) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22DBN516; 43) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22FBD583; 44) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22GBS172; 45) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22GBS173; 46) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 22IBE748; 47) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABG351; 48) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23ABH483; 49) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23BBP326; 50) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23CBT556; 51) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23EBK295; 52) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23HBB398; 53) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23HBV259; 54) CDS980833R, UDI-DI: 10195327064501(each), 40195327064502(case), Lot Number: 23IBA514; 55) CDS980865V, UDI-DI: 10193489304206(each), 40193489304207(case), Lot Number: 21ABX303; 56) CDS980865V, UDI-DI: 10193489304206(each), 40193489304207(case), Lot Number: 21FBH003; 57) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21IBP619; 58) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21JBA988; 59) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21JBA989; 60) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 21VBC540; 61) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22DBD515; 62) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22FBO257; 63) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22GBM603; 64) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22IMH323; 65) CDS980865W, UDI-DI: 10193489929157(each), 40193489929158(case), Lot Number: 22OBL581; 66) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23CMA547; 67) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23DMB940; 68) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23DMF336; 69) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23EME858; 70) CDS980865X, UDI-DI: 10195327243135(each), 40195327243136(case), Lot Number: 23FMI799; 71) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23HMF744; 72) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23IMA365; 73) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 23KMH550; 74) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24AMG267; 75) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24CMC328; 76) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24EMC301; 77) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24FMG472; 78) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24HMH806; 79) CDS980865Y, UDI-DI: 10195327446529(each), 40195327446520(case), Lot Number: 24IMH073; 80) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23CMB403; 81) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23CMF017; 82) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23DMH779; 83) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23EMG643; 84) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 23FME101; 85) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24AMB397; 86) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24BMF152; 87) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24CMC092; 88) CDS981454K, UDI-DI: 10195327310455(each), 40195327310456(case), Lot Number: 24DMH818; 89) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24FMG576; 90) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24GMI703; 91) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24JMB660; 92) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24KMB678; 93) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 24LMC844; 94) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 25AMJ535; 95) CDS981454L, UDI-DI: 10195327689445(each), 40195327689446(case), Lot Number: 25BMJ035; 96) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 23JMI916; 97) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 24CMF795; 98) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 24GMI705; 99) CDS981502K, UDI-DI: 10195327295370(each), 40195327295371(case), Lot Number: 25BMH022; 100) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22ABS105; 101) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22FBS293; 102) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22HBN671; 103) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22HBQ624; 104) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22IBJ349; 105) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22JBK515; 106) CDS981505O, UDI-DI: 10195327071202(each), 40195327071203(case), Lot Number: 22KBM385; 107) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23ABG957; 108) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23CBI326; 109) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23CBU775; 110) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23DBL133; 111) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23GBG193; 112) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23IBV136; 113) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23KBQ737; 114) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 23KBQ766; 115) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 24ABU182; 116) CDS981505P, UDI-DI: 10195327286231(each), 40195327286232(case), Lot Number: 24BBH054; 117) CDS981505Q, UDI-DI: 10198459028991(each), 40198459028992(case), Lot Number: 24GBJ138; 118) CDS981505Q, UDI-DI: 10198459028991(each), 40198459028992(case), Lot Number: 24IBD132; 119) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 24KBP956; 120) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 24LBR473; 121) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 25BBD996; 122) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 25DBO814; 123) CDS981505R, UDI-DI: 10198459177361(each), 40198459177362(case), Lot Number: 25FBS513; 124) CDS981858F, UDI-DI: 10889942973682(each), 40889942973683(case), Lot Number: 21ABH636; 125) CDS982035W, UDI-DI: 10193489999259(each), 40193489999250(case), Lot Number: 22GBI202; 126) CDS982035W, UDI-DI: 10193489999259(each), 40193489999250(case), Lot Number: 22HBP524; 127) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 22CME430; 128) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 22LMB455; 129) CDS982281O, UDI-DI: 10888277787223(each), 40888277787224(case), Lot Number: 23AMB059; 130) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 23HME111; 131) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 23IMF292; 132) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24AMD955; 133) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24CMB637; 134) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24DMI184; 135) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24EMD974; 136) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24FMA426; 137) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24HMD397; 138) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 24KMD431; 139) CDS982281P, UDI-DI: 10195327472801(each), 40195327472802(case), Lot Number: 25BMC093; 140) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25KMJ877; 141) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25LMD470; 142) CDS982281S, UDI-DI: 10198459561238(each), 40198459561239(case), Lot Number: 25LME952; 143) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24CDA088; 144) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24CDB619; 145) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24FDB730; 146) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24IDA960; 147) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24JDA511; 148) CDS982559K, UDI-DI: 10195327624125(each), 40195327624126(case), Lot Number: 24KDA429; 149) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 24GDB706; 150) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 24HDB820; 151) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 24IDB388; 152) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 24IDB716; 153) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 25ADB572; 154) CDS982561K, UDI-DI: 10195327624491(each), 40195327624492(case), Lot Number: 25BDA748; 155) CDS983036K, UDI-DI: 10193489842012(each), 40193489842013(case), Lot Number: 23KBN289; 156) CDS983042K, UDI-DI: 10193489842029(each), 40193489842020(case), Lot Number: 23KBN273; 157) CDS983049K, UDI-DI: 10193489841657(each), 40193489841658(case), Lot Number: 23JBX571; 158) CDS983049K, UDI-DI: 10193489841657(each), 40193489841658(case), Lot Number: 23JBY028; 159) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 20LBW136; 160) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21CBT358; 161) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21DBH504; 162) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21HBK744; 163) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21JBD011; 164) CDS983182G, UDI-DI: 10193489362480(each), 40193489362481(case), Lot Number: 21JBY974; 165) CDS983297M, UDI-DI: 10193489425857(each), 40193489425858(case), Lot Number: 21FDA407; 166) CDS983344J, UDI-DI: 10195327416171(each), 40195327416172(case), Lot Number: 23HBE028; 167) CDS983344J, UDI-DI: 10195327416171(each), 40195327416172(case), Lot Number: 23IBG219; 168) CDS983480G, UDI-DI: 10195327497842(each), 40195327497843(case), Lot Number: 24CDB716; 169) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 23IMA904; 170) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 24AME511; 171) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 24CMA539; 172) CDS983581K, UDI-DI: 10195327295448(each), 40195327295449(case), Lot Number: 25BMI188; 173) CDS983581L, UDI-DI: 10198459282263(each), 40198459282264(case), Lot Number: 25EMG021; 174) CDS983581L, UDI-DI: 10198459282263(each), 40198459282264(case), Lot Number: 25FMI149; 175) CDS983611O, UDI-DI: 10198459154034(each), 40198459154035(case), Lot Number: 24LLA763; 176) CDS983611O, UDI-DI: 10198459154034(each), 40198459154035(case), Lot Number: 25BLA570; 177) CDS983611O, UDI-DI: 10198459154034(each), 40198459154035(case), Lot Number: 25DLA165; 178) CDS983611O, UDI-DI: 10198459154034(each), 40198459154035(case), Lot Number: 25FLA484; 179) CDS983611O, UDI-DI: 10198459154034(each), 40198459154035(case), Lot Number: 25GLB024; 180) CDS983838G, UDI-DI: 10193489600360(each), 40193489600361(case), Lot Number: 21ADB812; 181) CDS983840I, UDI-DI: 10193489405668(each), 40193489405669(case), Lot Number: 21FDA409; 182) CDS983863D, UDI-DI: 10889942570980(each), 40889942570981(case), Lot Number: 22JDB465; 183) CDS983863D, UDI-DI: 10889942570980(each), 40889942570981(case), Lot Number: 22KDB188; 184) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21CBC111; 185) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21CBF370; 186) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21DBU793; 187) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21IBA056; 188) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 21JBK128; 189) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22ABP297; 190) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22ABS153; 191) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22EBV412; 192) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22HBT635; 193) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22KBS560; 194) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 22LBO557; 195) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 23ABM823; 196) CDS983879C, UDI-DI: 10193489435054(each), 40193489435055(case), Lot Number: 23EBK642; 197) CDS983903D, UDI-DI: 10198459053474(each), 40198459053475(case), Lot Number: 24HBQ596; 198) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21CBJ921; 199) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21DBI790; 200) CDS984147K, UDI-DI: 10193489245424(each), 40193489245425(case), Lot Number: 21EBR750; 201) CDS984156L, UDI-DI: 10193489245370(each), 40193489245371(case), Lot Number: 21CBJ888; 202) CDS984156L, UDI-DI: 10193489245370(each), 40193489245371(case), Lot Number: 21DBQ627; 203) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21CBD679; 204) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21FBL924; 205) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21GBU901; 206) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21HBM083; 207) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 21KBY807; 208) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22BBX356; 209) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22CBW600; 210) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22EBJ581; 211) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22GBO260; 212) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22HBP143; 213) CDS984254K, UDI-DI: 10193489289091(each), 40193489289092(case), Lot Number: 22KBP500; 214) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23ABK592; 215) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23BBS246; 216) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23CBR791; 217) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23DBD812; 218) CDS984254L, UDI-DI: 10195327298159(each), 40195327298150(case), Lot Number: 23EBJ573; 219) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 23HBD301; 220) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 23HBV822; 221) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 23JBT902; 222) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24ABT691; 223) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24BBS781; 224) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24DBS099; 225) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24EBF862; 226) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24GBB449; 227) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24GBU116; 228) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24IBB732; 229) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24IBP093; 230) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 24KBJ956; 231) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25BBI005; 232) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25CBE421; 233) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25DBL471; 234) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25FBD448; 235) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25GBG607; 236) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25GBT220; 237) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25IBW208; 238) CDS984254M, UDI-DI: 10195327415242(each), 40195327415243(case), Lot Number: 25KBK734; 239) CDS984346Q, UDI-DI: 10195327404383(each), 40195327404384(case), Lot Number: 23EBQ808; 240) CDS984500K, UDI-DI: 10193489655551(each), 40193489655552(case), Lot Number: 23BBU289; 241) CDS984500K, UDI-DI: 10193489655551(each), 40193489655552(case), Lot Number: 23CBA146; 242) CDS984500K, UDI-DI: 10193489655551(each), 40193489655552(case), Lot Number: 23CBX444; 243) CDS984527I, UDI-DI: 10193489376739(each), 40193489376730(case), Lot Number: 21BDA355; 244) CDS984607G, UDI-DI: 10889942719389(each), 40889942719380(case), Lot Number: 21ABV551; 245) CDS984607G, UDI-DI: 10889942719389(each), 40889942719380(case), Lot Number: 21EBC029; 246) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21IBK997; 247) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21JBM719; 248) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21LBE361; 249) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 21VBC530; 250) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22CBP419; 251) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBQ393; 252) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBR470; 253) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22EBS549; 254) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22GBF580; 255) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22HBJ382; 256) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22IBO809; 257) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22IBP797; 258) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22JBH295; 259) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 22JBQ863; 260) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23ABA744; 261) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23ABE701; 262) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23BBU944; 263) CDS984607I, UDI-DI: 10193489959482(each), 40193489959483(case), Lot Number: 23DBK552; 264) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 21IBV618; 265) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 21LBQ977; 266) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22BBL459; 267) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22FBK619; 268) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22GBU468; 269) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22KBB895; 270) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 22OBI284; 271) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 23ABF816; 272) CDS984610C, UDI-DI: 10193489959505(each), 40193489959506(case), Lot Number: 23BBG636; 273) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 23JMI892; 274) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 23LMG325; 275) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24AME371; 276) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24BMI129; 277) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24DMD924; 278) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24EMC388; 279) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24FMA276; 280) CDS985030K, UDI-DI: 10193489377804(each), 40193489377805(case), Lot Number: 24FMF570; 281) CDS985030L, UDI-DI: 10198459046087(each), 40198459046088(case), Lot Number: 24GMJ763; 282) CDS985030L, UDI-DI: 10198459046087(each), 40198459046088(case), Lot Number: 24IMH079; 283) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 24KMD040; 284) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 24LMG952; 285) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 25CMB603; 286) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 25CMJ169; 287) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 25GMB151; 288) CDS985030M, UDI-DI: 10198459137501(each), 40198459137502(case), Lot Number: 25GMK359; 289) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 23JME572; 290) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 23KMI005; 291) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 23LMD171; 292) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 24AME376; 293) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 24CMH553; 294) CDS985031J, UDI-DI: 10193489377828(each), 40193489377829(case), Lot Number: 24FMH322; 295) CDS985031K, UDI-DI: 10198459046049(each), 40198459046040(case), Lot Number: 24HMD413; 296) CDS985031K, UDI-DI: 10198459046049(each), 40198459046040(case), Lot Number: 24IME699; 297) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 24KME783; 298) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 24LMC521; 299) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25AMF781; 300) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25BMB200; 301) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25CMD903; 302) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25EMB855; 303) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25GMC417; 304) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25GMC800; 305) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25GMH432; 306) CDS985031L, UDI-DI: 10198459137518(each), 40198459137519(case), Lot Number: 25HMH553; 307) CDS985032K, UDI-DI: 10193489377880(each), 40193489377881(case), Lot Number: 23LMC694; 308) CDS985032K, UDI-DI: 10193489377880(each), 40193489377881(case), Lot Number: 24AMB539; 309) CDS985032K, UDI-DI: 10193489377880(each), 40193489377881(case), Lot Number: 24AMJ389; 310) CDS985032K, UDI-DI: 10193489377880(each), 40193489377881(case), Lot Number: 24DMF901; 311) CDS985032K, UDI-DI: 10193489377880(each), 40193489377881(case), Lot Number: 24EMG131; 312) CDS985032L, UDI-DI: 10198459046056(each), 40198459046057(case), Lot Number: 24HMG433; 313) CDS985032L, UDI-DI: 10198459046056(each), 40198459046057(case), Lot Number: 24IMH497; 314) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 24KMA640; 315) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 24KMG999; 316) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 25BMD439; 317) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 25CMF075; 318) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 25EMD972; 319) CDS985032M, UDI-DI: 10198459137525(each), 40198459137526(case), Lot Number: 25GME758; 320) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23BBS726; 321) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23FBE479; 322) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23FBV622; 323) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23HBH071; 324) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23JBP419; 325) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23KBB103; 326) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23LBI902; 327) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 23LBV047; 328) CDS985202B, UDI-DI: 10889942942053(each), 40889942942054(case), Lot Number: 24BBB251; 329) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24DBD819; 330) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24EBR586; 331) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24JBM801; 332) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24JBU521; 333) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 24LBI139; 334) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 25BBD069; 335) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 25CBJ749; 336) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 25FBG461; 337) CDS985202C, UDI-DI: 10195327661908(each), 40195327661909(case), Lot Number: 25GBD241; 338) CDS985279K, UDI-DI: 10195327497460(each), 40195327497461(case), Lot Number: 24CDA763; 339) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21ABP208; 340) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21CBM829; 341) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21GBL257; 342) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21LBH556; 343) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 21LBI665; 344) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBB121; 345) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBM692; 346) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22FBQ646; 347) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22GBU431; 348) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22JBR384; 349) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22LBT331; 350) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 22NBE575; 351) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23ABC226; 352) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23CBP302; 353) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23DBD828; 354) CDS985316A, UDI-DI: 10888277077553(each), 40888277077554(case), Lot Number: 23GBT765; 355) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 23IBA494; 356) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 23JBS895; 357) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24ABT609; 358) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24BBO443; 359) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24DBF304; 360) CDS985317C, UDI-DI: 1019 5327477837(each), 40195327477838(case), Lot Number: 24FBN572; 361) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24GBW059; 362) CDS985317C, UDI-DI: 10195327477837(each), 40195327477838(case), Lot Number: 24IBU020; 363) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 24LBD251; 364) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 25BBD080; 365) CDS985317D, UDI-DI: 10198459152450(each), 40198459152451(case), Lot Number: 25CBV211; 366) CDS985317F, UDI-DI: 10198459309311(each), 40198459309312(case), Lot Number: 25FBK413; 367) CDS985317F, UDI-DI: 10198459309311(each), 40198459309312(case), Lot Number: 25HBD051; 368) CDS985317G, UDI-DI: 10198459509063(each), 40198459509064(case), Lot Number: 25LBC622; 369) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 21CBH306; 370) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 21CBU636; 371) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 22KBC308; 372) CDS985425J, UDI-DI: 10193489318890(each), 40193489318891(case), Lot Number: 23ABE885; 373) CDS985432N, UDI-DI: 10195327336127(each), 40195327336128(case), Lot Number: 23BBE587; 374) CDS985432O, UDI-DI: 10195327382025(each), 40195327382026(case), Lot Number: 23EBE298; 375) CDS985432O, UDI-DI: 10195327382025(each), 40195327382026(case), Lot Number: 23EBP900; 376) CDS985432O, UDI-DI: 10195327382025(each), 40195327382026(case), Lot Number: 23GBG465; 377) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 23KBS506; 378) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24CBB250; 379) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24CBN817; 380) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24DBD900; 381) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24FBI944; 382) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24GBJ137; 383) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24HBC597; 384) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24IBR801; 385) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24JBH797; 386) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 24KBE902; 387) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 25ABI525; 388) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 25CBI656; 389) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 25DBH920; 390) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 25EBI924; 391) CDS985432Q, UDI-DI: 10195327527198(each), 40195327527199(case), Lot Number: 25HBK350; 392) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21ADA639; 393) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21ADB694; 394) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 21HDC208; 395) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22EDA009; 396) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22FDB552; 397) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 22JDB343; 398) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23ADB631; 399) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23BDB741; 400) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23CDB828; 401) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 23JDB638; 402) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24ADA064; 403) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24HDA224; 404) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 24IDA344; 405) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25ADB448; 406) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25DDB938; 407) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25GDA910; 408) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25HDB068; 409) CDSOPM834W, UDI-DI: 10193489748758(each), 40193489748759(case), Lot Number: 25KDA137; 410) DYKM1161, UDI-DI: 10888277838529(each), 40888277838520(case), Lot Number: 23ABS895; 411) DYKM2106, UDI-DI: 10195327020736(each), 40195327020737(case), Lot Number: 22FBT157; 412) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21ABM611; 413) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21EBE678; 414) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21HBF503; 415) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 21HBG593; 416) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22CBB535; 417) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22DBM212; 418) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22FBF302; 419) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22FBI537; 420) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22IBS840; 421) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22JBP430; 422) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 22KBO875; 423) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24ABH888; 424) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24BBJ839; 425) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24FBB726; 426) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24GBC828; 427) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 24HBH168; 428) DYKMBNDL137, UDI-DI: 10193489478143(each), 40193489478144(case), Lot Number: 25CBT503; 429) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 22JBF639; 430) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 22JBG015; 431) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23ABK726; 432) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23DBL148; 433) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23EBL742; 434) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23FBE416; 435) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23GBJ242; 436) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23HBH773; 437) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23HBY495; 438) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23JBB704; 439) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23KBV904; 440) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 23LBO287; 441) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24GBH578; 442) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24GBV669; 443) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24IBE136; 444) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 24KBQ931; 445) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25ABQ126; 446) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25ABS680; 447) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25BBI466; 448) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25EBK537; 449) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25FBK659; 450) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25FBU008; 451) DYKMBNDL138A, UDI-DI: 10195327240691(each), 40195327240692(case), Lot Number: 25GBP632; 452) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21ABL153; 453) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21BBF813; 454) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBB270; 455) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBF326; 456) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21CBJ601; 457) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21EBM705; 458) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21FBF908; 459) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21GBJ360; 460) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21GBO489; 461) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21IBJ315; 462) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21IBK151; 463) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 21VBB503; 464) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22ABL303; 465) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22BBR667; 466) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22CBV519; 467) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBF963; 468) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBQ935; 469) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBQ939; 470) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22EBS811; 471) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22FBS797; 472) DYKMBNDL141, UDI-DI: 10193489478181(each), 40193489478182(case), Lot Number: 22HBQ237; 473) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22IBK345; 474) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22JBD221; 475) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22KBD324; 476) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 22LBL868; 477) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23FBI710; 478) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23GBP488; 479) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23IBA610; 480) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23JBK019; 481) DYKMBNDL141A, UDI-DI: 10195327247768(each), 40195327247769(case), Lot Number: 23LBA277; 482) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24GBK949; 483) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBA088; 484) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBG682; 485) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24HBV609; 486) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 24JBF929; 487) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25ABL332; 488) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25BBS248; 489) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25CBI784; 490) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25DBI356; 491) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25EBR525; 492) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25FBQ856; 493) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25GBF719; 494) DYKMBNDL141B, UDI-DI: 10198459025921(each), 40198459025922(case), Lot Number: 25HBA792; 495) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25BBH467; 496) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25EBN171; 497) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25FBC129; 498) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25HBO175; 499) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25HBP136; 500) DYKMBNDL142C, UDI-DI: 10198459252969(each), 40198459252960(case), Lot Number: 25LBL432; 501) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21ABU298; 502) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21BBN599; 503) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21CBS017; 504) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 21CBT399; 505) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 23CBJ799; 506) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 23CBN837; 507) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 23CBS209; 508) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 23CBV575; 509) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 24JBD655; 510) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 25CBF563; 511) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 25GBV808; 512) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 25HBA302; 513) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 25HBB031; 514) DYKMBNDL143, UDI-DI: 10193489482607(each), 40193489482608(case), Lot Number: 25HBK504; 515) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ABI784; 516) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ABU821; 517) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21BLA414; 518) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CBH443; 519) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA295; 520) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA414; 521) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21CLA448; 522) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21DLA128; 523) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21DLA534; 524) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21FLA052; 525) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21GLA640; 526) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21GLA871; 527) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21HLA178; 528) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21HLA660; 529) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21ILA240; 530) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21JLA461; 531) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21KLB158; 532) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21LLA431; 533) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 21LLA651; 534) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22BLA947; 535) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22CLA771; 536) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22DLA727; 537) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22DLA896; 538) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22ELB110; 539) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22FLA327; 540) DYKMBNDL144, UDI-DI: 10193489478358(each), 40193489478359(case), Lot Number: 22HLA742; 541) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22ILA386; 542) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22JLB035; 543) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 22LLA124; 544) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23BLA654; 545) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23DLA832; 546) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23FLA815; 547) DYKMBNDL144A, UDI-DI: 10195327241032(each), 40195327241033(case), Lot Number: 23GLA491; 548) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 23HLA606; 549) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 23JLB143; 550) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALA066; 551) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALA759; 552) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ALB064; 553) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24DLA205; 554) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24FLA724; 555) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24GLA978; 556) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24HLA916; 557) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24ILA591; 558) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24KLA175; 559) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 24LLA520; 560) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25ALA559; 561) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25CLA509; 562) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25DLA246; 563) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25FLA991; 564) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25GLA271; 565) DYKMBNDL144B, UDI-DI: 10195327481346(each), 40195327481347(case), Lot Number: 25HLA637; 566) DYKMBNDL145A, UDI-DI: 10193489846454(each), 40193489846455(case), Lot Number: 21CBF520; 567) DYKMBNDL145A, UDI-DI: 10193489846454(each), 40193489846455(case), Lot Number: 21DBK033; 568) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ALA530; 569) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21CLA198; 570) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21DLA040; 571) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ELA429; 572) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21GLB011; 573) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21GLB130; 574) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21HLA172; 575) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21ILA264; 576) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21JLA153; 577) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21JLA429; 578) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21KLA020; 579) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 21LLA721; 580) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22BLB218; 581) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22CLA549; 582) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22DLA064; 583) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22FLA364; 584) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22ILA954; 585) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22ILB139; 586) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 22KLA317; 587) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLA145; 588) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLA750; 589) DYKMBNDL146, UDI-DI: 10193489478372(each), 40193489478373(case), Lot Number: 23BLB013; 590) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21ABI979; 591) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21EBK914; 592) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21FBR384; 593) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 21JBM753; 594) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22ABD080; 595) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBA085; 596) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBA086; 597) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22BBR619; 598) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22CBZ658; 599) DYKMBNDL162, UDI-DI: 10193489477375(each), 40193489477376(case), Lot Number: 22EBH424; 600) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23ABV362; 601) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBH227; 602) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBH774; 603) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBN659; 604) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23HBY496; 605) DYKMBNDL162A, UDI-DI: 10195327240790(each), 40195327240791(case), Lot Number: 23KBJ631; 606) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21ABL344; 607) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21CBE515; 608) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21CBM832; 609) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21GBR322; 610) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 21JBQ867; 611) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 22BBA226; 612) DYKMBNDL18, UDI-DI: 10193489478662(each), 40193489478663(case), Lot Number: 22BBN302; 613) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 21ABM596; 614) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 21ABM598; 615) DYKMBNDL181, UDI-DI: 10193489478471(each), 40193489478472(case), Lot Number: 22GBJ697; 616) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 22KBA921; 617) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 24CBA390; 618) DYKMBNDL181A, UDI-DI: 10195327240868(each), 40195327240869(case), Lot Number: 24JBH678; 619) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21ABJ952; 620) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21BBK654; 621) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBC342; 622) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBO357; 623) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21EBR810; 624) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21GBG496; 625) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21HBD948; 626) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21HBP647; 627) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 21JBM262; 628) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22CBT349; 629) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22DBP875; 630) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22EBH625; 631) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22EBQ040; 632) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22FBX130; 633) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22GBO583; 634) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22HBM469; 635) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22JBO386; 636) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 22KBR441; 637) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23ABK213; 638) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23EBB390; 639) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23EBC184; 640) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23GBE575; 641) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23GBJ244; 642) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23IBK826; 643) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23JBY280; 644) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 23KBX885; 645) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24ABJ776; 646) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBJ797; 647) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBK954; 648) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24DBM619; 649) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24EBE950; 650) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24GBV124; 651) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBE721; 652) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBT803; 653) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 24LBU036; 654) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBA179; 655) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBA208; 656) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25EBR608; 657) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25FBF593; 658) DYKMBNDL186, UDI-DI: 10193489478525(each), 40193489478526(case), Lot Number: 25LBB412; 659) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 22JBV648; 660) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 22JBW065; 661) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23ABK212; 662) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23ABK704; 663) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23CBS133; 664) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23FBA473; 665) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23IBV210; 666) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 23KBA330; 667) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24BBO524; 668) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24CBB397; 669) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24DBB757; 670) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 24IBB922; 671) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 25CBL224; 672) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 25EBH375; 673) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 25EBI083; 674) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 25GBA267; 675) DYKMBNDL18A, UDI-DI: 10195327240851(each), 40195327240852(case), Lot Number: 25GBG592; 676) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 20LBY961; 677) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21ABI053; 678) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21ABP202; 679) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21CBM818; 680) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21DBI983; 681) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 21HBK339; 682) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22ABB447; 683) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22ABL294; 684) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22FBH375; 685) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22GBG264; 686) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22GBW967; 687) DYKMBNDL24, UDI-DI: 10193489479195(each), 40193489479196(case), Lot Number: 22HBA912; 688) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 22JBD779; 689) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 22JBM009; 690) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 23HBW431; 691) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 24EBB702; 692) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25GBA035; 693) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25GBD166; 694) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25HBP139; 695) DYKMBNDL24A, UDI-DI: 10195327240974(each), 40195327240975(case), Lot Number: 25HBU246; 696) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21ABR425; 697) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21CBZ652; 698) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21DBH744; 699) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21GBO491; 700) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 21KBA483; 701) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22ABL305; 702) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22ABO110; 703) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22GBV355; 704) DYKMBNDL25, UDI-DI: 10193489479201(each), 40193489479202(case), Lot Number: 22HBQ415; 705) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22IBD939; 706) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22JBV591; 707) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 22JBW781; 708) DYKMBNDL25A, UDI-DI: 10195327240981(each), 40195327240982(case), Lot Number: 23HBD671; 709) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21ABJ588; 710) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21BBF814; 711) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21BBH640; 712) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21DBJ864; 713) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21GBR317; 714) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(c ase), Lot Number: 21GBR325; 715) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 21JBJ745; 716) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 22HBS997; 717) DYKMBNDL41, UDI-DI: 10193489478877(each), 40193489478878(case), Lot Number: 22IBC493; 718) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 22JBV772; 719) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 22KBB454; 720) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23EBT501; 721) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23KBK074; 722) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 23LBN530; 723) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 24HBA089; 724) DYKMBNDL41A, UDI-DI: 10195327241315(each), 40195327241316(case), Lot Number: 24HBS364; 725) DYKMBNDL41B, UDI-DI: 10198459194221(each), 40198459194222(case), Lot Number: 24KBO741; 726) DYKMBNDL41B, UDI-DI: 10198459194221(each), 40198459194222(case), Lot Number: 25FBA949; 727) DYKMBNDL44, UDI-DI: 10193489478907(each), 40193489478908(case), Lot Number: 21ABI958; 728) DYKMBNDL44, UDI-DI: 10193489478907(each), 40193489478908(case), Lot Number: 21BBB228; 729) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21CBM521; 730) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21EBH613; 731) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21FBP768; 732) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21HBC599; 733) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21IBL058; 734) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21JBZ344; 735) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21KBC792; 736) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 21LBL111; 737) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22CBP632; 738) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22CBZ659; 739) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22DBS009; 740) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22EBH425; 741) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22FBS026; 742) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22GBS120; 743) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22GBW926; 744) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22HBB977; 745) DYKMBNDL44A, UDI-DI: 10193489867114(each), 40193489867115(case), Lot Number: 22HBN316; 746) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 22JBC209; 747) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 22JBM605; 748) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23ABP556; 749) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23DBD291; 750) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23DBM061; 751) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23IBP913; 752) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23JBN677; 753) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23KBA266; 754) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 23LBD748; 755) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24CBU335; 756) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24FBG799; 757) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24FBL254; 758) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24HBA816; 759) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24IBT870; 760) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBE585; 761) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBH752; 762) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24JBR177; 763) DYKMBNDL44B, UDI-DI: 10195327241049(each), 40195327241040(case), Lot Number: 24KBH583; 764) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25DBI596; 765) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25DBL691; 766) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25EBA283; 767) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25FBF961; 768) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25FBQ306; 769) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25GBM364; 770) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25HBD384; 771) DYKMBNDL44C, UDI-DI: 10198459194375(each), 40198459194376(case), Lot Number: 25HBS633; 772) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 20LBK848; 773) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21ABC436; 774) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21ABJ546; 775) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21BBN603; 776) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBF517; 777) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBM480; 778) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21CBS659; 779) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21DBG619; 780) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21FBH155; 781) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21GBA955; 782) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21GBC744; 783) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21HBE131; 784) DYKMBNDL45, UDI-DI: 10193489478914(each), 40193489478915(case), Lot Number: 21IBO778; 785) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21ABI788; 786) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21ABT642; 787) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21BBP819; 788) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21DBL007; 789) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21FBF707; 790) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21FBO536; 791) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21GBN132; 792) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21GBU931; 793) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21HBK334; 794) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21IBO769; 795) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21JBX804; 796) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 21LBL251; 797) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22ABK146; 798) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22ABL284; 799) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22CBB053; 800) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22DBB583; 801) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22DBN318; 802) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22EBV520; 803) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22GBG265; 804) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22GBW927; 805) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22HBB979; 806) DYKMBNDL57, UDI-DI: 10193489478853(each), 40193489478854(case), Lot Number: 22HBK580; 807) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 22JBW403; 808) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 22LBL894; 809) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23BBS867; 810) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23CBM230; 811) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23CBP087; 812) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 23JBK351; 813) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24EBU783; 814) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24HBT009; 815) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24LBH373; 816) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 24LBR198; 817) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25ABH726; 818) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25CBU771; 819) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25GBY473; 820) DYKMBNDL57A, UDI-DI: 10195327241087(each), 40195327241088(case), Lot Number: 25HBE319; 821) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21CBD336; 822) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21EBG521; 823) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21KBC488; 824) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 21LBD743; 825) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22ABG023; 826) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22CBX205; 827) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22DBE579; 828) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22DBM671; 829) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22EBA077; 830) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22FBC970; 831) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22GBQ603; 832) DYKMBNDL5A, UDI-DI: 10193489846485(each), 40193489846486(case), Lot Number: 22OBI934; 833) DYKS2050, UDI-DI: 10198459345104(each), 40198459345105(case), Lot Number: 25HBL690; 834) DYNJ0130759J, UDI-DI: 10195327001438(each), 40195327001439(case), Lot Number: 23BMG804; 835) DYNJ0151234D, UDI-DI: 10193489207439(each), 40193489207430(case), Lot Number: 21ADA115; 836) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 20LKB286; 837) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21CKA772; 838) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21DKA380; 839) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21DKA745; 840) DYNJ0190795O, UDI-DI: 10193489442618(each), 40193489442619(case), Lot Number: 21GBF067; 841) DYNJ0192492, UDI-DI: 10889942053728(each), 40889942053729(case), Lot Number: 21ADB354; 842) DYNJ0395166K, UDI-DI: 10195327292942(each), 40195327292943(case), Lot Number: 23DBE442; 843) DYNJ0395879J, UDI-DI: 10193489220261(each), 40193489220262(case), Lot Number: 21ADB355; 844) DYNJ0395959T, UDI-DI: 10195327206314(each), 40195327206315(case), Lot Number: 23JDA178; 845) DYNJ0415915K, UDI-DI: 10193489803105(each), 40193489803106(case), Lot Number: 23CMA970; 846) DYNJ0537209Y, UDI-DI: 10195327418229(each), 40195327418220(case), Lot Number: 23IBO050; 847) DYNJ0595455I, UDI-DI: 10193489873689(each), 40193489873680(case), Lot Number: 23KMA210; 848) DYNJ0606945I, UDI-DI: 10193489456318(each), 40193489456319(case), Lot Number: 23KMA213; 849) DYNJ0665273F, UDI-DI: 10193489394931(each), 40193489394932(case), Lot Number: 23BBL864; 850) DYNJ0665273F, UDI-DI: 10193489394931(each), 40193489394932(case), Lot Number: 23EBA055; 851) DYNJ0665273F, UDI-DI: 10193489394931(each), 40193489394932(case), Lot Number: 23EBS696; 852) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 23GMC994; 853) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 24ABS869; 854) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 24CBN254; 855) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 24EBB577; 856) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 24GBF608; 857) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 24KBB181; 858) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 25ABA245; 859) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 25ABB091; 860) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 25ABC316; 861) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 25BBA996; 862) DYNJ0665821D, UDI-DI: 10195327460976(each), 40195327460977(case), Lot Number: 25CBC843; 863) DYNJ0773815M, UDI-DI: 10193489270976(each), 40193489270977(case), Lot Number: 23JBU875; 864) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 23GME079; 865) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 23HMH357; 866) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 23LBC731; 867) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24ABQ862; 868) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24CBN277; 869) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24EBE219; 870) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24EBV627; 871) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24GBE734; 872) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24HBT575; 873) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24KBC551; 874) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 24LBS409; 875) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 25BBH582; 876) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 25CBI453; 877) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 25FBO051; 878) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 25GBP391; 879) DYNJ0800049I, UDI-DI: 10195327461010(each), 40195327461011(case), Lot Number: 25HBM137; 880) DYNJ0843201G, UDI-DI: 10193489345421(each), 40193489345422(case), Lot Number: 21ADA098; 881) DYNJ15364J, UDI-DI: 10193489707557(each), 40193489707558(case), Lot Number: 21BBD339; 882) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 20LMF531; 883) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 21CMC537; 884) DYNJ16747L, UDI-DI: 10889942748594(each), 40889942748595(case), Lot Number: 21EMD491; 885) DYNJ17452S, UDI-DI: 10193489448726(each), 40193489448727(case), Lot Number: 23CBV576; 886) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21BMA742; 887) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21CMB012; 888) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21FMB644; 889) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21GMF267; 890) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21HME696; 891) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21IMG131; 892) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 21JMH116; 893) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22AMA539; 894) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22BMB087; 895) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22DMA855; 896) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22EME207; 897) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22GMG472; 898) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22IMG491; 899) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 22LMA654; 900) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23AMD537; 901) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23DMF905; 902) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23FMF588; 903) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23GBT067; 904) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23HBP801; 905) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23IBL292; 906) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 23KBA883; 907) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24ABA326; 908) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24BBD246; 909) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24CBN315; 910) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24DBS717; 911) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24EBQ041; 912) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24FBS074; 913) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24HBO070; 914) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24IBW181; 915) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24KBA593; 916) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 24LBK214; 917) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 25BBA503; 918) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 25BBA542; 919) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 25EBS641; 920) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 25GBF549; 921) DYNJ19732D, UDI-DI: 10889942590018(each), 40889942590019(case), Lot Number: 25HBT477; 922) DYNJ19733C, UDI-DI: 10889942757794(each), 40889942757795(case), Lot Number: 22HMD917; 923) DYNJ19993O, UDI-DI: 10889942854110(each), 40889942854111(case), Lot Number: 23HBI690; 924) DYNJ21811I, UDI-DI: 10889942525959(each), 40889942525950(case), Lot Number: 23BBR769; 925) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21BBJ353; 926) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21CBY834; 927) DYNJ21826Q, UDI-DI: 10889942883349(each), 40889942883340(case), Lot Number: 21EBD225; 928) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22BBO487; 929) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22JBB065; 930) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 22JBG214; 931) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23BBE709; 932) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23BBJ010; 933) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23DBH622; 934) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23FBD594; 935) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23GBA464; 936) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23GBN955; 937) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23HBO965; 938) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23JBF848; 939) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 23LBA701; 940) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24BBF877; 941) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24CBH730; 942) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24GBF658; 943) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24GBQ410; 944) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24IBA042; 945) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 24JBS253; 946) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 25BBM063; 947) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 25DBV203; 948) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 25FBJ321; 949) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 25GBF226; 950) DYNJ21826R, UDI-DI: 10195327104412(each), 40195327104413(case), Lot Number: 25GBV110; 951) DYNJ22359D, UDI-DI: 10193489552997(each), 40193489552998(case), Lot Number: 21ADC025; 952) DYNJ22359D, UDI-DI: 10193489552997(each), 40193489552998(case), Lot Number: 21HDA942; 953) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 21EBN507; 954) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 21LBU859; 955) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 22ABM026; 956) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 22BBS420; 957) DYNJ22501I, UDI-DI: 10193489916157(each), 40193489916158(case), Lot Number: 23CBE334; 958) DYNJ22662L, UDI-DI: 10193489650808(each), 40193489650809(case), Lot Number: 21DMB128; 959) DYNJ22662L, UDI-DI: 10193489650808(each), 40193489650809(case), Lot Number: 21DMC725; 960) DYNJ22662L, UDI-DI: 10193489650808(each), 40193489650809(case), Lot Number: 21CMB744; 961) DYNJ24222Q, UDI-DI: 10195327514136(each), 40195327514137(case), Lot Number: 24CBC278; 962) DYNJ24231C, UDI-DI: 10198459357626(each), 40198459357627(case), Lot Number: 25FDB701; 963) DYNJ24231C, UDI-DI: 10198459357626(each), 40198459357627(case), Lot Number: 25HDA559; 964) DYNJ24231C, UDI-DI: 10198459357626(each), 40198459357627(case), Lot Number: 25LDB494; 965) DYNJ24636I, UDI-DI: 10193489982145(each), 40193489982146(case), Lot Number: 23CBT957; 966) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23AMB651; 967) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23AMD158; 968) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23AMJ578; 969) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23DMC135; 970) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23EMC437; 971) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23GMD733; 972) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23JMC857; 973) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23KMG970; 974) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 23LMD075; 975) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 24BME115; 976) DYNJ24714L, UDI-DI: 10193489238143(each), 40193489238144(case), Lot Number: 24EMH666; 977) DYNJ24715L, UDI-DI: 10195327165772(each), 40195327165773(case), Lot Number: 23LMD460; 978) DYNJ24716G, UDI-DI: 10193489506389(each), 40193489506380(case), Lot Number: 23DMB573; 979) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21IBK016; 980) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21JBM247; 981) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21JBX549; 982) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 21KBQ972; 983) DYNJ24906L, UDI-DI: 10193489962734(each), 40193489962735(case), Lot Number: 22BBB879; 984) DYNJ25094Q, UDI-DI: 10889942372393(each), 40889942372394(case), Lot Number: 21OBA672; 985) DYNJ25796I, UDI-DI: 10889942415670(each), 40889942415671(case), Lot Number: 22LBR363; 986) DYNJ26277A, UDI-DI: 10080196979752(each), 40080196979753(case), Lot Number: 22HMC684; 987) DYNJ26425K, UDI-DI: 10193489341034(each), 40193489341035(case), Lot Number: 21DDB922; 988) DYNJ26425K, UDI-DI: 10193489341034(each), 40193489341035(case), Lot Number: 21DDB987; 989) DYNJ27524G, UDI-DI: 10193489549010(each), 40193489549011(case), Lot Number: 23GBJ533; 990) DYNJ28570I, UDI-DI: 10888277742277(each), 40888277742278(case), Lot Number: 22IMB678; 991) DYNJ30429K, UDI-DI: 10193489281682(each), 40193489281683(case), Lot Number: 21AKA573; 992) DYNJ30429L, UDI-DI: 10193489844221(each), 40193489844222(case), Lot Number: 22HMG195; 993) DYNJ30430O, UDI-DI: 10195327091132(each), 40195327091133(case), Lot Number: 22IMH835; 994) DYNJ30493C, UDI-DI: 10193489442250(each), 40193489442251(case), Lot Number: 22HME641; 995) DYNJ30880J, UDI-DI: 10195327223748(each), 40195327223749(case), Lot Number: 22KMB747; 996) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21AMB065; 997) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21BMC333; 998) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21CMA929; 999) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21DME178; 1000) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21HMC610; 1001) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21IME223; 1002) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21JMG523; 1003) DYNJ31283L, UDI-DI: 10193489462227(each), 40193489462228(case), Lot Number: 21LMD700; 1004) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22CMC426; 1005) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23JMD197; 1006) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22AMB915; 1007) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22BME125; 1008) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22CME381; 1009) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22EMI458; 1010) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22HMB139; 1011) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22HMG533; 1012) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22IME048; 1013) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22JMD006; 1014) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22KMC453; 1015) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 22LMF819; 1016) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23BMA655; 1017) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23CMG182; 1018) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23CMH896; 1019) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23EMJ424; 1020) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23GMI083; 1021) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23IMB296; 1022) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23IMG579; 1023) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMA381; 1024) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMF470; 1025) DYNJ31283M, UDI-DI: 10195327074593(each), 40195327074594(case), Lot Number: 23KMJ727; 1026) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24BMK012; 1027) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24CMD605; 1028) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24DMG373; 1029) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24EMD713; 1030) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24GMA034; 1031) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24HMB126; 1032) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24HMI467; 1033) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24IME048; 1034) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24JME593; 1035) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 24KMG666; 1036) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25AMI942; 1037) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25BMI478; 1038) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25DMH101; 1039) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25EMF404; 1040) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25FMF966; 1041) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25HMH052; 1042) DYNJ31283N, UDI-DI: 10195327601805(each), 40195327601806(case), Lot Number: 25JMK962; 1043) DYNJ31287G, UDI-DI: 10193489468724(each), 40193489468725(case), Lot Number: 21BMA106; 1044) DYNJ31287G, UDI-DI: 10193489468724(each), 40193489468725(case), Lot Number: 21BMB509; 1045) DYNJ31287J, UDI-DI: 10195327074517(each), 40195327074518(case), Lot Number: 23KMF471; 1046) DYNJ31290L, UDI-DI: 10195327074524(each), 40195327074525(case), Lot Number: 23HMF196; 1047) DYNJ31290L, UDI-DI: 10195327074524(each), 40195327074525(case), Lot Number: 23KMA382; 1048) DYNJ31290L, UDI-DI: 10195327074524(each), 40195327074525(case), Lot Number: 23KMF472; 1049) DYNJ31290L, UDI-DI: 10195327074524(each), 40195327074525(case), Lot Number: 23LMB228; 1050) DYNJ32200F, UDI-DI: 10889942415649(each), 40889942415640(case), Lot Number: 22LBR364; 1051) DYNJ32554, UDI-DI: 10884389217726(each), 40884389217727(case), Lot Number: 22JMB735; 1052) DYNJ32724M, UDI-DI: 10193489734584(each), 40193489734585(case), Lot Number: 21ADB802; 1053) DYNJ33670F, UDI-DI: 10193489220698(each), 40193489220699(case), Lot Number: 21ADA278; 1054) DYNJ34683D, UDI-DI: 10193489508284(each), 40193489508285(case), Lot Number: 21ADC027; 1055) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 24JMF348; 1056) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 24LMB128; 1057) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25AMB882; 1058) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25BMI129; 1059) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25CMF473; 1060) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25EMG346; 1061) DYNJ35176K, UDI-DI: 10198459160998(each), 40198459160999(case), Lot Number: 25GMF870; 1062) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 24JME432; 1063) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25AMG411; 1064) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25AMI234; 1065) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25BME369; 1066) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25CMI416; 1067) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25DMC946; 1068) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25EMD264; 1069) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25FMA192; 1070) DYNJ35182J, UDI-DI: 10198459160981( each), 40198459160982(case), Lot Number: 25HMB720; 1071) DYNJ35182J, UDI-DI: 10198459160981(each), 40198459160982(case), Lot Number: 25JML524; 1072) DYNJ35512F, UDI-DI: 10889942930517(each), 40889942930518(case), Lot Number: 21ABR627; 1073) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21DBO076; 1074) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21FBS907; 1075) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21GBS876; 1076) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21IBB712; 1077) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 21LBI599; 1078) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22ABU211; 1079) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22DBB588; 1080) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22EBQ927; 1081) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22FBC561; 1082) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 22JBK351; 1083) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23GBH314; 1084) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23JBD889; 1085) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 23LBG099; 1086) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24BBU887; 1087) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24EBO185; 1088) DYNJ35744A, UDI-DI: 10889942600595(each), 40889942600596(case), Lot Number: 24FBK949; 1089) DYNJ35835G, UDI-DI: 10195327412005(each), 40195327412006(case), Lot Number: 24FMB572; 1090) DYNJ36637N, UDI-DI: 10193489796933(each), 40193489796934(case), Lot Number: 23CMH794; 1091) DYNJ36638O, UDI-DI: 10193489796834(each), 40193489796835(case), Lot Number: 23CMH795; 1092) DYNJ36639L, UDI-DI: 10193489796940(each), 40193489796941(case), Lot Number: 23CMB824; 1093) DYNJ36812F, UDI-DI: 10195327371180(each), 40195327371181(case), Lot Number: 23HBP365; 1094) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23EBU238; 1095) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23GBD566; 1096) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23HBP073; 1097) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23IBB162; 1098) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 23IBS456; 1099) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 24BBD638; 1100) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 24DBC101; 1101) DYNJ37103F, UDI-DI: 10195327388256(each), 40195327388257(case), Lot Number: 24EBK442; 1102) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 23DBM581; 1103) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 23EBB768; 1104) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 23KBA024; 1105) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 24BBD136; 1106) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 24CBD041; 1107) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 24HBT248; 1108) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 25BBB902; 1109) DYNJ37105C, UDI-DI: 10195327375195(each), 40195327375196(case), Lot Number: 25CBO436; 1110) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 23EBG542; 1111) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 23KBO206; 1112) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24ABK293; 1113) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24ABK294; 1114) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24FBJ665; 1115) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24FBT664; 1116) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24GBA045; 1117) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24JBH149; 1118) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24KBC144; 1119) DYNJ37106D, UDI-DI: 10195327375096(each), 40195327375097(case), Lot Number: 24KBO024; 1120) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23DBM582; 1121) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23EBD178; 1122) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23IBK102; 1123) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23IBV399; 1124) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23JBQ413; 1125) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 23LBI544; 1126) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24BBG286; 1127) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24EBK443; 1128) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24HBD378; 1129) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24JBE066; 1130) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 24KBM633; 1131) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25ABI114; 1132) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25BBB514; 1133) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25GBD299; 1134) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25HBM467; 1135) DYNJ37107F, UDI-DI: 10195327375102(each), 40195327375103(case), Lot Number: 25KBL533; 1136) DYNJ37123D, UDI-DI: 10193489459302(each), 40193489459303(case), Lot Number: 22BBM514; 1137) DYNJ37428F, UDI-DI: 10889942164516(each), 40889942164517(case), Lot Number: 21CBQ635; 1138) DYNJ37635L, UDI-DI: 10195327176303(each), 40195327176304(case), Lot Number: 23FBL430; 1139) DYNJ37683L, UDI-DI: 10198459439049(each), 40198459439040(case), Lot Number: 25HMC572; 1140) DYNJ37689J, UDI-DI: 10198459439056(each), 40198459439057(case), Lot Number: 25GMJ466; 1141) DYNJ37696G, UDI-DI: 10193489294866(each), 40193489294867(case), Lot Number: 21AMA190; 1142) DYNJ37696G, UDI-DI: 10193489294866(each), 40193489294867(case), Lot Number: 21DMB435; 1143) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21EMC373; 1144) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21FMB538; 1145) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21GMC277; 1146) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21IMD490; 1147) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21JMA051; 1148) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21JMA643; 1149) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 21LMF755; 1150) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22BMB144; 1151) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22DMH440; 1152) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 22GMF679; 1153) DYNJ37696I, UDI-DI: 10193489898453(each), 40193489898454(case), Lot Number: 23AMB157; 1154) DYNJ37707F, UDI-DI: 10195327394745(each), 40195327394746(case), Lot Number: 23FMJ667; 1155) DYNJ37883F, UDI-DI: 10195327527068(each), 40195327527069(case), Lot Number: 23KMJ277; 1156) DYNJ38234G, UDI-DI: 10195327327651(each), 40195327327652(case), Lot Number: 23EBR720; 1157) DYNJ38247C, UDI-DI: 10888277746299(each), 40888277746290(case), Lot Number: 22HMC530; 1158) DYNJ38397J, UDI-DI: 10193489646719(each), 40193489646710(case), Lot Number: 21CBJ903; 1159) DYNJ38397J, UDI-DI: 10193489646719(each), 40193489646710(case), Lot Number: 23CBR314; 1160) DYNJ38404J, UDI-DI: 10195327373955(each), 40195327373956(case), Lot Number: 23IME110; 1161) DYNJ38406I, UDI-DI: 10195327373900(each), 40195327373901(case), Lot Number: 23IMA431; 1162) DYNJ38529D, UDI-DI: 10195327525545(each), 40195327525546(case), Lot Number: 24BBJ214; 1163) DYNJ38529D, UDI-DI: 10195327525545(each), 40195327525546(case), Lot Number: 24DBC120; 1164) DYNJ38529G, UDI-DI: 10198459222429(each), 40198459222420(case), Lot Number: 25ABK835; 1165) DYNJ38529G, UDI-DI: 10198459222429(each), 40198459222420(case), Lot Number: 25BBS058; 1166) DYNJ38529G, UDI-DI: 10198459222429(each), 40198459222420(case), Lot Number: 25BBS173; 1167) DYNJ38529G, UDI-DI: 10198459222429(each), 40198459222420(case), Lot Number: 25DBH090; 1168) DYNJ39002F, UDI-DI: 10889942951796(each), 40889942951797(case), Lot Number: 23JMC033; 1169) DYNJ39881F, UDI-DI: 10193489704068(each), 40193489704069(case), Lot Number: 23GMI050; 1170) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBL153; 1171) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBS675; 1172) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21EBS723; 1173) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21HBA944; 1174) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21HBH566; 1175) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 21IBR355; 1176) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22BBO207; 1177) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22BBU071; 1178) DYNJ39934D, UDI-DI: 10193489475203(each), 40193489475204(case), Lot Number: 22EBK821; 1179) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 22FBS460; 1180) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 22HBJ054; 1181) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23ABM892; 1182) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23BBQ109; 1183) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23FBL144; 1184) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23GBJ167; 1185) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23HBO846; 1186) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 23JBC963; 1187) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24ABR822; 1188) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24BBP203; 1189) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24DBA107; 1190) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24DBM519; 1191) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24GBF384; 1192) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24HBT612; 1193) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 24KBA611; 1194) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 25ABS136; 1195) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 25DBM175; 1196) DYNJ39934F, UDI-DI: 10195327186999(each), 40195327186990(case), Lot Number: 25FBU505; 1197) DYNJ40309L, UDI-DI: 10193489913811(each), 40193489913812(case), Lot Number: 22BBK400; 1198) DYNJ40557, UDI-DI: 10884389868850(each), 40884389868851(case), Lot Number: 21ABN203; 1199) DYNJ40557, UDI-DI: 10884389868850(each), 40884389868851(case), Lot Number: 21CBG563; 1200) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 21IBX363; 1201) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 21IBX396; 1202) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 22BBT297; 1203) DYNJ40557A, UDI-DI: 10193489925906(each), 40193489925907(case), Lot Number: 22DBK226; 1204) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 22EBR223; 1205) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 22GBG734; 1206) DYNJ40557B, UDI-DI: 10195327156701(each), 40195327156702(case), Lot Number: 23ABB877; 1207) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23HBP214; 1208) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23IBH883; 1209) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23JBX673; 1210) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 23LBA343; 1211) DYNJ40557C, UDI-DI: 10195327479961(each), 40195327479962(case), Lot Number: 24BBB096; 1212) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24GBJ691; 1213) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24HBE197; 1214) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24HBL231; 1215) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24KBQ488; 1216) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 24LBL572; 1217) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25BBL392; 1218) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25GBP428; 1219) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25HBL160; 1220) DYNJ40557D, UDI-DI: 10195327691370(each), 40195327691371(case), Lot Number: 25IBF313; 1221) DYNJ40557F, UDI-DI: 10198459589157(each), 40198459589158(case), Lot Number: 25KBR002; 1222) DYNJ40597C, UDI-DI: 10193489974225(each), 40193489974226(case), Lot Number: 22DBK958; 1223) DYNJ40699C, UDI-DI: 10195327279936(each), 40195327279937(case), Lot Number: 23JMH882; 1224) DYNJ41030K, UDI-DI: 10195327156763(each), 40195327156764(case), Lot Number: 22HME079; 1225) DYNJ41030K, UDI-DI: 10195327156763(each), 40195327156764(case), Lot Number: 23GMB280; 1226) DYNJ41394C, UDI-DI: 10888277937208(each), 40888277937209(case), Lot Number: 21ABW205; 1227) DYNJ41692F, UDI-DI: 10193489662993(each), 40193489662994(case), Lot Number: 22LBR516; 1228) DYNJ41808C, UDI-DI: 10195327565664(each), 40195327565665(case), Lot Number: 23KBT722; 1229) DYNJ41808C, UDI-DI: 10195327565664(each), 40195327565665(case), Lot Number: 24ABF458; 1230) DYNJ41808C, UDI-DI: 10195327565664(each), 40195327565665(case), Lot Number: 24DBJ076; 1231) DYNJ41857C, UDI-DI: 10193489662658(each), 40193489662659(case), Lot Number: 22IBW059; 1232) DYNJ42051F, UDI-DI: 10193489705317(each), 40193489705318(case), Lot Number: 21DDA607; 1233) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21ABU488; 1234) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21GBB636; 1235) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21JBZ483; 1236) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 21LBN641; 1237) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22BBJ794; 1238) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22CBZ149; 1239) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22GBD911; 1240) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22HBZ293; 1241) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22IBU352; 1242) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22KBR014; 1243) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 22LBP906; 1244) DYNJ42215B, UDI-DI: 10888277738720(each), 40888277738721(case), Lot Number: 23DBB335; 1245) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23EBD832; 1246) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23GBL372; 1247) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23IBG121; 1248) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23JBT608; 1249) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 23LBB918; 1250) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24ABK305; 1251) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24ABT170; 1252) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24BBI050; 1253) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24FBK367; 1254) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24GBM991; 1255) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24HBF332; 1256) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24IBA105; 1257) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24JBL213; 1258) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 24KBM235; 1259) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25BBC870; 1260) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25CBP208; 1261) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25DBB075; 1262) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25FBG165; 1263) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25FBS104; 1264) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25GBF601; 1265) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25JBB592; 1266) DYNJ42215C, UDI-DI: 10195327391669(each), 40195327391660(case), Lot Number: 25LBM037; 1267) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22FME394; 1268) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21BMA680; 1269) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21BMD156; 1270) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21CMD909; 1271) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21HMC587; 1272) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21JMG016; 1273) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21KME766; 1274) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21KMG901; 1275) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 21LMD228; 1276) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22AMD130; 1277) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22CME903; 1278) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22DMI958; 1279) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22GMA359; 1280) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22HMH141; 1281) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 22KMD575; 1282) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23BMG639; 1283) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23CMH177; 1284) DYNJ42234B, UDI-DI: 10193489801071(each), 40193489801072(case), Lot Number: 23EMJ755; 1285) DYNJ42793D, UDI-DI: 10193489427547(each), 40193489427548(case), Lot Number: 23HMF999; 1286) DYNJ42888B, UDI-DI: 10193489335002(each), 40193489335003(case), Lot Number: 21BBB540; 1287) DYNJ42944C, UDI-DI: 10193489660593(each), 40193489660594(case), Lot Number: 21CBJ906; 1288) DYNJ43006I, UDI-DI: 10193489927160(each), 40193489927161(case), Lot Number: 22JMG800; 1289) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23GMB626; 1290) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23GMG538; 1291) DYNJ43012J, UDI-DI: 10195327412777(each), 40195327412778(case), Lot Number: 23IMD412; 1292) DYNJ43186C, UDI-DI: 10193489475210(each), 40193489475211(case), Lot Number: 21BBC547; 1293) DYNJ43186C, UDI-DI: 10193489475210(each), 40193489475211(case), Lot Number: 21CBO118; 1294) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 21IBR381; 1295) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 21KBL792; 1296) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 22BBL717; 1297) DYNJ43186D, UDI-DI: 10193489870848(each), 40193489870849(case), Lot Number: 22FBB767; 1298) DYNJ43186F, UDI-DI: 10195327186982(each), 40195327186983(case), Lot Number: 22LBR903; 1299) DYNJ43186F, UDI-DI: 10195327186982(each), 40195327186983(case), Lot Number: 23BBC787; 1300) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23ABO776; 1301) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23CBB473; 1302) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23CBX218; 1303) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23DBQ604; 1304) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23FBA479; 1305) DYNJ43186G, UDI-DI: 10195327325091(each), 40195327325092(case), Lot Number: 23GBO024; 1306) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21GBR214; 1307) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21HBT086; 1308) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 21KBL461; 1309) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22ABH986; 1310) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22BBU712; 1311) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22EBH854; 1312) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBB221; 1313) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBE821; 1314) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22GBQ330; 1315) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22HBW601; 1316) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22LBN719; 1317) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22OBB442; 1318) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 22OBI176; 1319) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23BBC162; 1320) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23BBS734; 1321) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23DBB718; 1322) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23DBB880; 1323) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23EBW551; 1324) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23FBT488; 1325) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23GBV872; 1326) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23IBM911; 1327) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23KBN758; 1328) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 23LBU954; 1329) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 24ABR199; 1330) DYNJ43421B, UDI-DI: 10193489901412(each), 40193489901413(case), Lot Number: 24CBF071; 1331) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21FMD820; 1332) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21HMB776; 1333) DYNJ43432C, UDI-DI: 10193489879001(each), 40193489879002(case), Lot Number: 21IME539; 1334) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 21LMD524; 1335) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMA059; 1336) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMA613; 1337) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22AMB868; 1338) DYNJ43432D, UDI-DI: 10195327039011(each), 40195327039012(case), Lot Number: 22BMC971; 1339) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22FMH502; 1340) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22FMH503; 1341) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22IMH856; 1342) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 22KMD710; 1343) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23BMC530; 1344) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23BMC726; 1345) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23DMG739; 1346) DYNJ43432F, UDI-DI: 10195327126698(each), 40195327126699(case), Lot Number: 23EMI628; 1347) DYNJ44020A, UDI-DI: 10193489806649(each), 40193489806640(case), Lot Number: 23DBI587; 1348) DYNJ44365I, UDI-DI: 10195327463328(each), 40195327463329(case), Lot Number: 24BBF260; 1349) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21ABB322; 1350) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21BBR226; 1351) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21DBC978; 1352) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21EBN450; 1353) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21GBC952; 1354) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21GBT043; 1355) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21KBA071; 1356) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 21LBQ228; 1357) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22BBG493; 1358) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22GBC896; 1359) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBC461; 1360) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBK861; 1361) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22HBQ924; 1362) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22OBD081; 1363) DYNJ44730D, UDI-DI: 10193489716672(each), 40193489716673(case), Lot Number: 22OBM108; 1364) DYNJ45062C, UDI-DI: 10193489733389(each), 40193489733380(case), Lot Number: 21KBP326; 1365) DYNJ45145F, UDI-DI: 10193489824889(each), 40193489824880(case), Lot Number: 21BBA717; 1366) DYNJ45892C, UDI-DI: 10193489272017(each), 40193489272018(case), Lot Number: 22HMD408; 1367) DYNJ46384N, UDI-DI: 10193489937275(each), 40193489937276(case), Lot Number: 23HMG225; 1368) DYNJ46534J, UDI-DI: 10193489337778(each), 40193489337779(case), Lot Number: 21FDB447; 1369) DYNJ46534J, UDI-DI: 10193489337778(each), 40193489337779(case), Lot Number: 21FDC712; 1370) DYNJ46628D, UDI-DI: 10193489756678(each), 40193489756679(case), Lot Number: 24IBP739; 1371) DYNJ46977B, UDI-DI: 10193489554069(each), 40193489554060(case), Lot Number: 23JBS354; 1372) DYNJ46980D, UDI-DI: 10889942852895(each), 40889942852896(case), Lot Number: 23EBB740; 1373) DYNJ47104C, UDI-DI: 10889942489190(each), 40889942489191(case), Lot Number: 21ABW902; 1374) DYNJ47608B, UDI-DI: 10193489731019(each), 40193489731010(case), Lot Number: 23ABH658; 1375) DYNJ47631A, UDI-DI: 10888277843103(each), 40888277843104(case), Lot Number: 23BBN379; 1376) DYNJ47715I, UDI-DI: 10193489887525(each), 40193489887526(case), Lot Number: 22HME316; 1377) DYNJ47784G, UDI-DI: 10193489887532(each), 40193489887533(case), Lot Number: 22HMF248; 1378) DYNJ48063G, UDI-DI: 10193489497830(each), 40193489497831(case), Lot Number: 22JMB326; 1379) DYNJ48091I, UDI-DI: 10195327500122(each), 40195327500123(case), Lot Number: 23JMI555; 1380) DYNJ48151I, UDI-DI: 10195327261870(each), 40195327261871(case), Lot Number: 23HMG449; 1381) DYNJ48178, UDI-DI: 10888277559226(each), 40888277559227(case), Lot Number: 23HME394; 1382) DYNJ48237J, UDI-DI: 10193489988185(each), 40193489988186(case), Lot Number: 22HMD410; 1383) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23BBD636; 1384) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23DBG666; 1385) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23EBN348; 1386) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23EBN371; 1387) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23FBQ576; 1388) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23IBJ632; 1389) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23JBW593; 1390) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23JBY697; 1391) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 23LBS921; 1392) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24CBC160; 1393) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24EBU758; 1394) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24FBK165; 1395) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24GBP662; 1396) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24JBP483; 1397) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24KBM177; 1398) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24LBA195; 1399) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 24LBM126; 1400) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 25ABJ265; 1401) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 25BBK576; 1402) DYNJ48473G, UDI-DI: 10193489742336(each), 40193489742337(case), Lot Number: 25EBA458; 1403) DYNJ48473I, UDI-DI: 10198459349829(each), 40198459349820(case), Lot Number: 25FBB300; 1404) DYNJ48473I, UDI-DI: 10198459349829(each), 40198459349820(case), Lot Number: 25FBU511; 1405) DYNJ48474I, UDI-DI: 10198459349843(each), 40198459349844(case), Lot Number: 25EBM309; 1406) DYNJ48474I, UDI-DI: 10198459349843(each), 40198459349844(case), Lot Number: 25HBC063; 1407) DYNJ49105G, UDI-DI: 10193489927382(each), 40193489927383(case), Lot Number: 23FBA447; 1408) DYNJ49247B, UDI-DI: 10193489504446(each), 40193489504447(case), Lot Number: 22HMD932; 1409) DYNJ49247B, UDI-DI: 10193489504446(each), 40193489504447(case), Lot Number: 22KMB758; 1410) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21ABL108; 1411) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21BBQ667; 1412) DYNJ49269F, UDI-DI: 10193489315035(each), 40193489315036(case), Lot Number: 21DBI826; 1413) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21EBN532; 1414) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21FBK536; 1415) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21KBJ103; 1416) DYNJ49269G, UDI-DI: 10193489890792(each), 40193489890793(case), Lot Number: 21LBG756; 1417) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 21LBW673; 1418) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22DBI721; 1419) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22DMC969; 1420) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22GME891; 1421) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22IMA085; 1422) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22KMC642; 1423) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 22LMC345; 1424) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Nu mber: 23CME921; 1425) DYNJ49269I, UDI-DI: 10195327068103(each), 40195327068104(case), Lot Number: 23DMB606; 1426) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23GMF313; 1427) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23IMA323; 1428) DYNJ49269J, UDI-DI: 10195327401962(each), 40195327401963(case), Lot Number: 23KMH889; 1429) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 21HDB682; 1430) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 21IDA623; 1431) DYNJ49670F, UDI-DI: 10193489917192(each), 40193489917193(case), Lot Number: 22EDA621; 1432) DYNJ49782I, UDI-DI: 10195327259105(each), 40195327259106(case), Lot Number: 23CBK571; 1433) DYNJ49782J, UDI-DI: 10195327406240(each), 40195327406241(case), Lot Number: 23EBU897; 1434) DYNJ50357C, UDI-DI: 10193489826791(each), 40193489826792(case), Lot Number: 21BKA485; 1435) DYNJ50360C, UDI-DI: 10193489826821(each), 40193489826822(case), Lot Number: 23HMC674; 1436) DYNJ50360C, UDI-DI: 10193489826821(each), 40193489826822(case), Lot Number: 23HMC675; 1437) DYNJ50429D, UDI-DI: 10195327043506(each), 40195327043507(case), Lot Number: 23BBB098; 1438) DYNJ50450C, UDI-DI: 10193489595307(each), 40193489595308(case), Lot Number: 21BBE826; 1439) DYNJ50450C, UDI-DI: 10193489595307(each), 40193489595308(case), Lot Number: 21DBK486; 1440) DYNJ50622I, UDI-DI: 10193489467000(each), 40193489467001(case), Lot Number: 21ABW206; 1441) DYNJ50710D, UDI-DI: 10193489716689(each), 40193489716680(case), Lot Number: 20KBZ974; 1442) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 21DBM047; 1443) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 21GBU067; 1444) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 22HBC296; 1445) DYNJ50710F, UDI-DI: 10193489851915(each), 40193489851916(case), Lot Number: 22OBJ527; 1446) DYNJ50812B, UDI-DI: 10889942368068(each), 40889942368069(case), Lot Number: 23CBV559; 1447) DYNJ51036B, UDI-DI: 10889942229475(each), 40889942229476(case), Lot Number: 21AMA118; 1448) DYNJ51036B, UDI-DI: 10889942229475(each), 40889942229476(case), Lot Number: 21AMD560; 1449) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21CMB429; 1450) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21DMC480; 1451) DYNJ51036C, UDI-DI: 10193489849868(each), 40193489849869(case), Lot Number: 21FME965; 1452) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22DMF167; 1453) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22DMH641; 1454) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22IMD862; 1455) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22JMA473; 1456) DYNJ51036F, UDI-DI: 10195327113346(each), 40195327113347(case), Lot Number: 22JMC109; 1457) DYNJ51036G, UDI-DI: 10195327332365(each), 40195327332366(case), Lot Number: 23FBQ872; 1458) DYNJ51088A, UDI-DI: 10193489985313(each), 40193489985314(case), Lot Number: 21IDA625; 1459) DYNJ51091A, UDI-DI: 10193489985306(each), 40193489985307(case), Lot Number: 22DDC213; 1460) DYNJ51095A, UDI-DI: 10193489985436(each), 40193489985437(case), Lot Number: 22DDC333; 1461) DYNJ51096B, UDI-DI: 10193489985290(each), 40193489985291(case), Lot Number: 22EDA486; 1462) DYNJ51176C, UDI-DI: 10889942724642(each), 40889942724643(case), Lot Number: 21ADA263; 1463) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 24FBF369; 1464) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 24LBN605; 1465) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 25ABK343; 1466) DYNJ51249B, UDI-DI: 10195327679583(each), 40195327679584(case), Lot Number: 25CBJ143; 1467) DYNJ51774C, UDI-DI: 10889942816200(each), 40889942816201(case), Lot Number: 21CBI319; 1468) DYNJ51802I, UDI-DI: 10193489476217(each), 40193489476218(case), Lot Number: 21ADA367; 1469) DYNJ51803G, UDI-DI: 10193489476224(each), 40193489476225(case), Lot Number: 21ADA120; 1470) DYNJ52111A, UDI-DI: 10195327571856(each), 40195327571857(case), Lot Number: 24ABH906; 1471) DYNJ52207B, UDI-DI: 10193489635614(each), 40193489635615(case), Lot Number: 23BMH298; 1472) DYNJ52330C, UDI-DI: 10195327434076(each), 40195327434077(case), Lot Number: 23JBK736; 1473) DYNJ52418F, UDI-DI: 10193489249705(each), 40193489249706(case), Lot Number: 21ADA541; 1474) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21ABE527; 1475) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21CBN962; 1476) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21DBM019; 1477) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21GBK596; 1478) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21IBG396; 1479) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 21JBS706; 1480) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 22DBA960; 1481) DYNJ52489D, UDI-DI: 10193489475227(each), 40193489475228(case), Lot Number: 22EBP123; 1482) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 22GBE349; 1483) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 22IBA379; 1484) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23BBH694; 1485) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23DBE733; 1486) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23HBC482; 1487) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23IBW198; 1488) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 23KBQ185; 1489) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24ABG587; 1490) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24BBE314; 1491) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24BBP202; 1492) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24DBS978; 1493) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24GBU005; 1494) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 24KBP098; 1495) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 25ABL375; 1496) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 25BBB594; 1497) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 25CBN850; 1498) DYNJ52489F, UDI-DI: 10195327187019(each), 40195327187010(case), Lot Number: 25FBC616; 1499) DYNJ52491F, UDI-DI: 10195327435899(each), 40195327435890(case), Lot Number: 25LME675; 1500) DYNJ52523A, UDI-DI: 10889942502462(each), 40889942502463(case), Lot Number: 21BBK699; 1501) DYNJ52607C, UDI-DI: 10195327134631(each), 40195327134632(case), Lot Number: 23DBG719; 1502) DYNJ52720C, UDI-DI: 10195327480356(each), 40195327480357(case), Lot Number: 23KBX843; 1503) DYNJ52950C, UDI-DI: 10193489745337(each), 40193489745338(case), Lot Number: 21BBK701; 1504) DYNJ52957F, UDI-DI: 10193489315721(each), 40193489315722(case), Lot Number: 23FBR704; 1505) DYNJ52958D, UDI-DI: 10193489740349(each), 40193489740340(case), Lot Number: 21BBA721; 1506) DYNJ53212D, UDI-DI: 10193489946871(each), 40193489946872(case), Lot Number: 21GBD684; 1507) DYNJ53212D, UDI-DI: 10193489946871(each), 40193489946872(case), Lot Number: 22BBE369; 1508) DYNJ53212D, UDI-DI: 10193489946871(each), 40193489946872(case), Lot Number: 23FBE999; 1509) DYNJ53287D, UDI-DI: 10193489552980(each), 40193489552981(case), Lot Number: 21BDA322; 1510) DYNJ53420D, UDI-DI: 10193489975949(each), 40193489975940(case), Lot Number: 23CBE656; 1511) DYNJ53537B, UDI-DI: 10195327158880(each), 40195327158881(case), Lot Number: 23HBS114; 1512) DYNJ53539D, UDI-DI: 10195327158903(each), 40195327158904(case), Lot Number: 23KBK120; 1513) DYNJ53686A, UDI-DI: 10193489386882(each), 40193489386883(case), Lot Number: 21ADA123; 1514) DYNJ53788A, UDI-DI: 10889942909421(each), 40889942909422(case), Lot Number: 21ADB618; 1515) DYNJ53825, UDI-DI: 10889942443963(each), 40889942443964(case), Lot Number: 24DBJ816; 1516) DYNJ53970K, UDI-DI: 10195327532741(each), 40195327532742(case), Lot Number: 23KMF315; 1517) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21HMA349; 1518) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21HMF506; 1519) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21IME798; 1520) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21JMC220; 1521) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21JMH447; 1522) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 21KME883; 1523) DYNJ53976D, UDI-DI: 10193489951103(each), 40193489951104(case), Lot Number: 22FMH898; 1524) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23EMK219; 1525) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23AMB679; 1526) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23CMH919; 1527) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23FMH056; 1528) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23HME945; 1529) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23HMG008; 1530) DYNJ53976F, UDI-DI: 10195327282851(each), 40195327282852(case), Lot Number: 23JMD007; 1531) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 23KMI515; 1532) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24AMA137; 1533) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24DMJ360; 1534) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24EME642; 1535) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24FMI100; 1536) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24HMI416; 1537) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 24KMH993; 1538) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25AMH782; 1539) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25BME348; 1540) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25EMF753; 1541) DYNJ53976G, UDI-DI: 10195327532772(each), 40195327532773(case), Lot Number: 25FMJ582; 1542) DYNJ54481A, UDI-DI: 10193489944914(each), 40193489944915(case), Lot Number: 23DME204; 1543) DYNJ54912B, UDI-DI: 10193489317077(each), 40193489317078(case), Lot Number: 23IBP189; 1544) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 24HBT914; 1545) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 24JBJ473; 1546) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25ABD151; 1547) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25BBM551; 1548) DYNJ54961A, UDI-DI: 10198459033469(each), 40198459033460(case), Lot Number: 25EBP981; 1549) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 24FBM712; 1550) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 24HBJ227; 1551) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 24IBI557; 1552) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 24IBQ388; 1553) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 25ABD766; 1554) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 25BBT123; 1555) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 25DBI205; 1556) DYNJ54972A, UDI-DI: 10198459033384(each), 40198459033385(case), Lot Number: 25EBN879; 1557) DYNJ54977B, UDI-DI: 10198459033414(each), 40198459033415(case), Lot Number: 24IBE274; 1558) DYNJ54977B, UDI-DI: 10198459033414(each), 40198459033415(case), Lot Number: 24JBG693; 1559) DYNJ54977B, UDI-DI: 10198459033414(each), 40198459033415(case), Lot Number: 24LBR616; 1560) DYNJ54977B, UDI-DI: 10198459033414(each), 40198459033415(case), Lot Number: 25ABK090; 1561) DYNJ54977B, UDI-DI: 10198459033414(each), 40198459033415(case), Lot Number: 25EBI965; 1562) DYNJ54988C, UDI-DI: 10195327007003(each), 40195327007004(case), Lot Number: 23CMC077; 1563) DYNJ55071F, UDI-DI: 10195327410957(each), 40195327410958(case), Lot Number: 23LBC872; 1564) DYNJ55285F, UDI-DI: 10195327284862(each), 40195327284863(case), Lot Number: 23CBE662; 1565) DYNJ55317F, UDI-DI: 10193489692389(each), 40193489692380(case), Lot Number: 23EBU898; 1566) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23GBC804; 1567) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23IBA353; 1568) DYNJ55317G, UDI-DI: 10195327429270(each), 40195327429271(case), Lot Number: 23JBH543; 1569) DYNJ55317I, UDI-DI: 10195327534202(each), 40195327534203(case), Lot Number: 23LBO119; 1570) DYNJ55324F, UDI-DI: 10195327120726(each), 40195327120727(case), Lot Number: 23FBA486; 1571) DYNJ55351B, UDI-DI: 10195327173760(each), 40195327173761(case), Lot Number: 23FBQ171; 1572) DYNJ55353C, UDI-DI: 10195327173487(each), 40195327173488(case), Lot Number: 23FBR554; 1573) DYNJ55441C, UDI-DI: 10889942828449(each), 40889942828440(case), Lot Number: 23JBB387; 1574) DYNJ55744A, UDI-DI: 10889942688937(each), 40889942688938(case), Lot Number: 21EMA571; 1575) DYNJ56257C, UDI-DI: 10193489825114(each), 40193489825115(case), Lot Number: 23IBE172; 1576) DYNJ56627B, UDI-DI: 10195327283025(each), 40195327283026(case), Lot Number: 25BBC873; 1577) DYNJ56766C, UDI-DI: 10193489648485(each), 40193489648486(case), Lot Number: 21ABG397; 1578) DYNJ56800F, UDI-DI: 10193489353709(each), 40193489353700(case), Lot Number: 21BKA247; 1579) DYNJ56864D, UDI-DI: 10193489970944(each), 40193489970945(case), Lot Number: 23BBH676; 1580) DYNJ56864D, UDI-DI: 10193489970944(each), 40193489970945(case), Lot Number: 23DBG861; 1581) DYNJ56864D, UDI-DI: 10193489970944(each), 40193489970945(case), Lot Number: 23FBI956; 1582) DYNJ56864D, UDI-DI: 10193489970944(each), 40193489970945(case), Lot Number: 23FBR514; 1583) DYNJ56864F, UDI-DI: 10195327464417(each), 40195327464418(case), Lot Number: 23HBD783; 1584) DYNJ56864F, UDI-DI: 10195327464417(each), 40195327464418(case), Lot Number: 23IBL784; 1585) DYNJ56864F, UDI-DI: 10195327464417(each), 40195327464418(case), Lot Number: 23KBV146; 1586) DYNJ56889, UDI-DI: 10889942691739(each), 40889942691730(case), Lot Number: 23CBO165; 1587) DYNJ56889, UDI-DI: 10889942691739(each), 40889942691730(case), Lot Number: 23JBT092; 1588) DYNJ56986F, UDI-DI: 10193489201901(each), 40193489201902(case), Lot Number: 21ABM378; 1589) DYNJ56986F, UDI-DI: 10193489201901(each), 40193489201902(case), Lot Number: 21BBL376; 1590) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21FBQ044; 1591) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21GBP729; 1592) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21HBN583; 1593) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21LBS167; 1594) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21LBS202; 1595) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 21VBB773; 1596) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22BBT830; 1597) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22DBD469; 1598) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22EBQ164; 1599) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22KBR892; 1600) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 22LBR917; 1601) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 23KMB418; 1602) DYNJ56986G, UDI-DI: 10193489884265(each), 40193489884266(case), Lot Number: 24CMC651; 1603) DYNJ57001A, UDI-DI: 10193489479836(each), 40193489479837(case), Lot Number: 21BBA507; 1604) DYNJ57080A, UDI-DI: 10193489234664(each), 40193489234665(case), Lot Number: 21CMB611; 1605) DYNJ57266F, UDI-DI: 10193489498189(each), 40193489498180(case), Lot Number: 21BBA724; 1606) DYNJ57615D, UDI-DI: 10195327326166(each), 40195327326167(case), Lot Number: 23DME211; 1607) DYNJ57768, UDI-DI: 10889942776719(each), 40889942776710(case), Lot Number: 22LBS131; 1608) DYNJ57768, UDI-DI: 10889942776719(each), 40889942776710(case), Lot Number: 23CBW160; 1609) DYNJ57768, UDI-DI: 10889942776719(each), 40889942776710(case), Lot Number: 23DBD114; 1610) DYNJ57768, UDI-DI: 10889942776719(each), 40889942776710(case), Lot Number: 23DBP275; 1611) DYNJ57806B, UDI-DI: 10193489643749(each), 40193489643740(case), Lot Number: 24ABI521; 1612) DYNJ57824G, UDI-DI: 10193489237528(each), 40193489237529(case), Lot Number: 21IDB163; 1613) DYNJ57824G, UDI-DI: 10193489237528(each), 40193489237529(case), Lot Number: 21KDB870; 1614) DYNJ58044B, UDI-DI: 10193489915846(each), 40193489915847(case), Lot Number: 23JBY451; 1615) DYNJ58082A, UDI-DI: 10195327096748(each), 40195327096749(case), Lot Number: 22FDB387; 1616) DYNJ58082A, UDI-DI: 10195327096748(each), 40195327096749(case), Lot Number: 22KDC032; 1617) DYNJ58085C, UDI-DI: 10195327096779(each), 40195327096770(case), Lot Number: 22EDA279; 1618) DYNJ58329B, UDI-DI: 10193489485684(each), 40193489485685(case), Lot Number: 24BBK453; 1619) DYNJ58372A, UDI-DI: 10193489345247(each), 40193489345248(case), Lot Number: 21ADA509; 1620) DYNJ58374F, UDI-DI: 10195327095062(each), 40195327095063(case), Lot Number: 23IBU240; 1621) DYNJ58443C, UDI-DI: 10193489778588(each), 40193489778589(case), Lot Number: 23IMA198; 1622) DYNJ58563C, UDI-DI: 10193489460025(each), 40193489460026(case), Lot Number: 21BBA725; 1623) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21IMD510; 1624) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21IME158; 1625) DYNJ58567B, UDI-DI: 10193489420371(each), 40193489420372(case), Lot Number: 21KMG511; 1626) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 21LMF341; 1627) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22AMF287; 1628) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22AMF288; 1629) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22CMA966; 1630) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22GME707; 1631) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22IMG805; 1632) DYNJ58567C, UDI-DI: 10195327061098(each), 40195327061099(case), Lot Number: 22JME349; 1633) DYNJ58713, UDI-DI: 10889942880980(each), 40889942880981(case), Lot Number: 21BBB510; 1634) DYNJ58817A, UDI-DI: 10195327312114(each), 40195327312115(case), Lot Number: 23HBQ114; 1635) DYNJ59782, UDI-DI: 10888277783799(each), 40888277783790(case), Lot Number: 21CBE064; 1636) DYNJ59915D, UDI-DI: 10195327042288(each), 40195327042289(case), Lot Number: 23ELA911; 1637) DYNJ59929C, UDI-DI: 10195327036942(each), 40195327036943(case), Lot Number: 23LBF701; 1638) DYNJ60062K, UDI-DI: 10195327322540(each), 40195327322541(case), Lot Number: 23DBG359; 1639) DYNJ60103A, UDI-DI: 10193489321364(each), 40193489321365(case), Lot Number: 22HMB861; 1640) DYNJ60485C, UDI-DI: 10195327493486(each), 40195327493487(case), Lot Number: 23LBQ563; 1641) DYNJ60730A, UDI-DI: 10195327157302(each), 40195327157303(case), Lot Number: 22HBU496; 1642) DYNJ61116B, UDI-DI: 10193489226768(each), 40193489226769(case), Lot Number: 21ABV416; 1643) DYNJ61244A, UDI-DI: 10193489811032(each), 40193489811033(case), Lot Number: 21ABW945; 1644) DYNJ61425B, UDI-DI: 10195327408275(each), 40195327408276(case), Lot Number: 23LMA607; 1645) DYNJ61480A, UDI-DI: 10193489963533(each), 40193489963534(case), Lot Number: 23JBA340; 1646) DYNJ61581C, UDI-DI: 10195327152147(each), 40195327152148(case), Lot Number: 23EME363; 1647) DYNJ61687, UDI-DI: 10193489589702(each), 40193489589703(case), Lot Number: 23IMB431; 1648) DYNJ61692D, UDI-DI: 10195327472580(each), 40195327472581(case), Lot Number: 23KMH106; 1649) DYNJ61735A, UDI-DI: 10193489469387(each), 40193489469388(case), Lot Number: 22LMF337; 1650) DYNJ61739A, UDI-DI: 10193489469417(each), 40193489469418(case), Lot Number: 23CMF304; 1651) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20LBT867; 1652) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20LBX795; 1653) DYNJ61836B, UDI-DI: 10193489444421(each), 40193489444422(case), Lot Number: 20XBC904; 1654) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21BBD938; 1655) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21BBS149; 1656) DYNJ61836C, UDI-DI: 10193489435825(each), 40193489435826(case), Lot Number: 21CBH759; 1657) DYNJ61836D, UDI-DI: 10193489863345(each), 40193489863346(case), Lot Number: 21DBR351; 1658) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21IBU114; 1659) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21JBF773; 1660) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21JBT954; 1661) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21KBY178; 1662) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 21LBW709; 1663) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22BBO063; 1664) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22GBC809; 1665) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22HBT804; 1666) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22JBI959; 1667) DYNJ61836F, UDI-DI: 10193489972665(each), 40193489972666(case), Lot Number: 22KBG479; 1668) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23ABJ028; 1669) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23BBJ107; 1670) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23DBJ016; 1671) DYNJ61836G, UDI-DI: 10195327330477(each), 40195327330478(case), Lot Number: 23FBC115; 1672) DYNJ61845, UDI-DI: 10193489605358(each), 40193489605359(case), Lot Number: 21DDA680; 1673) DYNJ61846, UDI-DI: 10193489605518(each), 40193489605519(case), Lot Number: 21ADB041; 1674) DYNJ61846, UDI-DI: 10193489605518(each), 40193489605519(case), Lot Number: 21DDC139; 1675) DYNJ61849, UDI-DI: 10193489605181(each), 40193489605182(case), Lot Number: 21DDB382; 1676) DYNJ62122A, UDI-DI: 10193489909708(each), 40193489909709(case), Lot Number: 23CBW279; 1677) DYNJ62201, UDI-DI: 10193489633528(each), 40193489633529(case), Lot Number: 21CDC428; 1678) DYNJ62368C, UDI-DI: 10193489407525(each), 40193489407526(case), Lot Number: 21BBA729; 1679) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22GBF339; 1680) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22GBQ160; 1681) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22HBM930; 1682) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22HBW557; 1683) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 22IBK987; 1684) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23BBD249; 1685) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23BBD864; 1686) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23CBO687; 1687) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23EBE814; 1688) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23GBN992; 1689) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23HBI392; 1690) DYNJ62650B, UDI-DI: 10195327121846(each), 40195327121847(case), Lot Number: 23IBN349; 1691) DYNJ62668B, UDI-DI: 10193489945690(each), 40193489945691(case), Lot Number: 22DBD976; 1692) DYNJ62743B, UDI-DI: 10193489800784(each), 40193489800785(case), Lot Number: 22IBK765; 1693) DYNJ62999B, UDI-DI: 10193489833461(each), 40193489833462(case), Lot Number: 21GDA095; 1694) DYNJ63001A, UDI-DI: 10193489245639(each), 40193489245630(case), Lot Number: 21ADB042; 1695) DYNJ63016C, UDI-DI: 10193489833515(each), 40193489833516(case), Lot Number: 21CDC230; 1696) DYNJ63030A, UDI-DI: 10193489245820(each), 40193489245821(case), Lot Number: 21BDA062; 1697) DYNJ63031B, UDI-DI: 10193489833492(each), 40193489833493(case), Lot Number: 21EDB802; 1698) DYNJ63031B, UDI-DI: 10193489833492(each), 40193489833493(case), Lot Number: 21FDA942; 1699) DYNJ63035C, UDI-DI: 10193489833539(each), 40193489833530(case), Lot Number: 21EDB803; 1700) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22CMD192; 1701) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22GMC785; 1702) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22HMA990; 1703) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22IMA803; 1704) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22IMA805; 1705) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 22LMJ344; 1706) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23AMH685; 1707) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23CMI070; 1708) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23EMI678; 1709) DYNJ63082B, UDI-DI: 10195327102135(each), 40195327102136(case), Lot Number: 23GMG812; 1710) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23KMC330; 1711) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23KMH957; 1712) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 23LMG702; 1713) DYNJ63082C, UDI-DI: 10195327505615(each), 40195327505616(case), Lot Number: 24BMA651; 1714) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21ABP388; 1715) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21CBC907; 1716) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21FBQ059; 1717) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21HBM419; 1718) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21JBO125; 1719) DYNJ63091A, UDI-DI: 10193489293562(each), 40193489293563(case), Lot Number: 21KBJ343; 1720) DYNJ63201B, UDI-DI: 10195327291006(each), 40195327291007(case), Lot Number: 23ABB913; 1721) DYNJ63202B, UDI-DI: 10195327291082(each), 40195327291083(case), Lot Number: 23ABB886; 1722) DYNJ63510A, UDI-DI: 10193489295818(each), 40193489295819(case), Lot Number: 21ADA838; 1723) DYNJ63603B, UDI-DI: 10193489901566(each), 40193489901567(case), Lot Number: 23DBP360; 1724) DYNJ63639B, UDI-DI: 10193489900323(each), 40193489900324(case), Lot Number: 23BBM590; 1725) DYNJ63802A, UDI-DI: 10193489799637(each), 40193489799638(case), Lot Number: 21DBU650; 1726) DYNJ63802A, UDI-DI: 10193489799637(each), 40193489799638(case), Lot Number: 21EBA575; 1727) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 23DBC378; 1728) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 23KBU440; 1729) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24ABK520; 1730) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24EBP263; 1731) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24FBK285; 1732) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 24JBC327; 1733) DYNJ63802B, UDI-DI: 10195327292867(each), 40195327292868(case), Lot Number: 25CBC682; 1734) DYNJ63813C, UDI-DI: 10193489914214(each), 40193489914215(case), Lot Number: 23DBG674; 1735) DYNJ63955, UDI-DI: 10193489774276(each), 40193489774277(case), Lot Number: 21CDC326; 1736) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 23JDB401; 1737) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 23LDB441; 1738) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24CDB433; 1739) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24EDA419; 1740) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24HDA013; 1741) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24KDB121; 1742) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 24LDB383; 1743) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25ADB757; 1744) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25CDB456; 1745) DYNJ64009D, UDI-DI: 10195327483920(each), 40195327483921(case), Lot Number: 25FDB361; 1746) DYNJ64164, UDI-DI: 10193489792034(each), 40193489792035(case), Lot Number: 21ABH841; 1747) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21BBM393; 1748) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21DBV412; 1749) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21FBK920; 1750) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21GBU418; 1751) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21IBG214; 1752) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 21OBA733; 1753) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22BBJ435; 1754) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22BBM524; 1755) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22DMD228; 1756) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22EMA712; 1757) DYNJ64164A, UDI-DI: 10193489825930(each), 40193489825931(case), Lot Number: 22FMG702; 1758) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 22JMI296; 1759) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23BMC271; 1760) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23EMB852; 1761) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23FME303; 1762) DYNJ64164B, UDI-DI: 10195327254834(each), 40195327254835(case), Lot Number: 23GMH157; 1763) DYNJ64165C, UDI-DI: 10193489881097(each), 40193489881098(case), Lot Number: 23CMC301; 1764) DYNJ64171F, UDI-DI: 10195327687731(each), 40195327687732(case), Lot Number: 24DMI715; 1765) DYNJ64171F, UDI-DI: 10195327687731(each), 40195327687732(case), Lot Number: 24JMB595; 1766) DYNJ64171F, UDI-DI: 10195327687731(each), 40195327687732(case), Lot Number: 24JMF979; 1767) DYNJ64171F, UDI-DI: 10195327687731(each), 40195327687732(case), Lot Number: 24KMC852; 1768) DYNJ64171F, UDI-DI: 10195327687731(each), 40195327687732(case), Lot Number: 24LMA492; 1769) DYNJ64171G, UDI-DI: 10198459223082(each), 40198459223083(case), Lot Number: 25AMB240; 1770) DYNJ64171G, UDI-DI: 10198459223082(each), 40198459223083(case), Lot Number: 25AMI225; 1771) DYNJ64171G, UDI-DI: 10198459223082(each), 40198459223083(case), Lot Number: 25BMF725; 1772) DYNJ64171G, UDI-DI: 10198459223082(each), 40198459223083(case), Lot Number: 25EMI364; 1773) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21FME760; 1774) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21GMC645; 1775) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21HME922; 1776) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21HMG002; 1777) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 21IMC839; 1778) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 22AMG273; 1779) DYNJ64175, UDI-DI: 10193489792539 (each), 40193489792530(case), Lot Number: 22BMF445; 1780) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 22HMF357; 1781) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 23AMF292; 1782) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 23EMK235; 1783) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 23IMC744; 1784) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 24AMA298; 1785) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 24BMD087; 1786) DYNJ64175, UDI-DI: 10193489792539(each), 40193489792530(case), Lot Number: 24FMA231; 1787) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21ABN353; 1788) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21BBA813; 1789) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21CBS359; 1790) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21DBB375; 1791) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21FBG721; 1792) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21FMF696; 1793) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21HMF811; 1794) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21IMD744; 1795) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21JMG559; 1796) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 21LMD287; 1797) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 22AMA774; 1798) DYNJ64176A, UDI-DI: 10193489499599(each), 40193489499590(case), Lot Number: 22BME410; 1799) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22EMA913; 1800) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22GMC042; 1801) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22HMD635; 1802) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 22JMH309; 1803) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23BMI540; 1804) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23DMF252; 1805) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23GMB321; 1806) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 23LMD889; 1807) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 24CMB056; 1808) DYNJ64176B, UDI-DI: 10195327157388(each), 40195327157389(case), Lot Number: 24DMI142; 1809) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 24HME754; 1810) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 24IMC876; 1811) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 24KME193; 1812) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 24LME469; 1813) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 25EMA046; 1814) DYNJ64176C, UDI-DI: 10198459076411(each), 40198459076412(case), Lot Number: 25FMH125; 1815) DYNJ64510A, UDI-DI: 10193489971057(each), 40193489971058(case), Lot Number: 23CBC285; 1816) DYNJ64527D, UDI-DI: 10198459234729(each), 40198459234720(case), Lot Number: 25DBN471; 1817) DYNJ64527D, UDI-DI: 10198459234729(each), 40198459234720(case), Lot Number: 25FBK195; 1818) DYNJ64603B, UDI-DI: 10195327284749(each), 40195327284740(case), Lot Number: 23CBA277; 1819) DYNJ64733, UDI-DI: 10193489216455(each), 40193489216456(case), Lot Number: 21DDB517; 1820) DYNJ64741A, UDI-DI: 10193489456004(each), 40193489456005(case), Lot Number: 21DDA364; 1821) DYNJ64741A, UDI-DI: 10193489456004(each), 40193489456005(case), Lot Number: 21DDA609; 1822) DYNJ65065A, UDI-DI: 10193489317244(each), 40193489317245(case), Lot Number: 21BBC473; 1823) DYNJ65066B, UDI-DI: 10193489317183(each), 40193489317184(case), Lot Number: 21ABW917; 1824) DYNJ65108A, UDI-DI: 10193489822144(each), 40193489822145(case), Lot Number: 21BBD365; 1825) DYNJ65164C, UDI-DI: 10195327246693(each), 40195327246694(case), Lot Number: 22IBE647; 1826) DYNJ65319B, UDI-DI: 10193489869040(each), 40193489869041(case), Lot Number: 23EBR728; 1827) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23ABT634; 1828) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23BBM829; 1829) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23EBJ166; 1830) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23FBV950; 1831) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23HBA343; 1832) DYNJ65322B, UDI-DI: 10193489869071(each), 40193489869072(case), Lot Number: 23IBG967; 1833) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 23LBT407; 1834) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24BBP255; 1835) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24CBO129; 1836) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24EBP995; 1837) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24GBS754; 1838) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24JBJ097; 1839) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24KBD923; 1840) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 24LBK259; 1841) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 25BBC387; 1842) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 25CBT826; 1843) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 25EBT577; 1844) DYNJ65322C, UDI-DI: 10195327565275(each), 40195327565276(case), Lot Number: 25GBL352; 1845) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21FBI078; 1846) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21IBT090; 1847) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 21KBG570; 1848) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 22ABV290; 1849) DYNJ65395, UDI-DI: 10193489271607(each), 40193489271608(case), Lot Number: 22DBB600; 1850) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22GBW912; 1851) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22IBB740; 1852) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 22IBV229; 1853) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23ABG837; 1854) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23EBT831; 1855) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23IBG477; 1856) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 23JBA736; 1857) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24ABM648; 1858) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24CBB587; 1859) DYNJ65395A, UDI-DI: 10195327174156(each), 40195327174157(case), Lot Number: 24FBQ247; 1860) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24HLA694; 1861) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24HLA763; 1862) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 24ILA811; 1863) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25BLA191; 1864) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLA213; 1865) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLA456; 1866) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25CLB119; 1867) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25GLA117; 1868) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25HLA834; 1869) DYNJ65395B, UDI-DI: 10198459066443(each), 40198459066444(case), Lot Number: 25JBX795; 1870) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21ABJ639; 1871) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21EBE294; 1872) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21GBD027; 1873) DYNJ65603, UDI-DI: 10193489281934(each), 40193489281935(case), Lot Number: 21HBN607; 1874) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21ABS730; 1875) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21LBU977; 1876) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 21VBC713; 1877) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22ABM177; 1878) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22JMD985; 1879) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22LMB610; 1880) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 22OBD050; 1881) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23BMH039; 1882) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23EMB961; 1883) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23EMC290; 1884) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 23KMB736; 1885) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24CMH438; 1886) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24FMG725; 1887) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24IMD115; 1888) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24JMD057; 1889) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 24JMJ624; 1890) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 25CME931; 1891) DYNJ65622, UDI-DI: 10193489281613(each), 40193489281614(case), Lot Number: 25HMH305; 1892) DYNJ65730A, UDI-DI: 10193489958270(each), 40193489958271(case), Lot Number: 22OBC403; 1893) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 23JMI202; 1894) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 23KMI663; 1895) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24AMB121; 1896) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24BMC898; 1897) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24BMJ136; 1898) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24CME387; 1899) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24DMD371; 1900) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24DMK513; 1901) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24GMC894; 1902) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24GMJ544; 1903) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24HMH127; 1904) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24IMI316; 1905) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24JMI243; 1906) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 24KMF425; 1907) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25AMI183; 1908) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25BMD281; 1909) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25BMD281A; 1910) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25CMG189; 1911) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25DMF044; 1912) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25EMB299; 1913) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25FMC631; 1914) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25GMJ308; 1915) DYNJ65826A, UDI-DI: 10195327516192(each), 40195327516193(case), Lot Number: 25KMA265; 1916) DYNJ65934A, UDI-DI: 10193489455359(each), 40193489455350(case), Lot Number: 21BBF325; 1917) DYNJ65958A, UDI-DI: 10193489481624(each), 40193489481625(case), Lot Number: 21BBN445; 1918) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 22LBO308; 1919) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23BBF990; 1920) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23CBB374; 1921) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23DBM741; 1922) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23GBP815; 1923) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23HBL398; 1924) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23HBY749; 1925) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23IBT769; 1926) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23JBA683; 1927) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23KBB176; 1928) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 23LBT033; 1929) DYNJ66023A, UDI-DI: 10195327151102(each), 40195327151103(case), Lot Number: 24BBQ965; 1930) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24DBT598; 1931) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24DBU552; 1932) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24FBN384; 1933) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24HBF662; 1934) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24JBN810; 1935) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24KBK253; 1936) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 24KBR232; 1937) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 25ABQ752; 1938) DYNJ66023B, UDI-DI: 10195327680749(each), 40195327680740(case), Lot Number: 25BBF183; 1939) DYNJ66138B, UDI-DI: 10198459482359(each), 40198459482350(case), Lot Number: 25HMG871; 1940) DYNJ66572, UDI-DI: 10193489360936(each), 40193489360937(case), Lot Number: 21BBS046; 1941) DYNJ66572, UDI-DI: 10193489360936(each), 40193489360937(case), Lot Number: 21BBS830; 1942) DYNJ66572A, UDI-DI: 10193489869460(each), 40193489869461(case), Lot Number: 21LBJ616; 1943) DYNJ66572A, UDI-DI: 10193489869460(each), 40193489869461(case), Lot Number: 22CBF350; 1944) DYNJ66572B, UDI-DI: 10195327244187(each), 40195327244188(case), Lot Number: 23ABS704; 1945) DYNJ66572B, UDI-DI: 10195327244187(each), 40195327244188(case), Lot Number: 23CBL488; 1946) DYNJ66682, UDI-DI: 10193489370294(each), 40193489370295(case), Lot Number: 21CBJ914; 1947) DYNJ66940B, UDI-DI: 10195327305253(each), 40195327305254(case), Lot Number: 23IBB170; 1948) DYNJ66940B, UDI-DI: 10195327305253(each), 40195327305254(case), Lot Number: 23IBT566; 1949) DYNJ66945B, UDI-DI: 10195327305246(each), 40195327305247(case), Lot Number: 23CBD537; 1950) DYNJ66945B, UDI-DI: 10195327305246(each), 40195327305247(case), Lot Number: 23EBR636; 1951) DYNJ67141A, UDI-DI: 10195327024215(each), 40195327024216(case), Lot Number: 23HBS080; 1952) DYNJ67241, UDI-DI: 10193489392227(each), 40193489392228(case), Lot Number: 23HBH208; 1953) DYNJ67241, UDI-DI: 10193489392227(each), 40193489392228(case), Lot Number: 23LBD765; 1954) DYNJ67278, UDI-DI: 10193489394757(each), 40193489394758(case), Lot Number: 23HBM356; 1955) DYNJ67303A, UDI-DI: 10195327015664(each), 40195327015665(case), Lot Number: 23JBV806; 1956) DYNJ67371, UDI-DI: 10193489889758(each), 40193489889759(case), Lot Number: 22DBE238; 1957) DYNJ67444B, UDI-DI: 10195327134648(each), 40195327134649(case), Lot Number: 23HBS081; 1958) DYNJ67502, UDI-DI: 10193489404173(each), 40193489404174(case), Lot Number: 21BBL566; 1959) DYNJ67509, UDI-DI: 10193489404388(each), 40193489404389(case), Lot Number: 21BBP851; 1960) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23EMI872; 1961) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23HME139; 1962) DYNJ67652A, UDI-DI: 10195327274542(each), 40195327274543(case), Lot Number: 23HMG482; 1963) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 23KMF780; 1964) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24BMD330; 1965) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24CMF453; 1966) DYNJ67652B, UDI-DI: 10195327491246(each), 40195327491247(case), Lot Number: 24FMJ102; 1967) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 24LMA719; 1968) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 25AMF026; 1969) DYNJ67652C, UDI-DI: 10198459170201(each), 40198459170202(case), Lot Number: 25DMA359; 1970) DYNJ67711A, UDI-DI: 10195327026370(each), 40195327026371(case), Lot Number: 23BBP132; 1971) DYNJ67814B, UDI-DI: 10195327135676(each), 40195327135677(case), Lot Number: 23KBT429; 1972) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 23KBH486; 1973) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 23LBU562; 1974) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24ABV046; 1975) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24CBD349; 1976) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24EBD903; 1977) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24FBS410; 1978) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24JBK383; 1979) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24KBB774; 1980) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25BBQ467; 1981) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25BBT615; 1982) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25GBL311; 1983) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25HBK166; 1984) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25IBJ392; 1985) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25JBX784; 1986) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 25LBN320; 1987) DYNJ67908A, UDI-DI: 10193489873702(each), 40193489873703(case), Lot Number: 23KMF500; 1988) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21BBB519; 1989) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21CBK754; 1990) DYNJ67961, UDI-DI: 10193489460179(each), 40193489460170(case), Lot Number: 21CBO800; 1991) DYNJ68052B, UDI-DI: 10195327588441(each), 40195327588442(case), Lot Number: 24CLA962; 1992) DYNJ68308, UDI-DI: 10193489821413(each), 40193489821414(case), Lot Number: 23IBD739; 1993) DYNJ68320, UDI-DI: 10193489822991(each), 40193489822992(case), Lot Number: 21BBB618; 1994) DYNJ68430, UDI-DI: 10193489465211(each), 40193489465212(case), Lot Number: 23IBB159; 1995) DYNJ68720, UDI-DI: 10193489843071(each), 40193489843072(case), Lot Number: 23GBG686; 1996) DYNJ68980, UDI-DI: 10193489870817(each), 40193489870818(case), Lot Number: 23CBU626; 1997) DYNJ69131B, UDI-DI: 10195327173678(each), 40195327173679(case), Lot Number: 23EBP597; 1998) DYNJ69317A, UDI-DI: 10195327224332(each), 40195327224333(case), Lot Number: 23BMH343; 1999) DYNJ69335B, UDI-DI: 10195327248734(each), 40195327248735(case), Lot Number: 23BMC736; 2000) DYNJ69335B, UDI-DI: 10195327248734(each), 40195327248735(case), Lot Number: 23BMD638; 2001) DYNJ69335B, UDI-DI: 10195327248734(each), 40195327248735(case), Lot Number: 23DMC718; 2002) DYNJ69335C, UDI-DI: 10195327552534(each), 40195327552535(case), Lot Number: 23LMG536; 2003) DYNJ69335C, UDI-DI: 10195327552534(each), 40195327552535(case), Lot Number: 24AME063; 2004) DYNJ69335C, UDI-DI: 10195327552534(each), 40195327552535(case), Lot Number: 24BMC413; 2005) DYNJ69371, UDI-DI: 10193489935028(each), 40193489935029(case), Lot Number: 23KBN032; 2006) DYNJ69430C, UDI-DI: 10195327481810(each), 40195327481811(case), Lot Number: 23HBI271; 2007) DYNJ69430C, UDI-DI: 10195327481810(each), 40195327481811(case), Lot Number: 23JBM683; 2008) DYNJ69430C, UDI-DI: 10195327481810(each), 40195327481811(case), Lot Number: 23LBP211; 2009) DYNJ69563A, UDI-DI: 10195327226138(each), 40195327226139(case), Lot Number: 23KBA130; 2010) DYNJ69915B, UDI-DI: 10195327597184(each), 40195327597185(case), Lot Number: 24JMI703; 2011) DYNJ69934, UDI-DI: 10193489997811(each), 40193489997812(case), Lot Number: 22HME118; 2012) DYNJ80071, UDI-DI: 10195327010126(each), 40195327010127(case), Lot Number: 23CBV608; 2013) DYNJ80085, UDI-DI: 10195327010065(each), 40195327010066(case), Lot Number: 23CBV727; 2014) DYNJ80130A, UDI-DI: 10195327410896(each), 40195327410897(case), Lot Number: 23GMF021; 2015) DYNJ80192, UDI-DI: 10195327014247(each), 40195327014248(case), Lot Number: 21KBI313; 2016) DYNJ80205, UDI-DI: 10195327016470(each), 40195327016471(case), Lot Number: 24BBK505; 2017) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23JBS811; 2018) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23KBR833; 2019) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 23LBP849; 2020) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24ABB561; 2021) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24DBD641; 2022) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24HBF212; 2023) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24IBQ276; 2024) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 24KBU010; 2025) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 25ABB722; 2026) DYNJ80410A, UDI-DI: 10195327528300(each), 40195327528301(case), Lot Number: 25CBT101; 2027) DYNJ80464, UDI-DI: 10195327039639(each), 40195327039630(case), Lot Number: 23CBH588; 2028) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 23JBK792; 2029) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 23KBI250; 2030) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 24BBF015; 2031) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 24DBJ635; 2032) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 25BBN761; 2033) DYNJ80645A, UDI-DI: 10195327528010(each), 40195327528011(case), Lot Number: 25CBB594; 2034) DYNJ80879D, UDI-DI: 10195327428976(each), 40195327428977(case), Lot Number: 23JBW728; 2035) DYNJ80879D, UDI-DI: 10195327428976(each), 40195327428977(case), Lot Number: 23LBM359; 2036) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 22DBA249; 2037) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 22KBI778; 2038) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23CBF631; 2039) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23CBG415; 2040) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23EBN355; 2041) DYNJ81014, UDI-DI: 10195327082895(each), 40195327082896(case), Lot Number: 23FBR818; 2042) DYNJ81014A, UDI-DI: 10195327495848(each), 40195327495849(case), Lot Number: 23JBH396; 2043) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23FBV730; 2044) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23IBB042; 2045) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23JBV322; 2046) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 23KBS890; 2047) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 24DBH527; 2048) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 24GBW743; 2049) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 24KBI826; 2050) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 24LBN989; 2051) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 25CBL521; 2052) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 25EBQ081; 2053) DYNJ81336A, UDI-DI: 10195327450304(each), 40195327450305(case), Lot Number: 25FBS126; 2054) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22EBG823; 2055) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22GBZ212; 2056) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22HBW957; 2057) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 22JBM301; 2058) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 23ABU291; 2059) DYNJ81518, UDI-DI: 10195327120986(each), 40195327120987(case), Lot Number: 23CBD541; 2060) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22EME413; 2061) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22HMB592; 2062) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 22HMB593; 2063) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23CMI084; 2064) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23EME110; 2065) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23FMF387; 2066) DYNJ81859, UDI-DI: 10195327152291(each), 40195327152292(case), Lot Number: 23HME214; 2067) DYNJ81905, UDI-DI: 10195327154998(each), 40195327154999(case), Lot Number: 22GBA833; 2068) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23DBP302; 2069) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23GBS218; 2070) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23IBO987; 2071) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23JBU264; 2072) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23KMB630; 2073) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 23LMB675; 2074) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24AME898; 2075) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24BME257; 2076) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMJ076; 2077) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMK399; 2078) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24DMK399A; 2079) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24EMG933; 2080) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24HMH952; 2081) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24IMI184; 2082) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 24KMG661; 2083) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25AMH197; 2084) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25BMB151; 2085) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25CMH914; 2086) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25EMA396; 2087) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25GMF869; 2088) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25IMD543; 2089) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25JMH386; 2090) DYNJ81905B, UDI-DI: 10195327377564(each), 40195327377565(case), Lot Number: 25LME657; 2091) DYNJ81921, UDI-DI: 10195327155544(each), 40195327155545(case), Lot Number: 22HBD171; 2092) DYNJ81947, UDI-DI: 10195327158118(each), 40195327158119(case), Lot Number: 23BBC236; 2093) DYNJ81947, UDI-DI: 10195327158118(each), 40195327158119(case), Lot Number: 23CBS299; 2094) DYNJ81947A, UDI-DI: 10195327401917(each), 40195327401918(case), Lot Number: 23EBL091; 2095) DYNJ82742, UDI-DI: 10195327207588(each), 40195327207589(case), Lot Number: 24AMI098; 2096) DYNJ82784, UDI-DI: 10195327210267(each), 40195327210268(case), Lot Number: 22IBT259; 2097) DYNJ82785, UDI-DI: 10195327210274(each), 40195327210275(case), Lot Number: 22JBG012; 2098) DYNJ83008, UDI-DI: 10195327226237(each), 40195327226238(case), Lot Number: 23GBT347; 2099) DYNJ83181, UDI-DI: 10195327240080(each), 40195327240081(case), Lot Number: 22LBD470; 2100) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 22KBI755; 2101) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23EBC088; 2102) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23FBC442; 2103) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23GBF023; 2104) DYNJ83234, UDI-DI: 10195327247782(each), 40195327247783(case), Lot Number: 23IBL690; 2105) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 22KBR110; 2106) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23BBS888; 2107) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23GBM802; 2108) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23IBO006; 2109) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 23LBG771; 2110) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 24EBE829; 2111) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 24HBR116; 2112) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 24JBK393; 2113) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 25BBU750; 2114) DYNJ83236, UDI-DI: 10195327247843(each), 40195327247844(case), Lot Number: 25EBN288; 2115) DYNJ83236A, UDI-DI: 10198459364297(each), 40198459364298(case), Lot Number: 25GBV516; 2116) DYNJ83267B, UDI-DI: 10195327342029(each), 40195327342020(case), Lot Number: 23EBD858; 2117) DYNJ83267B, UDI-DI: 10195327342029(each), 40195327342020(case), Lot Number: 25FBO888; 2118) DYNJ83315, UDI-DI: 10195327289362(each), 40195327289363(case), Lot Number: 23HBT636; 2119) DYNJ83323, UDI-DI: 10195327260293(each), 40195327260294(case), Lot Number: 23IBB023; 2120) DYNJ83401, UDI-DI: 10195327264314(each), 40195327264315(case), Lot Number: 23ADA122; 2121) DYNJ83401, UDI-DI: 10195327264314(each), 40195327264315(case), Lot Number: 23CDC415; 2122) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23DDB319; 2123) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23EDA014; 2124) DYNJ83401A, UDI-DI: 10195327379568(each), 40195327379569(case), Lot Number: 23KDB925; 2125) DYNJ83549, UDI-DI: 10195327278687(each), 40195327278688(case), Lot Number: 23JBT121; 2126) DYNJ83683, UDI-DI: 10195327302498(each), 40195327302499(case), Lot Number: 23DBE880; 2127) DYNJ83761, UDI-DI: 10195327301750(each), 40195327301751(case), Lot Number: 23BBP136; 2128) DYNJ83948, UDI-DI: 10195327318154(each), 40195327318155(case), Lot Number: 23KBT912; 2129) DYNJ83999, UDI-DI: 10195327325428(each), 40195327325429(case), Lot Number: 23DLB004; 2130) DYNJ84473, UDI-DI: 10195327368975(each), 40195327368976(case), Lot Number: 23ILA147; 2131) DYNJ84540, UDI-DI: 10195327375720(each), 40195327375721(case), Lot Number: 23GBJ944; 2132) DYNJ84648A, UDI-DI: 10195327449940(each), 40195327449941(case), Lot Number: 23HBQ821; 2133) DYNJ84937B, UDI-DI: 10198459089510(each), 40198459089511(case), Lot Number: 24KMB597; 2134) DYNJ84972, UDI-DI: 10195327442651(each), 40195327442652(case), Lot Number: 25BBC532; 2135) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 23IBH150; 2136) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 24DBT813; 2137) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25ABM165; 2138) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25GBK279; 2139) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25IBQ365; 2140) DYNJ85007, UDI-DI: 10195327445348(each), 40195327445349(case), Lot Number: 25KBF640; 2141) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 23IBK215; 2142) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 24KBB114; 2143) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25CBL547; 2144) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25EBK149; 2145) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25FBH505; 2146) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25HBA955; 2147) DYNJ85008, UDI-DI: 10195327445362(each), 40195327445363(case), Lot Number: 25KBL958; 2148) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 23IBK219; 2149) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 24DBO515; 2150) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 24LBA279; 2151) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25ABM160; 2152) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25DBL816; 2153) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25GBN428; 2154) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25GBX104; 2155) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25IBP658; 2156) DYNJ85009, UDI-DI: 10195327445355(each), 40195327445356(case), Lot Number: 25JBM584; 2157) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 23IBH149; 2158) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24DBA418; 2159) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24GBC603; 2160) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24HBS004; 2161) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 24JBE493; 2162) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25CBD913; 2163) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25EBH468; 2164) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25GBU913; 2165) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25KBJ312; 2166) DYNJ85030, UDI-DI: 10195327448288(each), 40195327448289(case), Lot Number: 25LBM028; 2167) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 23IBL567; 2168) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24ABU575; 2169) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24ABW604; 2170) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24DBE599; 2171) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24FBE223; 2172) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24GBA885; 2173) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24GBJ336; 2174) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24HBP882; 2175) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 24KBM440; 2176) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25ABS847; 2177) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25DBD521; 2178) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25FBN620; 2179) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25GBR600; 2180) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25JBJ844; 2181) DYNJ85155, UDI-DI: 10195327469146(each), 40195327469147(case), Lot Number: 25LBQ771; 2182) DYNJ85246, UDI-DI: 10195327482794(each), 40195327482795(case), Lot Number: 23KBS394; 2183) DYNJ85246, UDI-DI: 10195327482794(each), 40195327482795(case), Lot Number: 24JBA875; 2184) DYNJ85453, UDI-DI: 10195327504380(each), 40195327504381(case), Lot Number: 23LBD246; 2185) DYNJ85453, UDI-DI: 10195327504380(each), 40195327504381(case), Lot Number: 24GBB738; 2186) DYNJ85902, UDI-DI: 10195327570224(each), 40195327570225(case), Lot Number: 24ABW109; 2187) DYNJ85904, UDI-DI: 10195327569204(each), 40195327569205(case), Lot Number: 24ABW111; 2188) DYNJ86358, UDI-DI: 10195327623920(each), 40195327623921(case), Lot Number: 24DBT962; 2189) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 24LBD035; 2190) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25BBG889; 2191) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25CBD157; 2192) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25CBL656; 2193) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25DBI319; 2194) DYNJ86358A, UDI-DI: 10198459203091(each), 40198459203092(case), Lot Number: 25EBR951; 2195) DYNJ87523, UDI-DI: 10198459040511(each), 40198459040512(case), Lot Number: 24HBC472; 2196) DYNJ87526, UDI-DI: 10198459040177(each), 40198459040178(case), Lot Number: 24HBQ893; 2197) DYNJ87530, UDI-DI: 10198459040207(each), 40198459040208(case), Lot Number: 24HBO727; 2198) DYNJ87536, UDI-DI: 10198459042782(each), 40198459042783(case), Lot Number: 24IBQ456; 2199) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 24HBB599; 2200) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 25CBS110; 2201) DYNJ87589, UDI-DI: 10198459046148(each), 40198459046149(case), Lot Number: 25CBT519; 2202) DYNJ87922, UDI-DI: 10198459078668(each), 40198459078669(case), Lot Number: 24HBH778; 2203) DYNJ87922, UDI-DI: 10198459078668(each), 40198459078669(case), Lot Number: 24JBC382; 2204) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 24LBK468; 2205) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25CBC626; 2206) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25GBG555; 2207) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25GBX075; 2208) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25JBR829; 2209) DYNJ87922A, UDI-DI: 10198459211713(each), 40198459211714(case), Lot Number: 25LBI670; 2210) DYNJ88343, UDI-DI: 10198459146541(each), 40198459146542(case), Lot Number: 25FBN319; 2211) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25BBO562; 2212) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25CBS371; 2213) DYNJ88795, UDI-DI: 10198459236532(each), 40198459236533(case), Lot Number: 25JBW661; 2214) DYNJ900312F, UDI-DI: 10195327698010(each), 40195327698011(case), Lot Number: 24LMH420; 2215) DYNJ900473F, UDI-DI: 10195327448141(each), 40195327448142(case), Lot Number: 23IMD804; 2216) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24DBR216; 2217) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24EBI630; 2218) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24HBR222; 2219) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24IBP307; 2220) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 24LBH537; 2221) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25ABA922; 2222) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25BBM255; 2223) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25BBU433; 2224) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25EBN039; 2225) DYNJ900511O, UDI-DI: 10195327658304(each), 40195327658305(case), Lot Number: 25GBJ998; 2226) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21BDB150; 2227) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21DDB318; 2228) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21FDA216; 2229) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21GDA613; 2230) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21HDB766; 2231) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21JDB219; 2232) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21LDA249; 2233) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 21LDB775; 2234) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22DDA354; 2235) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22DDC460; 2236) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22EDB848; 2237) DYNJ900637D, UDI-DI: 10193489298321(each), 40193489298322(case), Lot Number: 22FDB319; 2238) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 22IDB505; 2239) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 22KDA621; 2240) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23ADB159; 2241) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23BDB808; 2242) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23CDC445; 2243) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23EDB016; 2244) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23FDB393; 2245) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23GDC145; 2246) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23IDA598; 2247) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 23KDB595; 2248) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24ADB608; 2249) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24CDA343; 2250) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24DDB816; 2251) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24FDA470; 2252) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24GDB290; 2253) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24JDA457; 2254) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24KDB535; 2255) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 24LDA688; 2256) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25DDA252; 2257) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25DDA830; 2258) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25EDC048; 2259) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25FDB179; 2260) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25FDB252; 2261) DYNJ900637F, UDI-DI: 10195327194345(each), 40195327194346(case), Lot Number: 25HDA524; 2262) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21BDB155; 2263) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21FDB070; 2264) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21GDA051; 2265) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21JDB654; 2266) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 21KDB540; 2267) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22ADA779; 2268) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22DDB174; 2269) DYNJ900638D, UDI-DI: 10193489298338(each), 40193489298339(case), Lot Number: 22EDA396; 2270) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23LDA701; 2271) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22GDB004; 2272) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22GDB557; 2273) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22IDA935; 2274) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22KDA430; 2275) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 22LDA717; 2276) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23BDA103; 2277) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23CDB246; 2278) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23EDA531; 2279) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23EDC178; 2280) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23GDA643; 2281) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23JDA516; 2282) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 23LDC079; 2283) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24ADA491; 2284) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24ADC253; 2285) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24DDA394; 2286) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24GDB592; 2287) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24IDA052; 2288) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24JDB500; 2289) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 24LDA617; 2290) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25BDB289; 2291) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25CDB600; 2292) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25DDB169; 2293) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25EDB071; 2294) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25FDA212; 2295) DYNJ900638F, UDI-DI: 10195327194352(each), 40195327194353(case), Lot Number: 25GDA546; 2296) DYNJ901293C, UDI-DI: 10193489210934(each), 40193489210935(case), Lot Number: 20LBO414; 2297) DYNJ901293C, UDI-DI: 10193489210934(each), 40193489210935(case), Lot Number: 21BBQ036; 2298) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21DBP543; 2299) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21GBF747; 2300) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21HBI249; 2301) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21JBK076; 2302) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21KBR503; 2303) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21LBB250; 2304) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 21LBQ931; 2305) DYNJ901293D, UDI-DI: 10193489876512(each), 40193489876513(case), Lot Number: 22GBA672; 2306) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBB795; 2307) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBK367; 2308) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 22KBK712; 2309) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23BBF811; 2310) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23BBU922; 2311) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23DBE024; 2312) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23EBD058; 2313) DYNJ901293F, UDI-DI: 10195327194000(each), 40195327194001(case), Lot Number: 23EBF123; 2314) DYNJ901293G, UDI-DI: 10195327405267(each), 40195327405268(case), Lot Number: 23JBM583; 2315) DYNJ901293G, UDI-DI: 10195327405267(each), 40195327405268(case), Lot Number: 23KBX231; 2316) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24CBM164; 2317) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24EBO749; 2318) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24HBK597; 2319) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24IBM824; 2320) DYNJ901293I, UDI-DI: 10195327628444(each), 40195327628445(case), Lot Number: 24KBB223; 2321) DYNJ901293J, UDI-DI: 10198459210525(each), 40198459210526(case), Lot Number: 25ABB962; 2322) DYNJ901293J, UDI-DI: 10198459210525(each), 40198459210526(case), Lot Number: 25FBV678; 2323) DYNJ901294D, UDI-DI: 10193489201055(each), 40193489201056(case), Lot Number: 21CBM996; 2324) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 22JBI903; 2325) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23BBF812; 2326) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23DBS344; 2327) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23FBO393; 2328) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23HBJ256; 2329) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23IBI898; 2330) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23JBN868; 2331) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 23KBW620; 2332) DYNJ901294F, UDI-DI: 10195327194017(each), 40195327194018(case), Lot Number: 24ABO438; 2333) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24DBE522; 2334) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24EBM171; 2335) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24FBL612; 2336) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24IBB759; 2337) DYNJ901294G, UDI-DI: 10195327628420(each), 40195327628421(case), Lot Number: 24KBP826; 2338) DYNJ901294I, UDI-DI: 10198459210471(each), 40198459210472(case), Lot Number: 25ABO543; 2339) DYNJ901294I, UDI-DI: 10198459210471(each), 40198459210472(case), Lot Number: 25FBR246; 2340) DYNJ901294I, UDI-DI: 10198459210471(each), 40198459210472(case), Lot Number: 25HBO825; 2341) DYNJ901312G, UDI-DI: 10195327462543(each), 40195327462544(case), Lot Number: 23IMA558; 2342) DYNJ901364G, UDI-DI: 10193489656060(each), 40193489656061(case), Lot Number: 24JLA825; 2343) DYNJ901364I, UDI-DI: 10198459224874(each), 40198459224875(case), Lot Number: 25ILA425; 2344) DYNJ901485J, UDI-DI: 10195327284411(each), 40195327284412(case), Lot Number: 23HMI283; 2345) DYNJ901486F, UDI-DI: 10195327284428(each), 40195327284429(case), Lot Number: 23HMH007; 2346) DYNJ901515C, UDI-DI: 10193489732436(each), 40193489732437(case), Lot Number: 21CBV915; 2347) DYNJ901752M, UDI-DI: 10193489604580(each), 40193489604581(case), Lot Number: 21ADB750; 2348) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24GBC321; 2349) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24HBU924; 2350) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 24LBN814; 2351) DYNJ901883K, UDI-DI: 10195327687618(each), 40195327687619(case), Lot Number: 25ABN022; 2352) DYNJ901883M, UDI-DI: 10198459288807(each), 40198459288808(case), Lot Number: 25DBR925; 2353) DYNJ901883M, UDI-DI: 10198459288807(each), 40198459288808(case), Lot Number: 25GBI877; 2354) DYNJ902019J, UDI-DI: 10195327394424(each), 40195327394425(case), Lot Number: 23IBP386; 2355) DYNJ902019J, UDI-DI: 10195327394424(each), 40195327394425(case), Lot Number: 23IBP535; 2356) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 24GBP777; 2357) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 24HBR275; 2358) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 24KBP846; 2359) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 25BBI987; 2360) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 25DBG926; 2361) DYNJ902022J, UDI-DI: 10195327659301(each), 40195327659302(case), Lot Number: 25GBC117; 2362) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21BBQ998; 2363) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21DBC079; 2364) DYNJ902247D, UDI-DI: 10193489332544(each), 40193489332545(case), Lot Number: 21GBE547; 2365) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22ABT877; 2366) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22DBM174; 2367) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22FBG142; 2368) DYNJ902247F, UDI-DI: 10193489948417(each), 40193489948418(case), Lot Number: 22OBA319; 2369) DYNJ902247G, UDI-DI: 10195327189587(each), 40195327189588(case), Lot Number: 22IBN886; 2370) DYNJ902247G, UDI-DI: 10195327189587(each), 40195327189588(case), Lot Number: 22KBS876; 2371) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23ABP539; 2372) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23FBF178; 2373) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23GBI471; 2374) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23JBX532; 2375) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 23LBN371; 2376) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24BBJ347; 2377) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24DBB200; 2378) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24IBB754; 2379) DYNJ902247I, UDI-DI: 10195327314675(each), 40195327314676(case), Lot Number: 24JBS825; 2380) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 24LBP686; 2381) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25ABC841; 2382) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25BBM532; 2383) DYNJ902247J, UDI-DI: 10198459188428(each), 40198459188429(case), Lot Number: 25FBL542; 2384) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25JBD155; 2385) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25JBO058; 2386) DYNJ902247K, UDI-DI: 10198459494772(each), 40198459494773(case), Lot Number: 25KBE472; 2387) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 21IBC819; 2388) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 21LBC621; 2389) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22ABH933; 2390) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22ABH935; 2391) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22DBL831; 2392) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22FBW219; 2393) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22GBF980; 2394) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22JBE981; 2395) DYNJ902308C, UDI-DI: 10193489242669(each), 40193489242660(case), Lot Number: 22OBM158; 2396) DYNJ902308D, UDI-DI: 10195327276737(each), 40195327276738(case), Lot Number: 22LBF699; 2397) DYNJ902308F, UDI-DI: 10195327326937(each), 40195327326938(case), Lot Number: 23DBF023; 2398) DYNJ902308F, UDI-DI: 10195327326937(each), 40195327326938(case), Lot Number: 23DBJ927; 2399) DYNJ902308G, UDI-DI: 10195327412173(each), 40195327412174(case), Lot Number: 23FBH153; 2400) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23IBJ191; 2401) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23KBT585; 2402) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 23LBT893; 2403) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24ABR435; 2404) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24BBG964; 2405) DYNJ902308I, UDI-DI: 10195327458522(each), 40195327458523(case), Lot Number: 24DBU297; 2406) DYNJ902310D, UDI-DI: 10193489315257(each), 40193489315258(case), Lot Number: 21ADB189; 2407) DYNJ902310D, UDI-DI: 10193489315257(each), 40193489315258(case), Lot Number: 21ADB455; 2408) DYNJ902321M, UDI-DI: 10193489863956(each), 40193489863957(case), Lot Number: 21DKA251; 2409) DYNJ902348M, UDI-DI: 10195327510527(each), 40195327510528(case), Lot Number: 23KDB412; 2410) DYNJ902352D, UDI-DI: 10889942804177(each), 40889942804178(case), Lot Number: 21BDA474; 2411) DYNJ902355L, UDI-DI: 10195327307097(each), 40195327307098(case), Lot Number: 25BDB788; 2412) DYNJ902362I, UDI-DI: 10195327510497(each), 40195327510498(case), Lot Number: 24ADB157; 2413) DYNJ902362I, UDI-DI: 10195327510497(each), 40195327510498(case), Lot Number: 24BDA089; 2414) DYNJ902366A, UDI-DI: 10193489616255(each), 40193489616256(case), Lot Number: 21CBM977; 2415) DYNJ902366B, UDI-DI: 10195327194024(each), 40195327194025(case), Lot Number: 23EBD057; 2416) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23FBR425; 2417) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23GBU902; 2418) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23IBA505; 2419) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23IBU154; 2420) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23JBQ231; 2421) DYNJ902366C, UDI-DI: 10195327405274(each), 40195327405275(case), Lot Number: 23LBH351; 2422) DYNJ902438J, UDI-DI: 10193489761450(each), 40193489761451(case), Lot Number: 21ADB962; 2423) DYNJ902438J, UDI-DI: 10193489761450(each), 40193489761451(case), Lot Number: 21CDB481; 2424) DYNJ902473L, UDI-DI: 10195327261436(each), 40195327261437(case), Lot Number: 23KLA364; 2425) DYNJ902473L, UDI-DI: 10195327261436(each), 40195327261437(case), Lot Number: 23KLA717; 2426) DYNJ902567I, UDI-DI: 10195327388577(each), 40195327388578(case), Lot Number: 23HMG997; 2427) DYNJ902665O, UDI-DI: 10195327421380(each), 40195327421381(case), Lot Number: 23IMI035; 2428) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 21FMB735; 2429) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 21GMF509; 2430) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22CMD603; 2431) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22EMD118; 2432) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22FMB393; 2433) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22IMH146; 2434) DYNJ902695A, UDI-DI: 10889942245673(each), 40889942245674(case), Lot Number: 22LMB722; 2435) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23GMC011; 2436) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23HME545; 2437) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23JMA960; 2438) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23JME975; 2439) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 23LMG552; 2440) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24CMG427; 2441) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24CMJ425; 2442) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24EMC395; 2443) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24EMH532; 2444) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24HMC925; 2445) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24IMF329; 2446) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 24JMA066; 2447) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25AME975; 2448) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25CMD002; 2449) DYNJ902695B, UDI-DI: 10195327362799(each), 40195327362790(case), Lot Number: 25FME478; 2450) DYNJ902916I, UDI-DI: 10195327356989(each), 40195327356980(case), Lot Number: 23JDB783; 2451) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23HBH459; 2452) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23IBN678; 2453) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23IBS140; 2454) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23JBF477; 2455) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23LBJ444; 2456) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 23LBP763; 2457) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 24BBG041; 2458) DYNJ902960K, UDI-DI: 10195327466589(each), 40195327466580(case), Lot Number: 24HBJ888; 2459) DYNJ902960L, UDI-DI: 10198459154058(each), 40198459154059(case), Lot Number: 24KBO359; 2460) DYNJ902960L, UDI-DI: 10198459154058(each), 40198459154059(case), Lot Number: 24LBB009; 2461) DYNJ902960L, UDI-DI: 10198459154058(each), 40198459154059(case), Lot Number: 25GBH874; 2462) DYNJ902960L, UDI-DI: 10198459154058(each), 40198459154059(case), Lot Number: 25HBS066; 2463) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23HBN095; 2464) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23JBH983; 2465) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23KBA719; 2466) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23LBP761; 2467) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 23LBV796; 2468) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 24BBJ136; 2469) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 24BBO491; 2470) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 24GBF149; 2471) DYNJ902961L, UDI-DI: 10195327466664(each), 40195327466665(case), Lot Number: 24HBR135; 2472) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 24KBK453; 2473) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 24LBH394; 2474) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 24LBM379; 2475) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 25ABV810; 2476) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 25BBJ598; 2477) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 25DBQ956; 2478) DYNJ902961M, UDI-DI: 10198459154065(each), 40198459154066(case), Lot Number: 25GBO123; 2479) DYNJ903053D, UDI-DI: 10193489712568(each), 40193489712569(case), Lot Number: 23AMA115; 2480) DYNJ903175C, UDI-DI: 10195327435912(each), 40195327435913(case), Lot Number: 23LBT799; 2481) DYNJ903477J, UDI-DI: 10195327565206(each), 40195327565207(case), Lot Number: 24EDA936; 2482) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23BBK637; 2483) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23EBA643; 2484) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23FBB377; 2485) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23HBE353; 2486) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23HBV125; 2487) DYNJ 903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23JBJ224; 2488) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23JBU430; 2489) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23KBM048; 2490) DYNJ903522F, UDI-DI: 10195327248444(each), 40195327248445(case), Lot Number: 23LBJ214; 2491) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24ABV428; 2492) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24CBD528; 2493) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24DBJ336; 2494) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24FBD407; 2495) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24HBP041; 2496) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24IBE099; 2497) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24IBV357; 2498) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 24JBY080; 2499) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25ABA361; 2500) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25CBU462; 2501) DYNJ903522G, UDI-DI: 10195327581121(each), 40195327581122(case), Lot Number: 25DBP020; 2502) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25JBM545; 2503) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25JBP549; 2504) DYNJ903522I, UDI-DI: 10198459566158(each), 40198459566159(case), Lot Number: 25KBT140; 2505) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21DBO218; 2506) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21FBE274; 2507) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21GBG879; 2508) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21IBP908; 2509) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21IBP909; 2510) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21KBR505; 2511) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 21LBT148; 2512) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22DBR197; 2513) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22GBE114; 2514) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22IBO773; 2515) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 22IBP136; 2516) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23CBG054; 2517) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23CBU917; 2518) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23EBL640; 2519) DYNJ903567C, UDI-DI: 10193489459517(each), 40193489459518(case), Lot Number: 23EBR769; 2520) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23HBU819; 2521) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23IBI992; 2522) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23LBN776; 2523) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 23LBV018; 2524) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 24ABF212; 2525) DYNJ903567D, UDI-DI: 10195327416164(each), 40195327416165(case), Lot Number: 24BBD430; 2526) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24CBU456; 2527) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24DBO100; 2528) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24KBG005; 2529) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24KBJ518; 2530) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24KBS228; 2531) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 24LBD933; 2532) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 25CBE607; 2533) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 25CBI827; 2534) DYNJ903567F, UDI-DI: 10195327612566(each), 40195327612567(case), Lot Number: 25DBJ920; 2535) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21ABI686; 2536) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21EBD387; 2537) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21IBK686; 2538) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 21JBQ513; 2539) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22FBW884; 2540) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22HBH661; 2541) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 22IBA639; 2542) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 23CBQ348; 2543) DYNJ903929, UDI-DI: 10889942468737(each), 40889942468738(case), Lot Number: 23EBW544; 2544) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21CBN613; 2545) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21DBM942; 2546) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21FBI614; 2547) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21FBM616; 2548) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 21IBK141; 2549) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22EBA227; 2550) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22GBS690; 2551) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22IBO536; 2552) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 22NBF998; 2553) DYNJ903930, UDI-DI: 10889942468621(each), 40889942468622(case), Lot Number: 23CBN882; 2554) DYNJ904021Q, UDI-DI: 10198459140440(each), 40198459140441(case), Lot Number: 24JDC025; 2555) DYNJ904073D, UDI-DI: 10195327597191(each), 40195327597192(case), Lot Number: 24LMF733; 2556) DYNJ904082J, UDI-DI: 10195327062002(each), 40195327062003(case), Lot Number: 22GBH238; 2557) DYNJ904083C, UDI-DI: 10193489282832(each), 40193489282833(case), Lot Number: 22KDA657; 2558) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23ABS351; 2559) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23ABS790; 2560) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23CBK134; 2561) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23DBD010; 2562) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23DBL273; 2563) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23EBS007; 2564) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23HBF689; 2565) DYNJ904130I, UDI-DI: 10195327278489(each), 40195327278480(case), Lot Number: 23HBH736; 2566) DYNJ904202D, UDI-DI: 10193489219807(each), 40193489219808(case), Lot Number: 21BDA737; 2567) DYNJ904202D, UDI-DI: 10193489219807(each), 40193489219808(case), Lot Number: 21BDB502; 2568) DYNJ904232K, UDI-DI: 10193489834277(each), 40193489834278(case), Lot Number: 22FDA699; 2569) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21DBU882; 2570) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21GBI161; 2571) DYNJ904240F, UDI-DI: 10193489832952(each), 40193489832953(case), Lot Number: 21HBQ139; 2572) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22BBD028; 2573) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22CBD647; 2574) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22DBH466; 2575) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22DBI520; 2576) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22EBG657; 2577) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO760; 2578) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO761; 2579) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22GBO762; 2580) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22HBS687; 2581) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 22KBK520; 2582) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23ABC466; 2583) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23BBL710; 2584) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23CBD237; 2585) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23EBM510; 2586) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23GBL919; 2587) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23GBM309; 2588) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23HBO357; 2589) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23IBA527; 2590) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23JBW627; 2591) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23LBJ439; 2592) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 23LBP282; 2593) DYNJ904240G, UDI-DI: 10193489999587(each), 40193489999588(case), Lot Number: 24ABI434; 2594) DYNJ904256D, UDI-DI: 10195327173814(each), 40195327173815(case), Lot Number: 24HBJ793; 2595) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21DBF301; 2596) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21EBC430; 2597) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21EBN308; 2598) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21GBJ327; 2599) DYNJ904310D, UDI-DI: 10193489825459(each), 40193489825450(case), Lot Number: 21JBU284; 2600) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 22DBJ147; 2601) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 22IBD724; 2602) DYNJ904310F, UDI-DI: 10195327116187(each), 40195327116188(case), Lot Number: 23CBI402; 2603) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23EBL642; 2604) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23GBM378; 2605) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23IBU342; 2606) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23KBG974; 2607) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 23LBM413; 2608) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24BBD690; 2609) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24FBD344; 2610) DYNJ904310G, UDI-DI: 10195327406356(each), 40195327406357(case), Lot Number: 24GBM228; 2611) DYNJ904458G, UDI-DI: 10193489286366(each), 40193489286367(case), Lot Number: 21ADB432; 2612) DYNJ904458G, UDI-DI: 10193489286366(each), 40193489286367(case), Lot Number: 21CDB169; 2613) DYNJ904498I, UDI-DI: 10195327105334(each), 40195327105335(case), Lot Number: 22DBT224; 2614) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 23LBF268; 2615) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 23LBF492; 2616) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24ABV367; 2617) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24CBK474; 2618) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24DBI357; 2619) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24EBM288; 2620) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24GBD132; 2621) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24HBF630; 2622) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24IBE071; 2623) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24JBH202; 2624) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24JBR375; 2625) DYNJ904660G, UDI-DI: 10195327552046(each), 40195327552047(case), Lot Number: 24LBC617; 2626) DYNJ904660I, UDI-DI: 10198459198700(each), 40198459198701(case), Lot Number: 25BBD970; 2627) DYNJ904660I, UDI-DI: 10198459198700(each), 40198459198701(case), Lot Number: 25CBT644; 2628) DYNJ904660I, UDI-DI: 10198459198700(each), 40198459198701(case), Lot Number: 25EBM716; 2629) DYNJ904660I, UDI-DI: 10198459198700(each), 40198459198701(case), Lot Number: 25GBG723; 2630) DYNJ904660I, UDI-DI: 10198459198700(each), 40198459198701(case), Lot Number: 25HBM753; 2631) DYNJ904681F, UDI-DI: 10193489803815(each), 40193489803816(case), Lot Number: 21BBB752; 2632) DYNJ904777C, UDI-DI: 10195327349349(each), 40195327349340(case), Lot Number: 23EBU756; 2633) DYNJ904777C, UDI-DI: 10195327349349(each), 40195327349340(case), Lot Number: 23FBC935; 2634) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 23IBT081; 2635) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 23JBV743; 2636) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 24BBA952; 2637) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 24BBN394; 2638) DYNJ904777D, UDI-DI: 10195327431259(each), 40195327431250(case), Lot Number: 24BBP708; 2639) DYNJ904777F, UDI-DI: 10198459036149(each), 40198459036140(case), Lot Number: 24LBM668; 2640) DYNJ904777F, UDI-DI: 10198459036149(each), 40198459036140(case), Lot Number: 25DBF368; 2641) DYNJ904777F, UDI-DI: 10198459036149(each), 40198459036140(case), Lot Number: 25FBB525; 2642) DYNJ904777F, UDI-DI: 10198459036149(each), 40198459036140(case), Lot Number: 25GBI874; 2643) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21HBB541; 2644) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21BBK007; 2645) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21CBM824; 2646) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBA001; 2647) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBC377; 2648) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBL903; 2649) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21FBL904; 2650) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21GBH416; 2651) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 21KBZ698; 2652) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22ABW913; 2653) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22CBZ135; 2654) DYNJ904784, UDI-DI: 10889942725335(each), 40889942725336(case), Lot Number: 22GBA676; 2655) DYNJ904784A, UDI-DI: 10195327190422(each), 40195327190423(case), Lot Number: 22IBA582; 2656) DYNJ904784D, UDI-DI: 10195327350864(each), 40195327350865(case), Lot Number: 23EBL415; 2657) DYNJ904784F, UDI-DI: 10195327431266(each), 40195327431267(case), Lot Number: 23HBK149; 2658) DYNJ904784F, UDI-DI: 10195327431266(each), 40195327431267(case), Lot Number: 23HBU587; 2659) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 24BBR413; 2660) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25CBU452; 2661) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25EBQ792; 2662) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25GBI837; 2663) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25KBE198; 2664) DYNJ904784G, UDI-DI: 10195327526245(each), 40195327526246(case), Lot Number: 25KBI026; 2665) DYNJ904827G, UDI-DI: 10195327189389(each), 40195327189380(case), Lot Number: 23FBG232; 2666) DYNJ904827G, UDI-DI: 10195327189389(each), 40195327189380(case), Lot Number: 23HBN523; 2667) DYNJ904827G, UDI-DI: 10195327189389(each), 40195327189380(case), Lot Number: 23JBJ561; 2668) DYNJ904843J, UDI-DI: 10193489928938(each), 40193489928939(case), Lot Number: 23JBV222; 2669) DYNJ904843J, UDI-DI: 10193489928938(each), 40193489928939(case), Lot Number: 23KBK541; 2670) DYNJ904995D, UDI-DI: 10193489735307(each), 40193489735308(case), Lot Number: 21ADB018; 2671) DYNJ905067C, UDI-DI: 10195327098681(each), 40195327098682(case), Lot Number: 24AMF906; 2672) DYNJ905161Q, UDI-DI: 10198459369407(each), 40198459369408(case), Lot Number: 25FMH759; 2673) DYNJ905304A, UDI-DI: 10193489633634(each), 40193489633635(case), Lot Number: 23FLA163; 2674) DYNJ905334A, UDI-DI: 10889942885923(each), 40889942885924(case), Lot Number: 23IBT639; 2675) DYNJ905350B, UDI-DI: 10195327039264(each), 40195327039265(case), Lot Number: 23JLA091; 2676) DYNJ905662B, UDI-DI: 10193489926200(each), 40193489926201(case), Lot Number: 24BMA171; 2677) DYNJ905687C, UDI-DI: 10195327361624(each), 40195327361625(case), Lot Number: 24BMH064; 2678) DYNJ905750B, UDI-DI: 10193489246858(each), 40193489246859(case), Lot Number: 21ADB693; 2679) DYNJ905779, UDI-DI: 10889942971824(each), 40889942971825(case), Lot Number: 21ABT001; 2680) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23FBD281; 2681) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23FBT546; 2682) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23JBI205; 2683) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 23LBF708; 2684) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABG424; 2685) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABK092; 2686) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24ABV903; 2687) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24GBI990; 2688) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 24JBG311; 2689) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25BBR363; 2690) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25CBQ416; 2691) DYNJ905779A, UDI-DI: 10193489929508(each), 40193489929509(case), Lot Number: 25DBQ990; 2692) DYNJ905821F, UDI-DI: 10198459133244(each), 40198459133245(case), Lot Number: 24JBG308; 2693) DYNJ905821F, UDI-DI: 10198459133244(each), 40198459133245(case), Lot Number: 25GBE392; 2694) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21EBO481; 2695) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21GBM188; 2696) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21HBQ156; 2697) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21KBK295; 2698) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 21LBV265; 2699) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 22ABF205; 2700) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 22GBM712; 2701) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 23ABN481; 2702) DYNJ905902C, UDI-DI: 10193489461275(each), 40193489461276(case), Lot Number: 23DBG524; 2703) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23KBK365; 2704) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23LBA464; 2705) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 23LBV082; 2706) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24BBA227; 2707) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24BBV164; 2708) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24CBU599; 2709) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24EBU549; 2710) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24GBL459; 2711) DYNJ905902D, UDI-DI: 10195327478667(each), 40195327478668(case), Lot Number: 24JBB488; 2712) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21CBJ290; 2713) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21GBO014; 2714) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21HBQ064; 2715) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21JBZ318; 2716) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 21LBV306; 2717) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBA089; 2718) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBE604; 2719) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBF669; 2720) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22EBG539; 2721) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22JBC736; 2722) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 22LBT258; 2723) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23ABG370; 2724) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23BBG666; 2725) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23DBO515; 2726) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23EBV687; 2727) DYNJ905903C, UDI-DI: 10193489461350(each), 40193489461351(case), Lot Number: 23GBF834; 2728) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 23LBB395; 2729) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 23LBK530; 2730) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 24EBT891; 2731) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 24HBP101; 2732) DYNJ905903D, UDI-DI: 10195327478674(each), 40195327478675(case), Lot Number: 24JBX485; 2733) DYNJ905975A, UDI-DI: 10195327030278(each), 40195327030279(case), Lot Number: 24KLA858; 2734) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23ADB941; 2735) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23KDB088; 2736) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23KDC349; 2737) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23LDA751; 2738) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 23LDB587; 2739) DYNJ905981D, UDI-DI: 10193489916553(each), 40193489916554(case), Lot Number: 24ADB700; 2740) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23CDC201; 2741) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23EDA506; 2742) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23IDA655; 2743) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23JDC244; 2744) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 23LDA290; 2745) DYNJ905986F, UDI-DI: 10195327194185(each), 40195327194186(case), Lot Number: 24ADA142; 2746) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24BDA786; 2747) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24FDA008; 2748) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24GDA322; 2749) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24HDB013; 2750) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24JDB377; 2751) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 24KDB077; 2752) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 25BDB392; 2753) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 25FDA905; 2754) DYNJ905986G, UDI-DI: 10195327606374(each), 40195327606375(case), Lot Number: 25FDB061; 2755) DYNJ905986I, UDI-DI: 10198459387814(each), 40198459387815(case), Lot Number: 25GDA834; 2756) DYNJ905986I, UDI-DI: 10198459387814(each), 40198459387815(case), Lot Number: 25HDB205; 2757) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23ADB959; 2758) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23CDA312; 2759) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23FDA701; 2760) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23GDB583; 2761) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23HDB808; 2762) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23JDB771; 2763) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 23KDA515; 2764) DYNJ905987C, UDI-DI: 10193489916867(each), 40193489916868(case), Lot Number: 24ADA155; 2765) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24BDA965; 2766) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24BDB171; 2767) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24CDA041; 2768) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24CDB516; 2769) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24DDA311; 2770) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24EDC137; 2771) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24GDA542; 2772) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24JDA321; 2773) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24JDB553; 2774) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 24KDA955; 2775) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 25ADA300; 2776) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 25CME460; 2777) DYNJ905987D, UDI-DI: 10195327606343(each), 40195327606344(case), Lot Number: 25DME523; 2778) DYNJ905987F, UDI-DI: 10198459387807(each), 40198459387808(case), Lot Number: 25GMD864; 2779) DYNJ905987F, UDI-DI: 10198459387807(each), 40198459387808(case), Lot Number: 25HMB501; 2780) DYNJ905987G, UDI-DI: 10198459505225(each), 40198459505226(case), Lot Number: 25IMG396; 2781) DYNJ905987G, UDI-DI: 10198459505225(each), 40198459505226(case), Lot Number: 25LMC401; 2782) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23DDB385; 2783) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23FDA469; 2784) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23GDA912; 2785) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23IDB800; 2786) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23LDA254; 2787) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 23LDB456; 2788) DYNJ905988F, UDI-DI: 10195327194192(each), 40195327194193(case), Lot Number: 24ADA098; 2789) DYNJ906027F, UDI-DI: 10195327419578(each), 40195327419579(case), Lot Number: 23IBI649; 2790) DYNJ906061F, UDI-DI: 10193489279146(each), 40193489279147(case), Lot Number: 21IDA821; 2791) DYNJ906065F, UDI-DI: 10193489861075(each), 40193489861076(case), Lot Number: 21IDB654; 2792) DYNJ906065K, UDI-DI: 10198459051463(each), 40198459051464(case), Lot Number: 24IDB713; 2793) DYNJ906066G, UDI-DI: 10193489861082(each), 40193489861083(case), Lot Number: 22DDB504; 2794) DYNJ906118, UDI-DI: 10193489668568(each), 40193489668569(case), Lot Number: 21BDB916; 2795) DYNJ906122D, UDI-DI: 10198459162671(each), 40198459162672(case), Lot Number: 25ADA326; 2796) DYNJ906156F, UDI-DI: 10198459257360(each), 40198459257361(case), Lot Number: 25DMH512; 2797) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 21IBU423; 2798) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 22KBA603; 2799) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 22KBM272; 2800) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23BBE573; 2801) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23DBH806; 2802) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23EBL829; 2803) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23FBN758; 2804) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23GBE754; 2805) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23KBK823; 2806) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 23LBA471; 2807) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 24BBF086; 2808) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 24CBA354; 2809) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 24FBI195; 2810) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 24HBQ156; 2811) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 24IBH067; 2812) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 25CBJ657; 2813) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 25EBB696; 2814) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 25FBP008; 2815) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 25HBD966; 2816) DYNJ906163B, UDI-DI: 10193489819410(each), 40193489819411(case), Lot Number: 25HBO476; 2817) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 21JBP608; 2818) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 23JBS718; 2819) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 24CBF256; 2820) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 24DBS359; 2821) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 24GBY565; 2822) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 24KBD458; 2823) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 24KBP806; 2824) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25ABA923; 2825) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25ABU290; 2826) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25CBS451; 2827) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25EBC635; 2828) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25EBN024; 2829) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25GBC536; 2830) DYNJ906165B, UDI-DI: 10193489849318(each), 40193489849319(case), Lot Number: 25HBN527; 2831) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 21CBB037; 2832) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 21GBM697; 2833) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 22FBF437; 2834) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 22JBU059; 2835) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 23ABC574; 2836) DYNJ906361, UDI-DI: 10193489742244(each), 40193489742245(case), Lot Number: 23IBO924; 2837) DYNJ906369A, UDI-DI: 10193489286991(each), 40193489286992(case), Lot Number: 21CBZ745; 2838) DYNJ906385F, UDI-DI: 10195327094751(each), 40195327094752(case), Lot Number: 23LBK526; 2839) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22EBC265; 2840) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22EBD169; 2841) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22GBX091; 2842) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22HBD336; 2843) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22HBS418; 2844) DYNJ906507D, UDI-DI: 10195327143220(each), 40195327143221(case), Lot Number: 22JBM029; 2845) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23BBJ904; 2846) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23EBV656; 2847) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23GBM188; 2848) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23HBF888; 2849) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23HBV241; 2850) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 23JBR108; 2851) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24ABU196; 2852) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24DBU228; 2853) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24GBV890; 2854) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24HBP008; 2855) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24IBC264; 2856) DYNJ906507F, UDI-DI: 10195327300104(each), 40195327300105(case), Lot Number: 24IBS335; 2857) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25CBF607; 2858) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25DBH514; 2859) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25DBJ963; 2860) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25FBM372; 2861) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25GBU376; 2862) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25HBD259; 2863) DYNJ906507G, UDI-DI: 10198459149818(each), 40198459149819(case), Lot Number: 25HBP826; 2864) DYNJ906529B, UDI-DI: 10195327162085(each), 40195327162086(case), Lot Number: 23HMF718; 2865) DYNJ906529B, UDI-DI: 10195327162085(each), 40195327162086(case), Lot Number: 23KMB560; 2866) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21ADA649; 2867) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21ADA953; 2868) DYNJ906612F, UDI-DI: 10193489364866(each), 40193489364867(case), Lot Number: 21CDC011; 2869) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21GBG878; 2870) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21GBQ428; 2871) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21IBA121; 2872) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 21LBU091; 2873) DYNJ906670C, UDI-DI: 10193489918427(each), 40193489918428(case), Lot Number: 22CBP273; 2874) DYNJ906891A, UDI-DI: 10193489390773(each), 40193489390774(case), Lot Number: 21HBK630; 2875) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 21IBI423; 2876) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22DBK648; 2877) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22EBO622; 2878) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22GBF054; 2879) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22IBK691; 2880) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22JBL535; 2881) DYNJ906891B, UDI-DI: 10193489972023(each), 40193489972024(case), Lot Number: 22OBA312; 2882) DYNJ906925A, UDI-DI: 10193489410709(each), 40193489410700(case), Lot Number: 21BBP039; 2883) DYNJ906932A, UDI-DI: 10193489411812(each), 40193489411813(case), Lot Number: 21ABE629; 2884) DYNJ906932A, UDI-DI: 10193489411812(each), 40193489411813(case), Lot Number: 21CBE124; 2885) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21DBP476; 2886) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21EBC594; 2887) DYNJ906932B, UDI-DI: 10193489465402(each), 40193489465403(case), Lot Number: 21FBF204; 2888) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22GBG496; 2889) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22GBL768; 2890) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22HBW697; 2891) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22IBA642; 2892) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 22JBR766; 2893) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23CBG971; 2894) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23CBW421; 2895) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23DBE078; 2896) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23FBU011; 2897) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23FBU054; 2898) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23IBW104; 2899) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 23LBO468; 2900) DYNJ906932C, UDI-DI: 10193489921304(each), 40193489921305(case), Lot Number: 24ABI594; 2901) DYNJ906934A, UDI-DI: 10193489411942(each), 40193489411943(case), Lot Number: 21CBE125; 2902) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21GBN206; 2903) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21HBO551; 2904) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21IBJ189; 2905) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21KBD951; 2906) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 21LBO447; 2907) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22EBV617; 2908) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22GBG314; 2909) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22JBF979; 2910) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22LBF478; 2911) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22NBF781; 2912) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 22OBD787; 2913) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23ABG140; 2914) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23BBB416; 2915) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23BBU919; 2916) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23EBD628; 2917) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23FBU012; 2918) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23FBU055; 2919) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23IBI732; 2920) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23JBQ233; 2921) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 23LBQ687; 2922) DYNJ906934C, UDI-DI: 10193489921311(each), 40193489921312(case), Lot Number: 24ABK789; 2923) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBK693; 2924) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBL573; 2925) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23CBP814; 2926) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23EBQ838; 2927) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23EBR992; 2928) DYNJ906975B, UDI-DI: 10195327271558(each), 40195327271559(case), Lot Number: 23FBE543; 2929) DYNJ906975C, UDI-DI: 10195327429324(each), 40195327429325(case), Lot Number: 23GLA651; 2930) DYNJ906975D, UDI-DI: 10195327502874(each), 40195327502875(case), Lot Number: 23JLA461; 2931) DYNJ906975D, UDI-DI: 10195327502874(each), 40195327502875(case), Lot Number: 23KLA951; 2932) DYNJ906975D, UDI-DI: 10195327502874(each), 40195327502875(case), Lot Number: 23LLA276; 2933) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24ELA573; 2934) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24ELA675; 2935) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24FLA271; 2936) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24FLA785; 2937) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24HLA203; 2938) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24HLA860; 2939) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24HLA937; 2940) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24HLB010; 2941) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24JLA212; 2942) DYNJ906975F, UDI-DI: 10195327561697(each), 40195327561698(case), Lot Number: 24LLA094; 2943) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 24LLB071; 2944) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25ALA269; 2945) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25ALA739; 2946) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25ALB018; 2947) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25BLA643; 2948) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25CLA585; 2949) DYNJ906975G, UDI-DI: 10198459202216(each), 40198459202217(case), Lot Number: 25ELA065; 2950) DYNJ906975I, UDI-DI: 10198459351334(each), 40198459351335(case), Lot Number: 25FLA517; 2951) DYNJ906986A, UDI-DI: 10193489448481(each), 40193489448482(case), Lot Number: 22EBR476; 2952) DYNJ907145A, UDI-DI: 10193489411614(each), 40193489411615(case), Lot Number: 21DBA527; 2953) DYNJ907169A, UDI-DI: 10193489431865(each), 40193489431866(case), Lot Number: 22DBN969; 2954) DYNJ907169A, UDI-DI: 10193489431865(each), 40193489431866(case), Lot Number: 22GBF388; 2955) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 23GBU779; 2956) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 23HBC009; 2957) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 23JBW513; 2958) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 23LBF289; 2959) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24ABD872; 2960) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24CBA405; 2961) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24DBE531; 2962) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24EBR583; 2963) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24FBM587; 2964) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24FBT101; 2965) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24IBA402; 2966) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24JBB971; 2967) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 24KBO342; 2968) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25ABL263; 2969) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25CBA070; 2970) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25FBJ778; 2971) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25FBW037; 2972) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25GBS771; 2973) DYNJ907354B, UDI-DI: 10195327451998(each), 40195327451999(case), Lot Number: 25HBL274; 2974) DYNJ907366A, UDI-DI: 10195327474669(each), 40195327474660(case), Lot Number: 23ILA762; 2975) DYNJ907384, UDI-DI: 10193489400083(each), 40193489400084(case), Lot Number: 23AMC133; 2976) DYNJ907384A, UDI-DI: 10195327359102(each), 40195327359103(case), Lot Number: 23HMF336; 2977) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21ABB064; 2978) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21CBU592; 2979) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21FME071; 2980) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 21IMG285; 2981) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22AMG437; 2982) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22CMB538; 2983) DYNJ907387, UDI-DI: 10193489401073(each), 40193489401074(case), Lot Number: 22IMA627; 2984) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23EMF510; 2985) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23FMG424; 2986) DYNJ907387A, UDI-DI: 10195327359119(each), 40195327359110(case), Lot Number: 23GMC399; 2987) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23BMA220; 2988) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23BMI414; 2989) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23DMA497; 2990) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23HMC702; 2991) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23JMB338; 2992) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 23LMA623; 2993) DYNJ907388, UDI-DI: 10193489401080(each), 40193489401081(case), Lot Number: 24AMG458; 2994) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 24CMB602; 2995) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 24DMF795; 2996) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 24HMB473; 2997) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 25BMF925; 2998) DYNJ907388A, UDI-DI: 10195327637569(each), 40195327637560(case), Lot Number: 25CMI159; 2999) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21CBO856; 3000) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21GBD502; 3001) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 21LMG466; 3002) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22BMC147; 3003) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22CMG917; 3004) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22EMA492; 3005) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22GMA663; 3006) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22HMC207; 3007) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 22IME107; 3008) DYNJ907394, UDI-DI: 10193489402469(each), 40193489402460(case), Lot Number: 23AMF670; 3009) DYNJ907394B, UDI-DI: 10195327359225(each), 40195327359226(case), Lot Number: 23DMH233; 3010) DYNJ907394B, UDI-DI: 10195327359225(each), 40195327359226(case), Lot Number: 23EMD892; 3011) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 23KMB103; 3012) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21BBM403; 3013) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21DBK092; 3014) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21DBU998; 3015) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21FBO458; 3016) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21HMH140; 3017) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21JMG857; 3018) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 21LME763; 3019) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22AMG438; 3020) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22BMA354; 3021) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22CMG107; 3022) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22DMD242; 3023) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22DME010; 3024) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22EMG662; 3025) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22KMA556; 3026) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 22LMI031; 3027) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 23AMF672; 3028) DYNJ907402, UDI-DI: 10193489405767(each), 40193489405768(case), Lot Number: 23CMD411; 3029) DYNJ907402A, UDI-DI: 10195327324544(each), 40195327324545(case), Lot Number: 23FMB745; 3030) DYNJ907402A, UDI-DI: 10195327324544(each), 40195327324545(case), Lot Number: 23FMI433; 3031) DYNJ907402A, UDI-DI: 10195327324544(each), 40195327324545(case), Lot Number: 23IMF119; 3032) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21BBM404; 3033) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21BBM502; 3034) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21DBQ728; 3035) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21EBC422; 3036) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21FBC533; 3037) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21HMC149; 3038) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21JMA294; 3039) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21LMA712; 3040) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 21LMF667; 3041) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AMC615; 3042) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AMD448; 3043) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22AME949; 3044) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22CMD161; 3045) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22FMI039; 3046) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22HME124; 3047) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22IMB402; 3048) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 22LMC940; 3049) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23AMA389; 3050) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMF091; 3051) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMG263; 3052) DYNJ907403, UDI-DI: 10193489406030(each), 40193489406031(case), Lot Number: 23BMH367; 3053) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23EMC989; 3054) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23GMA340; 3055) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23GMH562; 3056) DYNJ907403A, UDI-DI: 10195327359218(each), 40195327359219(case), Lot Number: 23IMF120; 3057) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23BMD980; 3058) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23CMD940; 3059) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23FMG426; 3060) DYNJ907413, UDI-DI: 10193489407136(each), 40193489407137(case), Lot Number: 23LMD520; 3061) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 24BME984; 3062) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 24FMF042; 3063) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 24KMF521; 3064) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 25BMI451; 3065) DYNJ907413A, UDI-DI: 10195327637552(each), 40195327637553(case), Lot Number: 25FME364; 3066) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21CBF261; 3067) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21CBS019; 3068) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21EBE793; 3069) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21EBN820; 3070) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21GMB733; 3071) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21HMF279; 3072) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 21JMA677; 3073) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22AMG111; 3074) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22BME206; 3075) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22DMG429; 3076) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22EMG725; 3077) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22HMC453; 3078) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22IMD821; 3079) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22JMI412; 3080) DYNJ907419, UDI-DI: 10193489409130(each), 40193489409131(case), Lot Number: 22LMI456; 3081) DYNJ907419A, UDI-DI: 10195327358754(each), 40195327358755(case), Lot Number: 23DME759; 3082) DYNJ907419A, UDI-DI: 10195327358754(each), 40195327358755(case), Lot Number: 23FMA440; 3083) DYNJ907419A, UDI-DI: 10195327358754(each), 40195327358755(case), Lot Number: 23FMD834; 3084) DYNJ907419A, UDI-DI: 10195327358754(each), 40195327358755(case), Lot Number: 23GMI779; 3085) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21CBR937; 3086) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21EBQ749; 3087) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21FBH069; 3088) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21FBK863; 3089) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21GBD601; 3090) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21GBG470; 3091) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21JMB209; 3092) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 21KMC798; 3093) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22AMD977; 3094) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22AME098; 3095) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22BMD527; 3096) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22BME152; 3097) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22DME804; 3098) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22GME949; 3099) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22JMG905; 3100) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22KME421; 3101) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 22LMD329; 3102) DYNJ907420, UDI-DI: 10193489409055(each), 40193489409056(case), Lot Number: 23BMG352; 3103) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23DMF108; 3104) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23DMI890; 3105) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23EMK083; 3106) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23FME339; 3107) DYNJ907420A, UDI-DI: 10195327358761(each), 40195327358762(case), Lot Number: 23HMA290; 3108) DYNJ907544, UDI-DI: 10193489497298(each), 40193489497299(case), Lot Number: 21DBM325; 3109) DYNJ907581A, UDI-DI: 10195327045296(each), 40195327045297(case), Lot Number: 24HBE816; 3110) DYNJ907681A, UDI-DI: 10195327346980(each), 40195327346981(case), Lot Number: 23CMC353; 3111) DYNJ907681A, UDI-DI: 10195327346980(each), 40195327346981(case), Lot Number: 23CME289; 3112) DYNJ907681A, UDI-DI: 10195327346980(each), 40195327346981(case), Lot Number: 23DMF127; 3113) DYNJ907681B, UDI-DI: 10195327440718(each), 40195327440719(case), Lot Number: 23HME554; 3114) DYNJ907681C, UDI-DI: 10198459207648(each), 40198459207649(case), Lot Number: 25CMH214; 3115) DYNJ907681C, UDI-DI: 10198459207648(each), 40198459207649(case), Lot Number: 25FME289; 3116) DYNJ907681C, UDI-DI: 10198459207648(each), 40198459207649(case), Lot Number: 25HMF144; 3117) DYNJ907681C, UDI-DI: 10198459207648(each), 40198459207649(case), Lot Number: 25HMH992; 3118) DYNJ907683C, UDI-DI: 10195327536480(each), 40195327536481(case), Lot Number: 24ALA185; 3119) DYNJ907683D, UDI-DI: 10195327641849(each), 40195327641840(case), Lot Number: 24DLA934; 3120) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22CBD802; 3121) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22DBI117; 3122) DYNJ907701B, UDI-DI: 10195327054441(each), 40195327054442(case), Lot Number: 22DBO377; 3123) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23BBB632; 3124) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23BBF774; 3125) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23BBU308; 3126) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23CBL575; 3127) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23CBN304; 3128) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23DBG749; 3129) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23FBH106; 3130) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23IBF949; 3131) DYNJ907718A, UDI-DI: 10193489976205(each), 40193489976206(case), Lot Number: 23KBD094; 3132) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23JBK535; 3133) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23JBW637; 3134) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 23KBG210; 3135) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24CBI490; 3136) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24DBB570; 3137) DYNJ907831G, UDI-DI: 10195327513702(each), 40195327513703(case), Lot Number: 24EBC033; 3138) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22HBG360; 3139) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22HBS664; 3140) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 22LBP490; 3141) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23ABE629; 3142) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23BBN968; 3143) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23EBJ494; 3144) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23FBD177; 3145) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23HBI650; 3146) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23JBN816; 3147) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23JBV731; 3148) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23KBE321; 3149) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 23LBV060; 3150) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24CBR860; 3151) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24DBL919; 3152) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24EBT841; 3153) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24GBP839; 3154) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24HBP511; 3155) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24IBS859; 3156) DYNJ907842B, UDI-DI: 10195327192563(each), 40195327192564(case), Lot Number: 24KBJ963; 3157) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25CBA960; 3158) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25CBE672; 3159) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25EBF761; 3160) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25GBE390; 3161) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25HBJ943; 3162) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25IBS487; 3163) DYNJ907842C, UDI-DI: 10198459219917(each), 40198459219918(case), Lot Number: 25KBT033; 3164) DYNJ907880A, UDI-DI: 10195327019204(each), 40195327019205(case), Lot Number: 25ELA138; 3165) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 23IBV289; 3166) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 23LBF416; 3167) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24ABD885; 3168) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24ABS275; 3169) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24CBD690; 3170) DYNJ907908A, UDI-DI: 10195327487041(each), 40195327487042(case), Lot Number: 24EBA383; 3171) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24HBM062; 3172) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24IBH056; 3173) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 24KBD526; 3174) DYNJ907908B, UDI-DI: 10198459033339(each), 40198459033330(case), Lot Number: 25ABF310; 3175) DYNJ907969B, UDI-DI: 10198459025778(each), 40198459025779(case), Lot Number: 24JBM796; 3176) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 22JBO771; 3177) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 22KBK450; 3178) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 22LBE941; 3179) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23ABE898; 3180) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23BBG123; 3181) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23DBI178; 3182) DYNJ908091A, UDI-DI: 10195327015343(each), 40195327015344(case), Lot Number: 23HBY687; 3183) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 21KBL408; 3184) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 21LBI471; 3185) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22CBG743; 3186) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22DBG490; 3187) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22FBM782; 3188) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22IBK801; 3189) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), L ot Number: 22IBT662; 3190) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22JBF721; 3191) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBH049; 3192) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBO246; 3193) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22KBW398; 3194) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 22OBM193; 3195) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23ABL473; 3196) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23BBG422; 3197) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23CBE990; 3198) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23CBG930; 3199) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23GBC887; 3200) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23HBA413; 3201) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23IBA498; 3202) DYNJ908151, UDI-DI: 10193489982800(each), 40193489982801(case), Lot Number: 23KBP882; 3203) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22ELA049; 3204) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 21LLA171; 3205) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22ALA136; 3206) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22CLB151; 3207) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22DLA667; 3208) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA116; 3209) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA293; 3210) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLA782; 3211) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22FLB062; 3212) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22GLB042; 3213) DYNJ908208, UDI-DI: 10195327011468(each), 40195327011469(case), Lot Number: 22GLB112; 3214) DYNJ908232, UDI-DI: 10195327013523(each), 40195327013524(case), Lot Number: 21KDB711; 3215) DYNJ908232, UDI-DI: 10195327013523(each), 40195327013524(case), Lot Number: 22ADB298; 3216) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23CBG248; 3217) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23GBS766; 3218) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23HBK245; 3219) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 23HBN967; 3220) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24DBK634; 3221) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24DBV190; 3222) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24EBN304; 3223) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24GBN485; 3224) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24HBR104; 3225) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24LBN862; 3226) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 24LBO070; 3227) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25CBE077; 3228) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25FBE954; 3229) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25GBU152; 3230) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25IBF012; 3231) DYNJ908261A, UDI-DI: 10195327337759(each), 40195327337750(case), Lot Number: 25LBR103; 3232) DYNJ908305, UDI-DI: 10195327030308(each), 40195327030309(case), Lot Number: 22HBY634; 3233) DYNJ908317A, UDI-DI: 10195327066826(each), 40195327066827(case), Lot Number: 24CDC148; 3234) DYNJ908404B, UDI-DI: 10195327410506(each), 40195327410507(case), Lot Number: 24JDB083; 3235) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22BBC124; 3236) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22HBA389; 3237) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22HBT451; 3238) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22IBE835; 3239) DYNJ908424, UDI-DI: 10195327071158(each), 40195327071159(case), Lot Number: 22IBF959; 3240) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23EBE317; 3241) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23HBW257; 3242) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23IBP270; 3243) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23KBR860; 3244) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23LBS778; 3245) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 23LBT932; 3246) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 24BBV260; 3247) DYNJ908424A, UDI-DI: 10195327286811(each), 40195327286812(case), Lot Number: 24EBD084; 3248) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24FBS775; 3249) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24HBI970; 3250) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24HBV288; 3251) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24IBM825; 3252) DYNJ908424B, UDI-DI: 10198459029011(each), 40198459029012(case), Lot Number: 24JBS437; 3253) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 24LBH067; 3254) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 24LBU118; 3255) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25CBD533; 3256) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25DBL478; 3257) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25FBW186; 3258) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25HBF941; 3259) DYNJ908424C, UDI-DI: 10198459177354(each), 40198459177355(case), Lot Number: 25HBI395; 3260) DYNJ908515B, UDI-DI: 10195327527112(each), 40195327527113(case), Lot Number: 24BMF113; 3261) DYNJ908515B, UDI-DI: 10195327527112(each), 40195327527113(case), Lot Number: 24BMJ733; 3262) DYNJ908515F, UDI-DI: 10198459327070(each), 40198459327071(case), Lot Number: 25EMB416; 3263) DYNJ908515F, UDI-DI: 10198459327070(each), 40198459327071(case), Lot Number: 25GMD105; 3264) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 22DBG283; 3265) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 22GBD311; 3266) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23ABC148; 3267) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23ABM615; 3268) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23BBU902; 3269) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23DBD850; 3270) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 23GBB423; 3271) DYNJ908531, UDI-DI: 10195327085100(each), 40195327085101(case), Lot Number: 24ABV074; 3272) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23EBT798; 3273) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23GBI824; 3274) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23HBO429; 3275) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23IBF496; 3276) DYNJ908642D, UDI-DI: 10195327404291(each), 40195327404292(case), Lot Number: 23JBY557; 3277) DYNJ908644D, UDI-DI: 10195327404307(each), 40195327404308(case), Lot Number: 23GBG730; 3278) DYNJ908644D, UDI-DI: 10195327404307(each), 40195327404308(case), Lot Number: 23HBF890; 3279) DYNJ908650B, UDI-DI: 10198459087394(each), 40198459087395(case), Lot Number: 25CMD819; 3280) DYNJ908650B, UDI-DI: 10198459087394(each), 40198459087395(case), Lot Number: 25CMI540; 3281) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24EBD242; 3282) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24GBC844; 3283) DYNJ908703A, UDI-DI: 10195327567491(each), 40195327567492(case), Lot Number: 24HBH035; 3284) DYNJ908703B, UDI-DI: 10198459091599(each), 40198459091590(case), Lot Number: 24JBW648; 3285) DYNJ908703C, UDI-DI: 10198459229633(each), 40198459229634(case), Lot Number: 25ABV138; 3286) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25CBC210; 3287) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25DBE997; 3288) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25EBS805; 3289) DYNJ908703D, UDI-DI: 10198459252518(each), 40198459252519(case), Lot Number: 25GBO748; 3290) DYNJ908727A, UDI-DI: 10195327330644(each), 40195327330645(case), Lot Number: 23EBJ928; 3291) DYNJ908727C, UDI-DI: 10195327473211(each), 40195327473212(case), Lot Number: 23JBP656; 3292) DYNJ908727C, UDI-DI: 10195327473211(each), 40195327473212(case), Lot Number: 23KBE428; 3293) DYNJ908727C, UDI-DI: 10195327473211(each), 40195327473212(case), Lot Number: 23KBK382; 3294) DYNJ908727C, UDI-DI: 10195327473211(each), 40195327473212(case), Lot Number: 23LBQ416; 3295) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24ABS285; 3296) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24EBJ708; 3297) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24GBT134; 3298) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24IBS872; 3299) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24JBF640; 3300) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24KBD015; 3301) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24LBC648; 3302) DYNJ908727D, UDI-DI: 10195327603205(each), 40195327603206(case), Lot Number: 24LBM679; 3303) DYNJ908727F, UDI-DI: 10198459206382(each), 40198459206383(case), Lot Number: 25ABM638; 3304) DYNJ908727F, UDI-DI: 10198459206382(each), 40198459206383(case), Lot Number: 25CBP185; 3305) DYNJ908727G, UDI-DI: 10198459390982(each), 40198459390983(case), Lot Number: 25GBE403; 3306) DYNJ908766A, UDI-DI: 10195327267834(each), 40195327267835(case), Lot Number: 25GLB098; 3307) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 22IBO061; 3308) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 22LBJ720; 3309) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23ABC149; 3310) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23ABS838; 3311) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23EBJ221; 3312) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23GBU776; 3313) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23GBU940; 3314) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23JBY247; 3315) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23KBM022; 3316) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 23KBW175; 3317) DYNJ908774, UDI-DI: 10195327175122(each), 40195327175123(case), Lot Number: 24BBK238; 3318) DYNJ908800A, UDI-DI: 10195327520816(each), 40195327520817(case), Lot Number: 24BBT194; 3319) DYNJ908802A, UDI-DI: 10195327520793(each), 40195327520794(case), Lot Number: 24BBL052; 3320) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 24LBD495; 3321) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25ABF284; 3322) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25BBS821; 3323) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25CBG036; 3324) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25CBP163; 3325) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25HBR514; 3326) DYNJ908859A, UDI-DI: 10198459181740(each), 40198459181741(case), Lot Number: 25JBQ642; 3327) DYNJ908861, UDI-DI: 10195327195281(each), 40195327195282(case), Lot Number: 23CBW905; 3328) DYNJ908861, UDI-DI: 10195327195281(each), 40195327195282(case), Lot Number: 23GBH518; 3329) DYNJ909027B, UDI-DI: 10195327617301(each), 40195327617302(case), Lot Number: 24KLA186; 3330) DYNJ909027B, UDI-DI: 10195327617301(each), 40195327617302(case), Lot Number: 25ALA334; 3331) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 22LMB692; 3332) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMB428; 3333) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMF978; 3334) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23AMI315; 3335) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23BME735; 3336) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23CME124; 3337) DYNJ909142, UDI-DI: 10195327278632(each), 40195327278633(case), Lot Number: 23DMF511; 3338) DYNJ909142D, UDI-DI: 10195327696115(each), 40195327696116(case), Lot Number: 24IMH986; 3339) DYNJ909142D, UDI-DI: 10195327696115(each), 40195327696116(case), Lot Number: 24JMG326; 3340) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25AMA338; 3341) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25AMI497; 3342) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25BMI753; 3343) DYNJ909142F, UDI-DI: 10198459143755(each), 40198459143756(case), Lot Number: 25CMF877; 3344) DYNJ909142G, UDI-DI: 10198459354496(each), 40198459354497(case), Lot Number: 25FME290; 3345) DYNJ909142G, UDI-DI: 10198459354496(each), 40198459354497(case), Lot Number: 25HMH554; 3346) DYNJ909383C, UDI-DI: 10198459440168(each), 40198459440169(case), Lot Number: 25HBE999; 3347) DYNJ909383C, UDI-DI: 10198459440168(each), 40198459440169(case), Lot Number: 25HBO948; 3348) DYNJ909383C, UDI-DI: 10198459440168(each), 40198459440169(case), Lot Number: 25KBB193; 3349) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23DBG095; 3350) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23FBB682; 3351) DYNJ909432, UDI-DI: 10195327354077(each), 40195327354078(case), Lot Number: 23GBG727; 3352) DYNJ909433, UDI-DI: 10195327354084(each), 40195327354085(case), Lot Number: 23EBV082; 3353) DYNJ909433, UDI-DI: 10195327354084(each), 40195327354085(case), Lot Number: 23JBQ682; 3354) DYNJ909433, UDI-DI: 10195327354084(each), 40195327354085(case), Lot Number: 23LMG567; 3355) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23EME409; 3356) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23EMF267; 3357) DYNJ909505, UDI-DI: 10195327386030(each), 40195327386031(case), Lot Number: 23FMD557; 3358) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 23JME303; 3359) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 23LMA802; 3360) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24AMA487; 3361) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24BMD766; 3362) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24CMG434; 3363) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24GMH511; 3364) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24HMB850; 3365) DYNJ909505A, UDI-DI: 10195327448448(each), 40195327448449(case), Lot Number: 24IMA563; 3366) DYNJ909505B, UDI-DI: 10198459124440(each), 40198459124441(case), Lot Number: 24JMI945; 3367) DYNJ909505C, UDI-DI: 10198459173561(each), 40198459173562(case), Lot Number: 25DMH508; 3368) DYNJ909505C, UDI-DI: 10198459173561(each), 40198459173562(case), Lot Number: 25EMJ685; 3369) DYNJ909505C, UDI-DI: 10198459173561(each), 40198459173562(case), Lot Number: 25HMG745; 3370) DYNJ909505D, UDI-DI: 10198459594229(each), 40198459594220(case), Lot Number: 25LMB637; 3371) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 23GLA007; 3372) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 23GLA881; 3373) DYNJ909510, UDI-DI: 10195327387235(each), 40195327387236(case), Lot Number: 24ILA570; 3374) DYNJ909510A, UDI-DI: 10198459124457(each), 40198459124458(case), Lot Number: 24JLB326; 3375) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25DLA162; 3376) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25DLA665; 3377) DYNJ909510C, UDI-DI: 10198459211874(each), 40198459211875(case), Lot Number: 25ELA490; 3378) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24CBB520; 3379) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24GBI979; 3380) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24HBD480; 3381) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24JBI640; 3382) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24JBW034; 3383) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 24LBH562; 3384) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25ABK779; 3385) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25CBO176; 3386) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25EBS311; 3387) DYNJ909563A, UDI-DI: 10195327569136(each), 40195327569137(case), Lot Number: 25GBC518; 3388) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 23FLA838; 3389) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 23ILA442; 3390) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 23KLA244; 3391) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24ALB062; 3392) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24DLA175; 3393) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24DLA572; 3394) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24HLA301; 3395) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24JLB214; 3396) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24KLB020; 3397) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 24LLA402; 3398) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25BLA450; 3399) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25CLB033; 3400) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25ELA790; 3401) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25FLA990; 3402) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25HLA515; 3403) DYNJ909623, UDI-DI: 10195327414993(each), 40195327414994(case), Lot Number: 25HLA855; 3404) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 23FMD859; 3405) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 23IMA627; 3406) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 23KMC366; 3407) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24AMC924; 3408) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24DMA086; 3409) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24DMI521; 3410) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24DMJ862; 3411) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24GMD239; 3412) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24HMH300; 3413) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24JMF321; 3414) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 24LMA925; 3415) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25AMI296; 3416) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25BME883; 3417) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25CMH632; 3418) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25DMD557; 3419) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25EMI600; 3420) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25FMG653; 3421) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25HMF682; 3422) DYNJ909624, UDI-DI: 10195327411602(each), 40195327411603(case), Lot Number: 25IME399; 3423) DYNJ909627, UDI-DI: 10195327415198(each), 40195327415199(case), Lot Number: 23IBL905; 3424) DYNJ909630, UDI-DI: 10195327416843(each), 40195327416844(case), Lot Number: 23GBR009; 3425) DYNJ909630, UDI-DI: 10195327416843(each), 40195327416844(case), Lot Number: 23IBK293; 3426) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23IBR224; 3427) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23JBO869; 3428) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23JBV722; 3429) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 23LBC689; 3430) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24CBK434; 3431) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24CBP841; 3432) DYNJ909630A, UDI-DI: 10195327513740(each), 40195327513741(case), Lot Number: 24DBB572; 3433) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 23HBV261; 3434) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 23LBN237; 3435) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24ABH462; 3436) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24BBL043; 3437) DYNJ909646, UDI-DI: 10195327420864(each), 40195327420865(case), Lot Number: 24CBI514; 3438) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24DBR221; 3439) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24GBP217; 3440) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 24HBM093; 3441) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25ABF309; 3442) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25DBF495; 3443) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25EBL872; 3444) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25GBO024; 3445) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25HBA067; 3446) DYNJ909646A, UDI-DI: 10195327676414(each), 40195327676415(case), Lot Number: 25IBD880; 3447) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 23KBM218; 3448) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24CBG292; 3449) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24CBO933; 3450) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24DBI478; 3451) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24DBM437; 3452) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24EBB383; 3453) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24FBH757; 3454) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24GBM375; 3455) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 24JBN452; 3456) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABC089; 3457) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABC094; 3458) DYNJ909710, UDI-DI: 10195327447687(each), 40195327447688(case), Lot Number: 25ABS346; 3459) DYNJ909710A, UDI-DI: 10198459327674(each), 40198459327675(case), Lot Number: 25FBE958; 3460) DYNJ909710A, UDI-DI: 10198459327674(each), 40198459327675(case), Lot Number: 25GBE494; 3461) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25GBR226; 3462) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25HBM092; 3463) DYNJ909710B, UDI-DI: 10198459401909(each), 40198459401900(case), Lot Number: 25KBO393; 3464) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23HMH032; 3465) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23IMD312; 3466) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 23KMF164; 3467) DYNJ909728, UDI-DI: 10195327462451(each), 40195327462452(case), Lot Number: 24AMC915; 3468) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 23KME642; 3469) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 23LMA320; 3470) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24AMI270; 3471) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24BMH621; 3472) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24BMK233; 3473) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24CME898; 3474) DYNJ909729, UDI-DI: 10195327462536(each), 40195327462537(case), Lot Number: 24CMG034; 3475) DYNJ909750B, UDI-DI: 10195327543525(each), 40195327543526(case), Lot Number: 23LBG699; 3476) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24BBP450; 3477) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24CBQ758; 3478) DYNJ909750D, UDI-DI: 10195327627775(each), 40195327627776(case), Lot Number: 24EBC579; 3479) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 23LMB636; 3480) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24AMD056; 3481) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24BMF839; 3482) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24EMH866; 3483) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 24JMD236; 3484) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 25EMG060; 3485) DYNJ909971, UDI-DI: 10195327522933(each), 40195327522934(case), Lot Number: 25GMA576; 3486) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24AMA700; 3487) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24DMG271; 3488) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24FMA347; 3489) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24GMF289; 3490) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24HMG231; 3491) DYNJ909972, UDI-DI: 10195327522957(each), 40195327522958(case), Lot Number: 24IMI949; 3492) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24KBF862; 3493) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24KBK896; 3494) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24LBA993; 3495) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 24LBR437; 3496) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25CBB461; 3497) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25DBS303; 3498) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25FBB523; 3499) DYNJ910069A, UDI-DI: 10198459156878(each), 40198459156879(case), Lot Number: 25GBT129; 3500) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24CBP577; 3501) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24GBE022; 3502) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24HBR313; 3503) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24JBO154; 3504) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 24LBI444; 3505) DYNJ910307, UDI-DI: 10195327622770(each), 40195327622771(case), Lot Number: 25ABC156; 3506) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25CBT525; 3507) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25EBE470; 3508) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25FBP016; 3509) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25JBH697; 3510) DYNJ910307A, UDI-DI: 10198459264405(each), 40198459264406(case), Lot Number: 25JBW820; 3511) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24CBN438; 3512) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24FBL642; 3513) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24GBR782; 3514) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24JBK741; 3515) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24KBS304; 3516) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 24LBP825; 3517) DYNJ910308, UDI-DI: 10195327622787(each), 40195327622788(case), Lot Number: 25BBA245; 3518) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25DBB387; 3519) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25DBI472; 3520) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25FBF532; 3521) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25GBO042; 3522) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25KBC349; 3523) DYNJ910308A, UDI-DI: 10198459264412(each), 40198459264413(case), Lot Number: 25LBS897; 3524) DYNJ910326, UDI-DI: 10195327628154(each), 40195327628155(case), Lot Number: 24DBM380; 3525) DYNJ910326B, UDI-DI: 10198459075001(each), 40198459075002(case), Lot Number: 24HBL975; 3526) DYNJ910326B, UDI-DI: 10198459075001(each), 40198459075002(case), Lot Number: 24KBQ357; 3527) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25CBT572; 3528) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25CBT588; 3529) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25DBC255; 3530) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25DBN328; 3531) DYNJ910326C, UDI-DI: 10198459266751(each), 40198459266752(case), Lot Number: 25DBR010; 3532) DYNJ910358, UDI-DI: 10195327648015(each), 40195327648016(case), Lot Number: 24DDC162; 3533) DYNJ910416, UDI-DI: 10195327680954(each), 40195327680955(case), Lot Number: 24EBA717; 3534) DYNJ910416, UDI-DI: 10195327680954(each), 40195327680955(case), Lot Number: 24HBJ616; 3535) DYNJ910416, UDI-DI: 10195327680954(each), 40195327680955(case), Lot Number: 24JBH673; 3536) DYNJ910416A, UDI-DI: 10198459176715(each), 40198459176716(case), Lot Number: 25BBI989; 3537) DYNJ910416A, UDI-DI: 10198459176715(each), 40198459176716(case), Lot Number: 25CBF067; 3538) DYNJ910416A, UDI-DI: 10198459176715(each), 40198459176716(case), Lot Number: 25CBL025; 3539) DYNJ910416A, UDI-DI: 10198459176715(each), 40198459176716(case), Lot Number: 25DBS724; 3540) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24FMD548; 3541) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24GMG658; 3542) DYNJ9104 31, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24KMD003; 3543) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24KMH139; 3544) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 24LMH172; 3545) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25AMA256; 3546) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25BMD967; 3547) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25DMB038; 3548) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25DMI838; 3549) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25GMB870; 3550) DYNJ910431, UDI-DI: 10195327680886(each), 40195327680887(case), Lot Number: 25IMG659; 3551) DYNJ910431A, UDI-DI: 10198459588051(each), 40198459588052(case), Lot Number: 25LMG122; 3552) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 24FBP464; 3553) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25BBD805; 3554) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25BBD954; 3555) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25DBI967; 3556) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25EBK628; 3557) DYNJ910433, UDI-DI: 10195327682880(each), 40195327682881(case), Lot Number: 25GBQ032; 3558) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 24JBR694; 3559) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 24LBH399; 3560) DYNJ910578A, UDI-DI: 10198459095658(each), 40198459095659(case), Lot Number: 25BBC824; 3561) DYNJ910578C, UDI-DI: 10198459272653(each), 40198459272654(case), Lot Number: 25CBG904; 3562) DYNJ910578C, UDI-DI: 10198459272653(each), 40198459272654(case), Lot Number: 25DBB506; 3563) DYNJ910678A, UDI-DI: 10198459336737(each), 40198459336738(case), Lot Number: 25FBG095; 3564) DYNJ910678A, UDI-DI: 10198459336737(each), 40198459336738(case), Lot Number: 25GBN594; 3565) DYNJ910732C, UDI-DI: 10198459494062(each), 40198459494063(case), Lot Number: 25JML588; 3566) DYNJ910732C, UDI-DI: 10198459494062(each), 40198459494063(case), Lot Number: 25LME962; 3567) DYNJ910733C, UDI-DI: 10198459494079(each), 40198459494070(case), Lot Number: 25JBB540; 3568) DYNJ910733C, UDI-DI: 10198459494079(each), 40198459494070(case), Lot Number: 25LME692; 3569) DYNJ910819, UDI-DI: 10198459141140(each), 40198459141141(case), Lot Number: 24LBH782; 3570) DYNJ910819, UDI-DI: 10198459141140(each), 40198459141141(case), Lot Number: 25ABS475; 3571) DYNJ910819, UDI-DI: 10198459141140(each), 40198459141141(case), Lot Number: 25BBE745; 3572) DYNJ910819, UDI-DI: 10198459141140(each), 40198459141141(case), Lot Number: 25BBK414; 3573) DYNJ910819, UDI-DI: 10198459141140(each), 40198459141141(case), Lot Number: 25BBO651; 3574) DYNJ910828, UDI-DI: 10198459144875(each), 40198459144876(case), Lot Number: 24JBY026; 3575) DYNJ910864, UDI-DI: 10198459161308(each), 40198459161309(case), Lot Number: 24KBB660; 3576) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25AMJ318; 3577) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25EMF898; 3578) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25HME905; 3579) DYNJ910928, UDI-DI: 10198459204333(each), 40198459204334(case), Lot Number: 25LMA455; 3580) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25EBB646; 3581) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25FBD829; 3582) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25GBM729; 3583) DYNJ911026A, UDI-DI: 10198459338946(each), 40198459338947(case), Lot Number: 25IBJ122; 3584) DYNJ911026B, UDI-DI: 10198459530531(each), 40198459530532(case), Lot Number: 25KBC353; 3585) DYNJ911032, UDI-DI: 10198459234460(each), 40198459234461(case), Lot Number: 25BBA642; 3586) DYNJ911032A, UDI-DI: 10198459338229(each), 40198459338220(case), Lot Number: 25FBK365; 3587) DYNJ911032A, UDI-DI: 10198459338229(each), 40198459338220(case), Lot Number: 25GBN709; 3588) DYNJ911032A, UDI-DI: 10198459338229(each), 40198459338220(case), Lot Number: 25KBM944; 3589) DYNJ911056, UDI-DI: 10198459238413(each), 40198459238414(case), Lot Number: 25CBF074; 3590) DYNJ911149, UDI-DI: 10198459273223(each), 40198459273224(case), Lot Number: 25CBR521; 3591) DYNJ911355, UDI-DI: 10198459326288(each), 40198459326289(case), Lot Number: 25EBV320; 3592) DYNJ911356, UDI-DI: 10198459326509(each), 40198459326500(case), Lot Number: 25JBA531; 3593) DYNJ911356, UDI-DI: 10198459326509(each), 40198459326500(case), Lot Number: 25KBT505; 3594) DYNJ911485, UDI-DI: 10198459360855(each), 40198459360856(case), Lot Number: 25FBV096; 3595) DYNJ911485, UDI-DI: 10198459360855(each), 40198459360856(case), Lot Number: 25GBR982; 3596) DYNJ911644, UDI-DI: 10198459502200(each), 40198459502201(case), Lot Number: 25IBP089; 3597) DYNJ911838, UDI-DI: 10198459552786(each), 40198459552787(case), Lot Number: 25KBE192; 3598) DYNJ911838, UDI-DI: 10198459552786(each), 40198459552787(case), Lot Number: 25LBM762; 3599) DYNJ911956, UDI-DI: 10198459587726(each), 40198459587727(case), Lot Number: 25LBA968; 3600) DYNJ911956, UDI-DI: 10198459587726(each), 40198459587727(case), Lot Number: 25LBF532; 3601) DYNJ9682800V, UDI-DI: 10195327257248(each), 40195327257249(case), Lot Number: 22KDA037; 3602) DYNJC9306L, UDI-DI: 10195327478148(each), 40195327478149(case), Lot Number: 23IMD838; 3603) DYNJG901000A, UDI-DI: 10195327677077(each), 40195327677078(case), Lot Number: 25DLA778; 3604) DYNJS3013, UDI-DI: 10884389201572(each), 40884389201573(case), Lot Number: 22DBO761; 3605) DYNJT2144R, UDI-DI: 10889942399192(each), 40889942399193(case), Lot Number: 20LMF883; 3606) DYNJT3352, UDI-DI: 10198459324123(each), 40198459324124(case), Lot Number: 25EBJ794; 3607) DYNJT3397, UDI-DI: 10198459324918(each), 40198459324919(case), Lot Number: 25FBN276; 3608) DYNJT3411, UDI-DI: 10198459325052(each), 40198459325053(case), Lot Number: 25FBN247; 3609) DYNJT3528, UDI-DI: 10198459332654(each), 40198459332655(case), Lot Number: 25EBD255; 3610) DYNJT5184, UDI-DI: 10198459467592(each), 40198459467593(case), Lot Number: 25HMG034; 3611) DYNJT7166, UDI-DI: 10198459602740(each), 40198459602741(case), Lot Number: 25KMI656; 3612) DYNJV0056AA, UDI-DI: 10198459489112(each), 40198459489113(case), Lot Number: 25LMC937; 3613) DYNJV0154AB, UDI-DI: 10193489336191(each), 40193489336192(case), Lot Number: 21BBM420; 3614) DYNJVB91045, UDI-DI: 10193489284478(each), 40193489284479(case), Lot Number: 21BDB808; 3615) DYNJVB91128D, UDI-DI: 10195327367916(each), 40195327367917(case), Lot Number: 23JLA241; 3616) HM280B, UDI-DI: 10195327481179(each), 40195327481170(case), Lot Number: 23JBT647; 3617) OPS767M, UDI-DI: 10193489552966(each), 40193489552967(case), Lot Number: 21BDA327; 3618) OPS767M, UDI-DI: 10193489552966(each), 40193489552967(case), Lot Number: 21HDA963; 3619) PHS389811I, UDI-DI: 10193489333657(each), 40193489333658(case), Lot Number: 22AME448; 3620) PHS389811J, UDI-DI: 10195327134051(each), 40195327134052(case), Lot Number: 23CMG855; 3621) PHS389811J, UDI-DI: 10195327134051(each), 40195327134052(case), Lot Number: 23FMB784; 3622) PHS389913T, UDI-DI: 10193489333671(each), 40193489333672(case), Lot Number: 20LMF727; 3623) PHS389913U, UDI-DI: 10193489453287(each), 40193489453288(case), Lot Number: 21BMC221; 3624) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21FMB052; 3625) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21HMB975; 3626) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21JMB398; 3627) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 21KME416; 3628) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22AMD873; 3629) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22DMC685; 3630) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22EMB748; 3631) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22HMF956; 3632) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 22HMF957; 3633) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 23AMG879; 3634) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 23CMG491; 3635) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 23HMA239; 3636) PHS389913V, UDI-DI: 10193489914962(each), 40193489914963(case), Lot Number: 23HMI661; 3637) PHS393071004C, UDI-DI: 10193489222623(each), 40193489222624(case), Lot Number: 21ADA094; 3638) PHS393205012C, UDI-DI: 10193489222852(each), 40193489222853(case), Lot Number: 21ADA066; 3639) PHS518775J, UDI-DI: 10195327034627(each), 40195327034628(case), Lot Number: 22HMG648; 3640) PHS807386J, UDI-DI: 10193489927078(each), 40193489927079(case), Lot Number: 23HBM378
What to do
Check the model, lot or serial numbers above against your device. If yours is affected, contact the manufacturer or your healthcare provider before you stop using a device you rely on.
Recall facts
| Recall number | Z-2114-2026 |
| Category | Medical device |
| Agency | U.S. Food and Drug Administration (FDA) |
| Severity | Class II |
| Status | Ongoing |
| Recall started | January 7, 2026 |
| Reported | May 13, 2026 |
| Company | Medline Industries, LP |
| Company location | Northfield, IL, United States |
| Initiated by | Voluntary: Firm initiated |
| Quantity | 160980 units |
More recalls from Medline Industries, LP · all medical device recalls · recalls from January 2026